K Number
K962906

Validate with FDA (Live)

Date Cleared
1996-08-23

(29 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Scripps Laboratories hCG One-Step is a rapid, one step immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy. For Laboratory and Professional Use.

Device Description

HCG is a hormone produced and secreted by the placenta following implantation. Its early presence and rapid increase in maternal urine make it an excellent marker for use in the diagnosis of pregnancy.

Scripps Laboratories hCG One-Step uses monoclonal and polyclonal antibody technology to specifically detect hCG in urine.

The product is configured in a dipstick format. When the product is partially submerged in urine, the sample travels through the Test Strip. If hCG is present at levels of 25 mIU/ml (W.H.O. 3rd I.S. 75/537) or greater, a positive result, two distinct red bands will appear. If no hCG is present, a negative result, one red band, will appear.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Scripps Laboratories hCG One-Step device, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a separate, pre-defined set of targets. Instead, it presents performance characteristics for the Scripps Laboratories hCG One-Step and compares them to legally marketed devices. The implied acceptance criteria are that the device's performance should be comparable or superior to these predicate devices.

Performance CharacteristicAcceptance Criteria (Implied from predicate devices)Reported Device Performance (Scripps Laboratories hCG One-Step)
Sensitivity99.5% - 99.6%100%
Specificity99.6%99.6%
Accuracy99.5% - 99.6%99.8%
Detection Limit25 mIU/ml25 mIU/ml
Timing1-3 minutes5 minutes
StandardizationWHO 3rd IS 75/537WHO 3rd IS 75/537

Note on Timing: While the Scripps product takes 5 minutes, compared to 1 or 3 minutes for predicate devices, the other performance metrics (sensitivity, specificity, accuracy) are equal or better, suggesting this timing difference was not considered a significant hindrance to substantial equivalence.

2. Sample Size and Data Provenance for Test Set

  • Sample Size: 507 urine specimens.
  • Data Provenance: The data was collected in a "multicenter field study." The document does not explicitly state the country of origin but implies a clinical laboratory or professional setting within the US, as it's for 510(k) approval by the FDA. The study was prospective in the sense that the Scripps device was being evaluated against a commercial benchmark using new specimens presented for testing.

3. Number of Experts and Qualifications for Ground Truth

  • The document states that the ground truth for the test set was established by a "commercially available visual hCG test" (Quidel® Quick Vue® One-Step hCG Urine Test).
  • The actual individuals performing the initial interpretations are described as "individuals with different educational backgrounds and training." There is no specific number of "experts" defined for establishing the ground truth beyond the performance of the predicate device.

4. Adjudication Method for the Test Set

The adjudication method was based on the results of a single, commercially available predicate device: the Quidel® Quick Vue® One-Step hCG Urine Test. There's no mention of multiple expert adjudicators or a specific voting/consensus method (e.g., 2+1, 3+1).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This study focuses on the standalone performance of the Scripps device compared to an existing device, not on the improvement of human readers with or without AI assistance. The device is a diagnostic test itself, not an AI-assisted interpretation tool for human readers.

6. Standalone Performance (Algorithm Only without Human-in-the-loop)

Yes, a standalone performance study was done. The Scripps Laboratories hCG One-Step is a rapid, one-step immunoassay, meaning it produces a visual result directly. The performance data (sensitivity, specificity, accuracy) reflect the device's ability to correctly identify hCG levels independently. "Individuals with different educational backgrounds and training" performed the testing, suggesting human interpretation of the visual result, but the performance metrics are attributed to the device itself.

7. Type of Ground Truth Used

The ground truth used was established by a predicate commercial device's result. Specifically, the Quidel® Quick Vue® One-Step hCG Urine Test was considered the reference standard against which the Scripps Laboratories device was compared for "qualitatively positive or negative for hCG."

8. Sample Size for the Training Set

The document does not provide information regarding a separate "training set" or its sample size. This type of immunoassay usually undergoes extensive internal R&D and validation with various contrived and clinical samples to optimize its formulation and design. The data presented in this 510(k) summary represents a clinical validation (test set) of the final product.

9. How Ground Truth for the Training Set Was Established

As no dedicated "training set" is described, there's no information on how its ground truth would have been established. For the development of an immunoassay, the "ground truth" for internal optimization typically involves:

  • Using samples spiked with known concentrations of hCG.
  • Comparing results to highly accurate laboratory methods (e.g., quantitative immunoassay, mass spectrometry) or established reference standards.
  • Testing with known positive and negative clinical samples confirmed by other means.

{0}------------------------------------------------

K962906

AUG 2 3 1006

Scripps Laboratories hCG One-Step 510(k) Notification

SECTION VI.

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Scripps Laboratories. The logo consists of the word "SCRIPPS" in large, bold, sans-serif font, with a small circle above the "I". Below the word "SCRIPPS" is the word "LABORATORIES" in a smaller, sans-serif font. The logo is black and white.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS VI.

  • Submitter Identification A.
    • Name and Address l.

Scripps Laboratories, Inc. 6838 Flanders Drive San Diego, CA 92121

    1. Contact Person
      Noreen E. Leamon
    1. Phone Numbers
      Voice: (619) 546-5800 FAX: (619) 546-5812
    1. Date of Submission
      July 22, 1996
  • Names of Device B.

    • l. Proprietary Name

SCRIPPS LABORATORIES hCG One-Step

    1. Common Name
      A Rapid, One-Step Immunoassay for the Qualitative Detection of Human Chorionic Gonadotropin (hCG) in Urine for the Early Detection of Pregnancy.
    1. Classification Name
      Human Chorionic Gonadotropin (hCG) Test System.
  • Legally Marketed Devices to Which Substantial Equivalence is C. Claimed

    • 1 . Quidel® QuickVue® One-Step hCG Urine Test (ref. K925435)
      1. Wyntek Diagnostics OSOM™ hCG-Urine Test (ref. K953383)

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Scripps Laboratories. The logo consists of the word "SCRIPPS" in large, bold, sans-serif font, with a small circle above the "I". Below the word "SCRIPPS" is the word "LABORATORIES" in a smaller, sans-serif font. The logo is black and white.

  • D. Description
    HCG is a hormone produced and secreted by the placenta following implantation. Its early presence and rapid increase in maternal urine make it an excellent marker for use in the diagnosis of pregnancy.

Scripps Laboratories hCG One-Step uses monoclonal and polyclonal antibody technology to specifically detect hCG in urine.

The product is configured in a dipstick format. When the product is partially submerged in urine, the sample travels through the Test Strip. If hCG is present at levels of 25 mIU/ml (W.H.O. 3rd I.S. 75/537) or greater, a positive result, two distinct red bands will appear. If no hCG is present, a negative result, one red band, will appear.

  • E. Intended Use Statement
    Scripps Laboratories hCG One-Step is a rapid, one step immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy. For Laboratory and Professional Use.

  • F. Summary of Similarities to Legally Marketed Devices
    Scripps Laboratories hCG One-Step is substantially equivalent in methodology, intended use, technology and performance characteristics to Quidel® Quick Vue® One-Step hCG Urine Test and to Wyntek Diagnostics OSOM™ hCG-Urine Test. The similarities between the three tests are shown in the following table.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Scripps Laboratories. The word "SCRIPPS" is in a large, bold, sans-serif font, with a dot above the "I". Below that, the word "LABORATORIES" is in a smaller, sans-serif font. The logo is black and white.

Summary of Similarities Between Scripps Laboratories hCG One-Step. Quidel® QuickVue® One-Step hCG Urine Test and Wyntek Diagnostics OSOM™ hCG-Urine Test.

CharacteristicScrippsLaboratories hCGOne-StepQuidel®QuickVue® One-Step hCG UrineTestWyntekDiagnosticsOSOM™ hCG-Urine Test
Intended Use
Diagnostic UseEarly Detection ofPregnancyEarly Detection ofPregnancyEarly Detection ofPregnancy
Use SettingLaboratory andProfessionalLaboratory andProfessionalLaboratory andProfessional
Methodology
SampleUrineUrineUrine
Timing5 minutes3 minutes1 minute
FormatOne-StepOne-StepOne-Step
ResultsInterpretationVisual, QualitativeVisual, QualitativeVisual, Qualitative
Detection Limit25 mIU/ml25 mIU/ml25 mIU/ml
StandardizationWHO 3rd IS 75/537WHO 3rd IS 75/537WHO 3rd IS 75/537
Technology
Assay TypeChromatographicImmunoassayChromatographicImmunoassayChromatographicImmunoassay
AntibodiesMurine Monoclonaland CaprinePolyclonal Anti-hCGMurine Monoclonaland CaprinePolyclonal Anti-hCGMurine Monoclonaland RabbitPolyclonal Anti-hCG
Solid PhaseMembraneMembraneMembrane
Performance
Sensitivity100%99.5%99.6%
Specificity99.6%99.6%99.6%
Accuracy99.8%99.5%99.6%

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Scripps Laboratories. The word "SCRIPPS" is in large, bold, black letters, with a black dot above the "I". Below that, in smaller, black letters, is the word "LABORATORIES". The logo is simple and professional.

  • G. Performance Data
      1. Clinical Accuracy

Scripps Laboratories hCG One-Step was evaluated and compared to a commercially available visual hCG test in a multicenter field study. Five hundred seven urine specimens presented for pregnancy testing were run on Scripps Laboratories hCG One-Step and designated as qualitatively positive or negative for hCG based on the results of Quidel® Quick Vue® One-Step hCG Urine Test. All testing was performed in the intended use setting by individuals with different educational backgrounds and training.

The clinical accuracy of Scripps Laboratories hCG One-Step in the intended use setting was 99.8% when compared to a commercially available visual hCG test. The sensitivity was 100% and the specificity was 99.6%.

Clinical Accuracy of Scripps Laboratories hCG One-Step

Image /page/4/Figure/8 description: The image shows a table titled "Comparative Test" with data related to hCG testing. The table is a 2x2 matrix, with rows and columns labeled '+' and '-'. The cells contain the numbers 238, 1*, 0, and 268. A note indicates that the sample contained 17.9 mIU/ml hCG, and the test was performed by Scripps Laboratories using hCG One-Step.

2. Sensitivity

Urine samples with hCG concentrations of 25 mIU/ml (W.H.O. 3rd I.S. 75/537) or greater yielded positive results on Scripps Laboratories hCG One-Step.

    1. Interference
      The following substances were added to urine containing either 0 mIU/ml or 25 mIU/ml hCG and did not interfere with the performance of Scripps Laboratories hCG One-Step at the concentration indicated:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Scripps Laboratories. The word "SCRIPPS" is in large, bold, sans-serif font, with a dot above the second "I". Below that, in a smaller font, is the word "LABORATORIES". The logo is black and white.

  • Performance Data (continued) G.
    • 3. Interference (continued)
Drug/ChemicalConcentration
Acetaminophen20 mg/dl
Acetylsalicylic Acid20 mg/dl
Ampicillin4 mg/dl
Atropine20 mg/dl
Caffeine20 mg/dl
Ephedrine20 mg/dl
Ethylenediaminetetraacetic Acid80 mg/dl
Gentisic Acid20 mg/dl
Ibuprofen20 mg/dl
Ethanol10%
Phenylpropanolamine HCl4000 mg/dl
Tetracycline20 mg/dl
Urine AnalyteConcentration
Albumin, human2000 mg/dl
Ascorbic Acid20 mg/dl
Bilirubin1 mg/dl
Creatinine200 mg/dl
Estradiol25 ng/ml
Estriol25 ng/ml
Glucose2000 mg/dl
Hemoglobin500 mg/dl
β-Hyroxybutyric Acid100 mg/dl
Pregnanediol1000 ng/ml
Progesterone40 ng/ml
Riboflavin3750 µg/dl
Uric Acid10 mg/dl

4. Cross Reactivity

The addition of 500 mIU/ml LH, 1000 mIU/ml FSH, and 1000 μIU/ml TSH to samples containing either 0 mIU/ml or 25 mIU/ml hCG did not exhibit cross reactivity in Scripps Laboratories hCG One-Step.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Scripps Laboratories. The logo consists of the word "SCRIPPS" in large, bold, sans-serif font, with a small black circle above the "I". Below "SCRIPPS" is the word "LABORATORIES" in a smaller, sans-serif font. The logo is black and white.

  • G. Performance Data (continued)
    • Urine pH న.

Urine samples containing either 0 mIU/ml or 25 mIU/ml hCG adjusted to different pH levels from 4 to 9 did not affect the performance of Scripps Laboratories hCG One-Step.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.