K Number
K962472

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-08-27

(63 days)

Product Code
Regulation Number
876.1620
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rüsch Urodynamic Rectal Balloon 2-way Catheter is intended to provide a fluid conduit for urodynamic studies, where the detrusor pressure is to be calculated from rectal pressure.

Device Description

The Rusch Urodynamic Rectal Balloon 2-way Catheter consists of a double lumen polyvinyl chloride tube upon which is affixed a latex balloon, with a radiopaque stripe.

To use, the balloon is inflated with sterile water using a luer taper syringe. The catheter is then connected to a transducer, and can be used in conjunction with equipment for calculating the detrusor pressure necessary to open the urinary sphincter).

AI/ML Overview

The provided text describes a 510(k) summary for the Rüsch Urodynamic Rectal Balloon 2-way Catheter. However, it does not contain information about acceptance criteria or specific study results involving a device's performance against such criteria in the manner requested. Instead, it focuses on the device's characteristics, comparison to predicate devices, and a general statement about biocompatibility and bench testing without numerical performance data or detailed study methodology.

Therefore, I cannot fulfill the request for a table of acceptance criteria and reported device performance, nor details about sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies, as this information is not present in the provided text.

The closest information available that might be interpreted in the context of "performance" is the statement:

"The device's performance in these tests [Tensile Testing, Balloon Burst Testing] is comparable with that of the Rüsch Urodynamic Rectal Balloon Catheter (K952573)."

This indicates that some form of performance evaluation was done through bench testing, and the outcome was "comparable" to a predicate device. However, it does not provide the specific numerical acceptance criteria for these tests or the reported numerical performance of the device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a handwritten string of alphanumeric characters. The string appears to be "K962472". The characters are written in a bold, black font, and the handwriting is somewhat stylized. The image is a close-up of the string, and the background is white.

AUG 27 1996

Image /page/0/Picture/2 description: The image shows the logo for RUSCH, a company specializing in international group regulatory affairs. The logo features the company name in bold, sans-serif font, with the word "RUSCH" appearing prominently at the top. Below the company name, the words "INTERNATIONAL Group Regulatory Affairs" are written in a smaller font. The text "A Subsidiary of Teleflex Incorporated (USA)" is at the bottom of the logo.

all Pines Park ffrev, NH 03452 603) 532-7706 FAX (603) 532-8211 or 6108

510(k) SUMMARY

[As required by 21 CFR 807.92]

1. Submitter and Contact Person

Neil R. Armstrong Group Regulatory Affairs Manager Rüsch International Tall Pines Park Jaffrey NH 03452 (603) 532 7706 Tel: (603) 532 8211 Fax:

Device Name 2.

Trade Name Rüsch Urodynamic Rectal Balloon 2-way Catheter

Common Name Urodynamic Rectal 2-way Catheter

Classification Name Accessory to Urodynamics Measurement System 78 FAP; 21 CFR 876.1620; Class II medical device

3. Comparison Devices

Rüsch Urodynamic Rectal Balloon Catheter - K952573 Porges Urodynamic Rectal Balloon Catheter - Preamendment Cook Urological Urodynamic Rectal Balloon Catheter - K912451

4. Description of Device

The Rusch Urodynamic Rectal Balloon 2-way Catheter consists of a double lumen polyvinyl chloride tube upon which is affixed a latex balloon, with a radiopaque stripe.

To use, the balloon is inflated with sterile water using a luer taper syringe. The catheter is then connected to a transducer, and can be used in conjunction with equipment for calculating the detrusor pressure necessary to open the urinary sphincter).

{1}------------------------------------------------

Intended Use ર .

The Rüsch Urodynamic Rectal Balloon 2-way Catheter is intended to provide a fluid conduit for urodynamic studies, where the detrusor pressure is to be calculated from rectal pressure.

Summary of Technological Characteristics 6.

The Rusch Urodynamic Rectal Balloon 2-way Catheter is manufactured of the same materials and by the same processes (including sterilization) as the Rüsch Urodynamic Rectal Balloon Catheter - K952573.

The Rüsch Urodynamic Rectal Balloon 2-way Catheter has a double lumen shaft design similar to the Porges Urodynamic Rectal Balloon Catheter (Preamendment) and the Cook Urological Urodynamic Rectal Balloon Catheter (K912451).

7. Summary of Performance Data

Biocompatibility testing has been completed based on the ISO 10993 matrix modified by FDA Blue Book Memo #G95/1.

Laboratory bench testing has been completed to section #6 of the FDA "510(k) Checklist for Urodynamic Catheters":

Tensile Testing Balloon Burst Testing

The device's performance in these tests is comparable with that of the Rüsch Urodynamic Rectal Balloon Catheter (K952573).

§ 876.1620 Urodynamics measurement system.

(a)
Identification. A urodynamics measurement system is a device used to measure volume and pressure in the urinary bladder when it is filled through a catheter with carbon dioxide or water. The device controls the supply of carbon dioxide or water and may also record the electrical activity of the muscles associated with urination. The device system may include transducers, electronic signal conditioning and display equipment, a catheter withdrawal device to enable a urethral pressure profile to be obtained, and special catheters for urethral profilometry and electrodes for electromyography. This generic type of device includes the cystometric gas (carbon dioxide) device, the cystometric hydrualic device, and the electrical recording cystometer, but excludes any device that uses air to fill the bladder.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.