K Number
K961520

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-11-13

(205 days)

Product Code
Regulation Number
870.3680
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Not Found

Device Description

UNIPOLAR IPG GROUND CABLE, Model 5473

AI/ML Overview

The provided document describes performance testing for medical device, likely a component of a larger system. However, the information requested about a study proving a device meets acceptance criteria, particularly in the context of an AI-powered device or a study involving human readers/experts, is not present in the provided text.

The document details device integrity testing for the UNIPOLAR IPG GROUND CABLE, Model 5473, focusing on its physical and functional integrity. It does not describe an AI medical device, diagnostic performance, or a comparative effectiveness study involving human readers.

Here's an analysis of the provided text against your requested points:

1. A table of acceptance criteria and the reported device performance

TestAcceptance Criteria (Test Requirement)Reported Performance
Autoclave SterilizationDevices shall remain within specification following flash autoclave sterilization.Pass
Grounding Clip Life CyclingDevices shall be capable of grasping and holding a minimum of 2 ounce weight in the vertical direction, and shall have acceptable contact resistance.Pass
Rotational FlexDevices shall be reusablePass

Missing Information (Not present in the provided text):

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample sizes for device integrity testing are provided:
    • Autoclave Sterilization: 10 cable assemblies
    • Grounding Clip Life Cycling: 20 cable assemblies
    • Rotational Flex: 8 cable assemblies
  • Data provenance (country of origin, retrospective/prospective) is not mentioned. This type of information is typically irrelevant for device integrity testing of physical components.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This document describes physical device integrity testing, not a diagnostic study requiring expert ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. This describes physical device testing, not a diagnostic study requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC study was not done. This document describes physical device integrity testing, not an AI-powered diagnostic study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This document describes physical device integrity testing, not an algorithm's performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the device integrity testing, the "ground truth" is defined by the test requirements/specifications. For example, "remaining within specification," "grasping and holding a minimum of 2 ounce weight," and "acceptable contact resistance." These are engineering-defined metrics, not clinical ground truth.

8. The sample size for the training set

  • Not applicable. No training set is mentioned as this is physical device testing, not an AI model.

9. How the ground truth for the training set was established

  • Not applicable. No training set or ground truth in the context of machine learning is mentioned.

In summary, the provided text details engineering-focused device integrity testing for a physical medical device component. It does not contain any information related to AI devices, diagnostic studies, human reader performance, or typical ground truth establishment methods for such studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K961520" in a handwritten style at the top. Below this, the words "MEDTRONIC CONFIDENTIAL" are printed in a bold, sans-serif font. The text suggests that the document is confidential and related to Medtronic, possibly indicating a product or project identifier.

NOV 13 100

Section 7 - Performance Data Demonstrating Safety and Effectiveness

1. Device Integrity Testing

The following table identifies the tests which were performed to verify that the UNIPOLAR IPG GROUND CABLE, Model 5473 meets the device integrity requirements. The table includes a summary of the requirements for each test. Following the table, a summary of each test objective and method is provided.

2. Sample Size

A total sample size of thirty (30) cable assemblies were used for the Model 5473 performance testing. Twenty (20) cable assemblies were exposed to life cycle testing, providing a 90% confidence level with a minimum reliability of 89%. Ten (10) cable assemblies were exposed to autoclave sterilization testing, providing a 90% confidence level with a minimum reliability of 79%. Rotational flex testing required a sample size of eight (8) cable assemblies to gain a 90% confidence with a minimum reliability of 74%. Lower sample sizes for the sterilization and rotational flex testing were deemed adequate due to the proven history that these materials have in many other Medtronic surgical cable assemblies.

TestTest RequirementDevicesTestedTestResults
AutoclaveSterilizationDevices shall remain within specificationfollowing flash autoclave sterilization.10Pass
GroundingClip LifeCyclingDevices shall be capable of grasping andholding a minimum of 2 ounce weight in thevertical direction, and shall have acceptablecontact resistance.20Pass
RotationalFlexDevices shall be reusable8Pass

Table 4: Device Integrity Tests

3. Device Integrity Test Objectives and Methods

A. Autoclave Sterilization

Objective

To verify the device can be sterilized without loss of function.

{1}------------------------------------------------

Test Method

The samples are subjected to an initial visual inspection, a contact resistance measurement, and a grip force measurement. The samples are subjected to one hundred (100) flash autoclave sterilization at 2700 F for 10 minutes at a pressure of 21 psi. After sterilization, the samples are subjected to a visual inspection, a contact resistance measurement, and a grip force measurement.

B. Grounding Clip Life Cycling

Objective

To verify the device can be reused without loss of function.

Test Method

The samples are subjected to an initial visual inspection, a contact resistance measurement, and a grip force measurement. The samples are subjected to 250 cycles of opening and closing of the IPG grounding clip. After cycling, the samples are subjected to a visual inspection, a contact resistance measurement, and a grip force measurement.

C. Rotational Flex

Objective

To verify the device can be reused without loss of function.

Test Method

The samples are subjected to an initial visual inspection, and a contact resistance measurement. The samples are subjected to 1000 cycles of rotational flex by clamping the IPG ground clip to a rotating platform, with a one (1) ounce weight six (6) inches from the ground clip. After cycling, the samples are again subjected to visual inspection, and a contact resistance measurement.

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.