K Number
K964321

Validate with FDA (Live)

Date Cleared
1997-01-09

(72 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Description

Glove, Patient Examination, Vinyl (Powdered)

AI/ML Overview

The provided text describes the specifications and quality characteristics of Vinyl Patient Examination Gloves. It focuses on physical dimensions, material properties, and quality assurance for the gloves, referencing ASTM and JIS standards.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text, structured according to your request:

1. A table of acceptance criteria and the reported device performance

CriteriaAcceptance Criterion (Specification)Reported Device Performance (Example for Small Size)Test Code / Standard
Physical Dimensions
Circumference of PalmSmall: 178 mm (6-3/4 in)178 mm (6-3/4 in)JIS-S-2045.5.9
Total LengthSmall: 240 mm (± 5 mm)240 mmJIS-S-2045.5.9
Finger Lengths(e.g., Thumb: 55 mm, Index: 66 mm for Small)(e.g., Thumb: 55 mm, Index: 66 mm for Small)JIS-S-2045.5.9
Finger Circumference(e.g., Thumb: 63 mm, Index: 56 mm for Small)(e.g., Thumb: 63 mm, Index: 56 mm for Small)JIS-S-2045.5.9
Material Properties
Tensile StrengthMin. 10.0 MpaMin. 10.0 MpaJIS-2045.5.2
ElongationMin. 350%350%JIS-2045.5.2
Weight(e.g., Small: 7.0 g/pc ± 0.2%)7.0 g/pc±0.2%
ThicknessFinger Tip: 0.08mm ± 0.02mm0.08mm ± 0.02mm(Implicit in spec)
Cuff: 0.09mm ± 0.02mm0.09mm ± 0.02mm(Implicit in spec)
Palm: 0.15mm ± 0.02mm0.15mm ± 0.02mm(Implicit in spec)
Quality Assurance
Pinhole RateUnder 4% Pinhole RateUnder 4% Pinhole RateFDA GLOVE 1000 ML WATER LEAK TEST
Overall ComplianceMeets all requirements of ASTM Standard D5250Stated as metASTM Standard D5250

Note: The table shows the "Reported Device Performance" as matching the "Acceptance Criterion" based on the conclusion statement: "Shinjiazhuang Hongray Plastic Products Co., Ltd. Vinyl Patient Examination Gloves meet ASTM Standards - meets pinhole requirements and labeling claims." This implies that the device did meet all specified criteria. The document provides the specifications for each size and characteristic, and the conclusion then confirms compliance.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample size used for the test set for any of the listed quality characteristics or the pinhole rate test. It also does not explicitly state the data provenance (e.g., country of origin, retrospective/prospective). However, the company is based in China and Taiwan, and the testing laboratories (Consumer Products Testing Co., USA for biocompatibility) are mentioned, which gives an indication of where some testing might have occurred. The JIS standards are Japanese Industrial Standards, implying international applicability or adoption.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable in the context of this device and study. The "ground truth" here is defined by objective, measurable physical and material properties (e.g., dimension measurements, tensile strength values, water leak test results for pinholes) according to established international standards (ASTM, JIS). Expert consensus is not typically used to establish these types of physical standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where human interpretation or subjective assessment is involved, often for clinical endpoints or image analysis. For objective physical measurements and material testing, the results are derived directly from the test procedures outlined in the standards, not through adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The device is a patient examination glove, not an AI-powered diagnostic tool. Therefore, MRMC comparative effectiveness studies, human readers, or AI assistance are irrelevant to its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is not an algorithm. Its performance is measured directly through physical and material tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context refers to the objective measurements and pass/fail criteria defined by recognized international standards such as ASTM Standard D5250 and various JIS-S-2045 sub-standards. For example, for tensile strength, the ground truth is "Min. 10.0 Mpa," and for pinholes, it's "Under 4% pinhole rate" as determined by the FDA glove water leak test.

8. The sample size for the training set

This question is not applicable. There is no concept of a "training set" for physical products like gloves undergoing these types of standardized tests. Machine learning models use training sets; physical products are tested against a defined set of specifications.

9. How the ground truth for the training set was established

This question is not applicable, as there is no training set for this type of product evaluation. The "ground truth" (specifications) for the testing was established by the relevant national and international standards bodies (ASTM, JIS, FDA) that created and maintain these standards.

{0}------------------------------------------------

K964321

Image /page/0/Figure/1 description: This image contains the letterhead for ShiJiaZhuang HongRay Plastic Products Co., Ltd. The head office is located at NO. 135 XIN HUA WEST ROAD in SHIJIAZHUANG CITY, Hebei Province, China, with the postal code 050081. The branch office is located at RM3, 5TH FL., NO. 76 FUSHING SOUTH ROAD, SECTION 1 in Taipei, Taiwan, with the postal code 53-186.

1.0 APPLICANT: MR. LIN CHIN YU SHIJIAZHUANG HONGRAY PLASTIC PRODUCTS CO., LTD. NO. 135 XIN-HUA WEST RD. SHIJIAZHUANG, CHINA TEL. 886 2 772-0306-7 FAX. 886 2 771-2601

  • 2.0 CONTACT PERSON MR. LIN CHIN YU SHIJIAZHUANG HONGRAY PLASTIC PRODUCTS NO. 135 XIN-HUA WEST RD. SHIJIAZHUANG, CHINA TEL. 886 2 772-0306-7 FAX. 886 2 771-2601
    AGENT : MR. DON MORRIS 14 LEEWARD ISLAND CLEARWATER, FL. 34630 USA TEL. 800-366-9545 FAX. 813-443-3482

  • 3.0 Device Class: 1 Product Code: 80LYZ

  • 4.0 Specification: Glove, Patient Examination, Vinyl (Powdered) meets all of the requirements of ASTM Standard D5250

  • 5.0 Device Description: Glove, Patient Examination, Vinyl (Powdered)

  • 6.0 Intended Use: A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

  • 7.0 Inner Surface: Prepowdered

  • 8.0 Biocompatibility Testing: Primary Dermal Irritation in Rabbits and Guinea Pig Sensitization (Buehler) by Consumer Products Testing Co., USA.

{1}------------------------------------------------

Image /page/1/Figure/0 description: This image shows the letterhead for ShiJiaZhuang HongRay Plastic Products Co., Ltd. The letterhead includes the company's head office address, factory address, phone number, branch office address, and fax number. The head office is located at NO, 135 XINHUA WEST ROAD, SHIJIAZHUANG CITY, and the branch office is located at RM3, 5TH FL., NO. 76 FUSHING SOUTH ROAD, SECTION 1. TAIPEI, TAIWAN.

510(K) SUMMARY

Page 2 of 2

9.0 Quality Characteristics

SIZESPECSMAI.I.MEDIUM1.ARGEX-LARGETEST CODE
CIRCUMFERENCEOF PALH (H/W)(INCH)1786-3/42107-3/42188-1/22309JIS-S-2045.5.9
TOTAL LEENGTH(± 5M/H)240245245245JIS-S-2045.5.9
LNGTH OFFINGERS (H/H)THUMBSINDEX FINGERMIDDLE FINGERRING FINGERLITTLE FINGER5566746954577081745561758879606378928262JIS-2045.5.9
CIRCUMFERENCEFINGERSTHUMBINDEX FINGERMIDDLE FINGERRING FINGERLITTLE FINGER6356595650716466635778707269628374767366JIS-S-2045.5.9
TENSILE STRENGTH(Mpa)Min.10.0Min.10.00Min.10.00Min.10.00JIS-2045.5.2
ELONGATION(Min.)350%350%350%350%JIS-2045.5.2
WEIGHT (g/pc)7.08.09.09.5±0.2%
THICKNESSFinger Tip 0.08mm ± 0.02mmCuff 0.09mm ± 0.02mmPalm 0.15mm ± 0.02mm
QUALITYASSURANCEUNDER 4% PINHOLE RATEFDA GLOVE 1000ML WATER LEAKTEST

SPECIFICATION OF VINYL GLOVES (AMBIDEXTROUS) POUDER

  • 10.0 Conclusion: Shinjiazhuang Hongray Plastic Products Co., Ltd. Vinyl Patient Examination Gloves meet ASTM Standards - meets pinhole requirements and labeling claims.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.