(65 days)
The MR Workstations (including the ViStar, Montage and OnCall) are designed to provide fast and convenient review of DICOM compatible digital images from all six diagnostic imaging modalities (MR, CT, Nuclear Medicine, X-Ray, CR and Ultrasound) and secondary capture as well as post-processing of MR and CT images including multi planar reformatting, angiographic rendering, surface rendering, segmentation/disarticulation and cine display.
When used in conjunction with any of these diagnostic imaging modalities, the indications for use remain unchanged from those defined for that modality.
The Picker MR Workstations (ViStar, Montage and OnCall) integrate image review and processing software with a variety of hardware configurations. The software consists of a graphical user interface and post-processing capabilities for MR and CT images including: multi planar reformatting, angiographic rendering, surface rendering, segmentation / disarticulation and cine display. The hardware configurations can be divided into three major types, ViStar, Montage and Oncall. In general, the ViStar line of products are single-monitor workstations which can be equipped with all of the image manipulation and reformating capabilities described above. The Montage line of products are multi-monitor, multimodality viewing stations which have limited image manipulation and reformating capabilities. Finally, the OnCall systems are PC-based for remote viewing and provide image review and cine display.
The provided text does not describe a study that proves the device meets specific acceptance criteria in the manner usually associated with clinical validation studies for AI/medical devices. Instead, it relies on a Substantial Equivalence Chart to demonstrate that the MR Workstations (ViStar, Montage, and OnCall) are substantially equivalent to previously cleared predicate devices (ViStar K905070 and EDGE/VISTA K931544).
This document is from 1996, a time when AI in medical imaging was not yet a prominent field, and validation methods were different. The "device" in question here is an image viewing and processing workstation, not an AI algorithm. Therefore, the questions related to AI-specific validation (like standalone performance, MRMC studies for AI improvement, and sample sizes for training/test sets for AI models) are not directly applicable.
However, I can extract the "acceptance criteria" and "reported device performance" based on the substantial equivalence comparison. In this context, "acceptance criteria" can be interpreted as the characteristics and functionalities of the predicate devices, and "reported device performance" are the characteristics of the new device that align with or exceed those of the predicate.
Here's the information formatted as requested, with notes explaining the lack of certain AI-specific details:
Acceptance Criteria and Reported Device Performance
1. A table of acceptance criteria and the reported device performance:
| Parameter / Acceptance Criteria (Predicate Device) | Reported Device Performance (MR Workstations - ViStar/Montage/OnCall) |
|---|---|
| Computer: Stardent 3000 supercomputer using parallel RISC architecture. (K905070) | ViStar/Montage: DEC/Alpha supercomputer with RISC architecture. OnCall: Pentium class personal computer platform. |
| RAM Memory: 64 MBytes. Expandable to 128 MBytes. (K905070) | ViStar/Montage: Ranges from 32 to 256 MBytes. Expandable. OnCall: 24 MBytes. Expandable to 64 MBytes. |
| Image Storage Capabilities: 760 MBytes disk storage, 1/4" cartridge tape drive, magnetic tape drive. Optional optical disk drive. (K905070) | ViStar/Montage: Disk storage ranges from 0.5 to 2.1 GBytes, Additional Storage, optical disk drive, VCR Interface and 1/2" streaming tape drive available. OnCall: Minimum 760 MBytes of Disk Storage. |
| Networking: Ethernet TCP/IP. (K905070) | Ethernet TCP/IP. Supports the DICOM v3.0 imaging/networking protocols. |
| Filming: Direct and Remote Filming Capabilities. (K905070) | ViStar/Montage: Direct and Remote Filming Capabilities. OnCall: Limited formats available. |
| Monitor: 19 inch, 1280 x 1024 high resolution color monitor, 60 Hz, non-interlaced. (K905070) | ViStar/Montage: 17-21" color or greyscale high resolution (1280 x 1024) monitor. 72 Hz, non-interlaced. OnCall: Laptop- Color display with 800 x 600 resolution and SVGA graphics. Desktop- same as ViStar/Montage. |
| Graphical Interface: Image review, Database Directory, Draw, Filter, and Filming. (K931544) | Image review, Database Directory, Draw, Filter, and Filming. |
| Image Utilities: Archive, Delete, Camera Summary, HPQ Import (K931544) | Archive, Delete, CD Player, CT Import, HPQ Import, HPQ Tape Import, Export, Camera Summary, Film Summary, Log Problem, Image Processing, Video Review, Video Capture, System Analysis. |
| Image Manipulation & Reformatting Capabilities: Cine Display, Multi planar Reformatting, Angiographic Rendering, Surface Rendering & Cubeview. (K905070) | ViStar: Cine Display, Multi planar Reformatting, Angiographic Rendering, Surface Rendering & Cubeview, Segmentation / Disarticulation (Optional). Montage: Cine Display, Multi planar Reformatting, Angiographic Rendering (Optional). OnCall: Cine Display. |
| Intended Use: As defined for the predicate device's specific modality (NMR devices to produce images corresponding to proton distribution, NMR parameters, and soft tissue structure). (K905070) | Review of DICOM compatible digital images from all six diagnostic imaging modalities (MR, CT, Nuclear Medicine, X-Ray, CR and Ultrasound) and secondary capture, as well as post-processing of MR and CT images. Intended use remains unchanged from those defined for the specific modality. |
2. Sample size used for the test set and the data provenance:
- Not Applicable. This document pre-dates AI-specific validation studies involving test sets derived from patient data. The "study" here is a comparison of technical specifications and intended uses for substantial equivalence, not a performance evaluation on a patient dataset.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. No ground truth for a test set of medical images was established as part of this submission for substantial equivalence.
4. Adjudication method for the test set:
- Not Applicable. No test set or ground truth adjudication was involved.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI device. No MRMC study was performed comparing human performance with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is an image workstation, not an AI algorithm designed for standalone performance.
7. The type of ground truth used:
- Not Applicable. No ground truth data was used in the context of an algorithm's performance. The "ground truth" for this submission is implicitly the established functionality and safety of the predicate devices, against which the new device's specifications are compared.
8. The sample size for the training set:
- Not Applicable. This device is an image processing workstation, not an AI model that requires a training set.
9. How the ground truth for the training set was established:
- Not Applicable. There was no training set for an AI model.
{0}------------------------------------------------
SUMMARY OF SAFETY AND EFFECTIVENESS
JUL - 3 1996
General Information 1.
| Classification: | Class IIImage Processing System | |
|---|---|---|
| Common/Usual Name: | Image Viewing and Processing Workstation | |
| Proprietary Name: | MR Workstations (ViStar, Montage and OnCall) | |
| Establishment Registration: | Picker International, Inc.World Headquarters595 Miner RoadCleveland, Ohio 44143FDA Owner Number: #1580240FDA Registration Number: #1525965 | |
| Performance Standards: | Not Applicable |
Intended Uses: 2.
The MR Workstations (including the ViStar, Montage and OnCall) are designed to provide fast and convenient review of DICOM compatible digital images from all six diagnostic imaging modalities (MR, CT, Nuclear Medicine, X-Ray, CR and Ultrasound) and secondary capture as well as post-processing of MR and CT images including multi planar reformatting, angiographic rendering, surface rendering, segmentation/disarticulation and cine display.
When used in conjunction with any of these diagnostic imaging modalities, the indications for use remain unchanged from those defined for that modality.
3. Device Description:
The Picker MR Workstations (ViStar, Montage and OnCall) integrate image review and processing software with a variety of hardware configurations. The software consists of a graphical user interface and post-processing capabilities for MR and CT images including: multi planar reformatting, angiographic rendering, surface rendering, segmentation / disarticulation and cine display. The hardware configurations can be divided into three major types, ViStar, Montage and Oncall. In general, the ViStar line of products are single-monitor workstations which can be equipped with all of the image manipulation and reformating capabilities described above. The Montage line of products are multi-monitor, multimodality viewing stations which have limited image manipulation and reformating capabilities. Finally, the OnCall systems are PC-based for remote viewing and provide image review and cine display.
{1}------------------------------------------------
4. Safety and Effectiveness
The Picker MR Workstations (ViStar, Montage and OnCall) are substantially equivalent to the ViStar (K905070) and the software described with the EDGE/VISTA systems (K931544) in safety and effectiveness. The following chart has been compiled to demonstrate the MR Workstations' substantial equivalence to these devices.
| Parameter | MR Workstations(ViStar/Montage/OnCall) | Predicate Device(ViStar K905070,EDGE/VISTA K931544) |
|---|---|---|
| Computer | ViStar/Montage: DEC/Alphasupecomputer with RISCarchitecture.OnCall: Pentium classpersonal computer platform(e.g. IBM Thinkpad 760C) | Stardent 3000supercomputer using parallelRISC architecture.(K905070) |
| RAM Memory | ViStar/Montage: Rangesfrom 32 to 256 MBytes.Expandable for all models.OnCall: 24 MBytes.Expandable to 64 MBytes. | 64 MBytes. Expandable to128 MBytes.(K905070) |
| Image Storage Capabilities | ViStar/Montage: Diskstorage ranges from 0.5 to2.1 GBytes, AdditionalStorage, optical disk drive,VCR Interface and 1/2"streaming tape driveavailable.OnCall: Minimum 760MBytes of Disk Storage. | 760 MBytes disk storage,1/4" cartridge tape drive,magnetic tape drive.Optional optical disk drive.(K905070) |
| Parameter | MR Workstations(ViStar/Montage/OnCall) | Predicate Device(ViStar K905070,EDGE/VISTA K931544) |
| Networking | Ethernet TCP/IP. Supportsthe DICOM v3.0imaging/networkingprotocols. | Ethernet TCP/IP.(K905070) |
| Filming | ViStar/Montage: Direct andRemote Filming Capabilities.OnCall: Limited formatsavailable. | Direct and Remote FilmingCapabilities.(K905070) |
| Monitor | ViStar/Montage: 17-21"color or greyscale highresolution (1280 x 1024)monitor. 72 Hz, non-interlaced.OnCall: Laptop- Colordisplay with 800 x 600resolution and SVGAgraphics. Desktop- same asViStar/Montage. | 19 inch, 1280 x 1024 highresolution color monitor, 60Hz, non-interlaced.(K905070) |
| Graphical Interface | Image review, DatabaseDirectory, Draw, Filter, andFilming. | Image review, DatabaseDirectory, Draw, Filter, andFilming.(K931544) |
| Image Utilities | Archive, Delete, CD Player,CT Import, HPQ Import,HPQ Tape Import, Export,Camera Summary, FilmSummary, Log Problem,Image Processing, VideoReview, Video Capture,System Analysis. | Archive, Delete, CameraSummary, HPQ Import(K931544) |
| Parameter | MR Workstations(ViStar/Montage/OnCall) | Predicate Device(ViStar K905070,EDGE/VISTA K931544) |
| Image Manipulation &Reformatting Capabilities | ViStar: Cine Display, Multiplanar Reformatting,Angiographic Rendering,Surface Rendering &Cubeview, Segmentation /Disarticulation (Optional).Montage: Cine Display,Multi planar Reformatting,Angiographic Rendering(Optional).OnCall: Cine Display. | Cine Display, Multi planarReformatting, AngiographicRendering, SurfaceRendering & Cubeview.(K905070) |
| Intended Use | The ViStar systems used with HPQ and 2055HP systems donot change the existing indications which are: The PickerInternational Vista HPQ and 2055 HP systems are indicatedfor use as NMR devices that produce images that: (1)correspond to the distribution of protons exhibiting NMR, (2)depend upon the NMR parameters (proton density, flowvelocity, spin-lattice relaxation time (T1), and spin-spinrelaxation time (T2)) and (3) display the soft tissue structureof the head and whole body. When interpreted by a trainedphysician, these images yield information that can be useful inthe determination of a diagnosis. (K905070)The Picker MR Workstations (including the ViStar,Montage and OnCall) are designed to provide fast andconvenient review of DICOM compatible digital images fromall six diagnostic imaging modalities (MR, CT, NuclearMedicine, X-Ray, CR and Ultrasound) and secondary captureas well as post-processing of MR and CT images includingmulti planar reformatting, angiographic rendering, surfacerendering, segmentation/disarticulation and cine display.When used in conjunction with any of these diagnosticimaging modalities, the indications for use remain unchangedfrom those defined for that modality. |
SUBSTANTIAL EQUIVALENCE CHART
PICKER INTERNATIONAL, INC. (MRWKST)
4/23/96
I-2
{2}------------------------------------------------
PICKER INTERNATIONAL, INC. (MRWKST)
:
1
4/24/96
I-3
{3}------------------------------------------------
PICKER INTERNATIONAL, INC. (MRWKST) 4/23/96
)
)
: :
I-4
N/A