KSEA UNIMAT PLUS
Device Facts
| Record ID | K963960 |
|---|---|
| Device Name | KSEA UNIMAT PLUS |
| Applicant | KARL STORZ Endoscopy-America, Inc. |
| Product Code | HET · Obstetrics/Gynecology |
| Decision Date | Dec 20, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1720 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Karl Storz Unimat Plus system is designed to provide irrigation and aspiration of the surgical site during general and endoscopic plastic, reconstructive and aesthetic surgical procedures.
Device Story
Membrane-type irrigation/aspiration pump; provides fixed flow rate of 3.5 L/min and maximum irrigation pressure of 400 mm Hg. Used in surgical settings for site clearance during plastic, reconstructive, and aesthetic procedures. Operated by surgical staff to manage fluid dynamics at the operative site.
Clinical Evidence
Bench testing only.
Technological Characteristics
Membrane-type pump; fixed flow rate 3.5 L/min; max irrigation pressure 400 mm Hg.
Indications for Use
Indicated for patients undergoing general, endoscopic plastic, reconstructive, and aesthetic surgical procedures requiring irrigation and aspiration of the surgical site.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
Related Devices
- K962898 — KSEA UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Aug 27, 1996
- K962090 — KARL STORZ MODEL 203210 O1 UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Sep 30, 1996
- K963949 — KSEA UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Dec 19, 1996
- K040239 — SUPRAFLOW CONSOLE · Neomedix Corp. · Feb 17, 2004
- K013417 — MISONIX IRRIGATION SYSTEM MOEDL BC20P · Misonix, Inc. · Jan 10, 2002