K Number
K962580

Validate with FDA (Live)

Date Cleared
1996-08-29

(59 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EZ Sure OneStep Pregnancy Test is Excel Scientific, Inc.'s name for OTC use of solid phase migratory chromatography immunoassay in vitro diagnostic test kit for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine.

Device Description

EZ Sure OneStep Pregnancy Test is a sandwich solid phase dyeconjugated non-enzyme immunoassay. When the specimen is added to the reaction device, the gold-sol-antibody conjugate binds to the hCG molecules in the specimen forming an antibody-antigen complex. This complex binds to the anti-hCG antibody immobilized in the position zone in the viewing window and produces a purple-pink color band when the hCG concentration is at or greater than 25 mIU/mL. In the absence of detectable levels of hCG, there is no purple-pink band in the positive reaction mixture continues to flow through the absorbent device, pass the positive reaction zone, producing a purple-pink color band in the control zone, demonstrating that the reagents and the device are functioning correctly.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
AccuracyHigh agreement with predicate device (BeSure OneStep) and ability to correctly identify hCG < 25 mIU/mL as negative and >= 25 mIU/mL as positive.100% agreement with predicate device (90 positive, 70 negative). Identifies 0 & 20 mIU/mL as negative. Identifies 25, 100, 450,000 mIU/mL as positive.
SensitivityDetect hCG at 25 mIU/mL or greater.Can detect hCG at 25 mIU/mL or greater.
SpecificityNo false positives from menopausal urine or common interfering substances.20 postmenopausal urine samples tested negative. No interference from 12 common substances (e.g., Acetaminophen, Glucose, Hemoglobin, Hormones like hTSH, hLH, hFSH) at specified concentrations.
Precision (Intra-Assay/Within Run)Consistent and correct identification of negative and positive samples within a single run.100% correct identification of negative (0 mIU/mL) and positive (25, 100, 450,000 mIU/mL) values in 11 replicates.
Precision (Inter-Assay/Between Run)Consistent and correct identification of negative and positive samples across multiple runs and different lots.100% correct identification of negative (0 mIU/mL) and positive (25, 100, 450,000 mIU/mL) values in 11 independent assays using 3 different lots over a two-month period.

2. Sample Size Used for the Test Set and Data Provenance

  • Accuracy Study:

    • Sample Size: A total of 100 positive urine specimens. The table shows 90 positive and 70 negative results, implying a total of 160 specimens were included in the comparison with the predicate device.
    • Data Provenance: Not explicitly stated, but the context of the K962580 premarket notification suggests it's likely a retrospective study conducted by the manufacturer, Excel Scientific, Inc., or a contract lab. There is no information on the country of origin of the data.
  • Sensitivity (Known amounts of hCG added to urine pool): Test was done on one urine pool spiked with various concentrations of hCG.

  • Specificity (Menopausal Urines):

    • Sample Size: 20 urine specimens from postmenopausal women.
    • Data Provenance: Not explicitly stated, but likely a retrospective study.
  • Specificity (Interfering Substances): Test was done on urine samples (0 and 25 mIU/mL hCG) spiked with specified substances.

  • Precision Studies:

    • Sample Size: 11 replicates for intra-assay precision, and 11 independent assays with 3 different lots for inter-assay precision. Four specimens (0, 25, 100, 450,000 mIU/mL hCG) were used for each.
    • Data Provenance: Not explicitly stated, likely an internal study by the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

There is no mention of experts being used to establish the ground truth for this device. The ground truth appears to be established by:

  • Comparison to a predicate device (Syntron's BeSure OneStep hCG Test) for the accuracy study.
  • Spiking urine with known concentrations of hCG for sensitivity and some specificity testing.
  • The known status of urine samples (e.g., "postmenopausal women" for specificity).

4. Adjudication Method for the Test Set

No adjudication method is mentioned. The comparisons appear to be direct, without a separate review or consensus process for the test results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

No MRMC study was done, nor is this an AI-driven device. This is an in-vitro diagnostic test kit.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is a standalone in-vitro diagnostic test kit that produces a visual result (color band). The "standalone" performance is the core of the studies described. It's not an algorithm, but a chemical immunoassay.

7. The Type of Ground Truth Used

The ground truth used in these studies includes:

  • Reference Device Comparison: Syntron's BeSure OneStep hCG Test for accuracy.
  • Known Concentrations: hCG spiked into urine pools at specific mIU/mL values for sensitivity.
  • Biological State: Urine from "postmenopausal women" to assess specificity against other hormones.
  • Spiked Interfering Substances: Urine samples with known concentrations of potential interfering substances to assess specificity.

8. The Sample Size for the Training Set

There is no mention of a training set as this is not a machine learning or AI-based device. The device's "training" in a general sense would be the research and development phase where the assay components are optimized, but specifics of data used for that are not provided.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no explicitly defined "training set" in the context of an immunoassay like there would be for a machine learning model. The development of the assay would involve laboratory optimization to achieve the desired performance characteristics (sensitivity, specificity) against known standards and samples.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "EXCEL" in a bold, sans-serif font inside a black rounded rectangle. Below the rectangle, the phrase "Commitment to Excellence" is written in a smaller, italicized font. The word "EXCEL" is the primary visual element, emphasizing the brand or company name.

Attachment #5

K962580

13

AUG 29 1996

Summary of Safety and Effectiveness

EZQ Sure OneStep Pregnancy Test

EZ $ Sure OneStep Pregnancy Test is Excel Scientific, Inc.'s name for OTC use of solid phase migratory chromatography immunoassay in vitro diagnostic test kit for the qualitative detection of human Chorionic Gonadotropin (hCG) in urine. EZP Sure OneStep Pregnancy Test is based on comparative data with BeSure OneStep immuncassay, an assay for human Chorionic Gonadotropin currently being marketed.

The Principle of EZQ Sure OneStep Prognancy Test is as follows:

    1. Both the EZ9 Sure OneStep Pregnancy Test and the BeSure OneStep Prograncy Test are a sandwich solid phase dyeconjugated non-enzyme immunoassay. When the specimen is added to the reaction device, the gold-sol-antibody conjugate binds to the hCG molecules in the specimen forming an antibody-antigen complex. This complex binds to the anti-hCG artibody immobilized in the position zone in the viewing window and produces a purple-pink color band when the hCG concertration is at or greater than 25 mlUmL. In the absence of detectable levels of hCG, there is no purple-pink band in the positive reaction mixture continues to flow through the absorbent device, pass the positive reaction zone, producing a purple-pink color band in the control zone, demonstrating that the reagents and the device are functioning correctly.
      The Differences Between the EZQ Sure OneStep Pregnancy Test and the BeSure OneStep Pregnancy Test are as follows .
BeSure OneStepPregnancy TestEZ♀ Sure OneStep EarlyPregnancy Test
Specimen Volume6 drops4 drops of urine
Incubation5 minutes5 minutes
ReagentNoNo
Stepone stepone step
Cutoff25 mIU/mL25 mlU/mL
Urine CupYesYes
    1. For questions concerning the safety and effectiveness of the EZQ Sure OneStep Prognancy Test, please contact:
      Min-Lee Cheng, Ph.D. 23071 E. La Palma Avenue Yorba Linda, CA 92887 Tel: (714) 692-7151

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "EXCEL" in a bold, sans-serif font, set against a black, rounded rectangle. Below the rectangle, the phrase "Commitment to Excellence" is written in a smaller font, with quotation marks around the phrase. The text is in black and the background is white.

Performance Characteristics

Curacy :

A study was performed using a total 100 positive urine specimens. These specimens were assayed with EZ? Sure OneStep Early Pregnancy Test and Syntron's BeSure test according to the packaging insest.

EZ Sure OneStep Early Pregnancy Test
(Positive)(Negative)
Syntron's BeSure(Positive)900
OneStep hCG Test(Negative)070

The above data do not include two specimens which were weakly reactive with both BeSure OneStap hCG and EZ Sure OneStep Early Pregnancy Tests. The reactions of these specimens were considered to be includive). The accuracy study was also performed by determining the qualitative recovery of known amounts of hCG added to the urine pool.

Concentration of hCG in theUrine Pool (miU/mL)ExpectedObserved
0negativeone band (negative)
20negativeone band (negative).
25positivetwo bands (positive)
100positivetwo bands (positive)
450,000positivetwo bands (positive)

The EZ9 Sure OneStep Early Prognancy Test can detect hCG in urine sample at concertration of 25 mill/mL or greater. This senativity has been determined against the World Health Organization Third International Standard. In some cases, samples containing less than 25 mlU/mL of hCG can also test as positive.

Specificity :

Menopousal Urines :

A study was conducted using urine speciment from 20 postmenspeased women. These speciments were chosen because unhe from postmenopausal women frequently interfore with prograncy tests due to cross reactivity with other gonadoropin hormones. All twenty urine specimens were negative when teeted with the EZ? Sure OneStep Early Pregnancy Test.

Potentially interfering substances were added to urine samples which had hCG invels of 0 and 25 mlUmi. In each case. no interference was observed when these spiked urine samples were tested with the EZP Sure OneStep Early Prograncy Tost.

SubstanceConcentration in Urine
Acetaminophen20 mg/dl
Acetylsalicylic Acid20 mg/dl
Ampicillin20 mg/dl
Ascorbic Acid20 mg/dl
Atropine20 mg/dl
Caffeine20 mg/dl

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "EXCEL" in a bold, sans-serif font, enclosed in a rounded rectangle. Below the rectangle, the phrase "Commitment to Excellence" is written in a smaller, italicized font. The text and the rectangle are black, contrasting with what is presumably a white background.

Gentisic Acid20 mg/dl
Glucose2,000 mg/dl
Hemoglobin1 mg/dl
Human Serum Proteins2,000 mg/dl
Tetracycline20 mg/dl
Uric Acid10 mg/dl

Homologous Hormones at the concentrations indicated below were found not to interfere with the EZP Sure OneStep Early Pregnancy Test either:

Homologous HormonesConcentration in UrineReference
hTSH1,000 $ \mu $ IU/mLWHO 68/38
hLH500 mIU/mLWHO 2nd IRP HMG
hFSH1,000 mIU/mLWHO 2nd IRP HMG

Standardization :

The reference preparation and controls have been calibrated as follows :

1.0 mlU/mL Excel = 1.0 mlU/mL 3rd I.S.

ecision :

Intra-Assay : Within run precision was determined using 11 replicates of four specimens containing 0, 25, 100 and 450,000 mlU/mL hCG. The negative and positive values were correctly identified 100% of the time.

Inter-Assay : Between run precision was determined using the same four specimens of 0, 25, 100 and 450,000 mIU/mL of hCG in 11 independent assays, and with three different lots of reaction devices over a two-month period. Again, the negative and positive values were correctly identified 100% of the time.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.