K Number
K963427

Validate with FDA (Live)

Date Cleared
1996-10-07

(38 days)

Product Code
Regulation Number
866.5880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMMAGE Immunochemistry System Transferrin (TRF) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of transferrin concentrations in human serum samples by rate nephelometry.

Device Description

The IMMAGE Immunochemistry System Transferrin (TRF) Reagent is used in the determination of transferrin concentrations in human serum samples on Beckman's IMMAGE Immunochemistry System. The reagent kit contains one cartridge of reagent, evaporation caps and a lot specific bar code card. Calibrators and control materials are purchased separately.

AI/ML Overview

Here's an analysis of the provided text regarding the IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent, structured according to your request:

Acceptance Criteria and Study Details for IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent

The provided document describes the performance of the IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent by comparing it to a predicate device, the Beckman Immunochemistry Systems TRF Transferrin Reagent (on ARRAY). The acceptance criteria are implicitly defined by the performance of the predicate device and the new device's ability to demonstrate substantial equivalence through various studies.


1. A table of acceptance criteria and the reported device performance

The document doesn't explicitly state "acceptance criteria" but rather presents performance data and claims of substantial equivalence based on comparisons to a predicate device. The performance of the predicate device, for which the new device aims to be equivalent, serves as the de facto acceptance target.

Performance MetricImplicit Acceptance Criteria (based on predicate)Reported Device Performance (IMMAGE TRF Reagent)
Assay MethodRate nephelometryRate nephelometry
Analytic Range75 - 750 mg/dL75 - 750 mg/dL
Extended Range12.5 - 4,500 mg/dL12.5 - 4,500 mg/dL
Sample and Reagent RatiosSame as predicateSame as predicate
AntibodySame as predicateSame as predicate
Shelf Life24 months (stored at 2 - 8°C)24 months
Open Container Stability(Not explicitly stated for predicate, but typical for such reagents)14 day
Calibration Stability(Not explicitly stated for predicate)14 day
Method Comparison (vs. predicate)Slope close to 1, Intercept close to 0, r close to 1Slope: 0.990, Intercept: 0.692, r: 0.994
Within-Run Imprecision (TRF Reagent)(Implied to be comparable or better than predicate, or within acceptable clinical limits)Level 1: 3.1 %CV (Mean 241 mg/dL)Level 2: 3.9 %CV (Mean 361 mg/dL)Level 3: 2.9 %CV (Mean 653 mg/dL)

Note on Acceptance Criteria: The document primarily focuses on demonstrating "substantial equivalence" to the predicate device. Therefore, the "acceptance criteria" are largely met by showing comparable performance in the described studies. Specific numeric acceptance thresholds (e.g., maximum allowable %CV, r-value threshold) are not explicitly stated within this summary document, but are inherent to the regulatory process for demonstrating equivalence.


2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set (Method Comparison): Not explicitly stated. The "Method Comparison Study Results" table only provides the slope, intercept, and 'r' value, without detailing the number of individual samples or patients used to generate these statistics.
  • Data Provenance: Not explicitly stated. The submitting company, Beckman Instruments, Inc., is based in Brea, California, USA, but this does not confirm the origin of the samples. The study type is not specified as retrospective or prospective, but method comparison studies are typically prospective analyses of current samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is an in vitro diagnostic reagent for quantitative biochemical measurement. The "ground truth" for such devices is established by the performance of a reference or predicate method. Experts (like radiologists) are not typically involved in establishing ground truth for quantitative chemical assays.

  • The ground truth for the comparison was the Beckman's TRF Transferrin Reagent on the ARRAY® Systems. The implicit "expert" is the established and validated performance of this predicate device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Adjudication methods like "2+1" or "3+1" are relevant for subjective interpretations (e.g., image reading). This is a quantitative chemical assay; therefore, such adjudication methods are not applicable nor mentioned. The results are numerical values obtained from the instruments.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done. This is an in vitro diagnostic device, not an AI or imaging diagnostic tool that involves human readers/interpreters in the diagnostic process. Therefore, the concept of human reader improvement with or without AI assistance is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This device is a standalone (algorithm only) in the sense that the IMMAGE system performs the assay and generates a quantitative result without human subjective interpretation of the assay itself. The "algorithm" here refers to the instrument's programming for rate nephelometry and calculation, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for the performance evaluation was established by a predicate device: Beckman's TRF Transferrin Reagent on the ARRAY® Systems. This is a common form of "ground truth" or reference in establishing substantial equivalence for in vitro diagnostic devices in regulatory submissions.

8. The sample size for the training set

This document does not refer to a "training set" in the context of typical machine learning or AI development. For in vitro diagnostic reagent development, "development" or "validation" samples are used, but they are not termed "training sets" in the AI sense. The document does not specify the number of samples used during the development or initial validation phases (if separate from the comparison studies).


9. How the ground truth for the training set was established

As there is no mention of a "training set" in the AI context, this question is not applicable. For chemistry reagents, development and validation involve testing against known concentrations and established methods, ensuring linearity, accuracy, and precision across the assay range using reference materials.

{0}------------------------------------------------

OCT 7 1996

Beckman Instruments, Inc., Section 510(k) Notification IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent Summary of Safety & Effectiveness

K963427

Summary of Safety & Effectiveness IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent

1.0 Submitted By:

Margie George Proiect Manager, System Development Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 961-3765 FAX: (714) 961-3759

2.0 Date Submitted:

30 August 1996

3.0 Device Name(s):

3.1 Proprietary Names

IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent

3.2 Classification Names

Transferrin immunological test svstem (21 CFR 866.5880)

4.0 Predicate Device(s):

Beckman Immunochemistry Systems TRF Transferrin Reagent

5.0 Description:

The IMMAGE Immunochemistry System Transferrin (TRF) Reagent is used in the determination of transferrin concentrations in human serum samples on Beckman's IMMAGE Immunochemistry System. The reagent kit contains one cartridge of reagent, evaporation caps and a lot specific Calibrators and control materials are purchased bar code card. separately.

{1}------------------------------------------------

Beckman Instruments, Inc., Section 510(k) Notification IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent Summary of Safety & Effectiveness

6.0 Intended Use:

:

The IMMAGE Immunochemistry System Transferrin (TRF) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of transferrin concentrations in human serum samples by rate nephelometry.

7.0 Comparison to Predicate(s):

The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.

ReagentAspect/CharacteristicComments
SIMILARITIES
IMMAGESystem (TRF)ReagentAssay method - rate nephelometrySame as Beckman'sImmunochemistry SystemTransferrin TRF Reagentas run on the ARRAY®System
Analytic Range 75 - 750 mg/dLExtended Range 12.5 - 4,500 mg/dL
Sample and reagent ratios
Antibody
Shelf life of 24 months(stored at 2 -8°C)
DIFFERENCES
IMMAGESystem (TRF)ReagentReaction TemperatureIMMAGE System TRFassays run at 37°C andthe predicate runs at26.7°C
Reagent containerIMMAGE reagent isstored in a plasticcartridge where as thepredicate is stored inglass

{2}------------------------------------------------

8.0 Summary of Performance Data:

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from Beckman's Immunochemistry System TRF Transferrin (on ARRAY) Reagent to the IMMAGE System TRF Reagent.

Method Comparison Study Results

IMMAGE TRF Reagent vs. Beckman TRF Transferrin Reagent (on ARRAY)

AnalyteSlopeInterceptrPredicate
IMMAGETRF0.9900.6920.994Beckman's TRFTransferrin Reagent onthe ARRAY® Systems

Stability Study Results

ReagentProduct Claim
IMMAGE TRF Reagent System24 month shelf-life14 day open container stability14 day calibration stability

Estimated Within-Run Imprecision

MATERIALMEAN(mg/dL)SD(mg/dL)%CVNumberofResults
TRF Reagent
Level 12417.43.180
Level 236114.03.980
Level 365318.92.980

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

§ 866.5880 Transferrin immunological test system.

(a)
Identification. A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.(b)
Classification. Class II (performance standards).