K Number
K961824

Validate with FDA (Live)

Date Cleared
1996-07-23

(74 days)

Product Code
Regulation Number
866.5040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 2 COBAS INTEGRA Reagent Cassettes contained in this submission are intended for use with the COBAS INTEGRA Analyzer. These are the COBAS INTEGRA Cassette for Albumin-T in urine and the COBAS INTEGRA Cassette for HbA1c. The COBAS INTEGRA Cassette for Albumin (Turbidimetric) contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative immunological determination of human albumin in serum and urine. The COBAS INTEGRA Cassette for Hemoglobin Alc contains an in vitro diagnostic reagent system intended for the quantitative determination of percent hemoglobin (HbAlc%) in hemolysate. The COBAS INTEGRA Analyzer and COBAS INTEGRA Reagent Cassettes together provide an integrated system for in vitro diagnostic testing. The COBAS INTEGRA Reagent Cassettes are comprised of chemistry, drugs of abuse, immunology, therapeutic drug monitoring, and hematology assay systems. The COBAS INTEGRA Analyzer provides quantitative measurement of these analytes via three measuring principles, i.e., absorbance, fluorescence polarization and ion-selective electrodes.

This submission also contains a modification to the previously cleared COBAS INTEGRA Reagent Cassette for Digoxin. The COBAS INTEGRA Reagent Cassette for Digoxin has been modified to include the use of heparinized samples.

Device Description

The COBAS INTEGRA Reagent Cassettes are compact and preparation-free with the added convenience of long term on-board stability. Sixty-eight COBAS INTEGRA Reagent Cassettes can be stored on board, 24 hours a day at 2-8ºC. Each cassette is barcoded. This barcode label provides the analyzer with specific reagent information such as the lot number, the expiration date and the number of tests.

AI/ML Overview

Here's an analysis of the provided text, focusing on the requested information regarding acceptance criteria and supporting studies:

This document describes the 510(k) summary for Roche COBAS® INTEGRA Reagent Cassettes, specifically for Albumin (ALB-T), HbA1c, and a modification to Digoxin (DIG), claiming substantial equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Device/ParameterAcceptance Criteria (Implied/Predicate)COBAS INTEGRA Performance (New Device)
COBAS INTEGRA Albumin (ALB-T)
MethodologyImmunoturbidimetricImmunoturbidimetric
Sample typeSerum, umbilical cord serum, CSF, urineUrine
Assay rangeNot specified in labeling6-193 mg/L; 6-3860 mg/L w/postdilution
Precision (Within-run)Not specified in labeling4.3 % at 10 mg/L; 1.2 % at 223 mg/L
AccuracyNot specified in labelingN = 200; R = 0.997 vs. Behring Albumin (Predicate device: Behring N Antiserum to Human Albumin)
Sensitivity (Analytical)Not specified in labeling7 mg/L
COBAS INTEGRA HbA1c (HBA1C)
MethodologyImmunoturbidimetric for HbA1c; Colorimetric for Total HbImmunoturbidimetric test for HbA1c; Colorimetric test for Total Hb
Sample typeAnticoagulated venous or capillary whole blood (heparin, EDTA, citrate or oxalate/fluoride)Anticoagulated venous or capillary whole blood (heparin, EDTA, citrate or oxalate/fluoride)
ApplicationHemolysate or whole bloodHemolysate
Reported measuring units% HbA1c% HbA1c
Assay range2-25 %3-30.9 %
Precision (Total)Mean % CV %: 5.3 % (5.3%); 12.9 % (4.9%) (Predicate device: Roche Unimate HbA1c Reagent)Mean % CV %: 4.8 % (2.8%); 12.1 % (2.4%)
AccuracyN = 208; R = 0.943 vs. BM Tina-quant HbA1c ReagentN = 240; R = 0.994 vs. Roche Unimate Reagent (Predicate device: Roche Unimate HbA1c Reagent)
Sensitivity (Analytical)0.76 umol/L for hemoglobin; 0.78 umol/L for HbA1c0.90 umol/L for hemoglobin; 0.22 umol/L for HbA1c
COBAS INTEGRA Digoxin (Modified)
MethodologyKinetic interaction of microparticles in solutionKinetic interaction of microparticles in solution
Sample typeSerumSerum and heparinized plasma
Assay range0.17 - 5.0 ng/mL0.17 - 5.0 ng/mL
Precision (Total)14.4 % at 0.81 ng/mL; 5.3 % at 1.57 ng/mL; 3.8 % at 4.1 ng/mL (Predicate device: COBAS INTEGRA Cassette for Digoxin - Cleared)9.7 % at 0.87 ng/mL; 6.1 % at 1.64 ng/mL; 3.9 % at 2.82 ng/mL
AccuracyN = 189; R = 0.958 vs. TDx (FPIA)N = 63; R = 0.967 vs. TDx (FPIA) (Predicate device: COBAS INTEGRA Cassette for Digoxin - Cleared)
Sensitivity (Analytical)0.17 ng/mL0.17 ng/mL

Note on Acceptance Criteria: The document primarily relies on demonstrating substantial equivalence to predicate devices, rather than explicit numerical acceptance criteria for the new devices. The "performance characteristics" of the predicate devices implicitly serve as the benchmark for comparability. For the Digoxin modification, the cleared COBAS INTEGRA Cassette for Digoxin serves as its own predicate and its performance metrics represent the baseline for the modification.

2. Sample Sizes Used for the Test Set and Data Provenance

  • COBAS INTEGRA Albumin (ALB-T):
    • Accuracy: N = 200 samples.
    • Data Provenance: Not explicitly stated, but clinical and nonclinical studies are mentioned. It's likely a mix of laboratory-generated samples and potentially clinical samples from various sources. The document does not specify country of origin or whether prospective/retrospective.
  • COBAS INTEGRA HbA1c (HBA1C):
    • Accuracy: N = 240 samples.
    • Data Provenance: Not explicitly stated beyond "clinical and nonclinical studies."
  • COBAS INTEGRA Digoxin (Modified):
    • Accuracy: N = 63 samples.
    • Data Provenance: Not explicitly stated beyond "clinical and nonclinical studies."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This submission is for in vitro diagnostic reagents and an analyzer system. The "ground truth" for such devices is established through reference methods or comparison to legally marketed predicate devices, not through expert radiological or clinical interpretation in the traditional sense. Therefore:

  • Number of Experts: Not applicable in the context of expert consensus like for imaging devices. Ground truth is inherently derived from the analytical performance of the reference method.
  • Qualifications of Experts: Not applicable. The "experts" are the validated methodologies of the predicate devices or reference methods.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation or subjective assessments to resolve discrepancies. For in vitro diagnostic assays, the comparison is quantitative against a reference method or predicate device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done?: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging systems where human readers interpret cases, often with and without AI assistance to measure the impact of AI on reader performance. This submission is for laboratory reagents and an analyzer, where the output is quantitative and direct, not an interpretation by a human reader.
  • Effect size of human readers with vs. without AI assistance: Not applicable, as no MRMC study was conducted.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

  • Was it done?: Yes, the performance characteristics (assay range, precision, accuracy, sensitivity) presented in Tables 1-3 represent the standalone performance of the COBAS INTEGRA Reagent Cassettes when used with the COBAS INTEGRA Analyzer. There is no human-in-the-loop component for these specific assays; the analyzer performs the test and provides the quantitative result directly. The study essentially demonstrates the algorithm and instrument's performance in producing these quantitative values.

7. Type of Ground Truth Used

The ground truth for these in vitro diagnostic tests is established by:

  • For Albumin and HbA1c: Performance against a legally marketed predicate device (Behring N Antiserum to Human Albumin for Albumin, and Roche Unimate HbA1c Reagent for HbA1c). The reference standard for "accuracy" is the results obtained from these predicate devices.
  • For Digoxin (Modified): Performance against the previously cleared COBAS INTEGRA Cassette for Digoxin and comparison to TDx (FPIA), which is likely a well-established and accepted reference method for Digoxin measurement.

8. Sample Size for the Training Set

The document does not specify a training set size. For a 510(k) submission detailing reagent cassettes, particularly for established immunological/colorimetric methods, the focus is on validation and verification of the final product's performance against predicate devices or reference methods. The "training" of an AI algorithm in the modern sense is not applicable here, as these are chemical/biological assays, not machine learning algorithms in the typical AI context. The document describes clinical and nonclinical studies and performance characteristics, which refer to the validation data for the device.

9. How the Ground Truth for the Training Set was Established

As explained above, the concept of a "training set" and associated ground truth establishment (in the AI/ML sense) is not applicable to this type of in vitro diagnostic device submission, which relies on chemical/biological reactions and analytical performance validation rather than machine learning algorithm development. The "ground truth" for analytical performance is established by comparison to recognized reference methods or existing, legally marketed predicate devices.

{0}------------------------------------------------

K961824

JUL 23 1996

Rache

Roche Diagnostic Systems

A Member of the Roche Group

Roche Diagnostic Systems, Inc Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

Direct Dial Fax

510(k) Summary

Roche COBAS® INTEGRA Reagent Cassettes

In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.

Identification of 510(k) Sponsor: I.

Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771

510(k) Submission dated May 8, 1996

Contact: Maria Feijoo Regulatory Affairs Associate Phone: (908) 253-7310 (908) 253-7547 Fax:

II. Device Name:

The device name, including both the trade/proprietary name and the classification name are provided in the table below.

Product NameClassificationNameRegulatoryClassCFRClassificationNumberPredicateProductNameDatePredicateClearedPredicate510(k)Number
COBAS INTEGRAAlbumin (ALB-T)(Turbidimetric)Albuminimmunologicaltest systemClass II966.5040Behring NAntiserum toHuman Albumin4/15/86K860894
COBAS INTEGRAHbA1c (HBA1C)GlycosylatedhemoglobinassayClass II864.7470Roche UnimateHBA1c Reagent2/9/96K952337/S1
COBAS INTEGRADigoxin (DIG)modificationDigoxin testsystemClass II862.3320COBAS INTEGRADigoxin9/8/95K951595

{1}------------------------------------------------

Identification of the legally marketed device to which the 510(k) sponsor III. claims equivalence:

The following table identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.

Product NameClassificationNameRegulatoryClassCFRClassificationNumberPredicateProductNameDatePredicateClearedPredicate510(k)Number
COBAS INTEGRAAlbumin (ALB-T)(Turbidimetric)Albuminimmunologicaltest systemClass II966.5040Behring NAntiserum toHuman Albumin4/15/86K860894
COBAS INTEGRAHbA1c (HBA1C)GlycosylatedhemoglobinassayClass II864.7470Roche UnimateHBA1c Reagent2/9/96K952337/S1
COBAS INTEGRADigoxin (DIG)modificationDigoxin testsystemClass II862.3320COBAS INTEGRADigoxin9/8/95K951595

IV. Description of the Device/Statement of Intended Use:

Through this submission, it is the intention of Roche to gain clearance of an additional 2 COBAS Reagent Cassettes. The 2 COBAS INTEGRA Reagent Cassettes contained in this submission are intended for use with the COBAS INTEGRA Analyzer. These are the COBAS INTEGRA Cassette for Albumin-T in urine and the COBAS INTEGRA Cassette for HbA1c. The COBAS INTEGRA Cassette for Albumin (Turbidimetric) contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative immunological determination of human albumin in serum and urine. The COBAS INTEGRA Cassette for Hemoglobin Alc contains an in vitro diagnostic reagent system intended for the quantitative determination of percent hemoglobin (HbAlc%) in hemolysate. The COBAS INTEGRA Analyzer and COBAS INTEGRA Reagent Cassettes together provide an integrated system for in vitro diagnostic testing. The COBAS INTEGRA Reagent Cassettes are comprised of chemistry, drugs of abuse, immunology, therapeutic drug monitoring, and hematology assay systems. The COBAS INTEGRA Analyzer provides quantitative measurement of these analytes via three measuring principles, i.e., absorbance, fluorescence polarization and ion-selective electrodes. The COBAS INTEGRA Reagent Cassettes are compact and preparation-free with the added convenience of long term on-board stability. Sixty-eight COBAS INTEGRA Reagent Cassettes can be stored on board, 24 hours a day at 2-8ºC. Each cassette is barcoded. This barcode label provides the analyzer with specific reagent information such as the lot number, the expiration date and the number of tests.

This submission also contains a modification to the previously cleared COBAS INTEGRA Reagent Cassette for Digoxin. The COBAS INTEGRA Reagent Cassette for Digoxin has been modified to include the use of heparinized samples.

{2}------------------------------------------------

v. Summary of the technological characteristics of the new device in comparison to those of the predicate.

Tables 1, 2 and 3 attached to this summary outline the technological characteristics (methodologies) of the COBAS INTEGRA Reagents in comparison to those of legally marketed predicate products.

{3}------------------------------------------------

VI. Brief discussion of the clinical and nonclinical tests relied on for a determination of substantial equivalence:

Tables 1, 2 and 3 attached to this summary demonstrate the results of clinical and nonclincal studies performed using the COBAS INTEGRA Reagent Cassettes. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in this chart. This information concludes that the performance of this device is essentially equivalent to other legally marketed devices of a similar kind.

COBAS INTEGRACassette for Albumin(Turbidimetric) in urineBehring N Antiserum toHuman Albumin(BNA Nephelometer)
MethodologyImmunoturbidimetricImmunoturbidimetric
Sample typeUrineSerum, umbilical cordserum, CSF and urine
CalibratorRoche Serumproteins TStandardBehring N Protein StandardPY
Reagent (active ingredients)1. Anti-albumin Tantiserum (rabbit) inphosphate buffer1. Antiserum to humanalbumin2. Phosphate buffered saline
Performance Characteristics:
Assay range6-193 mg/L6-3860 mg/L w/postdilutionNot specified in labeling
Precision (Within-run)4.3 % at 10 mg/L1.2 % at 223 mg/LNot specified in labeling
AccuracyN = 200R = 0.997 vs. BehringAlbuminNot specified in labeling
Sensitivity (Analytical)7 mg/LNot specified in labeling

Table 1

{4}------------------------------------------------

Table 2

COBAS INTEGRACassette for HbA1cRoche Unimate HbA1cReagent on COBAS MIRA
MethodologyImmunoturbidimetric test for HbA1cColorimetric test for Total HbImmunoturbidimetric test for HbA1cColorimetric test for Total Hb
Sample typeAnticoagulated venous or capillary whole blood(heparin, EDTA, citrate or oxalate/fluoride)Anticoagulated venous or capillary whole blood(heparin, EDTA, citrate or oxalate/fluoride)
ApplicationHemolysateHemolysate or whole blood
Reported measuring units% HbA1c% HbA1c
CalibratorRoche HbA1c CalibratorRoche HbA1c Calibrator
Reagent (active ingredients)R1. Potassium phosphate bufferTensidesR2. Latex coated withmonoclonal (mouse) antibody specific for HbA1cBovine serum albuminNaClHEPPS BufferR3. Agglutinator (synthetic polyvalent antigen )Bovine serum albuminFormate bufferR1. Latex coated withmonoclonal (mouse)antibody specific for HbA1cBovine serum albuminNaClHEPPS BufferR2. Potassium phosphate bufferTensidesR3. Agglutinator (synthetic polyvalent antigen )Bovine serum albuminFormate buffer
Performance Characteristics:
Assay range3-30.9 %2-25 %
Precision (Total)Mean % CV %4.8 % 2.8 %12.1 % 2.4 %Mean % CV %5.3 % 4.0 %12.9 % 4.9 %
AccuracyN = 240R = 0.994 vs. RocheUnimate ReagentN = 208R = 0.943 vs. BM Tina-quant HbA1c Reagent
Sensitivity (Analytical)0.90 umol/L for hemoglobin0.22 umol/L for HbA1c0.76 umol/L for hemoglobin0.78 umol/L for HbA1c

{5}------------------------------------------------

Table 3

COBAS INTEGRACassette for Digoxin(Modified)COBAS INTEGRACassette for Digoxin(Cleared)
MethodologyKinetic interaction ofmicroparticles in solutionKinetic interaction ofmicroparticles in solution
Sample typeSerum and heparinizedplasmaSerum
CalibratorCOBAS-FP DigoxinCalibrators (K951595)COBAS-FP DigoxinCalibrators (K951595)
ControlsCOBAS-FP TDM Controls(K954992)COBAS-FP TDM Controls(K954992)
Reagent (active ingredients)0 Anti-digoxin monoclonalantibody (mouse) inbuffer1 Conjugated digoxinderivative microparticlesin buffer0 Anti-digoxin monoclonalantibody (mouse) inbuffer1 Conjugated digoxinderivative microparticlesin buffer
Performance Characteristics:
Assay range0.17 - 5.0 ng/mL0.17 - 5.0 ng/mL
Precision (Total)9.7 % at 0.87 ng/mL6.1 % at 1.64 ng/mL3.9 % at 2.82 ng/mL14.4 % at 0.81 ng/mL5.3 % at 1.57 ng/mL3.8 % at 4.1 ng/mL
AccuracyN = 63R = 0.967 vs. TDx (FPIA)N = 189R = 0.958 vs. TDx (FPIA)
Sensitivity (Analytical)0.17 ng/mL0.17 ng/mL

l

§ 866.5040 Albumin immunological test system.

(a)
Identification. An albumin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the albumin (a plasma protein) in serum and other body fluids. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.