K Number
K964694

Validate with FDA (Live)

Device Name
ELECSYS LH ASSAY
Date Cleared
1997-01-24

(63 days)

Product Code
Regulation Number
862.1485
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro quantitative determination of human luteinizing hormone in human serum and plasma.

Device Description

Sandwich principle. Total duration of assay: 18 minutes.
• 1st incubation (9min.): 20 µL of sample, a biotinylated monoclonal LH-specific antibody (75 µL) and a monoclonal LH-specific antibody labeled with a ruthenium complex (75 µL)** react to form a sandwich complex.
• 2nd incubation (9min.): after addition of streptavidin-coated microparticles (30 µL), the complex becomes bound to the solid phase via interaction of biotin and streptavidin.
**Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)3)2+)
•The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
•Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

AI/ML Overview

The provided text describes the performance characteristics of the Elecsys® LH Assay and compares it to a predicate device, the Enzymun-Test® LH. The document acts as a 510(k) Summary, indicating that the study is for regulatory submission to establish substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state "acceptance criteria" for each performance metric but rather presents "Performance Characteristics" for both the new Elecsys® LH and the predicate Enzymun-Test® LH. The implication is that the Elecsys® LH's performance should be comparable to or better than the predicate device for substantial equivalence.

FeatureAcceptance Criteria (Implied: Comparable to or better than Predicate)Elecsys® LH Reported Performance
PrecisionSimilar %CV values to predicate across low, mid, and high levels.Low: Within-Run Mean 0.54 mIU/mL, %CV 1.82%; Total Mean 0.54 mIU/mL, %CV 5.17% Mid: Within-Run Mean 9.38 mIU/mL, %CV 1.13%; Total Mean 9.38 mIU/mL, %CV 1.97% High: Within-Run Mean 50.72 mIU/mL, %CV 0.81%; Total Mean 50.72 mIU/mL, %CV 1.99%
Lower Detection LimitComparable to or lower than predicate (0.50 mIU/mL).0.10 mIU/mL
LinearityComparable to predicate range (0.5 - 150 mIU/mL) with a deviation of ±10%.0.1 - 200 mIU/mL (with a deviation from a linear line of ±10%)
Method ComparisonStrong correlation with the predicate device (r value close to 1).Vs Enzymun-Test® LH: Least Squares: y = 1.00x - 0.199, r = 0.993, SEE = 1.141 Passing/Bablok: y = 0.964x + 0.040, r = 0.993, SEE = 0.456
Interfering substancesNo interference at or above specific concentration limits for bilirubin, hemoglobin, lipemia, biotin, and rheumatoid factor.No interference at: Bilirubin: 25.0 mg/dL Hemoglobin: 1 g/dL Lipemia: 1500 mg/dL Biotin: 30 ng/mL Rheumatoid Factor: 1500 U/mL
Specificity (Cross-reactivity)Low or no cross-reactivity with FSH, HCG, TSH, HGH, HPL (e.g., < 0.1% or 0.00%).< 0.1% for: FSH (300 mIU/mL) HCG (600 IU/mL) TSH (300 µIU/mL) HGH (600 µIU/mL) HPL (13.80 nmol/mL)

2. Sample size used for the test set and the data provenance:

  • Precision Test Set Sample Size: For the Elecsys® LH, N=60 for each of the low, mid, and high levels.
  • Method Comparison Test Set Sample Size: For the Elecsys® LH vs Enzymun-Test® LH, N=166.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The document indicates it's part of a 510(k) submission, implying it's clinical performance data. It does not specify if it's retrospective or prospective, but clinical performance studies for diagnostic devices typically involve prospective sample collection or the use of well-characterized biobank samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the text. The device is an in vitro diagnostic assay that generates quantitative results for Luteinizing Hormone (LH). The "ground truth" for such devices is typically established through direct measurement on reference samples or comparison with a well-established reference method, rather than expert interpretation of images or clinical assessments by individual experts.

4. Adjudication method for the test set:

This information is not applicable for this type of in vitro diagnostic assay. Adjudication methods (like 2+1 or 3+1) are typically used in studies where human readers interpret data (e.g., medical images) and their interpretations need to be reconciled to establish a consensus ground truth. Here, the device produces a numerical result.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable as the document describes an in vitro diagnostic device for quantitative LH measurement, not an AI-assisted diagnostic tool that aids human readers. Therefore, an MRMC study or its effect size is irrelevant in this context.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

The device is an automated immunoassay system (Elecsys® LH Assay on the Elecsys® 2010 instrument). The performance characteristics reported (Precision, Lower Detection Limit, Linearity, Method Comparison, Interfering Substances, Specificity) represent the standalone performance of the algorithm/device system. It operates without human interpretation of the primary signal; results are quantitatively determined by the instrument and its built-in algorithms.

7. The type of ground truth used:

The "ground truth" for this type of immunoassay is:

  • Reference measurements: For precision, linearity, and lower detection limit, the "true" values of LH in control samples or serially diluted samples are used.
  • Comparison to a predicate device: For method comparison, the "true" values are derived from the results of the legally marketed predicate device (Enzymun-Test® LH), which serves as the comparative standard.
  • Known concentrations of interferents or related substances: For interfering substances and specificity, "true" known concentrations of these compounds are spiked into samples to assess their impact.

8. The sample size for the training set:

This information is not provided and is likely not directly applicable in the same way it would be for a machine learning model. This is a traditional immunoassay. The "training" of such a system involves extensive biochemical and engineering development, optimization of reagents, and establishment of calibration curves. The sample sizes for these developmental stages are not typically reported as a single "training set" like in AI/ML. The provided sample sizes are for performance validation (i.e., the "test set").

9. How the ground truth for the training set was established:

As mentioned above, the concept of a "training set" with an established ground truth in the AI/ML sense doesn't directly apply here. For a traditional immunoassay, the "ground truth" during development (analogous to training) is established through:

  • Known concentrations: Using highly purified and precisely quantified LH standards.
  • Reference methods: Comparing early versions of the assay to established reference methods or highly characterized internal standards.
  • Clinical panels: Testing the assay on panels of patient samples with confirmed clinical states or values from reference laboratories.
  • Reagent characterization: Thorough characterization of antibodies and other reagents for specificity and binding affinity.

{0}------------------------------------------------

JAN 2 4 1997

Image: Boehringer Mannheim Corporation logo and K964694
510(k) Summary
IntroductionAccording to the requirements of 21 CFR 807.92, the following informationprovides sufficient detail to understand the basis for a determination ofsubstantial equivalence.
1.Submittername,address,contactBoehringer Mannheim Corporation2400 Bisso LaneP.O. Box 4117Concord, CA 94524-4117(510) 674 - 0690 extension 8240FAX 510 687 - 1850
Contact Person: Betsy Soares-MaddoxDate Prepared: November 19, 1996
2.Device nameProprietary name: Elecsys® LH AssayCommon name: Electrochemiluminescence assay for the determination ofhuman luteinizing hormone (LH).
Classification name: System, Test, Human Luteinizing Hormone
3.PredicatedeviceWe claim substantial equivalence to the Enzymun® LH Assay (K900799).
4.DeviceDescriptionSandwich principle. Total duration of assay: 18 minutes.• 1st incubation (9min.): 20 µL of sample, a biotinylated monoclonal LH-specific antibody (75 µL) and a monoclonal LH-specific antibody labeled witha ruthenium complex (75 µL)** react to form a sandwich complex.• 2nd incubation (9min.): after addition of streptavidin-coated microparticles(30 µL), the complex becomes bound to the solid phase via interaction ofbiotin and streptavidin.
**Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)3)2+)
4.DeviceDescription•The reaction mixture is aspirated into the measuring cell where themicroparticles are magnetically captured onto the surface of the electrode.Unbound substances are then removed with ProCell. Application of a voltageto the electrode then induces chemiluminescent emission which is measuredby a photomultiplier (0.4 second read frame).•Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided viathe reagent bar code.
5.Intended useImmunoassay for the in vitro quantitative determination of human luteinizinghormone in human serum and plasma.
6.Comparisonto predicatedeviceThe Boehringer Mannheim Elecsys® LH Assay is substantially equivalent toother products in commercial distribution intended for similar use. Mostnotably it is substantially equivalent to the currently marketed Enzymun®LH Assay (K900779).
The following table compares the Elecsys® LH Assay with the predicatedevice, Enzymun® LH Assay. Specific data on the performance of the testhave been incorporated into the draft labeling in attachment 5. Labeling forthe predicate device in provided in attachment 6.
Similarities:
•Intended Use: Immunoassay for the in vitro quantitative determinationof human luteinizing hormone (LH)
•Sample type: Serum and plasma
• Antibody: Same pair of monoclonal mouse anti-LH antibodies
•Solid phase binding principle: Streptavidin/Biotin
Continued on next page

:
一一

..............................................................................................................................................................................

{1}------------------------------------------------

510(k) Summary, Continued

!

: . "

{2}------------------------------------------------

510(k) Summary, Continued

Differences:

Comparison
to predicate
device cont.

FeatureElecsys® LHEnzymun-Test® LH
AssayStandardizationEnzmun® LHWHO # 80/552
Detection methodElectrochemiluminescenceELISA/1-step sandwichassay using streptavidintechnology
InstrumentrequiredElecsys® 2010ES 300
CalibrationStabilityA calibration is recommendedevery 7 days if kits is notconsumed; 4 weeks with samereagent lot if reagent isconsumed within 7 days.Full calibration requiredevery 2 weeks. One-pointcalibration required everyrun.

mance Characteristics:

FeatureElecsys® LHEnzymun-Test® LH
PrecisionModified NCCLS (mIU/mL):Modified NCCLS (mIU/mL):
LevelLowMidHighLowMidHigh
N606060120120120
Within-Run Mean0.549.3850.723.613.559.5
%CV1.821.130.812.93.81.5
Total Mean0.549.3850.723.613.559.5
%CV5.171.971.994.44.73.9
Lower DetectionLimit0.10 mIU/mL0.50 mIU/mL

Continued on next page

{3}------------------------------------------------

510(k) Summary, Continued

l

Performance Characteristics:

.

Comparison
to predicate
device, (cont.)

FeatureElecsys® LHEnzymun-Test® LH
Linearity0.1 - 200 mIU/mL (with adeviation from a linear line of±10%)0.5 - 150 mIU/mL (with adeviation from a linear line of±10%)
MethodComparisonVs Enzymun-Test® LHLeast Squares$y = 1.00x - 0.199$$r = 0.993$$SEE = 1.141$$N = 166$Passing/Bablok$y = 0.964x + 0.040$$r = 0.993$$SEE = 0.456$$N = 166$Vs Enzymun-Test® LHLeast Squares$y = 0.93x + 0.42$$r = 0.953$$SEE = 5.079$$N = 62$
InterferingsubstancesNo interference at:No interference at:
Bilirubin25.0 mg/dL64.5 mg/dL
Hemoglobin1 g/dL1 g/dL
Lipemia1500 mg/dL1250 mg/dL
Biotin30 ng/mL40 ng/mL
RheumatoidFactor1500 U/mLno interference
SpecificityLevel tested     % Cross-reactivityLevel tested     % Cross-reactivity
FSH300 mIU/mL     < 0.1200 mU/mL     0.00
HCG600 IU/mL     < 0.1200 U/mL     0.00
TSH300 µIU/mL     < 0.1100 μU/mL     0.00
HGH600 µIU/mL     < 0.1200 ng/mL     0.00
HPL13.80 nmol/mL     < 0.1--

. -

29

§ 862.1485 Luteinizing hormone test system.

(a)
Identification. A luteinizing hormone test system is a device intended to measure luteinizing hormone in serum and urine. Luteinizing hormone measurements are used in the diagnosis and treatment of gonadal dysfunction.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.