K Number
K962404

Validate with FDA (Live)

Date Cleared
1997-01-17

(210 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use statement for the device is to measure heart rate, respiration rate, invasive pressure, non-invasive pressure, arrhythmia(only for adults), temperature, SpO2 and central apnea. The device will produce visual and aural alarms if any of the above parameters vary beyond preset limits and produce timed or alarm recordings.

Device Description

The SC 6000 and SC 6000P Portable patient monitors with neonatal monitoring enhancement are software enhanced versions of the SC 6000 and SC 6000P Portable patient monitors. The software modifications have been undertaken to expand the Intended Patient Population from pediatric and adults to also include neonatal patients.

AI/ML Overview

The provided text describes the Siemens SC 6000 & SC 6000 P Neonatal Monitoring Enhancement, which is a software upgrade to expand the intended patient population of existing monitors to include neonatal patients. The filing is a 510(k) submission, comparing the enhanced device to a predicate device (Siemens SIRECUST 1261).

Here's an analysis of the provided information against your requested criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here are implicitly the specifications of the predicate device, as the submission aims to demonstrate substantial equivalence. The reported device performance is the specifications of the SC 6000 & SC 6000 P.

ParameterAcceptance Criteria (Predicate Device: SIRECUST 1261 Specification)Reported Device Performance (SC6000 & SC6000 P Specification)
ECG & Heart Rate
Available leadsI, II, III, aVR, aVF, aVL, VI, II, III, aVR, aVF, aVL, V
Measuring range15 - 300 bpm15 - 300 bpm
Accuracy± 10% for 15 - 300 bpm± 5% for 15 - 200 bpm± 8% for 201 - 300 bpm
Respiration
MethodImpedance pneumographyImpedance pneumography
Measuring range2 - 155 breaths per min.2 - 155 breaths per min.
Measuring accuracy± 3 bpm± 3 bpm
Apnea Detection?YesYes
SpO2
Measuring methodAbsorption-spectrophotometryAbsorption-spectrophotometry
Measuring range1 - 100%1 - 100%
Pulse Rate30 - 300 bpm30 - 300 bpm
SpO2 Accuracy (Range 70 - 100%)± 2%± 2%
SpO2 Accuracy (Range 0 - 69%)not specifiednot specified
Pulse Rate Accuracy± 10%± 10%
Temperature
Measurement Range-5° C to 50° C0° C to 50° C
Accuracy (Range 30 - 50°C)± 0.1°C± 0.1°C
Accuracy (Range 0 - 30°C)± 0.2°C± 0.2°C
Non Invasive Blood Pressure (NBP)
Parameter displaySystolic, Diastolic, MeanSystolic, Diastolic, Mean
Measuring methodOscillometric techniqueOscillometric technique
Measurement range (heart rate)40-240 bpm40-240 bpm
Measurement range (systolic)30-130 mmHg30-130 mmHg
Measurement range (mean)20-110 mmHg20-110 mmHg
Measurement range (diastolic)10-100 mmHg10-100 mmHg
Initial inflation pressure110 mmHg +/- 15 mmHg110 mmHg +/- 15 mmHg
Static cuff accuracy± 3 mmHg± 3 mmHg
Invasive Blood Pressure (IBP)
Measuring methodresistive strain gauge transducerresistive strain gauge transducer
Measuring range-50 to + 399 mmHg-50 to + 399 mmHg
Accuracy (exclusive of transducer)± 2 mmHg± 2 mmHg
Transducer specificationsSiemens-approved transducers with a resistance of 200 to 3000 Ω and an equivalent pressure sensitivity of 5uV/V/mmHg + 10%Siemens-approved transducers with a resistance of 200 to 3000 Ω and an equivalent pressure sensitivity of 5uV/V/mmHg ± 10%

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The provided text mentions "validation results and other information in this submission" but does not specify the sample size, data provenance (country of origin), or whether the data was retrospective or prospective for any test sets used.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not provide any information regarding the number of experts used, their qualifications, or how ground truth was established for a test set. This type of information is typically related to clinical trials or performance testing involving human interpretation, which is not detailed here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not provide any information about an adjudication method. This is relevant for studies where multiple experts assess data, and discrepancies need to be resolved.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a patient monitoring system, not an AI-assisted diagnostic tool that human readers would use to interpret cases. The submission is for a software enhancement to expand the patient population, focusing on technical specifications for vital sign monitoring.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The provided text describes a "software enhanced version" of existing patient monitors. The performance metrics presented (e.g., accuracy of heart rate, SpO2, blood pressure) are inherently "standalone" in the sense that the device outputs these measurements automatically. The "validation results" mentioned implicitly refer to testing the device's ability to accurately measure these parameters without human intervention in the measurement process itself. However, the document does not explicitly describe a study designed to evaluate standalone algorithm performance in isolation from the hardware.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Given that this is a patient monitoring device, the ground truth for parameters like heart rate, respiration rate, SpO2, and blood pressure would typically be established using reference medical devices or gold standard measurement techniques. For example:

  • Heart Rate/Respiration: Manually counted by trained personnel, or derived from reference ECG/respiration waveforms.
  • SpO2: Measured by a highly accurate reference oximeter or blood gas analysis.
  • Blood Pressure: Measured directly via arterial line (for invasive) or by a validated reference oscillometric or auscultatory method (for non-invasive).
    However, the document does not explicitly state the type of ground truth used for its validation. It refers to "validation results," but doesn't detail the methodology.

8. The sample size for the training set

The document does not mention any training set or its sample size. This is common for 510(k) submissions for non-AI devices, where performance is often evaluated against engineering specifications and predicate device equivalence rather than machine learning model training.

9. How the ground truth for the training set was established

As no training set is mentioned (see point 8), there is no information on how its ground truth was established.

{0}------------------------------------------------

K962404

JAN 17 1997

510(k) SC 6000 & SC 6000 P Neonatal Monitoring Enhancement

Summary & Certification Section 2:

Summary per 807.92(c) 510 (k)

Subscribers Name & Address: (2)

Siemens Medical Systems, Inc. Electromedical Group 16 Electronics Avenue Danvers, MA 01923

Tele: 508 750 7500 Fax: 508 777 3398

Official Correspondent: Robert W. Becker Contact Person for this submission: Mark E. Kolnsberg

  • (3) Trade Name:

Siemens SC 6000 & SC 6000 P Portable Bedside Monitoring System

Common NameClassificationNumberClassRegulation Number
Cardiac monitor74DRTII21 CFR 870.2300
Pulse rate monitor74BWSII21 CFR 870.2300
Pulse oximeter74DQAII21 CFR 870.2700
Breathing frequencymonitor73BZQII21 CFR 868.2375
Clinical electronicthermometer80BWXII21 CFR 880.2910
Indwelling bloodpressure monitor74CAAII21 CFR 870.1110
Noninvasive bloodpressure monitor74DXNII21 CFR 870.1130
Heart Rate Monitor,Neonatal74FLOII21 CFR 870.2300
Monitor Bloodpressure, Neonatal,Invasive74FLPII21 CFR 870.1110
Arrhythmia detector& Alarm74DSIIII21 CFR 870.1025

Common name. Classification number. Class & Regulation Number:

(4) Predicate Device Identification:

Siemens SIRECUST 1261 granted premarket approval under 510 K file number K952054.

{1}------------------------------------------------

510(k) SC 6000 & SC 6000 P Neonatal Monitoring Enhancement

Device Description: (ર)

The SC 6000 and SC 6000P Portable patient monitors with neonatal monitoring enhancement are software enhanced versions of the SC 6000 and SC 6000P Portable patient monitors. The software modifications have been undertaken to expand the Intended Patient Population from pediatric and adults to also include neonatal patients.

Intended Use of the Device: (6)

The intended use statement for the device is to measure heart rate, respiration rate, invasive pressure, non-invasive pressure, arrhythmia(only for adults), temperature, SpO2 and central apnea. The device will produce visual and aural alarms if any of the above parameters vary beyond preset limits and produce timed or alarm recordings.

Summary of technological characteristics of Device and Predicate Device: (7)

The comparison of intended use and technological features of these devices to other legally marketed devices taken together with the validation results and other information in this submission indicate that these devices are substantially equivalent to legally marketed predicated devices in safety, effectiveness and intended use.

ParameterSC6000 & SC6000 P SpecificationPredicate Device:SIRECUST 1261 Specification((When equipped with Large IntegratedModule(LIM))
ECG & Heart Rate
Available leads:I, II, III, aVR, aVF, aVL, VI, II, III, aVR, aVF, aVL, V
Measuring range:15 - 300 bpm15 - 300 bpm
Accuracy:± 5% for 15 - 200 bpm± 8% for 201 - 300 bpm± 10% for 15 - 300 bpm
Respiration
Method:Impedance pneumographyImpedance pneumography
Measuring range:2 - 155 breaths per min.2 - 155 breaths per min.
Measuring accuracy:± 3 bpm± 3 bpm
Apnea Detection?:YesYes
SpO2
Measuring method:Absorption-spectrophotometryAbsorption-spectrophotometry
Measuring range:1 - 100%1 - 100%
SpO2:1 - 100%1 - 100%
Pulse Rate:30 - 300 bpm30 - 300 bpm
SpO2 Accuracy:
Range 70 - 100%± 2%± 2%
Range 0 - 69%:not specifiednot specified
Pulse Rate Accuracy:± 10%± 10%
Temperature

{2}------------------------------------------------

510(k) SC 6000 & SC 6000 P Neonatal Monitoring Enhancement


产品 - - - - - -

ParameterSC6000 & SC6000 P SpecificationPredicate Device:SIRECUST 1261 Specification((When equipped with Large IntegratedModule(LIM))
Measurement Range:0° C to 50° C-5° C to 50° C
Accuracy:
Range 30 - 50°C± 0.1°C± 0.1°C
Range 0 - 30°C± 0.2°C± 0.2°C
Non Invasive Blood Pressure (NBP)
Parameter display:Systolic, Diastolic, MeanSystolic, Diastolic, Mean
Measuring method:Oscillometric techniqueOscillometric technique
Measurement range:
heart rate:40-240 bpm40-240 bpm
systolic pressure30-130 mmHg30-130 mmHg
mean pressure20-110 mmHg20-110 mmHg
diastolic pressure10-100 mmHg10-100 mmHg
Initial inflation pressure:110 mmHg +/- 15 mmHg110 mmHg +/- 15 mmHg
Static cuff accuracy:± 3 mmHg± 3 mmHg
Invasive Blood Pressure (IBP)
Measuring method:resistive strain gauge transducerresistive strain gauge transducer
Measuring range:-50 to + 399 mmHg-50 to + 399 mmHg
Accuracy:± 2 mmHg exclusive of transducer± 2 mmHg exclusive of transducer
After transducer zeroing
TransducerspecificationsSiemens-approved transducers with aresistance of 200 to 3000 Ω and anequivalent pressure sensitivity of5uV/V/mmHg ± 10%Siemens-approved transducers with aresistance of 200 to 3000 Ω and anequivalent pressure sensitivity of5uV/V/mmHg + 10%

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.