K Number
K963292

Validate with FDA (Live)

Date Cleared
1996-10-31

(71 days)

Product Code
Regulation Number
862.1145
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The COBAS INTEGRA Analyzer and COBAS INTEGRA Reagent Cassettes together provide an integrated system for in vitro diagnostic testing. The COBAS INTEGRA Reagent Cassettes are comprised of chemistry, drugs of abuse, immunology, therapeutic drug monitoring, and hematology assay systems. The COBAS INTEGRA Analyzer provides quantitative measurement of these analytes via three measuring principles, i.e., absorbance, fluorescence polarization and ion-selective electrodes. The COBAS INTEGRA Reagent Cassettes are compact and preparation-free with the added convenience of long term on-board stability. Sixty-eight COBAS INTEGRA Reagent Cassettes can be stored on board, 24 hours a day at 2-8℃. Each cassette is barcoded. This barcode label provides the analyzer with specific reagent information such as the lot number. the expiration date and the number of tests.

Through this submission, it is the intention of Roche to gain clearance of an additional 4 COBAS Reagent Cassettes and 6 ancillary reagents. These are the COBAS INTEGRA Cassette for Antithrombin III, COBAS INTEGRA Cassette for Ferritin, COBAS INTEGRA Cassette for Myoglobin, COBAS INTEGRA Cassette for Rheumatoid Factors, Roche Plasmachrom Calibrator, Roche Plasmachrom N & P Controls, Roche FERR T Standard, Roche FERR/MYO T Control, Roche MYOT Standard, and Roche RF Standard II. The COBAS INTEGRA Cassette for Antithrombin III (AT III) contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative determination of human antithrombin III activity in plasma. The COBAS INTEGRA Cassette for Ferritin contains an in vitro diagnostic reagent system intended for the quantitative immunological determination of human ferritin in serum. The COBAS INTEGRA Cassette for Myoglobin contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative immunological determination of human myoglobin in serum and plasma. The COBAS INTEGRA Cassette for Rheumatoid Factors contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative immunological determination of human rheumatoid factors in serum. Plasmachrom Calibrator is intended for use as a calibrator in quantitative kinetic antithrombin III activity assays. Plasmachrom Control N is an assayed control intended for use to monitor the accuracy and precision at normal concentration levels in quantitative kinetic colorimetric tests for antithrombin III activity assays. Plasmachrom Control P is an assayed control intended for use to monitor the accuracy and precision at pathological concentration levels in quantitative kinetic colorimetric tests for antithrombin III activity assays. MYOT Standard is intended for use as a calibrator in quantitative determinations of human myoglobin. FERR T Standard is intended for use as a calibrator in quantitative determinations of human ferritin. FERR/MYO T Control is intended for use as a quality control material to monitor accuracy and precision in quantitative determinations of human ferritin and myoglobin. RF T Standard is intended for use as a calibrator in quantitative determinations of human rheumatoid factor.

This submission also contains a modification to the previously cleared COBAS INTEGRA Reagent Cassette for Calcium. The performance of the COBAS INTEGRA Reagent Cassette for Calcium has been improved due to the use of new instrument parameters.

Device Description

The COBAS INTEGRA Analyzer and COBAS INTEGRA Reagent Cassettes together provide an integrated system for in vitro diagnostic testing. The COBAS INTEGRA Reagent Cassettes are comprised of chemistry, drugs of abuse, immunology, therapeutic drug monitoring, and hematology assay systems. The COBAS INTEGRA Analyzer provides quantitative measurement of these analytes via three measuring principles, i.e., absorbance, fluorescence polarization and ion-selective electrodes. The COBAS INTEGRA Reagent Cassettes are compact and preparation-free with the added convenience of long term on-board stability. Sixty-eight COBAS INTEGRA Reagent Cassettes can be stored on board, 24 hours a day at 2-8℃. Each cassette is barcoded. This barcode label provides the analyzer with specific reagent information such as the lot number. the expiration date and the number of tests.

AI/ML Overview

The provided text describes several COBAS INTEGRA Reagent Cassettes and ancillary reagents. Each product has its own performance characteristics detailed in separate tables comparing it to a predicate device. I will break down the acceptance criteria and study information for each of the primary assay cassettes.

General Information Applicable to All Devices:

  • Data Provenance: Not explicitly stated, but clinical and non-clinical studies are mentioned. It is implicitly retrospective as it compares performance to predicate devices and existing methodologies.
  • Number of experts used to establish ground truth & qualifications: Not specified.
  • Adjudication method: Not specified.
  • Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable, as these are diagnostic assays, not image-based AI systems involving human readers.
  • Standalone (algorithm only) performance: This is the core of the evaluation for these devices, as they are automated diagnostic assays.
  • Type of ground truth: For accuracy, the ground truth is established by correlation to a legally marketed predicate device or a clinical method (e.g., enzyme immunoassay, radioimmunoassay).

1. COBAS INTEGRA Cassette for Antithrombin III

Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Predicate: Berichrom AT III)Reported Device Performance (COBAS INTEGRA AT III)
Assay Range0-140%0-150%
Precision (Day-to-day)Normal plasma: 1.0-2.5%4.1% at 51.9%
Pathological plasma: 1.5-3.0%2.8% at 105.7%
Accuracy (R-value)> 0.95 vs. Berichrom ReagentR = 0.973 vs. Berichrom Reagent
Sensitivity (Analytical)Not specified in labeling1.6 X 10⁻³ ΔA/min per % change in AT III activity

Study Details:

  • Sample size for test set: N = 200 (for Accuracy)
  • Sample size for training set: Not specified.
  • How ground truth for training set was established: Not specified.

2. COBAS INTEGRA Cassette for Ferritin

Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Predicate: Behring N Latex Ferritin)Reported Device Performance (COBAS INTEGRA Ferritin)
Assay Range5-320 ug/L0-300 ug/L (0-3000 ug/L with postdilution)
Precision (Within-run)3.8% at 5 ug/L1.5% at 15 ug/L9.9 % at 19 ug/L1.2 % at 260 ug/L
Accuracy (Regression)y = 0.80x + 2.9 ug/L vs.behring N Latex Ferritin; < 10 ug/L (vs. radioimmunoassay)y = 1.01x - 0.37 ug/L vs. enzyme immunoassay
Sensitivity (Analytical)25 ug/L (vs. radioimmunoassay)1.3% at 30 ug/L1.5% at 50 ug/L0.9% at 200 ug/L (These are precision measurements, not analytical sensitivity values as typically described)

Study Details:

  • Sample size for test set: N = 188 (for Accuracy)
  • Sample size for training set: Not specified.
  • How ground truth for training set was established: Not specified.

3. COBAS INTEGRA Cassette for Myoglobin

Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Predicate: Behring N Latex Myoglobin)Reported Device Performance (COBAS INTEGRA Myoglobin)
Assay Range25-400 ug/L0-500 ug/L (0-5000 IU/mL with postdilution)
Precision (Day-to-day)4.8% at 85 ug/L4.2% at 160 ug/L5.0% at 310 ug/L1.4% at 46 ug/L0.6% at 312 ug/L
Accuracy (Regression)y = 0.95X - 4.33 vs. radioimmunoassay (for predicate)N = 117 (for predicate)y = 0.91x + 15 ug/L vs. Behring N Latex Myoglobin
Sensitivity (Analytical)3 ug/L (This value appears to be the predicate's sensitivity, not specific acceptance criterion).Not explicitly stated in a comparable format for the device.

Study Details:

  • Sample size for test set: N = 230 (for Accuracy)
  • Sample size for training set: Not specified.
  • How ground truth for training set was established: Not specified.

4. COBAS INTEGRA Cassette for Rheumatoid Factors

Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Predicate: Behring N Latex RF)Reported Device Performance (COBAS INTEGRA Rheumatoid Factors)
Assay Range10-600 IU/mL10-120 IU/mL (10-600 IU/mL with postdilution)
Precision (Day-to-day)9.3% at 92 IU/mL6.2% at 155 IU/mL6.6% at 306 IU/mL8.0% at 32 IU/mL5.0% at 125 IU/mL
Accuracy (Regression)y = 0.99 x - 20 IU/mL vs. enzyme immunoassay (for predicate)N = 72 (for predicate)y = 0.70x + 12 IU/mL vs. Behring N Latex RF Reagent
Sensitivity (Analytical)10 IU/mL10 IU/mL

Study Details:

  • Sample size for test set: N = 244 (for Accuracy)
  • Sample size for training set: Not specified.
  • How ground truth for training set was established: Not specified.

5. COBAS INTEGRA Cassette for Calcium (Modified)

Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Predicate: Original COBAS INTEGRA Calcium or Boehringer Mannheim Calcium on Hitachi 911)Reported Device Performance (COBAS INTEGRA Calcium - Modified)
Assay Range (Serum/Plasma)Original: 0-20 mg/dLPredicate: 0-15 mg/dL (0-24.9 w/postdilution)0-20 mg/dL (0-140 mg/dL with postdilution)
Assay Range (Urine)Original: 0-100 mg/dLPredicate: 0-15 mg/dL (0-60 mg/dL w/postdilution)0-28 mg/dL (0-140 mg/dL with postdilution)
Precision (Total) Serum/PlasmaOriginal: 2.1 % at 7.9 mg/dL; 1.8 % at 10.2 mg/dLPredicate: 1.6 % at 8.9 mg/dL; 1.2 % at 9.7 mg/dL; 1.3% at 12.5 mg/dL3.5% at 9.2 mg/dL3.1% at 13.6 mg/dL
Precision (Total) UrineOriginal: 2.8% at 3.9 mg/dL; 0.77% at 18 mg/dLPredicate: 2.2% at 2.9 mg/dL; 1.7% at 7.8 mg/dL; 1.4% at 36.7 mg/dL0.91% at 5.6 mg/dL1.1% at 23.7 mg/dL
Accuracy (R-value)Original: R = 0.960 vs. BM Hitachi 911Predicate: R = 0.997 vs. BM Hitachi 717R = 0.987 vs. BM Hitachi 911
Sensitivity (Analytical)Original: 3.2 X 10⁻² AA per mg/dLPredicate: Not specified in labeling4.0 X 10⁻² AA per mg/dL of calcium

Study Details:

  • Sample size for test set: N = 196 (for Accuracy)
  • Sample size for training set: Not specified.
  • How ground truth for training set was established: Not specified.

{0}------------------------------------------------

K963292

Image /page/0/Picture/1 description: The image shows the logo for Roche Diagnostic Systems. The logo consists of the Roche symbol on the left, followed by the text "Roche Diagnostic Systems" in bold, sans-serif font. The Roche symbol is a stylized graphic, and the text is aligned horizontally.

A Member of the Roche Group

OCT 31 1988

Roche Diagnostic Systems, Inc. 1080 U.S. Highway 202 Branchourg, New Jersey 08876-1760

Direct Dial Fax

510(k) Summary

Roche COBAS® INTEGRA Reagent Cassettes

In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.

I. Identification of 510(k) Sponsor:

Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche. Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

510(k) Submission dated August 20, 1996

Contact: Andrea Casper Sr. Manager Regulatory and Clinical Affairs Phone: (908) 253-7332 Fax: (908) 253-7547

II. Device Name:

The device name, including both the trade/proprietary name and the classification name are provided in the table below.

Product NameClassificationNameRegulatoryClassCFRClassificationNumberPredicateProductNameDatePredicateClearedPredicate510(k)Number
Chemistry Panel
COBAS INTEGRACalcium(modification)Calcium testsystemClass II862.1145COBAS INTEGRACalciumBoehringerMannheimCalcium Reagent9/5/951/23/85K951595K850281
Immunology Panel
COBAS INTEGRAAntithrombin IIIAntithrombin IIIassayClass II866.7060BerichromAntithrombin IIIReagent4/19/94K933125

Table 1

{1}------------------------------------------------

COBAS INTEGRAFerritinFerritinimmunologicaltest systemClass II866.5340Behring N LatexFerritin Reagent3/27/95K950707
COBAS INTEGRAMyoglobinMyoglobinimmunologicaltest systemClass II866.5680Behring N LatexMyoglobinReagent6/01/96K902154
COBAS INTEGRARheumatoid FactorsRheumatoidfactorimmunologicaltest systemClass II866.5775Behring N LatexRF Reagent10/20/94K942328
PlasmachromCalibratorCalibratorClass II862.1150Behring N ProteinStandard Plasma4/7/88K883662
PlasmachromControlsQuality controlmaterial (assayedand unassayed)Class I862.1660N/T ProteinControl Py4/17/96K951012
Roche FERR TStandardCalibratorClass II862.1150Behring N FerritinStandard3/27/95K950707
Roche FERR/MYOT ControlQuality controlmaterial (assayedand unassayed)Class I862.1660Behring N FerritinControls3/27/95K950707
Behring NMyoglobinControls6/01/96K902154
Roche MYO TStandardCalibratorClass II862.1150Behring NMyoglobinStandard6/01/96K902154
Roche RF TStandardCalibratorClass II862.1150Behring N RFStandard10/20/94K942328

III. Identification of the legally marketed device to which the 510(k) sponsor claims equivalence:

Table 1, presented above, identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.

IV. Description of the Device/Statement of Intended Use:

The COBAS INTEGRA Analyzer and COBAS INTEGRA Reagent Cassettes together provide an integrated system for in vitro diagnostic testing. The COBAS INTEGRA Reagent Cassettes are comprised of chemistry, drugs of abuse, immunology, therapeutic drug monitoring, and hematology assay systems. The COBAS INTEGRA Analyzer provides quantitative measurement of these analytes via three measuring principles, i.e., absorbance, fluorescence polarization and ion-selective electrodes. The COBAS INTEGRA Reagent Cassettes are compact and preparation-free with the added convenience of long term on-board stability. Sixty-eight COBAS INTEGRA Reagent Cassettes can be stored on board, 24 hours a day at 2-8℃. Each cassette is barcoded. This barcode label provides the analyzer with specific reagent information such as the lot number. the expiration date and the number of tests.

{2}------------------------------------------------

Through this submission, it is the intention of Roche to gain clearance of an additional 4 COBAS Reagent Cassettes and 6 ancillary reagents. These are the COBAS INTEGRA Cassette for Antithrombin III, COBAS INTEGRA Cassette for Ferritin, COBAS INTEGRA Cassette for Myoglobin, COBAS INTEGRA Cassette for Rheumatoid Factors, Roche Plasmachrom Calibrator, Roche Plasmachrom N & P Controls, Roche FERR T Standard, Roche FERR/MYO T Control, Roche MYOT Standard, and Roche RF Standard II. The COBAS INTEGRA Cassette for Antithrombin III (AT III) contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative determination of human antithrombin III activity in plasma. The COBAS INTEGRA Cassette for Ferritin contains an in vitro diagnostic reagent system intended for the quantitative immunological determination of human ferritin in serum. The COBAS INTEGRA Cassette for Myoglobin contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative immunological determination of human myoglobin in serum and plasma. The COBAS INTEGRA Cassette for Rheumatoid Factors contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative immunological determination of human rheumatoid factors in serum. Plasmachrom Calibrator is intended for use as a calibrator in quantitative kinetic antithrombin III activity assays. Plasmachrom Control N is an assayed control intended for use to monitor the accuracy and precision at normal concentration levels in quantitative kinetic colorimetric tests for antithrombin III activity assays. Plasmachrom Control P is an assayed control intended for use to monitor the accuracy and precision at pathological concentration levels in quantitative kinetic colorimetric tests for antithrombin III activity assays. MYOT Standard is intended for use as a calibrator in quantitative determinations of human myoglobin. FERR T Standard is intended for use as a calibrator in quantitative determinations of human ferritin. FERR/MYO T Control is intended for use as a quality control material to monitor accuracy and precision in quantitative determinations of human ferritin and myoglobin. RF T Standard is intended for use as a calibrator in quantitative determinations of human rheumatoid factor.

This submission also contains a modification to the previously cleared COBAS INTEGRA Reagent Cassette for Calcium. The performance of the COBAS INTEGRA Reagent Cassette for Calcium has been improved due to the use of new instrument parameters .

  • V . Summary of the technological characteristics of the device new in comparison to those of the predicate.
    Tables 2-6 attached to this summary outline the technological characteristics (methodologies) of the COBAS INTEGRA Reagents in comparison to those of legally marketed predicate products.

  • VI. Brief discussion of the clinical and nonclinical tests relied on for a determination of substantial equivalence:
    Tables 2-6 attached to this summary demonstrate the results of clinical and nonclincal studies performed using the COBAS INTEGRA Reagent Cassettes. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in this chart. This information concludes that the performance of this device is essentially equivalent to other legally marketed devices of a similar kind.

{3}------------------------------------------------

Table 2

COBAS INTEGRA Cassette for AT III

COBAS INTEGRACassette forAntithrombin IIIBerichrom AT III (A)
MethodologyKinetic colorimetric testKinetic colorimetric test
Sample typePlasma (citrate)Plasma
Reported measuring units%%
CalibratorPlasmachrom Calibrator(lyophilized human plasma)Calibrator value: 102%Behring N Protein StandardPlasma (lyophilized humanplasma)Calibrator Value: 87%
ControlsPlasmachrom N & P Controls(lyophilized human plasma)Control N: 99.4%Control P: 61.1%N/T Protein Control Py(lyophilized human plasma)Control N: 92%Control P: 34%
Reagent (active ingredients)TRISNaClCitrateBSAHeparinThrombinD-phenylalanyl-prolyl-arginyl-4-acetylanilideThrombintos-gly-pro-arg-ANBA-IPATRIS
Performance Characteristics:
Assay range0-150%0-140%
Precision (Day-to-day)4.1% at 51.9%2.8% at 105.7%Normal plasma - 1.0-2.5%Pathological plasma-1.5-3.0%
AccuracyN = 200R = 0.973 vs. BerichromReagentR = > 0.95
Sensitivity (Analytical)1.6 X 10-3 ΔA/min per %change in AT III activityNot specified in labeling

{4}------------------------------------------------

Accuracy

Sensitivity (Analytical)

1.3% at 30 ug/L 1.5% at 50 ug/L 0.9% at 200 ug/L

y = 1.01x - 0.37 ug/L vs. enzyme immunoassay

N = 44

5 ug/L

Table 3

COBAS INTEGRA Cassette for Ferritin

COBAS INTEGRACassette for FerritinBehring N LatexFerritin Reagent onBehring Nephelometer
MethodologyImmunoturbidimetric test forFerritinImmunoturbidimetric test forFerritin
Sample typeSerumSerum
Reported measuring unitsug/Lug/L
CalibratorRoche FERR T Standard(human sourced)0, 12.5, 50, 100, 200, & 300ug/LBehring N Ferritin Standard(human sourced)344 ug/L
ControlsRoche FERR/MYO T Control(lyophilized human serumw/added BSA)Ferritin: 126 ± 13 ug/LMyoglobin: 115 ± 12 ug/LBehring N Ferritin Controls(human sourced)Control 1: 20.2 ug/LControl 2: 153 ug/L
Reagent (active ingredients)R1. Glycine bufferR2. Latex particles coatedwith anti-human ferritin(rabbit) in glycine bufferR1. Freeze dried polystyreneparticles coated with rabbitanti-human ferritinR2. Buffered solution ofrabbit serumR3. buffered solution ofdetergents
Performance Characteristics:
Assay range0-300 ug/L0-3000 ug/L w/postdilution5-320 ug/L
Precision (Within-run)9.9 % at 19 ug/L1.2 % at 260 ug/L3.8% at 5 ug/L1.5% at 15 ug/L

N = 188

Reagent

< 10 ug/L

y = 0.80x + 2.9 ug/L vs.
Behring N Latex Ferritin

{5}------------------------------------------------

radioimmunoassay

25 ug/L

Table 4

COBAS INTEGRA Cassette for Myoglobin

COBAS INTEGRACassette for MyoglobinBehring N LatexMyoglobin Reagent
MethodologyImmunoturbidimetric test forMyoglobinImmunoturbidimetric test forMyoglobin
Sample typeSerum and plasma (heparin,EDTA, & fluoride)Serum and plasma
Reported measuring unitsug/Lug/L
CalibratorRoche MYO T Standard(sourced from human skeletalmuscle in liquid form)0, 62.5, 125, 250, & 500ug/LBehring N MyoglobinStandard (lyophilized humanserum)785 ng/mL
ControlsRoche FERR/MYO T Control(lyophilized human serumw/added BSA)Ferritin: 126 ± 13 ug/LMyoglobin: 115 ± 12 ug/LBehring N MyoglobinControl (lyophilized humanserum)106 ng/mL
Reagent (active ingredients)R1. Glycine buffer w/ BSAand immunoglobulins (rabbit)R2. Latex particles in glycinebuffer coated w/ anti-myoglobin (rabbit)R1. lyophilisate ofpolystyrene particles coatedwith rabbit anti-humanmyoglobinR2. buffered rabbit serumsoln.R3. detergent soln.
Performance Characteristics:
Assay range0-500 ug/L0-5000 IU/mL w/postdilution25-400 ug/L
Precision (Day-to-day)1.4% at 46 ug/L0.6% at 312 ug/L4.8% at 85 ug/L4.2% at 160 ug/L5.0% at 310 ug/L
AccuracyN = 230y = 0.91x + 15 ug/L vs.Behring N Latex MyoglobinN = 117y = 0.95X - 4.33 vs.radioimmunoassay

Reagent

3 ug/L

Sensitivity (Analytical)

{6}------------------------------------------------

Table 5

COBAS INTEGRA Cassette for Rheumatoid Factors

COBAS INTEGRACassette for RheumatoidFactorsBehring N Latex RFReagent on BehringNephelometer
MethodologyImmunoturbidimetric test forRFImmunoturbidimetric test forRF
Sample typeSerumSerum
Reported measuring unitsIU/mLIU/mL
CalibratorRoche RF T Standard II(liquid human serum)10, 20, 40, 80, &160 IU/mLBehring N RF Standard(lyophilized human serum)73 IU/mL
ControlsRoche RF/ASO T Control II(K954992)Behring N RheumatologySL Controls
Reagent (active ingredients)R1. Acceleratorpolyethylene glycol in glycinebuffer with bovine serumalbuminR2. Latex particles coatedwith human IgG in glycinebufferR1. Lyophilisate ofpolystyrene particles coatedwith antigen-antibodycomplex of human gamma-globulin/ anti-humangamma-globulin from sheepR2. aqueous solution ofpolyethylene glycol
Performance Characteristics:
Assay range10-120 IU/mL10-600 IU/mL w/postdilution10-600 IU/mL
Precision (Day-to-day)8.0% at 32 IU/mL5.0% at 125 IU/mL9.3% at 92 IU/mL6.2% at 155 IU/mL6.6% at 306 IU/mL
AccuracyN = 244$y = 0.70x + 12 IU/mL$ vs.Behring N Latex RF ReagentN = 72$y = 0.99 x - 20 IU/mL$ vs.enzyme immunoassay
Sensitivity (Analytical)10 IU/mL10 IU/mL

{7}------------------------------------------------

Table 6

COBAS INTEGRA Cassette for Calcium

COBAS INTEGRACassette for Calcium(Modified)COBAS INTEGRACassette for Calcium(Cleared)BoehringerMannheimCalcium on Hitachi911
MethodologyColorimetrico-CresolphthaleincomplexoneColorimetrico-cresolphthaleincomplexoneColorimetrico-cresolphthaleincomplexone
Sample typeSerum, heparinizedplasma and urineSerum, heparinizedplasma and urineSerum, heparinizedplasma and urine
CalibratorRoche Calibrator(K942706)Roche Calibrator(K942706)Boehringer MannheimPrecical Calibrato r Serum
Reagent (activeingredients)1. o-Cresolphthaleincomplexone2. 8-Hydroxyquinoline3. CAPS4. NaOH1. o-Cresolphthaleincomplexone2. 8-Hydroxyquinoline3. CAPS4. NaOH1. 4-Aminobutyric acid2. o-Cresolphthaleincomplexone3. 8-Hydroxyquinoline4. HCI
Performance Characteristics:
Assay rangeSerum, plasma:0-20 mg/dLUrine:0-28 mg/dL0-140 mg/dLw/postdilutionSerum, plasma:0-20 mg/dLUrine:0-100 mg/dLSerum, plasma:0-15 mg/dL0-24.9 w/postdilutionUrine:0-15 mg/dL0-60 mg/dLw/postdilution
Precision(Total)Serum, plasma:3.5% at 9.2 mg/dL3.1% at 13.6 mg/dLUrine:0.91% at 5.6 mg/dL1.1% at 23.7 mg/dLSerum, plasma:2.1 % at 7.9 mg/dL1.8 % at 10.2 mg/dLUrine:2.8% at 3.9 mg/dL0.77% at 18 mg/dLSerum, plasma:1.6 % at 8.9 mg/dL1.2 % at 9.7 mg/dL1.3% at 12.5 mg/dLUrine:2.2% at 2.9 mg/dL1.7% at 7.8 mg/dL1.4% at 36.7 mg/dL
AccuracyN = 196R = 0.987 vs. BMHitachi 911N = 240R = 0.960 vs. BMHitachi 911N = 53R = 0.997 vs. BMHitachi 717
Sensitivity(Analytical)$4.0 X 10^{2}$ AA per mg/dLof calcium$3.2 X 10^{2}$ AA per mg/dLof calciumNot specified in labeling

§ 862.1145 Calcium test system.

(a)
Identification. A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).(b)
Classification. Class II.