K Number
K962762

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-08-16

(31 days)

Product Code
Regulation Number
862.2320
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SSMPD is intended as a general use clinical laboratory gamma radiation counter. The SSMPD is designed for use by operators qualified to measure gamma-ray emissions from radiolabelled reagents within particular energy bands or energy regions of interest (ROI) as part of diagnostic radioassay or radioimmunoassay methods. The standard quantitative units of measure reported are the measured count per minute (cpm) and the calculated disintegrations per minute (dpm).

Device Description

The ssMPD is a twelve sample, bench-top gamma radiation counter that can be used to measure counts per minute and calculate disintegrations per minute emitted by multi-photon emission isotopes, including 125I. In volumes of 0.05 to 2.0 mls contained in 12 mm x 75 mm or 13 mm x 100 mm sample tubes disintegrations are measured over the range of 1 to 106 dpms. The sample tubes must be capped to minimize the possibility of contaminating the detector chamber. The full detector assembly is a twin detector system consisting of two scintillator crystals--each coupled to a high resolution planar photomultiplier. The entire detector is encased in a composite lead-tin-copper shield to minimize environmental background. In addition to permitting efficient counting of gamma emitters, the sample holder acts as a separator to reduce crosstalk within the twin scintillator detector assembly. The complete SSMPD system is comprised of a standalone SSMPD connected directly to a personal computer (PC) configured with SSMPD read-out electronics, MPDATA™ control and data logger software, and optionally, a printer.

AI/ML Overview

The BioTraces Inc. SEQUENTIAL SAMPLE MULTI-PHOTON DETECTOR (SSMPD) is a gamma scintillation counter intended for general use in clinical laboratories to measure gamma-ray emissions from radiolabelled reagents.

Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

1. Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state quantitative acceptance criteria or report specific performance metrics in a structured table. Instead, it focuses on describing the device's capabilities and comparing them to a predicate device.

However, based on the Description and Intended Use sections, we can infer some general performance expectations:

Acceptance Criteria (Inferred)Reported Device Performance (Inferred)
Measurement Range: Accurate measurement of disintegrations per minute (dpm) for multi-photon emission isotopes.Disintegrations are measured over the range of 1 to 10^6 dpms.
Isotope Compatibility: Ability to measure common multi-photon emission isotopes.Includes 125I. (No other specific isotopes are mentioned, but "multi-photon emission isotopes" implies broader capability).
Sample Volume Compatibility: Ability to accommodate various sample volumes.In volumes of 0.05 to 2.0 mls contained in 12 mm x 75 mm or 13 mm x 100 mm sample tubes.
Output Units: Standard quantitative units of measure.Reports measured count per minute (cpm) and calculated disintegrations per minute (dpm).
Background Minimization: Effective reduction of environmental background noise.The full detector assembly is a twin detector system consisting of two scintillator crystals, each coupled to a high resolution planar photomultiplier. The entire detector is encased in a composite lead-tin-copper shield to minimize environmental background. The SSMPD uses a 2" primary lead shield and a secondary composite lead-tin-copper shield to minimize backscattering events contributed by the lead itself. The SSMPD performs pulse shape as well as pulse height analysis in software and hardware to increase the specificity of pulse acceptance and rejection and lower the environmental background.
Crosstalk Reduction: Minimization of signal interference within the detector.The sample holder acts as a separator to reduce crosstalk within the twin scintillator detector assembly.
Clinical Use Appropriateness: Suitable for use in diagnostic radioassay or radioimmunoassay methods.Intended as a general use clinical laboratory gamma radiation counter. Designed for use by operators qualified to measure gamma-ray emissions from radiolabelled reagents within particular energy bands or energy regions of interest (ROI) as part of diagnostic radioassay or radioimmunoassay methods.
Substantial Equivalence: Comparable to a predicate device in specifications and principles of operation.Substantially equivalent to the Gamma 5500 Counting System (K770088).

Note: The document primarily focuses on describing the technical features and intended use of the SSMPD and establishing substantial equivalence to a predicate device. It does not present a formal study with detailed statistical results to "prove" the device meets acceptance criteria. Instead, the argument for effectiveness is made through the device's technical specifications and a comparison to a previously cleared device.

2. Sample Size and Data Provenance for the Test Set

The provided text does not describe an explicit "test set" in the context of a performance study with a defined sample size or data provenance. The document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting raw clinical or performance data from novel studies.

The detailed descriptions of the device's capabilities (e.g., measuring dpms from 1 to 10^6, handling 125I) are likely derived from internal engineering evaluations and characterization studies performed by BioTraces Inc. However, no specific details about these studies (sample size, country of origin, retrospective/prospective nature) are provided in this summary.

3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

Because no explicit "test set" and a corresponding independent performance study are detailed, there is no information provided regarding experts used to establish ground truth for such a test set.

4. Adjudication Method (Test Set)

Given the lack of a described test set with human assessment, there is no adjudication method mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The provided text does not describe a multi-reader multi-case (MRMC) comparative effectiveness study. The SSMPD is a standalone gamma counter, meaning its performance is based on measurements of radioactivity rather than interpretation of images by human readers. Therefore, the concept of human readers improving with or without AI assistance is not applicable to this type of device.

6. Standalone (Algorithm Only) Performance Study

The entire description of the SSMPD can be considered a description of its standalone performance (i.e., algorithm/device only performance without human-in-the-loop performance). The device itself performs the measurement and calculation of cpm and dpm. The details provided, such as the measurement range of 1 to 10^6 dpms and the ability to detect 125I, characterize its standalone capabilities. However, a formal "study" with statistical results for standalone performance is not detailed in this summary.

7. Type of Ground Truth Used

For the characterization of the SSMPD's performance, the "ground truth" would be the known radioactivity levels or concentrations of the radiolabelled samples used during the device's development and testing. This would typically involve:

  • Reference Standards: Samples with precisely known activities or disintegrations per minute (dpm) measured using established gold-standard methods.
  • Known Isotope Samples: Samples containing specific isotopes (like 125I) at known concentrations to verify detection and energy windowing.

The text implies that the ground truth for its performance claims (e.g., 1 to 10^6 dpms measurement range) was established through such physical and chemical reference standards, but it does not explicitly state this.

8. Sample Size for the Training Set

The concept of a "training set" is typically associated with machine learning or AI algorithms where a model learns from data. While the SSMPD uses "pulse shape as well as pulse height analysis in software and hardware," which involves signal processing, it's not explicitly framed as a machine learning model requiring a distinct training set in the conventional sense.

Therefore, the provided text does not specify a "sample size for the training set."

9. How the Ground Truth for the Training Set Was Established

As there is no explicit mention of a "training set" in the context of a machine learning algorithm, there is no information provided on how ground truth for such a training set was established. Any internal calibration or tuning of the pulse analysis algorithms would have relied on the same fundamental principles of known radioactivity standards as described in point 7.

{0}------------------------------------------------

K962762

-17000

SAFETY AND EFFECTIVENESS SUMMARY IX.

BioTraces Inc.

10517-A West Drive Fairfax, VA 22030

tel: (703) 273-6941 Office: fax: (703) 273-6968

Premarket Notification [510(k) ] Summary

Summary submitted by:Address:Telephone:Facsimile:Contact person:Summary prepared on:E. James Wadiak, PresidentBIOTRACES INC., 10517-A West Drive, Fairfax, VA, 22030(703) 273-6941(703) 273-6968E. James Wadiak, PresidentJuly 8, 1996
Proprietary Name:Common/Usual Name:Classification Name:SEQUENTIAL SAMPLE MULTI-PHOTON DETECTOR (SSMPD)Gamma Scintillation CounterCounter (Beta, Gamma) For Clinical Use
Substantial Equivalence:The ssMPD is substantially equivalent in specifications andprinciples of operation to the Gamma 5500 Counting System(originally listed as Gamma 4000 Scintillation Counter),510(k) Number: K770088, manufactured by BeckmanInstruments, Inc., Scientific Instruments Division, Irvine,California 92713.
Description:The ssMPD is a twelve sample, bench-top gamma radiationcounter that can be used to measure counts per minute andcalculate disintegrations per minute emitted by multi-photonemission isotopes, including 125I. In volumes of 0.05 to 2.0mls contained in 12 mm x 75 mm or 13 mm x 100 mm sampletubes disintegrations are measured over the range of 1 to 106dpms. The sample tubes must be capped to minimize thepossibility of contaminating the detector chamber.
The full detector assembly is a twin detector system consistingof two scintillator crystals--each coupled to a high resolutionplanar photomultiplier. The entire detector is encased in a

Section IX - 1

{1}------------------------------------------------

composite lead-tin-copper shield to minimize environmentalbackground. In addition to permitting efficient counting ofgamma emitters, the sample holder acts as a separator toreduce crosstalk within the twin scintillator detector assembly.
The complete SSMPD system is comprised of a standaloneSSMPD connected directly to a personal computer (PC)configured with SSMPD read-out electronics, MPDATA™control and data logger software, and optionally, a printer.
Intended Use:The SSMPD is intended as a general use clinical laboratorygamma radiation counter. The SSMPD is designed for use byoperators qualified to measure gamma-ray emissions fromradiolabelled reagents within particular energy bands or energyregions of interest (ROI) as part of diagnostic radioassay orradioimmunoassay methods. The standard quantitative unitsof measure reported are the measured count per minute (cpm)and the calculated disintegrations per minute (dpm).
Device Comparison:Both the SSMPD and the Gamma 5500 Counting Systemutilize a 2" inorganic scintillating crystal, a photomultipliertube (PMT), various electronic systems to amplify, scale, andanalyze pulse heights, and a variety of output formats anddevices for display of results to quantify gamma emissions in

a sample.

However, the SSMPD uses two identical CaF2(Eu) scintillator crystals with PMT read-outs as a detector whereas the Gamma 5500 Counting System uses one Nal(T)) scintillator crystal with PMT read-out. Both inorganic scintillator crystals have emission maximums that are well matched to the sensitivity of the bialkali photocathode PMTs used in the ssMPD and a refractive index that is close to that of most PMT windows and optical coupling compounds. The photoelectron yield of Nal(Tl) produces the highest signal of all known scintillators and is twice that of CaF2(Eu). However, the light yield of CaF2(Eu) is easily within an acceptable range of performance.

NaI(Tl) is susceptible to mechanical, thermal, and radiation damage. It is hydroscopic and colors over time and must be housed in a sealed environment. It is toxic. In addition, it is soft and cannot be machined easily. CaF (Eu) is not

{2}------------------------------------------------

susceptible to mechanical, thermal, and radiation da is transparent and insoluble in water and mos solvents, permitting radioactive samples in soluti placed in direct contact with the crystal. It is nonaddition, it is hard and can be machined easily.

The ssMPD performs pulse shape as well as pul analysis in software and hardware to increase the spe pulse acceptance and rejection and lower the envir background. The Gamma 5500 Counting System pulse height analysis.

The ssMPD uses a 2" primary lead shield and a s composite lead-tin-copper shield to minimize back events contributed by the lead itself. The Gamm Counting System utilizes 1" of lead shielding secondary composite shield.

§ 862.2320 Beta or gamma counter for clinical use.

(a)
Identification. A beta or gamma counter for clinical use is a device intended to detect and count beta or gamma radiation emitted by clinical samples. Clinical samples are prepared by addition of a radioactive reagent to the sample. These measurements are useful in the diagnosis and treatment of various disorders.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.