K Number
K961985

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-09-05

(107 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elect Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The Elect System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor effectiveness of diabetes control.

Device Description

The Elect Blood Glucose Monitoring System comprises a glucose reagent test strip, a amperometric meter and a quality control solution. The user inserts a test strip in the Elect Meter and applies a small drop of blood from a fingerstick to the sample point on the strip. The meter measures the electrical current that is generated which is proportional to the concentration of glucose present in the blood sample. The blood glucose result is displayed in 20 seconds. The system is calibrated to give a result equivalent to a plasma value, the measurement used by most clinical laboratories.

AI/ML Overview

The provided text does not contain information about acceptance criteria, device performance, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth details for a study. The text primarily describes the "Elect Blood Glucose Monitoring System," its intended use, and its equivalence to other devices, focusing on product description rather than clinical study results or validation.

{0}------------------------------------------------

Selfcare, Inc.

200 Prospect Street Waltham, Massachusetts 02154 Telephone 617 647 3900 Facsimile 617 647 3939

K961985

Page 66

Image /page/0/Picture/4 description: The image shows the word "selfcare" in a simple, sans-serif font. To the left of the word is a small graphic consisting of a circle above a square divided into four smaller squares. The graphic is black, and the word "selfcare" is also black.

EP 5 1988

510(k) Summary

Manufacturer:

Selfcare, Inc. 200 Prospect Street Waltham, MA 02154 Telephone: +617 647 3900 Fax: +617 647 3939

Contact Person: Carol Adiletto

Date: 20th May 1996

Device Name: Elect Blood Glucose Monitoring System

Common Name: Blood glucose meter and reagent test strips for blood glucose

Classification: Class II as per 21 CFR 862.1345. Glucose test system; Code 75 CGA (Glucose oxidase, glucose) Classification panel - clinical chemistry

Equivalent Devices: Precision QID Blood Glucose Monitoring System Glucometer Elite Diabetes Care System

Product Description

The Elect Blood Glucose Monitoring System comprises a glucose reagent test strip, a amperometric meter and a quality control solution. The user inserts a test strip in the Elect Meter and applies a small drop of blood from a fingerstick to the sample point on the strip. The meter measures the electrical current that is generated which is proportional to the concentration of glucose present in the blood sample. The blood glucose result is displayed in 20 seconds. The system is calibrated to give a result equivalent to a plasma value, the measurement used by most clinical laboratories.

{1}------------------------------------------------

Intended Use:

The Elect Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The Elect System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor effectiveness of diabetes control.

Substantial Equivalence:

The Elect Blood Glucose Monitoring System is similar in technological characteristics, methodology and intended use to the predicate devices listed above. The Elect System utilizes similar biosensor methodology based on a standard enzymatic methodology employing the enzyme glucose oxidase coupled to an electrochemical detection system to measure glucose levels in whole blood.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.