K Number
K963803

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-05-23

(242 days)

Product Code
Regulation Number
866.6010
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Centocor CA 15-3 RIA is an in vitro test for the quantitative determination of DF3 antibodydefined antigen, encoded by the MUC 1 gene, in serum and plasma, Serial test results obtained with the Centocor CA 15-3 RIA, are useful in the early detection of recurrence in patients previously treated for stage III breast cancer, who are clinically free of disease. Early detection of recurrence cannot be readily accomplished by routine clinical or diagnostic studies alone. The use of a circulating serum tumor marker assay, such as Centocor CA 15-3 RIA, can be useful in the identification of these patients. The Centocor CA 15-3 RIA should be used in conjunction with all relevant information derived from diagnostic tests, physical examination and full medical history, in accordance with appropriate patient management procedures

Device Description

The Centocor CA 15-3 RIA is an in vitro diagnostic device based on the well established RIA "forward sandwich" principle. The device is in kit format and consists of murine monoclonal antibody 115D8-coated polystyrene beads, standards, diluent, positive control, and 122 radio-labeled, murine monoclonal antibody, DF3. Antibody coated beads are incubated with a specimen or the appropriate standards and controls. During this initial incubation step, DF3 antibody-defined antigen present in the specimen are bound by the coated beads. Unbound material present in the specimen are removed by aspiration and washing. In the second step, the beads are incubated with DF3 radio-labeled with "2". The tracer detects DF3 antibody-defined antigen bound to the beads during the initial incubation step. The unbound tracer antibody is removed by washing and aspiration. The bound radioactivity is determined by counting the beads in a gamma counter. The bound radioactivity is directly proportional to the concentration of the DF3 antibody-defined antigen in the specimen. A standard curve is obtained by plotting the DF3 antibody-defined antigen concentration of the standards against the bound radioactivity.

AI/ML Overview

Here's an analysis of the provided text to extract the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for substantial equivalence are established by comparing the Centocor CA 15-3 RIA to a predicate device, the Biomira TRUQUANT BR RIA. The key metric for this comparison is linear agreement and precision.

Acceptance Criteria (Implicit from Predicate Device's Claims)Reported Device Performance (Centocor CA 15-3 RIA)
Good linear agreement with predicate device (Biomira TRUQUANT BR RIA)Slope: 0.93Intercept: -2.04Spearman's Rank Correlation: 0.92
Precision comparable to predicate deviceIntra-assay CVs: 5.4% to 9.4%Inter-assay CVs: 7.0% to 11.5%CVs around 35 U/mL: 5.8% to 8.5%

Note: The document explicitly states "The Centocor CA 15-3 RIA and the TRUQUANT BR RIA demonstrated good linear agreement and precision comparable to that claimed in the TRUQUANT BR RIA product instructions." The exact numerical acceptance thresholds for "good linear agreement" and "comparable precision" are not explicitly stated, but the reported performance values met this unspecified standard.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Linear Agreement Study: 204 serum samples

  • Data Provenance: Not explicitly stated regarding the country of origin. The study appears to be a retrospective analysis of collected serum samples from breast cancer patients.

  • Sample Size for Inter and Intra-Assay Precision Study: 24 coded samples (2 sets of 12 samples) and 1 kit control. Tested repeatedly.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts

This information is not provided in the document. The study compares two diagnostic assays against each other rather than validating against a clinical ground truth established by experts.

4. Adjudication Method for the Test Set

This information is not provided. The study is a direct comparison between two assays, not a study involving expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Reader Improvement with AI vs. Without AI Assistance

This information is not applicable/provided. The device is an in-vitro diagnostic assay, not an AI-assisted diagnostic tool that would involve human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The performance data presented is for the Centocor CA 15-3 RIA device itself, which is an in-vitro diagnostic test. Therefore, the "standalone" performance is what is described. It measures the assay's analytical performance in terms of linearity and precision.

7. The Type of Ground Truth Used

The "ground truth" for the comparative study was the results obtained from the predicate device (Biomira TRUQUANT BR RIA). The Centocor CA 15-3 RIA's performance was evaluated for its agreement with this established method. For the precision studies, the ground truth was the expected value of the control samples.

8. The Sample Size for the Training Set

This information is not provided. The document describes a comparison study and assay validation, not the development or training of an algorithm in the machine learning sense. The assay works based on a biochemical principle, not a trained model.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided. As noted above, the device is an in-vitro diagnostic assay based on a biochemical principle, not a machine learning model that would require a "training set" with established ground truth in that context.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a handwritten string of characters, "K963803". The characters are written in a dark color, possibly black or dark gray, against a white background. The handwriting appears somewhat rough or stylized, with varying line thickness and slight imperfections, giving it a distinct, non-uniform appearance.

Image /page/0/Picture/2 description: The image shows the word "Centocor" in a bold, sans-serif font. To the left of the word is a stylized, circular logo. The logo is black with a white crescent shape cut out of the right side. The text and logo are all in black against a white background.

MAY 2 3 1997

510(K) SUMMARY CENTOCOR® CA 15-3TM RIA 10.0

Submitter Information

  • Centocor Incorporated Address: 200 Great Valley Parkway Malvern, PA 19355
    For information regarding this 510(k) Summary please contact, Centocor, Inc., Mr. Christopher D. Zalesky, (610) 651-6815

Summary Preparation Date: September 20, 1996

Name of the Device:

Trade/Proprietary Name:Centocor CA 15-3TM RIA
Common/Usual Name:Radioimmunoassay for the quantitative determination ofDF3-defined antigen in serum or plasma
Classification Name:In accordance with FDA's reclassification order issuedSeptember 19, 1996, Docket # 95P-0136, Serum TumorMarkers, Centocor CA 15-3 RIA has been classified as aClass II device, Tumor Associated Antigen ImmunoassaySystems.

{1}------------------------------------------------

Substantial Equivalence

The Centocor CA 15-3 RIA has been found to be substantially equivalent to Biomira Diagnostic's TRUQUANT BR RIA, P950011, approved by the Food and Drug Administration on March 29, 1996.

Device Description:

The Centocor CA 15-3 RIA is an in vitro diagnostic device based on the well established RIA "forward sandwich" principle. The device is in kit format and consists of murine monoclonal antibody 115D8-coated polystyrene beads, standards, diluent, positive control, and 122 radio-labeled, murine monoclonal antibody, DF3. Antibody coated beads are incubated with a specimen or the appropriate standards and controls. During this initial incubation step, DF3 antibody-defined antigen present in the specimen are bound by the coated beads. Unbound material present in the specimen are removed by aspiration and washing. In the second step, the beads are incubated with DF3 radio-labeled with "2". The tracer detects DF3 antibody-defined antigen bound to the beads during the initial incubation step. The unbound tracer antibody is removed by washing and aspiration. The bound radioactivity is determined by counting the beads in a gamma counter. The bound radioactivity is directly proportional to the concentration of the DF3 antibody-defined antigen in the specimen. A standard curve is obtained by plotting the DF3 antibody-defined antigen concentration of the standards against the bound radioactivity.

Intended Use

The Centocor CA 15-3 RIA is an in vitro test for the quantitative determination of DF3 antibody-defined antigen, encoded by the MUC 1 gene, in serum and plasma of patients previously treated for stage II or stage III breast cancer. Serial test results obtained with the Centocor CA 15-3 RIA, in patients who are clinically free of disease,

Page 54

{2}------------------------------------------------

should be used in conjunction with all relevant information derived from diagnostic tests, physical examination, and full medical history in accordance with appropriate patient management procedures used for early detection of recurrence.

Breast cancer is the most common malignancy among woman in the United States. It is estimated that approximately 184,000 new cases of breast cancer will be diagnosed each year and that approximately 44.000 woman will die of the disease. One of every nine woman in the U.S. will develop breast cancer and approximately 30% of woman who have this malignancy will die of the disease. Metastatic disease may be present at the time of initial diagnosis and can occur at any time following primary therapy. Up to 70% of patients with metastases will respond to systemic treatment with cytotoxic drugs or endocrine therapy; therefore, early detection of recurrence is important to patient management. The median survival following diagnosis of recurrent disease is approximately 2 years, but may range from a few months to decades.

In patients previously treated for stage II or stage III disease, early detection of recurrence cannot be readily accomplished by routine clinical or diagnostic studies alone. A circulating serum tumor marker assay, such as the Centocor CA 15-3 RIA, can be useful in the identification of these patients.

Technological Characteristics:

The Centocor CA 15-3 RIA and TRUQUANT BR RIA both utilize well established solid phase radioimmunoassay procedures for quantifying the amount of antigen in patient specimens. The Centocor CA 15-3 RIA is based on the "forward sandwich" principle. The amount of tracer bound to the assay solid phase is directly proportional to the amount of antigen present in the patient sample. The TRUQUANT BR RIA is based on the "competitive inhibition" principle. With this method, the amount of tracer bound to the solid phase is inversely proportional to the amount of antigen present in the patient sample.

{3}------------------------------------------------

Both the Centocor CA 15-3 RIA and the TRUQUANT BR RIA utilize monoclonal antibodies directed against mucinous antigens associated with carcinoma of the breast. These antigens are large molecular weight glycoproteins with O-linked oligosaccharide chains. Breast carcinoma-associated antigen, encoded by the human MUC 1 gene, is identified by several names including DF3 antibody-defined antigen, MAM 6 and milk mucin.

Performance Data

Analytical Performance in Breast Cancer Patients

Analytical performance of the CA 15-3 assay was evaluated in a study in which serum samples were obtained from 204 breast cancer patients. Each sample was analyzed using the Centocor CA 15-3 RIA and the TRUQUANT BR method. Samples were analyzed on two occasions, with both assays, using multiple product lots. The assays were compared for linear agreement using Bablock-Passing regression analysis, with the other commercial method as the reference method. The results are summarized below:

204 = Total samples tested 0.93 = Slope -2.04 = Intercept 0.92 = Spearman's Rank Correlation

Inter and Intra Assay Precision

Twenty-four coded samples (2 sets of twelve samples) and one kit control were tested in duplicate by six laboratories using three Centocor CA 15-3 RIA kit lots on each of three days. The order of the coded samples was randomized for each day's testing. The estimates of intra-assay precision (CV ) ranged from 5.4 to 9.4%. The estimates of inter-assay lot precision (CV .) ranged from 7.0% to 11.5%. Samples with a mean CA 15-3

{4}------------------------------------------------

assay value around the reference value of 35 U/mL had CV, s of in the range of 5.8% to 8.5%. Nonclinical Studies

Conclusion from Studies

The Centocor CA 15-3 RIA and the TRUQUANT BR RIA demonstrated good linear agreement and precision comparable to that claimed in the TRUQUANT BR RIA product instructions. The two products are substantially equivalent.

==============================================================================================================================================================================

|


|

056

第十分

Page 57

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 23 1997

Mr. Christopher D. Zalesky Director, Regulatory Affairs Centocor Incorporated 200 Great Valley Parkway --Malvern, Pennsylvania 19355

Re: K963803/S2 Trade Name: Centocor CA 15-3TM RIA Regulatory Class: II Product Code: MOI Dated: May 7, 1997 Received: May 8, 1997

Dear Mr. Zalesky:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

{6}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K963803

Device Name: Centocor ®CA 15-3TM RIA

INDICATIONS FOR USE

The Centocor CA 15-3 RIA is an in vitro test for the quantitative determination of DF3 antibodydefined antigen, encoded by the MUC 1 gene, in serum and plasma,

Serial test results obtained with the Centocor CA 15-3 RIA, are useful in the early detection of recurrence in patients previously treated for stage III breast cancer, who are clinically free of disease. Early detection of recurrence cannot be readily accomplished by routine clinical or diagnostic studies alone. The use of a circulating serum tumor marker assay, such as Centocor CA 15-3 RIA, can be useful in the identification of these patients. The Centocor CA 15-3 RIA should be used in conjunction with all relevant information derived from diagnostic tests, physical examination and full medical history, in accordance with appropriate patient management procedures

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Concurrence of CDRH, Office of Device Evaluation (ODE)

Gata E. Macini

(Division Sian-Off) Division of Clinical Laboratory Devices 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

or

Over-The-Counter Use

(Optional Format 1-2-96)

INDICAT.WPD

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.