Intended Use

The device is intended for oral or nasal intubation. It is intended for airway management.

Device Description

The Rusch Oral/Nasal Tracheal Tube, Cuffed, Magill/Murphy consists of a clear implant tested PVC tracheal tube with radiopaque stripe. The cuff is inflated via a pilot balloon using any lock-tip or luer taper syringe. The main tube is graduated with multiple centimeter markings to allow easy determination of intubated length. A black positioning ring is present on the tube to aid in visualization of the tube. An optional Murphy eye provides an alternative opening. The eye is the difference between the Murphy and the Magill Tube. The device is for Nasal/Oral use.

AI/ML Overview

This document is a 510(k) summary for a medical device (Rüsch Oral/Nasal Tracheal Tube, Cuffed- Magill/Murphy) and does not contain information about studies proving the device meets acceptance criteria in the context of AI/machine learning performance.

The document describes a physical medical device, not a software algorithm. Therefore, the questions about acceptance criteria for AI models, sample sizes for test/training sets, expert adjudication, MRMC studies, and ground truth establishment are not applicable.

Here's a breakdown of why the requested information cannot be provided from the given text:

  1. Table of acceptance criteria and reported device performance: This document focuses on demonstrating substantial equivalence to a predicate device based on manufacturing standards and biocompatibility, not performance metrics like accuracy, sensitivity, or specificity for a diagnostic AI.
  2. Sample size for test set and data provenance: Not applicable. There's no "test set" in the context of AI evaluation for this device.
  3. Number of experts and qualifications: Not applicable. There's no ground truth established by experts for AI evaluation.
  4. Adjudication method: Not applicable.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable. This is for evaluating changes in human performance with AI assistance, which isn't relevant here.
  6. Standalone (algorithm only) performance: Not applicable.
  7. Type of ground truth used: Not applicable. The "ground truth" for this device would relate to its physical properties meeting engineering specifications and biocompatibility, not clinical outcomes or human expert labels for an AI.
  8. Training set sample size: Not applicable.
  9. How ground truth for training set was established: Not applicable.

What the document does state regarding "testing" is:

  • Bench/Laboratory Testing: "Based on test data, the device conforms to ASTM 1242-89." and "The device is biocompatible per the requirements of ISO 10993."
  • Acceptance Criteria Implied: The acceptance criteria are conformity to ASTM 1242-89 (a standard for tracheal tubes) and biocompatibility per ISO 10993.
  • Study Proving Acceptance: The "test data" mentioned for conformity to ASTM 1242-89 and biocompatibility per ISO 10993 are the studies proving the device meets these implied acceptance criteria. However, no specifics about these studies (e.g., sample sizes, methodology beyond "bench/laboratory testing") are provided in this summary.

In summary, this document is for a traditional medical device's 510(k) clearance, not an AI/software as a medical device (SaMD), and therefore doesn't provide the type of information requested about AI model performance validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Rusch INTERNATIONAL. The logo is in black and white and features the word "Rusch" in a stylized font. Below the word "Rusch" is the word "INTERNATIONAL" in a smaller, sans-serif font. Underneath that, it says "A Subsidiary of Teleflex Incorporated (USA)".

K961837

Pines Park

AUG 1 6 1996

Substantial Equivalence 510(k) Summary Rüsch Oral/Nasal Tracheal Tube, Cuffed- Magill/Murphy

To: Whom it May Concern

Date: May 10, 1996

Submitter/Contact- Name and Address:

Miss Karenann J. Brozowski Group Requlatory Affairs Director Rüsch International Tall Pines Park Jaffrey, NH 03452

Telephone: (603) 532-7706 (603) 532-8211 Fax:

Device Details:

Trade Name: Rüsch Oral/Nasal Tracheal Tube, Cuffed- Magill/Murphy

Common Name: Tracheal Tube

Classification Name: Tube, Tracheal(w/wo Connector 73BTR)

Predicate Legally Marketed Device:

Sheridan HVT(TM) Tracheal Tube with X-Ray Line and 15mm. Connector High Volume Tapered Low Pressure Murphy Eye

{1}------------------------------------------------

page 2

Device Description:

The Rusch Oral/Nasal Tracheal Tube, Cuffed, Magill/Murphy consists of a clear implant tested PVC tracheal tube with radiopaque stripe. The cuff is inflated via a pilot balloon using any lock-tip or luer taper syringe. The main tube is graduated with multiple centimeter markings to allow easy determination of intubated length. A black positioning ring is present on the tube to aid in visualization of the tube. An optional Murphy eye provides an alternative opening. The eye is the difference between the Murphy and the Magill Tube. The device is for Nasal/Oral use.

Device Intended Use:

The device is intended for oral or nasal intubation. It is intended for airway management.

Technological Characteristics of the Device:

The device is manufactured according to ASTM 1242-89. It is equivalent in design and intended use with the Sheridan product. Both of the products are PVC with the same general appearance.

The Rusch Tracheal Tube comes in sizes 4.5-11.0 inclusive.

Bench/Laboratory Testing:

Based on test data, the device conforms to ASTM 1242-89.

The device is biocompatible per the requirements of ISO 10993.

§ 868.5730 Tracheal tube.

(a)
Identification. A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.(b)
Classification. Class II (performance standards).