The device is used to anchor suture into bone . The suture is subsequently used by the surgeon to reattach the repositioned/injured soft tissue to bone. The purpose of this 510(k) submission is to obtain clearance for the Mitek Micro Anchor for the lateral canthoplasty and collateral ligament(s) around the (PIP) joint. The intended use is the same as the predicate device, i.e., suture. when used to approximate soft tissue to bone during a period of rehabilitation.

Device Description

The Mitek Micro Anchor is 3.7mm in length by 1.3mm in diameter. It is manufactured from Titanium 6Al 4V and uses a Nitinol arc. It is similar in design to the Mitek GII Mini Anchor. The same methods of manufacturing and assembly are used for the Micro Anchor as those used for manufacturing all other Mitek metal anchors. The Mitek Mini anchor design was first cleared to market by US FDA in K915089 and K930892. The Mitek Micro Anchor delivery system consists of a Mitek Micro Anchor, a drill bit, and an Inserter.

AI/ML Overview

The provided text describes a medical device, the Mitek Micro Anchor, and its non-clinical and clinical testing for FDA 510(k) clearance. However, it does not describe an AI/ML powered device or a study involving an algorithm. Therefore, many of the requested categories in your prompt are not applicable to the given input.

I will provide the information that is available from the text, indicating where information is not present or not relevant to this type of device submission.

Here's the analysis based on the provided text:

Acceptance Criteria and Reported Device Performance

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are not explicitly stated in a quantitative manner as "If X, then Y." Instead, the submission implies acceptance by demonstrating superior strength compared to the predicate device (suture) and by referencing historical data for the identical surgical procedure in clinical tests. The "performance" is primarily shown through pull tests.

Metric (Implicit Acceptance Criteria and Reference)Reported Device Performance (Average Failure Load in lbs, SD)
Predicate Device (USP Class I knot pull synthetic sterilized #4/0 suture)2.64 lbs
Mitek Micro Anchor Performance Comparison:
Anchor pull strength (cadaver hand) - Should be greater than predicate9.67 lbs, SD 1.32
Anchor pull strength (cadaver skull) - Should be greater than predicate7.97 lbs, SD 1.92
Anchor pull strength (pig metacarpals) - Should be greater than predicate7.28 lbs, SD 0.72
Suture with bone tunnels - Compared as another fixation method5.73 lbs, SD 0.46

Note: The intent is to show that the Mitek Micro Anchor's pull strength in various tissues is significantly greater than the standard #4/0 suture's pull strength (2.64 lbs), thus demonstrating its effectiveness. The specific threshold for "acceptance" is not numerically defined beyond this comparative superiority.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Non-Clinical (Pull Tests): Specific sample sizes for each pull test (cadaver hand, skull, pig metacarpals, suture w/ bone tunnels) are not explicitly stated. The "average failure load" suggests multiple tests were performed for each category, but the number of samples per average is missing.
  • Clinical Tests: "clinical trials" are mentioned, conducted in accordance with IDE #G880026. The number of patients or procedures in these trials is not provided.
  • Data Provenance: Not specified (country of origin). The clinical tests were likely prospective as they refer to "clinical trials." The non-clinical tests are laboratory-based.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This device is a surgical implant (anchor), not an AI/ML diagnostic tool. Ground truth in the context of expert review for image interpretation or similar AI applications is not relevant here. The "ground truth" for non-clinical testing is the measured mechanical failure load, and for clinical testing, it's patient outcomes as assessed by the surgeons.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. See point 3.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI/ML device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Non-Clinical: Mechanical failure loads measured directly.
  • Clinical: "Safety and effectiveness was based upon the results obtained in the clinical trials and their equivalence to historical data for the identical surgical procedure." This implies a form of clinical outcomes data (e.g., success of reattachment, absence of complications) compared against historical benchmarks.

8. The sample size for the training set

  • Not Applicable. This is not an AI/ML device. There is no "training set."

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS MITEK MICRO ANCHOR

SUBMITTER:

NAME: Mitek Surgical Products, Inc. ADDRESS: 60 Glacier Drive, Westwood, MA 02090 TEL#: 617-461-9700 CONTACT: Edward F. Kent, Vice President, Regulatory Affairs DATE: June 28, 1996

NAME OF DEVICE

CLASSIFICATION NAME: Staple, Fixation, Bone COMMON NAME: Appliance for reconstruction of bone to soft tissue PROPRIETARY NAME: MITEK MICRO ANCHOR

PREDICATE DEVICE: SUTURE

DESCRIPTION OF DEVICE

FUNCTION

Fixation of USP #4/0 suture below bone surface.

DEVICE DESIGN

The Mitek Micro Anchor is 3.7mm in length by 1.3mm in diameter. It is manufactured from Titanium 6Al 4V and uses a Nitinol arc. It is similar in design to the Mitek GII Mini Anchor. The same methods of manufacturing and assembly are used for the Micro Anchor as those used for manufacturing all other Mitek metal anchors. The Mitek Mini anchor design was first cleared to market by US FDA in K915089 and K930892.

MATERIAL USED

The Mitek Micro Anchor delivery system consists of a Mitek Micro Anchor, a drill bit, and an Inserter.

INTENDED USE

The device is used to anchor suture into bone . The suture is subsequently used by the surgeon to reattach the repositioned/injured soft tissue to bone. The purpose of this 510(k) submission is to obtain clearance for the Mitek Micro Anchor for the lateral canthoplasty and collateral ligament(s) around the (PIP) joint. The intended use is the same as the predicate device, i.e., suture. when used to approximate soft tissue to bone during a period of rehabilitation.

COMPARISON TO PREDICATE DEVICE

It has greater strength than USP #4/0 suture, and is the same "design" as the Mitek Mini Anchor.

{1}------------------------------------------------

K962793

DESCRIPTION OF NON CLINICAL TESTS

The MitekMicro Anchor pull tests from cadaveric and porcine locations produced an average failure load (LBS) of:

1) CADAVER HAND9.67, SD 1.32
2) CADAVER SKULL7.97, SD 1.92
3) PIG METACARPALS7.28, SD 0.72
4) SUTURE W/ BONE TUNNELS5.73, SD 0.46

USP Class I knot pull synthetic sterilized #4/0 suture, double strand, single knot, is 2.64 lbs.

Suture tissue failure for tendinous tissue is 15.75 lbs, and capsular tissue is 21.80 Ibs. Reference: Soft Tissue Fixation to Bone, Daniel B. Robertson, M.D. et al, The American Journal of Sports Medicine, Vol. 14, No. 5, 1986.

DESCRIPTION OF CLINICAL TESTS

The clinical tests were done in accordance with IDE #G880026. Safety and effectiveness was based upon the results obtained in the clinical trials and their equivalence to historical data for the identical surgical procedure.

N/A