KSEA UNIMAT PLUS
Device Facts
| Record ID | K963949 |
|---|---|
| Device Name | KSEA UNIMAT PLUS |
| Applicant | KARL STORZ Endoscopy-America, Inc. |
| Product Code | HET · Obstetrics/Gynecology |
| Decision Date | Dec 19, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1720 |
| Device Class | Class 2 |
Intended Use
The Karl Storz Unimat Plus system is designed to provide irrigation and aspiration of the surgical site during gynecologic laparoscopic surgical procedures.
Device Story
Membrane-type suction/irrigation pump; provides fixed irrigation/aspiration flow rate of 3.5 L/min; maximum irrigation pressure 400 mm Hg. Used in gynecologic laparoscopic surgery to clear surgical site. Operated by surgical staff in OR environment. Device facilitates visualization and site maintenance during minimally invasive procedures.
Clinical Evidence
No clinical data.
Technological Characteristics
Membrane-type pump; fixed flow rate 3.5 L/min; max pressure 400 mm Hg.
Indications for Use
Indicated for irrigation and aspiration of the surgical site during gynecologic laparoscopic surgical procedures.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
Related Devices
- K962090 — KARL STORZ MODEL 203210 O1 UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Sep 30, 1996
- K963960 — KSEA UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Dec 20, 1996
- K962898 — KSEA UNIMAT PLUS · KARL STORZ Endoscopy-America, Inc. · Aug 27, 1996
- K063326 — CONMED LINVATEC 24K IRRIGATION SYSTEM · Conmed Linvatec · Jan 31, 2007
- K990732 — LAP-WAVE 3000 (P07) · W.O.M. World of Medicine GmbH · Jul 23, 1999