K Number
K963497

Validate with FDA (Live)

Date Cleared
1996-11-08

(66 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the modified Gammacell with version 1.0 control system remains unchanged. Both the improved and the predicate device effectively irradiate blood components to inactivate leukocytes and lymphocytes.

Device Description

The modified Gammacell E- series Blood Irradiator models, the 1000 Elite and the 3000 Elan use the new software version 1.0 control system. With the exception of the modified control system, the 1000 Elite and 3000 Elan are identical to previously cleared models.

The basic operation of an irradiator involves placing of the blood product into the beaker, loading the beaker into the Gammacell, closing the safety door and pressing the "start" button. Upon start of the cycle, the unit moves the beaker to irradiate position and rotates the blood product continuously at a uniform speed to provide a consistent dose of irradiation. Concurrently, the timer begins to count down (irradiation time is pre-programmed by the site supervisor) from the set time and when the time reaches zero, the sample returns to unload position and the audible alarm sounds. If the cycle is completed as pre-programmed the message "cycle completed OK." is displayed. Any problems in the irradiation cycle will be displayed specifically and clearly.

AI/ML Overview

This document describes a 510(k) summary for a modified Gammacell E-series Blood Irradiator, specifically the 1000 Elite and 3000 Elan models with a new software version 1.0 control system. The primary goal of the submission is to demonstrate substantial equivalence to previously cleared versions (Version 2.09).

Here's an analysis of the provided information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a formal table of acceptance criteria with specific quantitative thresholds. Instead, it describes general functional requirements and claims of equivalence to the predicate device.

Acceptance Criteria (Stated Concept)Reported Device Performance
Functional Requirements
Security FunctionsMet functional requirements
Safety FunctionsMet functional requirements
Operational FeaturesMet functional requirements
Timing FunctionsMet functional requirements
Counting FunctionsMet functional requirements
Ancillary Equipment: Independent Backup TimerMet functional requirements
Ancillary Equipment: Beaker Rotation Sensing SystemMet functional requirements
Operator InterfaceMet functional requirements; simplified and user-friendly with clear messages
Technological Equivalence
Basic Function (blood irradiation)Unchanged from predicate Gammacell 1000E and 3000E
Electrical SafetyComplies with Canadian Standards Association (CSA) standards; CSA certified
Electromagnetic Compatibility (EMC)Tested and demonstrated compliance; withstands electromagnetic interference
Safety (overall)Equivalent or better than the predicate device; irradiation source and radioactivity unchanged
Efficacy (overall)As safe and effective as the predicate device
Intended UseUnchanged

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a numerical sample size for the test set. It mentions "Validation testing results" but does not detail the number of units or test cases involved.

  • Data Provenance: The testing appears to be internal validation by the manufacturer, Nordion International Inc., based in Canada. The document states, "The performance of the device was tested against a set of functional specifications in an environment which simulated, as much as possible, the actual operating environment." This suggests controlled, internal testing rather than external clinical data. It is retrospective in the sense that it's based on internal validation performed by the developer.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The validation testing seems to be based on engineering and functional testing against predetermined specifications, rather than a clinical assessment requiring expert consensus on ground truth. The "predetermined specifications" may have been established by internal experts, but their number and qualifications are not disclosed.

4. Adjudication Method for the Test Set

This information is not provided and is likely not applicable given the nature of the validation described. The validation appears to be against functional requirements and engineering specifications rather than a subjective assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence through functional and safety testing, not a comparative clinical study involving human readers or clinicians.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

The device described is a blood irradiator, not a diagnostic algorithm. Therefore, the concept of an "algorithm only, standalone performance" study as typically applied to AI/ML diagnostic devices is not applicable. The "algorithm" here is the software control system of a physical medical device. The validation described is of this control system's functional performance.

7. The Type of Ground Truth Used

The "ground truth" for this device's validation is the functional specifications and engineering requirements for the blood irradiator's control system. This includes aspects like security functions, safety functions, operational features, timing, counting, and interface behavior, as well as compliance with electrical safety and EMC standards. It is not based on clinical outcomes, pathology, or expert consensus on diagnostic interpretations.

8. The Sample Size for the Training Set

The concept of a "training set" as understood in machine learning is not explicitly applicable here. The software control system was likely developed and verified using standard software engineering practices. No "training set" of data for an AI/ML model is mentioned or implied.

9. How the Ground Truth for the Training Set Was Established

As the concept of a training set for an AI/ML model is not applicable here, the method for establishing its ground truth is not relevant or described. The "ground truth" for engineering and functional validation is the set of design specifications and regulatory standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with a stylized key on the left and the text "A Tradition of Excellence" and "1946-1996" on the right. The key is abstract and appears to be made of thick, black lines. The text is in a serif font and is arranged in two lines, with the phrase "A Tradition of Excellence" above the years "1946-1996". The overall design is simple and classic, suggesting a sense of history and quality.

Image /page/0/Picture/1 description: The image shows the logo for Nordion International Inc. The logo features a stylized infinity symbol with a maple leaf incorporated into the left side. Below the logo is the text "K963497" in a handwritten style.

NOV - 8 1996

510K SUMMARY OF SAFETY AND EFFECTIVENESS

1. Submitter's Name:Nordion International Inc.
Submitter's Address:447 March RoadKanata, OntarioCANADA K2K 1X8
Submitter's Telephone No.:(613) 592-2790
Contact Person:Mr. Grant MalkoskeVice-PresidentQuality, Regulatory Affairs and OperationsServices
Date Summary Prepared:August 27, 1996
2. Common or Usual Name:Blood irradiator.
Proprietary Name:Gammacell 3000 ELAN, Version 1.0Gammacell 1000 ELITE, Version 1.0
3. Legally Marketed Device:Gammacell 3000 Elan, Version 2.09Gammacell 1000 Elite, Version 2.09
File Number of Last 510(k):K952291

447 March Road, Kanata. Ontario. Canada K2K 1X8 Tel.: (613) 592-2790 Fax: (613) 592-6937

{1}------------------------------------------------

4. Product description:

The modified Gammacell E- series Blood Irradiator models, the 1000 Elite and the 3000 Elan use the new software version 1.0 control system. With the exception of the modified control system, the 1000 Elite and 3000 Elan are identical to previously cleared models.

The basic operation of an irradiator involves placing of the blood product into the beaker, loading the beaker into the Gammacell, closing the safety door and pressing the "start" button. Upon start of the cycle, the unit moves the beaker to irradiate position and rotates the blood product continuously at a uniform speed to provide a consistent dose of irradiation. Concurrently, the timer begins to count down (irradiation time is pre-programmed by the site supervisor) from the set time and when the time reaches zero, the sample returns to unload position and the audible alarm sounds. If the cycle is completed as pre-programmed the message "cycle completed OK." is displayed. Any problems in the irradiation cycle will be displayed specifically and clearly.

ട. Intended Use:

The intended use of the modified Gammacell with version 1.0 control system remains unchanged. Both the improved and the predicate device effectively irradiate blood components to inactivate leukocytes and lymphocytes.

6. Technological Comparison:

The basic function of the Gammacell 1000 Elite and Gammacell 3000 Elan with the modified control system has not been changed from the predicate Gammacell 1000E and 3000E. However, some technological changes are being made to the control system primarily to make the unit more user friendly. Precisely, the operator display panel has been simplified, and the information and error messages displayed on the panel have been written so that they can be readily understood by the operator. The access to the machine is effectively controlled through enhanced security and access controls. In addition, the modified unit is proven to withstand any electromagnetic interference that may affect the function of the control system.

Safety and Efficacy: 7.

Safety:

The safety of the Gammacell is equivalent or better than the predicate device. The irradiation source and the radioactivity of the caesium-137 source(s) remains unchanged.

In terms of electrical safety, all units are designed to comply with applicable standards of the Canadian Standards Association and are certified as such by the CSA. In addition, the modified units are tested and demonstrated for compliance with Electromagnetic Compatibility (EMC).

Efficacy:

{2}------------------------------------------------

The performance of the device was tested against a set of functional specifications in an environment which simulated, as much as possible, the actual operating environment. The functional requirements included security and safety functions, operational features, timing and counting functions, ancillary equipment such as independent Backup Timer and Beaker Rotation Sensing System, operator interface etc. Validation testing results provide documented evidence that the equipment design tested, adequately meets the functional requirements and the device is as safe and effective as the predicate device.

8. Conclusion:

Although some new features have been added to the new control system, mainly for the purpose of operator conveniences, these new features in no way compromise the intended use, safety or effectiveness of the device. The modified device models, with Version 1.0, are substantially equivalent to the 510(k) cleared models with Version 2.09

N/A