K Number
K963197

Validate with FDA (Live)

Device Name
ON.SITE ALCOHOL
Manufacturer
Date Cleared
1996-09-19

(35 days)

Product Code
Regulation Number
862.3040
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ON•SITE Alcohol is recommended for professional use only and is not intended for overthe-counter sale to the general public. Professional use includes applications in healthcare, corrections and drug treatment.

ON•SITE Alcohol yields only a "qualitative" result. A positive result indicates that alcohol (ethanol) is present at a concentration of approximately 0.02% when performed according to the instructions.

Confirmation of ON•SITE Alcohol results using gas chromatography is recommended if quantitative results are desired.

Device Description

ON•SITE Alcohol is a self contained, disposable enzymatic test device that provides rapid detection of alcohol in saliva or urine. It includes all reagents and supplies necessary to perform the test.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study details:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated in a quantitative form, but rather implied by the design intention and the performance characteristics observed in the 'modified' device compared to the 'current' device. The key goal was to increase the reaction time and slightly raise the limit of detection.

Criteria (Implied)Current Device PerformanceModified Device PerformanceTarget/Acceptance (Implied)
Accuracy & Precision (Detection at 0.008% w/v EtOH)13/20 positives in 2 mins, Mean Reaction Time: 111 sec0/20 positives in 2 mins, Mean Reaction Time: 147 secReduced detection at 0.008% (i.e., fewer positives at 2 mins, increased reaction time)
Accuracy & Precision (Detection at 0.02% w/v EtOH)20/20 positives in 2 mins, Mean Reaction Time: 81 sec12/20 positives in 2 mins, Mean Reaction Time: 115 secMaintain detection at 0.02% (or slightly reduced as a side effect of shifted LOD), increased reaction time
Accuracy & Precision (Detection at 0.032% w/v EtOH)20/20 positives in 2 mins, Mean Reaction Time: 61 sec20/20 positives in 2 mins, Mean Reaction Time: 100 secMaintain detection at 0.032%, increased reaction time
Sensitivity (Reaction Times across various EtOH concentrations)Range of mean reaction times from 1 min 6 sec to 1 min 47 secRange of mean reaction times from 1 min 29 sec to 2 min 41 secConsistently increased reaction times across all tested concentrations

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size:
    • Accuracy and Precision: 20 replicates for each of the three ethanol controls (0.008%, 0.02%, 0.032% w/v). This means 20x3 = 60 tests for the current device and 60 tests for the modified device, totaling 120 tests for this part.
    • Sensitivity: 10 replicates for each of the eight ethanol controls (0.005% to 0.0225% w/v). This means 10x8 = 80 tests for the current device and 80 tests for the modified device, totaling 160 tests for this part.
  • Data Provenance: Not specified within the document (e.g., country of origin, type of samples used - saliva/urine). Given the context of a 510(k) summary, it's implied to be laboratory testing of controlled ethanol solutions, not human samples. It is a prospective laboratory study comparing a modified device to its predicate.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Not applicable. The ground truth was established by laboratory-prepared ethanol solutions of known concentrations. The "results were read" likely implies visual interpretation by a technician, but their number or qualifications are not specified as it's not an expert-driven ground truth.

4. Adjudication Method for the Test Set

Not applicable. The outcomes were objective measurements (time to positive result) and counts of positive/negative results at a specific time point, based on predefined ethanol concentrations. No adjudication process for differing interpretations of subjective data would be needed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC study was not done. This study is a technical performance comparison between two versions of a diagnostic device in a laboratory setting, not a study involving human readers interpreting cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, this was a standalone performance study. The device's reaction to known concentrations of alcohol was evaluated directly, without human interpretation of results beyond simple visual observation of a color change, and timing that change. There is no "algorithm" in the modern sense relevant to AI.

7. The Type of Ground Truth Used

The ground truth used was objectively known concentrations of ethanol in laboratory-prepared control solutions.

8. The Sample Size for the Training Set

Not applicable. This device is an enzymatic test, not an AI/machine learning model, so there is no training set in the computational sense. The "training" for the device's development would have been implicit in chemical formulation and calibration, not a data-driven training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set in the context of an AI/ML model. If interpreted very broadly as the development process for the chemical formulation, the ground truth would still be based on known chemical reactions and desired performance characteristics to achieve the intended "shift in reaction time and detection levels."

{0}------------------------------------------------

K963197

510(k) SUMMARY

1.Submitter:ANSYS, Inc.
Address:2 GoodyearIrvine, CA 92618
Telephone:(714) 770-9381 Ext. 7915
Contact Person:Lorna Gamboa
Date Prepared:7/9/96
2.Trade Name :ON•SITE Alcohol
Common Name:Alcohol Test System
Classification:Class II

3. Predicate Device Information:

The modified ON.SITE Alcohol test is being compared to its predecessor, the current ON.SITE Alcohol, both manufactured by ANSYS, Inc.

4. Description of the Device:

ON•SITE Alcohol is a self contained, disposable enzymatic test device that provides rapid detection of alcohol in saliva or urine. It includes all reagents and supplies necessary to perform the test.

5. Intended Use:

ON•SITE Alcohol is recommended for professional use only and is not intended for overthe-counter sale to the general public. Professional use includes applications in healthcare, corrections and drug treatment.

ON•SITE Alcohol yields only a "qualitative" result. A positive result indicates that alcohol (ethanol) is present at a concentration of approximately 0.02% when performed according to the instructions.

Confirmation of ON•SITE Alcohol results using gas chromatography is recommended if quantitative results are desired.

{1}------------------------------------------------

Comparison of technological Characteristics: 6.

The modified ON•SITE Alcohol uses the same technology as its predecessor. The only difference is in the formulation of one of its reagents. It has been modified to increase the reaction time, thus slightly raising the limit of detection.

Performance Characteristics and Supportive Data: 7.

Accuracy and Precision

Accuracy and precision of the current and modified device was compared by running twenty replicates of three ethanol controls (0.008%, 0.02% and 0.032% w/v). Tests were timed and results were read at exactly 2 minutes. Results were reported as either positive or negative. In addition, reaction times for each test (i.e., time required to produce an observable positive result) were recorded. The mean, standard deviation and coefficient of variations were calculated and compared.

Results confirmed the desired shift in reaction time and detection levels as follows:

% w/vEtOHNo. of PositivesMean Reaction TimeCoefficient of Variation
currentmodifiedcurrentmodifiedcurrentmodified
0.008130111 sec147 sec10.3%8.2%
0.020201281 sec115 sec8.3%8.8%
0.032202061 sec100 sec7.9%6.7%

Sensitivity

The sensitivity of the current and modified device were also compared. Ten replicates of eight controls were tested using both devices. Reaction times were compared. Again, results confirmed the targeted shift in reaction times.

% w/vMean Reaction TimeDifference
EtOHcurrentmodifiedsec
0.0051 min 47 sec2 min 41 sec54
0.00751 min 33 sec2 min 14 sec41
0.011 min 19 sec2 min 10 sec51
0.01251 min 21 sec1 min 53 sec32
0.0151 min 19 sec1 min 42 sec23
0.01751 min 12 sec1 min 47 sec35
0.021 min 6 sec1 min 41 sec35
0.02251 min 6 sec1 min 29 sec23

§ 862.3040 Alcohol test system.

(a)
Identification. An alcohol test system is a device intented to measure alcohol (e.g., ethanol, methanol, isopropanol, etc.) in human body fluids (e.g., serum, whole blood, and urine). Measurements obtained by this device are used in the diagnosis and treatment of alcohol intoxication and poisoning.(b)
Classification. Class II.