K Number
K961925

Validate with FDA (Live)

Date Cleared
1996-07-29

(73 days)

Product Code
Regulation Number
878.5000
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The polyester nonabsorbable surgical sutures USP are indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.

Device Description

Polyester nonabsorbable surgical sutures USP, manufactured by R.K. Medical, L.L.C. are equivalent to Ti-cron* polyester nonabsorbable surgical suture USP manufactured by Davis & Geck. The polyester nonabsorbable surgical sutures USP manufactured by R.K. Medical and Davis & Geck are braided and coated with silicone. The R.K. Medical polyester suture is dyed green (D&C Green #6). Davis & Geck's Ti-cron* is dyed blue (D&C Blue #6).

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and device performance:

This document describes a 510(k) submission for a medical device, but it fundamentally lacks the detailed information required to answer most of your detailed questions about AI device performance and studies. The device described is a Polyester Nonabsorbable Surgical Suture, which is a physical, non-AI medical device. The document predates the widespread use of AI in medical imaging and diagnostics, and thus, its content is focused on traditional medical device equivalency rather than AI performance.

Therefore, the following answers are based on the limited information provided for a non-AI device.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (R.K. Medical Polyester Suture)
Suture DiameterMeets or exceeds USP specifications
Suture LengthMeets or exceeds USP specifications
Knot Pull Tensile StrengthMeets or exceeds USP specifications
Needle Attachment StrengthMeets or exceeds USP specifications

Additional Note: The document states the R.K. Medical sutures are "equivalent to Ti-cron* polyester nonabsorbable surgical suture USP manufactured by Davis & Geck" in terms of the above parameters.


Regarding the remaining questions on AI device studies:

The provided document does not contain any information about an AI-powered device, image analysis, diagnostic algorithms, or any type of study that would involve:

  • AI performance metrics (sensitivity, specificity, AUC)
  • Sample sizes for test or training sets for an AI model
  • Data provenance for AI input data
  • Experts establishing ground truth for AI analysis
  • Adjudication methods for AI
  • Multi-reader multi-case (MRMC) comparative effectiveness studies with AI assistance
  • Standalone (algorithm-only) AI performance studies
  • Types of ground truth for AI (pathology, outcomes)
  • Training set details for AI algorithms

Therefore, it is impossible to provide answers to questions 2 through 9 based on the given text. This document is for a physical surgical suture, not an AI device.

{0}------------------------------------------------

JUL 29 1996

K961925

510 (k) SUMMAR Y

35 EAGLE ROAD • DANBURY, CT 06810 U.S.A. PHONE: (203) 797-8700 • FAX: (203) 792-1465

"This summary of 510 (k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. "

Applicant: R.K. Medical L.L.C. 35 Eagle Road Danbury. CT 06810 Tele: (203) 797-8700 Fax: (203) 792-1465

Contact:

Mr. John A. Hutter, Plant Manager or Mr. George R. Krebs, President

Date: 5/15/96

Name of Device:

Common or Usual Name: Polyester Nonabsorbable Surgical Sutures Classification Name: Suture, Nonabsorbable, Synthetic, Polyester

Polyester nonabsorbable surgical sutures USP, manufactured by R.K. Medical, L.L.C. are equivalent to Ti-cron* polyester nonabsorbable surgical suture USP manufactured by Davis & Geck.

The polyester nonabsorbable surgical sutures USP manufactured by R.K. Medical and Davis & Geck are braided and coated with silicone. The R.K. Medical polyester suture is dyed green (D&C Green #6). Davis & Geck's Ti-cron* is dyed blue (D&C Blue #6).

The polyester nonabsorbable surgical sutures USP are indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.

Testing of suture diameter, suture length, knot pull tensile strength and needle attachment strength according to methods described in USP XXIII demonstrates that R.K. Medical Polyester nonabsorbable surgical sutures meet or exceed USP specifications and are equivalent in terms of the above parameters to polyester nonabsorbable surgical sutures manufactured by Davis & Geck.

John A. Hutter, Plant Manager

-- 40 --

§ 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture.

(a)
Identification. Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.