K Number
K963120

Validate with FDA (Live)

Date Cleared
1997-02-28

(200 days)

Product Code
Regulation Number
870.5550
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Responder 2500 is intended to be used in a hospital environment by emergency personnel in evaluating and treating life-threatening arrhythmias.
♦ Responder 2500 is intended to be used by trained operators
♦ Responder 2500 is designed for external and internal defibrillation (including cardioversion)
♦ Responder 2500 is capable of monitoring the heart rate with adjustable alarm limits.
♦ Responder 2500 is designed for external pacing with adjustable current and rate.

Device Description

Responder 2500 and Cardiac Care System is an emergency cardiac monitor/defibrillator with an external pacing option.

AI/ML Overview

This 510(k) submission for the Marquette Electronics Responder 2500 Defibrillator and Cardiac Care System focuses on demonstrating substantial equivalence to predicate devices (Marquette Series 900 and Marquette Series 1500 Responder) rather than providing detailed acceptance criteria and a study proving device performance against those criteria in a typical AI/ML medical device context.

Therefore, many of the requested details regarding acceptance criteria, sample sizes, expert involvement, and ground truth establishment, which are standard for AI/ML device evaluations, are not present in this document. The submission relies on compliance with voluntary standards and general performance equivalency to predicate devices.

Here's an attempt to extract and interpret the available information according to your requested format, along with explicit statements about what information is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (from voluntary standards)Reported Device Performance
DefibrillationANSI/AAMI DF2-1996"Performs as well as the predicate devices" (Marquette Series 900 and 1500 Responder). Animal studies performed.
Cardiac MonitoringANSI/AAMI ES1-1993"Capable of monitoring the heart rate with adjustable alarm limits." "Performs as well as the predicate devices." Field tests on monitor function.
External PacingANSI/AAMI EC13-1992"Designed for external pacing with adjustable current and rate." "Performs as well as the predicate devices." Animal studies performed.
General SafetyIEC 601-1, UL 2601-1Quality assurance measures applied, including safety testing. "Is as safe...as the predicate devices."
Electromagnetic CompatibilityIEC 601-1-2Quality assurance measures applied. "Performs as well as the predicate devices."
Essential PerformanceIEC 601-2-4 (Defibrillators)"Performs as well as the predicate devices."
ECG PerformanceIEC 601-2-27 (ECG monitoring)"Performs as well as the predicate devices."
Intended UseIdentical to predicate device intended use"Intended use of Responder 2500 is identical to the intended use of the predicate devices."
TechnologyEmploys same technology as predicate devices"Responder 2500 employs the same technology as the predicate devices."

Study Description:

The document describes the "development" of the Responder 2500, which involved various testing and quality assurance measures to confirm compliance with listed voluntary standards and establish equivalence to predicate devices. These measures are:

  • Requirements specification reviews
  • Code inspections
  • Software and hardware testing
  • Safety testing
  • Environmental testing
  • Field tests on the monitor function
  • Animal studies on the defibrillation and pacing functions
  • Final validation

The results of these measurements "demonstrated that Responder 2500 is as safe, as effective, and performs as well as the predicate devices Marquette Series 900 and Marquette Series 1500 Responder."

Missing Information (Relative to AI/ML Device Evaluation):

Specific quantitative acceptance criteria (e.g., sensitivity, specificity, AUC thresholds) are not provided, as this is a traditional medical device submission focused on performance standards and predicate equivalence.


2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified. The document mentions "field tests on the monitor function" and "animal studies on the defibrillation and pacing functions," but provides no numbers for test subjects (human or animal) or the duration/scope of these tests.
  • Data Provenance: Not specified. No mention of country of origin or whether data was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not applicable/Not specified. The evaluation relies on compliance with established engineering standards and functional performance, not on expert adjudication of diagnostic outcomes like an AI/ML algorithm might.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable/Not specified. (See point 3)

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI/ML device and therefore no MRMC study, or impact on human readers, was conducted or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The device itself is a standalone functional unit. Performance was assessed based on its functional capabilities (monitoring, defibrillation, pacing) and compliance with standards. There isn't an "algorithm only" performance concept in the way it applies to AI/ML devices, as the device's function is inherently a standalone "performance."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Functional Ground Truth: The "ground truth" for this device's performance would be defined by the technical specifications outlined in the voluntary standards referenced (e.g., ANSI/AAMI DF2-1996 for defibrillator performance, IEC 601-2-27 for ECG performance). For animal studies, the "ground truth" would be established by physiological measurements and successful or unsuccessful defibrillation/pacing events as determined by veterinarians or other qualified personnel. It is not expert consensus for image interpretation or pathology.

8. The sample size for the training set

  • Not applicable. This device is a hardware/software system, not a machine learning model that undergoes a "training" phase with a dataset in the AI/ML sense. Its development involved traditional engineering, software development, and quality assurance processes.

9. How the ground truth for the training set was established

  • Not applicable. (See point 8)

{0}------------------------------------------------

K963120

FEB 2 8 1997

4

SECTION 2 - SUMMARY AND CERTIFICATION

2.1 510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter:Marquette Electronics, Inc.8200 W. Tower AvenueMilwaukee, WI 53223Telephone: (414) 355-5000FAX: (414) 362-3553
Contact Person:Kristin Pabst
Device:Trade Name: Responder 2500 Defibrillator and Cardiac Care System
Classification Name: DC-Defibrillator, low energy (including paddles)
Pacemaker, Cardiac, External Transcutaneous (Non-invasive)
Predicate Devices:Marquette Series 900Marquette Series 1500 Responder
Device Description:Responder 2500 and Cardiac Care System is an emergency cardiac monitor/defibrillator with an external pacing option.
Intended Use:Responder 2500 is intended to be used in a hospital environment by emergency personnel in evaluating and treating life-threatening arrhythmias.
♦ Responder 2500 is intended to be used by trained operators
♦ Responder 2500 is designed for external and internal defibrillation (including cardioversion)
♦ Responder 2500 is capable of monitoring the heart rate with adjustable alarm limits.
♦ Responder 2500 is designed for external pacing with adjustable current and rate.

510(k) Notification "Responder 2500" - August 9, 1996

{1}------------------------------------------------

5


The intended use of Responder 2500 is identical to the intended use of the predicate devices.

Responder 2500 employs the same technology as the predicate devices. Technology:

Performance:

Responder 2500 complies with the voluntary standards ANSI/AAMI DF2-1996, ANSI/AAMI ES1 1993, ANSVAAMI EC13 1992, IEC 601-1, IEC 601-1-2, IEC 601-2-4, IEC 601-2-27, UL 2601-1.

The following quality assurance measures were applied to the development of Responder 2500.

Requirements specification reviews, code inspections, software and hardware testing, safety testing, environmental testing, field tests on the monitor function and animal studies on the defibrillation and pacing functions, and final validation.

The results of these measurements demonstrated that Responder 2500 is as safe, as effective, and performs as well as the predicate devices Marquette Series 900 and Marquette Series 1500 Responder.

§ 870.5550 External transcutaneous cardiac pacemaker (noninvasive).

(a)
Identification. An external transcutaneous cardiac pacemaker (noninvasive) is a device used to supply a periodic electrical pulse intended to pace the heart. The pulse from the device is usually applied to the surface of the chest through electrodes such as defibrillator paddles.(b)
Classification. Class II. The special controls for this device are:(1) “American National Standards Institute/American Association for Medical Instrumentation's DF-21 ‘Cardiac Defibrillator Devices’ ” 2d ed., 1996, and
(2) “The maximum pulse amplitude should not exceed 200 milliamperes. The maximum pulse duration should not exceed 50 milliseconds.”