K Number
K964444

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-08-11

(278 days)

Product Code
Regulation Number
888.3020
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QuickLock Orthopaedic Targeting System is indicated for use whenever distal interlocking screws are placed in intramedullary nails. The system includes nail-specific, reusable accessory adapter components that enable mechanical attachment of the QuickLock Probe to the nail system proximal instrumentation and alignment of the QuickLock Guide during calibration.

Device Description

Not Found

AI/ML Overview

Here's an analysis of the provided text, focusing on acceptance criteria and study details for the QuickLock Orthopaedic Targeting System:

Acceptance Criteria and Device Performance

The document does not explicitly state formal "acceptance criteria" in a quantitative, pass/fail manner. Instead, it presents a comparative effectiveness study against predicate devices, highlighting improvements in various aspects. The "acceptance criteria" can be inferred from the areas where the QuickLock system demonstrates superior or equivalent performance to the predicate devices.

Table 1: Inferred Acceptance Criteria and Reported Device Performance

Inferred Acceptance CriteriaReported Device Performance (QuickLock System)Predicate Device Performance
Improved Efficiency / Reduced Procedure TimeAveraged 8.1 minutes (range 5.2 - 14.0 minutes) for distal targeting over 22 procedures (13 femoral, 9 tibial). Specifically, the average time using the final probe configuration was 6.6 minutes.Averaged 16.1 minutes (range 10.4 - 30.0 minutes) for distal targeting over 17 procedures (10 femoral, 7 tibial) using Richards or Synthes radiolucent drills.
Reduced Skill and Experience Required for UsersUsers have continuous visual control, making the task easier. Less training, skill, and experience are required. Surgeons learn the technique after a few minutes of training and proficiency is retained without frequent performance. (Qualitative improvement)Requires considerable skill and experience due to reliance on short x-ray exposures and intermittent visual control. Surgeons who infrequently perform the procedure become markedly less effective. (Qualitative assessment)
Improved Accuracy of Visual Cues and Drilling AccuracyThe probe and guide are mechanically aligned with target holes during calibration, electronically offsetting errors. Drilling accuracy is enhanced by high visual sensitivity of dot and cross-hair sight adjustments and continuous display of visual sighting information. Ensures correct hole size and shape for maximum screw holding strength and endurance. Reduces the occurrence of drill strikes. (Qualitative improvement with direct impact on outcome)Adjustment of C-arm position relies on visual perception of distal holes as true circles, limited by subjective judgment. Operator skill strongly affects drilling accuracy between visual snapshots. Holes started out of alignment must be elongated, reducing holding strength and endurance. Drills often strike edges, causing nicks that can lead to premature nail breakage. (Qualitative assessment with direct impact on outcome)
Enhanced Operator ComfortNo need to stay out of an x-ray beam, allowing surgeons to stand, hold the guide, and drill in a comfortable position. (Qualitative improvement)Surgeons work in awkward positions, reaching from the side to avoid the x-ray beam, due to the need to minimize x-ray exposure. (Qualitative assessment)
Minimization of Radiation Exposure to Surgeons and PatientsThe low-energy, 25 KHz sinusoidal, nonionizing electromagnetic field generated by the probe coils is a nonsignificant radiation hazard and produces negligible temperature rise of the probe surface. (Qualitative assessment based on expert opinion)Ionizing electromagnetic x-rays are generated by the C-arm image intensifier, posing a prominent risk. (Qualitative assessment)
Compliance with Safety Standards (Electrical, Biocompatibility)The display, probe, and guide are to be certified to UL Standard 544, 2d Edition. Risk of shock is eliminated by low voltage levels, isolation, insulation, and grounding. Components are made of surgical stainless steel, nickel-titanium alloy (Nitnol SE 508), USP Class VI biocompatible polysulfone, and USP Class III biocompatible polyphenylsulfone. (Certification planned, material biocompatibility stated)Predicate devices are described as using surgical stainless steel, radiolucent plastic, etc. The document implies predicate devices also meet general safety standards, but no specific comparative detail is provided; rather, the QuickLock system is presented as meeting safety norms appropriate to its design.

Study Details

The provided text describes a clinical trial that primarily focused on time required for distal targeting and provided qualitative assessments for other performance aspects.

  1. Sample size used for the test set and the data provenance:

    • Sample Size:
      • Predicate devices: 17 procedures (10 femoral, 7 tibial)
      • QuickLock system: 22 procedures (13 femoral, 9 tibial)
    • Data Provenance: Clinical trials performed at the University of Arizona Health Sciences Center, Tucson, Arizona, USA, from September 1995 to the present (at the time of the document). These trials were conducted under IRB classification as a nonsignificant risk device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number of experts used to establish a quantitative ground truth for the performance metrics (e.g., accuracy of drilling in mm).
    • The primary investigator for the clinical trials was John T. Ruth, M.D., Assistant Professor, Section of Orthopedic Surgery at the University of Arizona Health Sciences Center. He, as the operating surgeon and researcher, would be responsible for the data collection and judgment of "ground truth" performance during the actual procedures, particularly for the time measurement and qualitative observations.
    • For the radiation hazard assessment, the opinion was from Howard Bassen, Head, Electro-Physics Branch, FDA Office of Science and Technology.
  3. Adjudication method for the test set:

    • The document does not specify a formal adjudication method (e.g., 2+1 consensus) for establishing ground truth regarding the procedure time or qualitative aspects. The data appears to have been collected directly from the clinical procedures and reported by the investigators. For time measurements, the "ground truth" is a direct measurement. For qualitative aspects like skill required or operator comfort, it's based on observations and experience of the operating surgeon (Dr. Ruth) and perhaps the surgical team.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a MRMC comparative effectiveness study was not done in the context of "human readers" or "AI assistance." This device is a surgical targeting system, not an imaging diagnostic AI system. The comparison is between two surgical methods/devices – one "freehand" (predicate) and one "assisted" (QuickLock system) – in terms of surgical outcomes like time, accuracy, and comfort.
    • The "effect size" for time reduction is substantial:
      • QuickLock (final configuration): 6.6 minutes
      • Predicate devices: 16.1 minutes
      • This represents a reduction of approximately 59% in targeting time ( (16.1 - 6.6) / 16.1 * 100 ).
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this is not applicable. The QuickLock system is a human-in-the-loop surgical guidance device. It assists the surgeon; it does not operate autonomously.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Time Data: Direct quantitative measurement during live surgical procedures.
    • Accuracy/Quality/Nicking: Surgeon's direct observation and assessment of the quality of drilled holes, alignment, and occurrence of nicks during the procedure. This is a form of expert observation/assessment derived from surgical outcomes.
    • Skill/Comfort/Radiation: Expert qualitative assessment based on the surgeon's experience and comparison between the two methods.
    • Electrical Safety & Biocompatibility: Compliance with established safety standards (UL Standard 544) and use of biocompatible materials (based on material specifications and external certifications/approvals like FDA for stents).
  7. The sample size for the training set:

    • The document does not explicitly describe a "training set" in the context of machine learning. The QuickLock system is a mechanical/electronic guidance system, not an AI/ML algorithm that requires a separate training set for its development. The "training" for the device itself would have been its engineering design and manufacturing. The clinical trials served as validation/testing.
  8. How the ground truth for the training set was established:

    • Since there's no "training set" in the AI/ML sense, this question is not applicable.

{0}------------------------------------------------

K964444

Page B-1

Appendix B Summary of Safety and Effectiveness Information

AUG 1 1 1997

Safety

Predicate Devices

QuickLock System

Radiation Exposure

Ionizing electromagnetic x-rays are generated by the C-arm image intensifier. Concern for minimizing the risk of x-ray exposure to surgeons and patients is a prominant factor in the freehand technique currently employed by the predicate devices.

The low-energy, 25 KHz sinusoidal, nonionizing electromagnetic field generated by the probe coils is an nonsignificant radiation hazard* and produces negligible temperature rise of the probe surface.

  • Opinion of Howard Bassen, Head, Electro-Physics Branch, FDA Office of Science and Technology, telephone conversation July 29, 1994, in which Mr. Bassen confirmed that FDA does not impose standards for electromagnetic radiation unless the device causes heating of body tissue.

Electric Shock

The C-arm image intensifier is an electronic device. It can contact operating room personnel, surgical apparatus in contact with the patient, or the patient directly, because during the distal targeting procedure it is close to the surgical field and is moved about by a technician to align the x-ray beam with the axis of the target hole. Operating room personnel and patients must, therefore, be protected from electric shock by Carm compliance with safety standards for electronic medical devices.

The display, probe and guide come into contact with operating room personnel and patients, and are to be certified to UL Standard 544, 2d Edition. The display is in a sturdy, grounded, metal enclosure with 115 VRMS at 60 Hz input power, + /-12 VDC and +5 VDC internal power. Line power is limited by a 250V, 3AG UL 198.6, 2A, FAST type fuse. The AC to DC power supply and electroluminescent display are medical-grade components. Separation and insulation of internal display components prevent shock and fire hazardous conditions. Risk of shock by contact with the probe and guide is eliminated by low voltage levels, isolation, insulation and grounding. The probe coil voltage is below 12 VRMS with polyimide tube insulation breakddown above 1200 V. Probe coils are isolated from the display unit by transformer coupling, and the metal outer surface is grounded through the probe cable shield. Guide components reach voltages far less than 12 VRMS. The guide body is nonconductive, insulated by encapsulation of internal components, and is grounded through a wire in the guide cable.

{1}------------------------------------------------

Predicate Devices

QuickLock System

Biocompatibilitu

Stainless steel guide wires or rods are routinely placed inside the medullary canal in reduction of fractures, reaming of the medullary canal, and guiding insertion of the nail. They are made of surgical stainless steel, are reusable, and steam or Eto sterilizable.

Surgical stainless steel bone drills are supplied with nail manufacturer's instrumentation. They are reusable and steam or Eto sterilizable.

Drill bushings used in proximal targeting are surgical stainless steel. They are reusable and steam or Eto sterilizable.

Drill or trocar holders of radioluscent plastic are used in the Howmedica and Zimmer devices. Radioluscent plastic components are used to hold the bone drill in the Richards and Synthes powered distal tageting devices. These are reusable and steam or Eto sterilizable.

Screw drivers are surgical stainless steel, reusable and steam or Eto sterilizable.

Probes are placed in nails after surgical insertion into the patient. They are sealed tubes of nickeltitanium alloy (Nitnol SE 508, Nitnol Devices and Componens, Inc.), a material used in the manufacture of vascular stents that are FDA approved for unrestricted implantation. The probe is sealed by laser welding a Nitnol disk into the end inserted into the nail, so there is no patient tissue or fluid contact with probe materials other than Nitnol. Probes are reusable and steam or Eto sterilizable. Probe cables have a silicone jacket and metal connectors that will withstand the same sterilization.

Drills are nonmagnetic, surgical stainless steel (SCF-23, Carpenter Steel Co.). They are reusable and steam or Eto sterilizable.

Drill bushings are molded USP Class VI biocompatible polysulfone (UDEL P-1700 MG 11, Amoco Performance Products, Inc.). The bushings are reusable and steam or Eto sterilizable.

Guides are molded USP Class III biocompatible polyphenylsulfone (Radel R-5000, Amoco Performance Products, Inc.). Guides are reusable and steam or Eto sterilizable. Guide cables have a silicone jacket and metal connectors that will withstand the same sterilization.

Screw drivers are type 316 and 400 series stainless steel. They are reusable and steam or Eto sterilizable.

Safety Standards Certification

Certification of compliance with UL Standard 544, 2d Edition will be obtained from an FDA approved testing laboratory before the product is commercially distributed.

{2}------------------------------------------------

Effectiveness

Predicate Devices

QuickLock System

Intended Use

Locate incisions, guide bone drilling, and guide insertion of interlocking screws in the distal targeting procedure of intramedullary nailing.

Same as for predicate devices

Skill and Experience Required

Users must strive to hold the drill in alignment after short x-ray exposures that provide only intermittent visual control. This requires considerable skill and experience. Surgeons who infrequently perform the procedure fall out of practice and are markedly less effective.

Users have continuous visual control so the task is easier and less training, skill and experience are required. Surgeons learn the technique after a few minutes of training. Proficiency is retained without frequent performance of the procedure.

Accuracy of Visual Cues

Adjustment of C-arm position to align the x-ray beam with target holes depends on visual perception of distal holes as true circles on the image intensifier screen. Subjective judgement limits reliable accuracy of this alignment. Operator skill strongly affects accuracy of drilling between visual snapshots.

The probe and guide are aligned mechanically with target holes during the calibration process, which electronically offsets errors for each procedure. Drilling accuracy is enhanced by the high visual sensitivity of dot and cross-hair sight adjustments and the continuous display of visual sighting information.

Operator Comfort

The need to minimize exposure to x-ray radiation hampers the freehand technique used with the predicate devices. Surgeons work in awkward positions, reaching from the side to avoid the xray beam.

Because there is no need to stay out of an x-ray beam, surgeons can stand, hold the guide, and drill in a comfortable position.

Quality of Drilled Holes

Holes started out of alignment must be elongated to pass through the nail and both cortices. This reduces the holding strength and endurance of bone screws and compromises the objective of nail interlocking.

Accurate alignment throughout the drilling process ensures correct hole size and shape, for maximum screw holding strength and endurance.

Nicking of Intramedullary Nails

Drills often strike the edges of target holes, leaving nicks that can cause stress concentration in the nail material and lead to premature nail breakage due to fatigue fracture.

Improved accuracy reduces the occurance of drill strikes and the resultant risk of induced nail fatigue failure.

{3}------------------------------------------------

Predicate Devices

QuickLock System

Time Required for Distal Targeting*

In 17 procedures (10 femoral and 7 tibial) performed using Richards or Synthes radiolucent drills, the time required for distal targeting averaged 16.1 minutes with a range of 10.4 to 30.0 minutes.

In 22 procedures (13 femoral and 9 tibial) performed using the QuickLock system, the time required for distal targeting averaged 8.1 minutes with a range of 5.2 to 14.0 minutes. These data are scewed by changes in probe configuration during the trials; the average time over procedures performed with the final probe configuration was 6.6 minutes.

  • Results of clinical trials performed at the University of Arizon Health Sciences Center, Tucson, Arizona, from September 1995 to present. These trials are conducted under IRB classification as a nonsignificant risk device, with approval by FDA, Center for Devices and Radiological Health IDE section's Mr. Ted Stevens, team leader of orthopedic devices, non-implantable. The principal investigator was John T. Ruth, M.D., Assistant Professor, Section of Orthopedic Surgery.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Rov B. Hollstien, Ph.D. RDH Medical 5775 Webster Road Paso Robles, California 93446

AUG | | | |997

K964444 Re : QuickLock Orthopaedic Targeting System Trade Name: Requlatory Class: II Product Code: HSB Dated: June 27, 1997 Received: June 30, 1997

:

Dear Dr. Hollstien:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

J Collie

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page of |

K964444 510(k) Number (if known):

QuickLock Orthopaedic Targeting System Device Name:

Indications For Use:

ﻤﺴﻨﺴﻨﺪ

The QuickLock Orthopaedic Targeting System is indicated for use whenever distal inter-The QuickBook Orthepaans Cargo - 8 - 8 - 8 - 8 - 8 - 8 - 8 - 8 - 1 - 1 - 2 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 locking sorono are playilable accessory adapter components that enable mechanical attachment of the the QuickLock Probe to the nail system proximal instrumentation and alignment of the QuickLock Guide during calibration.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) NumberK964444
Prescription Use (Per 21 CFR 801.109)X
------------------------------------------

OR

Over-The-Counter Use___
(Optional Format 1-2-96)

N/A