K Number
K963286

Validate with FDA (Live)

Date Cleared
1996-10-07

(47 days)

Product Code
Regulation Number
866.5460
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMMAGE Immunochemistry System Haptoglobin (HPT) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of human haptoglobin concentrations in human serum samples by rate nephelometry. This assay is designed for use with Beckman's IMMAGE Immunochemistry System.

Device Description

The IMMAGE Immunochemistry System Haptoglobin (HPT) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of human haptoglobin concentrations in human serum samples.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the IMMAGE™ Immunochemistry System Haptoglobin (HPT) Reagent.

Please note: The provided document is a 510(k) summary, which often focuses on demonstrating substantial equivalence to a predicate device rather than presenting de novo acceptance criteria for a novel device. As such, some information (like standalone performance targets without a comparative predicate, or extensive details on ground truth setting for complex clinical scenarios) might be less explicit or not present, as the primary goal is to show the new device performs similarly to an already-approved device.


Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implicit from Predicate, or Explicit)Reported Device Performance (IMMAGE HPT Reagent)
Method ComparisonSubstantial equivalence to predicate device (Beckman's HPT Haptoglobin Reagent on Array® Systems).Slope: 1.004, Intercept: -6.80, r: 0.996
Stability - Shelf-lifeSame as predicate (24 months at 2-8°C)24 months
Stability - Open ContainerNot explicitly stated as a criterion, but a claim is made.14 day open container stability
Stability - CalibrationNot explicitly stated as a criterion, but a claim is made.14 day calibration stability
Imprecision (Within-Run)Not explicitly stated numerically as an acceptance criterion. Implicitly, performance should be acceptable for clinical use and comparable to predicate.Level 1 (60.8 mg/dL): %CV = 1.4Level 2 (136 mg/dL): %CV = 2.5Level 3 (322 mg/dL): %CV = 2.8

Explanation of Implicit Criteria:
For 510(k) submissions, the acceptance criteria are largely implicit in demonstrating "substantial equivalence" to a predicate device. This means the new device needs to perform as well as, or virtually identically to, the predicate device in relevant performance characteristics. The strong correlation (r = 0.996) and slope/intercept close to 1 and 0 respectively for method comparison are key indicators of this equivalence. Similarly, the stability and imprecision data are presented without explicit numerical targets but are intended to show acceptable performance for an in vitro diagnostic.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size:
    • Method Comparison: Not explicitly stated, but the summary refers to "method comparison" studies. The number of samples (e.g., patient samples) used to generate the slope, intercept, and 'r' value is not provided.
    • Imprecision:
      • For each of the three control levels, 80 results were used to calculate the mean, SD, and %CV.
  • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). This type of detail is often omitted in a 510(k) summary for in vitro diagnostic devices unless there are specific clinical or demographic implications. The samples are human serum, but their origin is not described.

3. Number of Experts and Qualifications for Ground Truth

  • Ground Truth Establishment: For an in vitro diagnostic (IVD) reagent like this, "ground truth" typically refers to the reference method (the predicate device in this case) or highly characterized control materials. It does not generally involve human expert interpretation in the same way as imaging or clinical decision support AI.
  • Experts: Not applicable in the context of human expert review for establishing ground truth for a quantitative immunoassay.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. This document describes performance characteristics of a quantitative immunoassay, not a subjective diagnostic interpretation requiring adjudication among human readers.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. These studies are typically relevant for AI-powered diagnostic imaging or interpretation tools where human readers are involved in the diagnostic process. This device is a reagent for a quantitative immunoassay system.
  • Effect Size with AI Assistance: Not applicable, as no human readers or AI assistance in interpretation are described for this IVD.

6. Standalone Performance Study

  • Standalone Performance: Yes, standalone performance was assessed for the IMMAGE HPT Reagent. The "Method Comparison Study Results" compare the IMMAGE HPT Reagent's measurements against the predicate device, effectively showing its performance in isolation (though compared to a reference). The "Estimated Within-Run Imprecision" also represents a standalone performance characteristic of the device.
    • Method Comparison: The IMMAGE HPT results were directly correlated with the predicate.
    • Imprecision: The device's internal variability was measured independently.

7. Type of Ground Truth Used

  • Ground Truth Type:
    • For Method Comparison: The "ground truth" was established by the predicate device (Beckman's HPT Haptoglobin Reagent on the Array® Systems). The IMMAGE HPT Reagent's values were compared to those obtained from the predicate.
    • For Imprecision: The "ground truth" for the mean values was derived from control materials with established concentrations (Level 1, Level 2, Level 3).

8. Sample Size for the Training Set

  • Training Set Sample Size: Not specified. For IVD reagents, the concept of a "training set" in the context of machine learning (where the term is most common) is generally not applicable. Reagent development involves extensive analytical validation (specificity, linearity, interference, matrix effects, etc.) rather than training an algorithm on a 'sample set'. The development process involves optimizing the chemical formulation and assay parameters based on performance against established analytical standards and reference methods.

9. How Ground Truth for the Training Set was Established

  • Ground Truth for Training Set: Not applicable in the machine learning sense. The "ground truth" during the development and optimization of such a reagent would be based on:
    • Chemical principles and established analytical methods for quantitative determination of haptoglobin.
    • Reference materials and calibrators with certified or highly accurate haptoglobin concentrations.
    • Performance characteristics required for clinical utility.

{0}------------------------------------------------

OCT 7 1996

K963286

Summary of Safety & Effectiveness IMMAGE™ Immunochemistry System Haptoglobin (HPT) Reagent

1.0 Submitted By:

100 m

Margie George Project Manager. System Development Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 961-3765 FAX: (714) 961-3759

2.0 Date Submitted:

23 Aug-1996

3.0 Device Name(s):

3.1 Proprietary Names

IMMAGE™ Immunochemistry System Haptoglobin (HPT) Reagent

3.2 Classification Names

Haptoglobin immunological test system(21 CFR 866.5460)

4.0 Predicate Device(s):

Beckman Immunochemistry Systems HPT Haptoglobin Reagent

5.0 Description:

The IMMAGE Immunochemistry System Haptoglobin (HPT) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of human haptoglobin concentrations in human serum samples.

6.0 Intended Use:

IMMAGE Immunochemistry System Haptoglobin (HPT) The reagent. in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of human haptoglobin concentrations in human serum samples by rate nephelometry. This assay is designed for use with Beckman's IMMAGE Immunochemistry System.

{1}------------------------------------------------

Beckman Instruments, Inc., Section 510(k) Notification IMMAGE™ Immunochemistry System Haptoglobin (HPT) Reagent Summary of Safety & Effectiveness

7.0 Comparison to Predicate(s):

The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.

ReagentAspect/CharacteristicComments
SIMILARITIES
IMMAGE System(HPT) ReagentAssay method - rate nephelometrySame as Beckman's HPTHaptoglobin Reagent
Analytic Range 35 - 350 mg/dLExtended Range 5.8 - 12,600 mg/dLSame as Beckman's HPTHaptoglobin Reagent
Sample and reagent ratiosSame as Beckman's HPTHaptoglobin Reagent
AntibodySame as Beckmna's HPTHaptoglobin Reagent
Shelf life of 24 months (stored at 2-8°C)Same as Beckman's HPTHaptoglobin Reagent
DIFFERENCES
IMMAGE System(HPT) ReagentReaction TemperatureIMMAGE System HPT assaysrun at 37°C and the predicateruns at 26.7°C
Antigen excess solution (AGXS)IMMAGE System HPTReagent AGXS is predilutedand present in the reagentcartridge while the predicaterequires off-line preparation ofthe solution

{2}------------------------------------------------

Beckman Instruments, Inc., Section 510(k) Notification IMMAGE™ Immunochemistry System Haptoglobin (HPT) Reagent Summary of Safety & Effectiveness

8.0 Summary of Performance Data:

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method companson, stability, and imprecision experiments that relate results obtained from the Beckman Haptoglobin Reagent to the IMMAGE HPT Reagent.

Method Comparison Study Results IMMAGE HPT Reagent vs. Beckman Haptoglobin Reagent

AnalyteSlopeInterceptrPredicate
IMMAGEHPT1.004-6.800.996Beckman's HPTHaptoglobinReagent on theArray®Systems

Stability Study Results

ReagentProduct Claim
IMMAGE HPT24 month shelf-life14 day open container stability14 day calibration stability

Estimated Within-Run Imprecision

MATERIALMEAN (mg/dL)SD (mg/dL)%CVNumber of Results
HPT Reagent
Level 160.80.871.480
Level 21363.42.580
Level 33229.72.880

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

§ 866.5460 Haptoglobin immunological test system.

(a)
Identification. A haptoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the haptoglobin (a protein that binds hemoglobin, the oxygen-carrying pigment in red blood cells) in serum. Measurement of haptoglobin may aid in the diagnosis of hemolytic diseases (diseases in which the red blood cells rupture and release hemoglobin) related to the formation of hemoglobin-haptoglobin complexes and certain kidney diseases.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.