(81 days)
Anatomical Region: Head, body, extremity, spine, neck, TMJ, Knee
Nuclei excited: Hydrogen
Diagnostic Use: 2D and 3D imaging, Contrast Media, MR Vascular Imaging
The 8.5CD Upgrade for MRT-35A adds a new sequence and software enhancements to the existing MRT-35A software version.
The provided document describes an upgrade to an existing Magnetic Resonance Imaging (MRI) device, the MRT-35A, with the new version 8.5CD. This upgrade primarily introduces a new 2D Time-Of-Flight (TOF) sequence and software enhancements for improved image reconstruction in Fast Spin Echo (FSE) sequences.
Here's an analysis of the acceptance criteria and the study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" for the device performance in terms of diagnostic accuracy or clinical utility. Instead, it details the device specifications and highlights the enhancements introduced by the 8.5CD upgrade compared to the previous 8.0CD version. The "reported device performance" refers to the capabilities and specifications of the upgraded system.
| Acceptance Criteria (Implied/Specified) | Reported Device Performance (8.5CD Upgrade) |
|---|---|
| Magnetic Field Strength | 0.35 Tesla (Static Field) |
| RF Power Deposition | 0.34 W/kg |
| Rate of Change of Magnetic Field | 6.83 Tesla/sec Axial, 6.69 Tesla/sec Transverse |
| Acoustic Noise Levels | L-peak: 101.6 db (peak impulse sound pressure)L-eq: 86.5 db (time integral of a-weighted sound pressure) |
| New 2DTOF Sequence Availability | Provided as "2D ANGIO TOF" (Seq #27) with Min Thick 2.0mm and Flow Comp (FC), and optional walking presat (superior or inferior). |
| Image Reconstruction (FSE) | Improved image reconstruction to reduce subtle acquisition artifacts previously found in TE=20 FSE sequences. |
| Clinical Benefit (2DTOF) | Allows image acquisitions of vascular structures with reduced motion artifacts. |
| Clinical Benefit (FSE) | Image quality is improved for Fast Spin Echo TE=20 (4 and 7 echo). |
| Signal-to-Noise (SNR) | Sample phantom images and clinical images were presented for all new sequences. SNR for each phantom image was given. Clinical studies were presented for all new sequences. (No specific SNR value stated as a criterion). |
| Slice Thickness | 0.8 mm to limits of FOV and sequence. |
| Interslice Spacing | 0 mm minimum. |
| Acquisition Matrix | 256x32 to 256x256 in steps of 1. |
| Display Matrix | 512x512. |
| Image Acquisition Time | 3 seconds for fluoro acquisitions typical, 6 minutes for 4 NEX FSE typical. |
| Hardware Configuration Changes | Version 8.5CD does not change the configuration of hardware for the MRT-35A. |
| Software Feature Changes | Version 8.5CD does not introduce new software features to the MRT-35A. |
2. Sample Size for Test Set and Data Provenance
The document states: "Sample phantom images and clinical images were presented for all new sequences. Signal-to-noise ratios for each phantom image was given. Clinical studies were presented for all new sequences."
- Sample Size for Test Set: Not specified, other than "sample phantom images and clinical images." The exact number of patients or images used in these "clinical studies" is not provided.
- Data Provenance: Not explicitly stated. Given that the manufacturer is "TOSHIBA AMERICA MRI, INC." located in California, USA, the clinical data would likely be from the USA or collaborating sites. Whether the data was retrospective or prospective is also not mentioned.
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not provide any information regarding the number or qualifications of experts used to establish ground truth for the test set. It mentions "clinical studies were presented," implying expert review and interpretation, but details are absent.
4. Adjudication Method
The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No information about an MRMC comparative effectiveness study is mentioned. The document focuses on the technical specifications and general improvements of the device rather than a study comparing human reader performance with and without AI assistance. The described enhancements are for the imaging device itself, not an AI-powered diagnostic aide.
6. Standalone (Algorithm Only) Performance
The device is an MRI machine with software upgrades, not an AI algorithm performing standalone diagnostic tasks. Therefore, a standalone performance study in the context of an AI algorithm is not applicable here. The "performance" refers to the image quality and sequence capabilities of the MRI system.
7. Type of Ground Truth Used
The concept of "ground truth" as typically defined for diagnostic AI (e.g., pathology, outcomes data) is not directly applicable to this document. The "ground truth" in this context would likely refer to:
- Physical Phantom Measurements: For parameters like SNR, slice thickness, uniformity, etc., using known physical standards.
- Qualitative Assessment by Clinicians: For "image quality" improvements and "reduced motion artifacts" in clinical images. This would be based on expert consensus, though the details of how this consensus was achieved are not provided.
8. Sample Size for Training Set
This information is not provided. As this is an upgrade to an MRI device and not an AI model requiring a separate "training set" in the machine learning sense, this question is not directly applicable. The software enhancements would have been developed and refined using internal R&D data and potentially prior versions' clinical data.
9. How Ground Truth for Training Set Was Established
Not applicable as this is an MRI device upgrade, not an AI model with a distinct training set. The development of the sequence and reconstruction algorithms would involve engineering specifications, phantom studies, and possibly iterative clinical feedback, but not a formally established "ground truth" for a training set in the AI paradigm.
{0}------------------------------------------------
APPENDIX A
510(k) SUMMARY
AUG - 2 1996
{1}------------------------------------------------
510(k) Summary
| 1. | Model Name:Device Name:Trade/Proprietary Name: | MRT-35AMagnetic Resonance Device AccessoryMRT-35A Upgrade 8.5CD |
|---|---|---|
| 2. | Establishment Registration: | #2936923 |
| 3. | Name and Address: | TOSHIBA AMERICA MRI, INC.280 Utah AvenueSouth San Francisco, CA 94080 |
| Contact Person: | Steven M. Kay(714) 730-5000 | |
| 4. | DATE OF SUBMISSION: | 26 April 1996 |
DEVICE DESCRIPTION 5.
The 8.5CD Upgrade for MRT-35A adds a new sequence and software enhancements to the existing MRT-35A software version.
5A. DEVICE CHARACTERISTICS:
The following describes the MRT-35A system with the 8.5CD Upgrade.
| Type of magnet | Superconductive |
|---|---|
| Static Field | 0.35 Tesla |
| Type of RF Coil | Not applicable |
| Data Acquisition Mode | 2D Multislice, Multiecho, 3DFT |
| Pulse sequences | Spin echo, Inversion recovery, Fast spin echo, CardiacGated, Partial Flip, Matched Bandwidth, 3D, inplanepresaturation, Spectre, Angiography, 2D Time-of-Flight |
| Reconstruction Technique | 2 Dimensional Fourier3 Dimensional Fourier3 Dimensional Slice shift,Oblique, MIP for Angio |
| Display Matrix: | 512x512 |
| Slice Selection method: | Selective gradient, "slab" select slice encode for 3D |
| Acquisition Matrix: | 256x32 to 256x256 in steps of 1 |
| Image Acquisition Time: | 3 seconds for fluoro acquisitions typical, 6 minutes for 4NEX FSE typical |
| Pre and Post Processing: | No baseline correction except for data conjugation ofAdvanced Fourier Imaging, Post processing filtering |
| Number of slices: | 64 for 3D acquisitions, protocol file limit of 256 slices |
| Slice Orientation: | unrestricted |
| Spacing: | gapless imaging |
{2}------------------------------------------------
New Hardware and Hardware Changes 5A.1
Version 8.5CD does not change the configuration of hardware for the MRT-35A.
5A.2 New Software Features
Version 8.5CD does not introduce new software features to the MRT-35A.
5A.3 New Sequences and Sequence Changes
2D Time of Flight(TOF) sequence 5A.3.1
The 8.5CD upgrade provides a 2DTOF sequence. 2DTOF includes optional walking presat (superior or inferior) to reduce the signal from adjacent tissue due to motion. The user interface for sequence selection and parameter specification common to other sequences is supported for 2DTOF also. There is no exceptional change in the manner that 2DTOF is selected when compared to pre-existing sequences. Acquisition orientation is restricted to transaxial orientation only. The sequence will not support oblique orientations nor gated acquisitions.
5A.3.2 Sequence specification summary
The new 2DTOF sequence, with its identification parameters, is given below. With the exception of this additional sequence, there is no change to the sequence software.
| Seq # | SequenceType | TE | FlowComp | Min Thick8.5CD | Differencefrom 8.0CD |
|---|---|---|---|---|---|
| 27 | 2D ANGIOTOF | 12 | FC | 2.0mm | NEW |
SAFETY PARAMETERS 6.
| Maximum static field strength: | 0.35 Tesla |
|---|---|
| Rate of change of magnetic field: | 6.83 Tesla/sec Axial6.69 Tesla/sec Transverse |
| Radio frequency power deposition: | 0.34 W/kg |
| Acoustic Noise levels: | L-peak: 101.6 db (peak impulse soundpressure)L-eq: 86.5 db (time integral of a-weightedsound pressure) |
{3}------------------------------------------------
7. IMAGING PERFORMANCE PARAMETERS
| Specification volume | 10 Cm dsv for head coil20 Cm dsv for body coil |
|---|---|
| Signal to noise: | Sample phantom images and clinical images werepresented for all new sequences. Signal-to-noiseratios for each phantom image was given. Clinicalstudies were presented for all new sequences. |
| Uniformity in three orthogonal planes | Not Applicable |
| Slice Profile in three orthogonal planes | Not Applicable |
| Geometric Distortion in threeorthogonal planes | Not Applicable |
| Slice Thickness | 0.8 mm to limits of FOV and sequence. |
| Interslice Spacing | 0 mm minimum |
8. Intended Use
1
| Anatomical Region: | Head, body, extremity, spine, neck, TMJ, Knee |
|---|---|
| Nuclei excited: | Hydrogen |
| Diagnostic Use: | 2D and 3D imaging, Contrast Media, MR VascularImaging |
Equivalency Information 9.
8.5CD Upgrade for MRT-35 is substantially equivalent to the 8.0CD Software release for MRT-35. Similarity and/or differences between this device and the predicate device are summarized below.
HARDWARE ENHANCEMENTS
8.5CD has no hardware component differences from the preceding MRT-35A upgrade, 8.0CD.
SOFTWARE ENHANCEMENTS
| Technical | Clinical |
|---|---|
| Improved Image Reconstruction for FSEImage reconstruction software is provided tofurther reduce subtle acquisition artifactspreviously found in TE=20 Fast Spin Echo (FSE)sequences. | Image quality is improved for Fast SpinEcho TE=20 (4 and 7 echo). |
SEQUENCE ENHANCEMENTS
| Technical | Clinical |
|---|---|
| 2D Time-Of-Flight (TOF) SequenceA new 2DTOF sequence is provided with optionalpre-saturation bands that are user selected. | Allows image acquisitions of vascularstructures with reduced motion artifacts. |
N/A