K Number
K963061

Validate with FDA (Live)

Date Cleared
1996-11-04

(89 days)

Product Code
Regulation Number
866.5270
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMMAGE Immunochemistry Systems C-reactive protein (CRP) reagent, in conjunction with Beckman CAL 5 Plus, is intended for use in the quantitative determination of human C-reactive protein concentrations in human serum and plasma samples by rate nephelometry. This assay is designed for use with Beckman's IMMAGE Immunochemistry System.

Device Description

The IMMAGE Immunochemistry Systems C-reactive protein (CRP) reagent, in conjunction with Beckman CAL 5 Plus, is intended for use in the quantitative determination of human C-reactive protein concentrations in human serum and plasma samples.

AI/ML Overview

The provided text describes the Beckman IMMAGE™ Immunochemistry System C-Reactive Protein (CRP) Reagent. Here's a breakdown of the requested information based on the document:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" for the method comparison in a clear quantitative manner (e.g., a specific R-value threshold). However, it implies that the device is substantially equivalent to the predicate through the presented performance data.

Performance MetricAcceptance Criteria (Implied/Direct)Reported Device Performance (IMMAGE CRP Reagent)
Method ComparisonSubstantial equivalence to Behring N Latex CRP Reagent.Demonstrated through method comparison study (details not fully provided in text).
Stability (Shelf-life)Not explicitly stated (typically defined internally by manufacturer).24 month shelf-life
Stability (Open Container)Not explicitly stated.14 days
Stability (Calibration)Not explicitly stated.14 days
Within-run ImprecisionNot explicitly stated (typically defined by industry standards or manufacturer's target).Level 1 (Mean 1.65 mg/dL): %CV = 8.1%
Level 2 (Mean 5.41 mg/dL): %CV = 3.1%
Level 3 (Mean 12.6 mg/dL): %CV = 1.8%

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not specify the sample size for the method comparison study (comparison between IMMAGE CRP Reagent and Behring N Latex CRP Reagent).
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • This information is not applicable to an immunochemistry reagent like the one described. "Ground truth" for such a device is established by comparing its quantitative results against a recognized reference method or a legally marketed predicate device, not by expert interpretation of images or clinical cases.

4. Adjudication Method for the Test Set

  • Not applicable as this is a quantitative chemical assay, not a diagnostic imaging or interpretive device requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is an in-vitro diagnostic (IVD) reagent for automated quantitative measurement, not an AI-assisted diagnostic tool that would involve human readers or MRMC studies.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, the performance data presented (method comparison, stability, imprecision) represents the standalone performance of the IMMAGE CRP Reagent system. The device itself performs the quantitative measurement without human interpretation of the results for its basic function.

7. The Type of Ground Truth Used

  • For the method comparison study, the "ground truth" was established by the results obtained from the predicate device (Behring N Latex CRP Reagent). The study aimed to demonstrate substantial equivalence to this legally marketed device.
  • For the stability and imprecision studies, the "ground truth" is typically derived from known control materials or calibrated reference solutions with assigned values.

8. The Sample Size for the Training Set

  • This information is not applicable as this is not a machine learning or AI device that requires a distinct "training set" in the context of the reported performance studies. The "training" of such a system involves internal development and optimization based on chemical and analytical principles.

9. How the Ground Truth for the Training Set was Established

  • Not applicable for the reasons stated in point 8.

{0}------------------------------------------------

K963061

Revised
10/4/96
Inc: Dated

BECKMAN

Summary of Safety & Effectiveness IMMAGE™ Immunochemistry System C-Reactive Protein (CRP) Reag

1.0 Submitted By:

Annette Hellie Regulatory Specialist, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 993-8767 FAX: (714) 961-4457

2.0 Date Submitted:

06 Aug-1996

3.0 Device Name(s):

3.1 Proprietary Names

IMMAGE™ Immunochemistry System C-Reactive Protein (CRP) Reagent

3.2 Classification Names

C-reactive protein immunological test system(21 CFR 866.5270)

4.0 Predicate Device(s):

Behring N Latex CRP

5.0 Description:

The IMMAGE Immunochemistry Systems C-reactive protein (CRP) reagent, in conjunction with Beckman CAL 5 Plus, is intended for use in the quantitative determination of human C-reactive protein concentrations in human serum and plasma samples.

6.0 Intended Use:

The IMMAGE Immunochemistry Systems C-reactive protein (CRP) reagent, in conjunction with Beckman CAL 5 Plus, is intended for use in the quantitative determination of human C-reactive protein concentrations in human serum and plasma samples by rate nephelometry. This assay is designed for use with Beckman's IMMAGE Immunochemistry System.

{1}------------------------------------------------

Comparison to Predicate(s): 7.0

The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.

ReagentAspect/CharacteristicComments
SIMILARITIES
IMMAGE System(CRP) ReagentIntended UseSame as BehringN Latex CRP
IMMAGE System(CRP) ReagentNephelometric methodologySame as BehringN Latex CRP
IMMAGE System(CRP) ReagentUse of Latex particle technologySame as BehringN Latex CRP
DIFFERENCES
IMMAGE System(CRP) ReagentForm of reagentthe IMMAGE CRP is liquidstable while Behring N LatexCRP reagent is lyophilized
IMMAGE System(CRP) ReagentAntibody sourceIMMAGE CRP uses goatwhile Behring N Latex CRPuses rabbit.
IMMAGE System(CRP) ReagentStabilityIMMAGE CRP is stable for 14days once opened, properlystored while Behring N LatexCRP reagent is stable for oneweek following reconstitution,.

{2}------------------------------------------------

Summary of Performance Data: 8.0

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the Behring N Latex CRP Reagent to the IMMAGE CRP Reagent.

Method Comparison Study Results IMMAGE CRP Reagent vs. Behring N Latex CRP Reagent

comments of a superior in a consideration a n a comprehensive a no a province a no a manager aThe State of Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children ChildrenCENTRAL CONSULTIONSTATIST COLLECTION CONTRACTORSANNE LCharacter Children and Children and Children and Charges of Children and Children of Children and.The Callery County of Children Collection Collection Collection Collection Collection Collection Come of Children(6) 03-2 10-2000Call CalleryCharles States and Camber 1991 - 1 - 1 accord 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -" which " don't does and " since " since " since "11151 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000Comments of the comments of the states of the country of the county of the county of the county of the county of the county of the county of the county of the county of the cA1
-namAlarennin
1

Stability Study Results

ReagentProduct Claim
IMMAGE CRP24 month shelf-life14 day open container stability
14 day calibration stability

Estimated Within-Run Imprecision

MATERIALMEAN (mg/dL)SD (mg/dL)%CVNumber of Results
CRP Reagent
Level 11.650.1348.180
Level 25.410.1653.180
Level 312.60.221.880

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).