(223 days)
Not Found
MIDCO has recently developed a quality Couchmount System (Patient Head Support System) for use in radiosurgery. This couchmount system complements the cass collimator set and enables MIDCO to provide its customers with a full set of instrumentation necessary for radiosurgery. The design of this couchmount is such that it will fit most linear accelerators with only slight modifications to the couchmount adapter assembly. The cass Couchmount System has six axes of adjustment which allows for easy setup with the Linac couch/gantry system. An additional feature of the system is a unique laser measurement apparatus for precisely measuring an isocenter of a target volume for Linac-based radiosurgery/stereotactic radiotherapy. These two features also allow for the correction of alignment errors during treatment planning setup, such as errors inherent in the use of room lasers and errors due to couch shifts during patient positioning. It will also give an index of gantry sag or misalignment. Most commercially available stereotactic headrings can be mounted to the cass Couchmount System by the use of a stereotactic frame holder made specifically for the headring. Although most headrings are similar, each has a slightly different adapter plate which rigidly holds the headring. The stereotactic localization features of most frames are similar, differing mainly in the organization of the frame's coordinate system and its mechanical dimensions.
Here's a breakdown of the acceptance criteria and study information for the cass Linac Radiosurgery Hardware Couchmount Stereotactic Frame Support System, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria (Target Accuracy) | Reported Device Performance (Accuracy) |
|---|---|
| ≤1.0 mm for gantry rotational intersection with rotational axis of the couch (for the base Linac) | The base Varian 2100 C Linac was determined to be ≤1.0 mm. |
| Generally accepted accuracy requirement for delivery of radiation beam to target volume isocenter: fractions of a millimeter. Some floorstand systems reported ≤1mm. | Overall system error range (Lutz method): 0.48 mm ± 0.262 mm |
| Established guidelines of many institutions using radiosurgery: ≤1 mm | Overall system error range (Lutz method): 0.48 mm ± 0.262 mm |
| Fischer System reported localization accuracy: +/- 0.5 mm | Total alignment error for in vivo verification tests (using Laser Measurement System and Precision Localization Box): 0.492 mm +/- 0.155 mm |
| Radionics System reported localization accuracy: 0.5 mm +/- 0.2 mm | Total alignment error for in vivo verification tests (using Laser Measurement System and Precision Localization Box): 0.492 mm +/- 0.155 mm |
Study Details
-
Sample Size used for the test set and the data provenance:
- Sample Size:
- For the Lutz method test: "random phantom targets throughout the extremes of the coordinate system of the Leksell frame and BRW frame" were set up. No specific number is given, but it implies a range of targets.
- For the overall system accuracy (humanoid phantom and dosimetry film): A single "humanoid phantom" was used, with a target point radiated.
- For port film tests: "various predetermined target sites" were used. No specific number is given.
- Data Provenance: The tests were conducted using a Varian 2100 C Linac, which is a common radiation therapy machine. The data is prospective, generated specifically for these tests by MIDCO and its consultants/instructors. There's no mention of specific countries of origin beyond the machine type.
- Sample Size:
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The document does not explicitly state the number or specific qualifications of experts used to establish the ground truth for the test set. However, it indicates involvement of:- "appropriate consultants and instructors" from MIDCO for on-site instructions.
- "radiation oncologists and radiation physicists" at each institution being "ultimately responsible for accurate measurements and gathering of data for the use of this system."
- The studies cited (Lutz, Serago, et al.) are foundational in the field, implying that the ground truth methodology is based on established scientific and medical consensus within the radiosurgery community, as recognized by qualified physicists and clinicians.
-
Adjudication method for the test set:
The document does not describe a formal "adjudication method" in the sense of multiple independent reviewers resolving discrepancies for the test set results. The results (e.g., error ranges for Dx, Dy, Dz displacements) appear to be direct measurements from the phantom studies. The methodology relies on established protocols from cited literature (Lutz method, Serago et al. method), which inherently incorporate methods for accurate measurement and calculation of discrepancies. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. This device is a hardware system (couchmount and collimators) for radiosurgery, not an AI or imaging diagnostic tool that would typically involve human readers interpreting output. -
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This question is not entirely applicable as the device is a hardware system. However, the performance reported (e.g., accuracy measurements like 0.48 mm ± 0.262 mm) represents the "standalone" mechanical accuracy of the system (Linac + collimators + couchmount) as configured and operated according to established protocols, without additional human interpretive input influencing the measurement of accuracy itself. The laser measurement system is an automated component for precise targeting. While humans operate the system, the reported accuracies are objective measurements of the system's physical performance. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth for the mechanical accuracy tests was established through phantom studies and dosimetry film analysis.- Lutz method: Uses a phantom pointer to simulate test targets at known stereotactic coordinates. Radiographic images are then taken, and x, y, z displacements are measured against the known phantom target positions.
- Overall accuracy test: A humanoid phantom with a known target point is radiated, and the isocenter of the multiple intersecting beams is compared to the known position of the phantom target point indicator.
- Port film tests: Predetermined target sites are set (aided by the Laser Measurement System), and then A-P and lateral port films are used to verify the actual versus intended target locations.
- This is a form of empirical ground truth based on physical measurements against precisely known phantom positions.
-
The sample size for the training set:
The document does not describe a "training set" in the context of machine learning. This device is a mechanical system, not an AI or software algorithm that requires a training set of data. The development process involved engineering design, prototyping, and testing, not machine learning model training. -
How the ground truth for the training set was established:
As there is no training set mentioned for this hardware device, this question is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for MIDCO Medical Instrumentation and Diagnostics Corporation. The logo features a stylized "M" above the word "MIDCO", with the full company name written below. The logo is contained within a rounded rectangle. There are also some horizontal lines on the left side of the logo.
| 1I HIMMANAGEMENT------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.Company of The First And Property of Any Andrew | A concess of consideration |
|---|---|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------well would and cannot come of the control of the control research formAndre Mallerial and 10 1000 100 1000 100 100 100------------------------------------------------------------------------------------------------------------------------------------------------------------------------------"W" Classesser " " Management" ( " " " " A s A s a separes | I more a minutes |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------PACITY'SNo. Annual program andA------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AND LABOR CHARACT CALLERY COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLEBy and an and there delivery comments of the | . |
| Iof the transmission superior superior superior and annual1--------------------------------------CARDE PARTY COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION CONTRACTOR COLLECTION CONTACTA BALL PRODUCT CALLERY CONTRACT CARTER FOR THE CONTRACT CARDED THE CONTRACT CARDED THE CONTRACT CARDED THE CONTRACT CARDED THE CONTRACT CARDED THE CONTRACT CARTER------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and and the collection of contract to be the contributed to the contribution of the contribution of the management of the contribution of the management of the management of | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A BAND ALL AND OLD AND OUR COLLECT OF COLOR STORES OF CONSULTION CONTRACT POSTA CONSUL COLL CONSULE OF CHARGE CONSULE | œ |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Maries In Concession of Children of Children and------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Car . Le and Commend Collection CollectionA Benefit Company Computers Collection Comments of Children Comments of ChildrenDATE COLORIES | 1 80 4000 |
| CHERE CONSULTION WAS NEW HISTORIAL ARRENT " " " " " " " " " " " " " " " " " " " " " " " " " "MACHOO TATEStatus and States and Accord Accommend Annual Property Andrew Annual Annual Annual Annual Annual Annual Annual Annual Annual Annual Annual Annual Annual Annon Annon Annon AnnSTORES POR REAL PLACE ALL PROPE S.L. LENGLE LEASE LEASE LECTRICAL FREENAME A. LEGENCESS AND A SECREED & SCHEART & STATUSon in con rates | |
Medical Instrumentation & Diagnostic Corporation (MIDCO) FDA Establishment Reg. #17211849 CASS System Reg. # K894263A
Image /page/0/Picture/4 description: The image shows the word "cass" in a bold, serif typeface. The letters are all lowercase except for the final "S", which is uppercase. The word is centered and takes up most of the frame. The background is plain white.
510(k) Summary As required by section 807.92(c)
cass Linac Radiosurgery Hardware Couchmount Stereotactic Frame Support System
NOV - 7 1996
Submitter's name: Dr. Tyrone L. Hardy 5995 Mira Mesa Blvd., Suite B San Diego, CA 92121 TEL. (619) 558-5880 FAX: (619) 558-5883 Contact: Dr. Tyrone L. Hardy Date of Preparation: November 3, 1996
{1}------------------------------------------------
510(k) Product Information for: cass Linac Radiosurgery Hardware Couchmount Stereotactic Frame Support System
Trade name: cass Photon Knife
Common name: Radiosurgery Couchmount System
Classification name: X-ray Radiation Therapy System (accessory) (per 21 CFR Section 892.550)
Device Class: Class II (90)IYE
FDA Document numbers of prior related submissions:
K894263/A
K911750/A
K9217401/B
Information on devices to which substantial equivalence is claimed:
- F.L. Fischer -- Liebinger LP a. 14540 Beltwood Parkway East Dallas, Texas 75244 Product Name: STP Stereotaxic Convergent Beam Irradiation System Document Control Number: K892425
- b. Radionics 76 Cambridge Street Burlington, Massachusetts 01803-0738 Product Name: X-Knife Radiosurgery System Document Control Number: K912630
{2}------------------------------------------------
Device Description: cass Linac Radiosurgery Hardware Couchmount Stereotactic Frame Support System
MIDCO has recently developed a quality Couchmount System (Patient Head Support System) for use in radiosurgery. This couchmount system complements the cass collimator set and enables MIDCO to provide its customers with a full set of instrumentation necessary for radiosurgery. The design of this couchmount is such that it will fit most linear accelerators with only slight modifications to the couchmount adapter assembly.
The cass Couchmount System has six axes of adjustment which allows for easy setup with the Linac couch/gantry system. An additional feature of the system is a unique laser measurement apparatus for precisely measuring an isocenter of a target volume for Linac-based radiosurgery/stereotactic radiotherapy. These two features also allow for the correction of alignment errors during treatment planning setup, such as errors inherent in the use of room lasers and errors due to couch shifts during patient positioning. It will also give an index of gantry sag or misalignment.
Most commercially available stereotactic headrings can be mounted to the cass Couchmount System by the use of a stereotactic frame holder made specifically for the headring. Although most headrings are similar, each has a slightly different adapter plate which rigidly holds the headring. The stereotactic localization features of most frames are similar, differing mainly in the organization of the frame's coordinate system and its mechanical dimensions.
This couchmount system is substantially equivalent to the couchmounts manufactured by F.L. Fischer and Radionics, Inc. The advantages of this couchmount system are the 6 axes of movement which allow for compensation for couch movement due to patient weight, the laser measurement system for accurate alignment to an isocenter, and an in vivo portal film method for verification. The system is designed for ease of use by the physicists who would normally do the setup and collimator beam profile measurements. MIDCO, through its appropriate consultants and instructors, will give the necessary on-site instructions for installation, measurements and quality control for the couchmount system. Each institution, with its radiation oncologists and radiation physicists, is ultimately responsible for accurate measurements and gathering of data for the use of this system.
{3}------------------------------------------------
SUMMARY OF THE SAFETY AND EFFECTIVENESS OF THE CASS LINAC RADIOSURGERY COUCHMOUNT SYSTEM
INTRODUCTION
Linac Radiosurgery is a method for the delivery of a high, controlled dose of radiation to a target volume within the head. The Linac Radiosurgery technique as originally described by Colombo, et al, 1985, and Hartmann, et al, 1985, requires the use of a rigid tertiary x-ray collimator system which is affixed to the Linear Accelerator for the purpose of controlling the collimation of x-ray beams. These systems have also required a means for rigid fixation of the patient's head for positioning, alignment and controlled delivery of focused radiation beams during the delivery of a high dose of ionizing radiation in single or multiple fractions. Most manufacturers of Radiosurgery hardware provide these two major components as parts of their radiosurgery systems. MIDCO has previously developed and obtained 510-K clearance for a system of radiosurgery collimators and floorstand (see K894263/A). As an additional complement to the system, MIDCO has recently developed a couchmount stereotactic frame positioning system which is the subject of this summary.
HISTORICAL BACKGROUND - PREDICATE DEVICES
Historically, several Linac-based techniques have been used for the delivery of high-energy photon beams as a means for radiosurgery/stereotactic radiotherapy. These various methods have been well reviewed by Podgorsak, et al., 1992, and, as therein noted, the system of multiple coplanar arcs is by far the most prevalent. These techniques have also varied with regard to the method of stereotactic head fixation (patient support) during treatment, wherein the aim is to accurately align a target volume (isocenter in the patient's head or body part) within the centers of rotation of the couch/gantry coordinate system of the linear accelerator. The most common methods employed include fixation of the stereotactic head frame to the radiotherapy couch ("Couch-Mount Systems"), or fixation to a rigid floorstand mounted in the revolving floor plate of the Linac radiotherapy couch ("Floorstand Systems") (see Podgorsak, 1992; Lutz, et al., 1988; and Colombo, et al, 1992). In the practice of the art, it is generally accepted that accuracy requirements for delivery of a radiation beam to a target volume isocenter should have fractions of a millimeter of accuracy, and some authors of floorstand systems have reported such accuracy in mechanical localization (see Friedman, 1989; Lutz, et al., 1988; and Winston and Lutz, 1988). In addition, floorstand systems have generally relied upon the use of phantom test targets placed at a given stereotactic coordinate, and then placed at the center of rotation of the Linac couch/gantry system. The Linac is then used to take test target x-rays with the couch and gantry in several different positions of rotation for the purpose of accurately aligning the stereotactic reference system (see Lutz, et al., 1988). Until recently, couchmount systems have generally not been as accurate as floorstand systems, and have relied upon fixed room laser lights projected on a phantom carriage device (see Fischer system device as described by Hartmann et al.,1985) for target localization within the Linac couch/gantry coordinate system.
{4}------------------------------------------------
Although head fixation to a floorstand mounted in the Linac floor turret has historically been regarded by many as the most reliable and accurate method, it greatly limits the accessibility to the patient's head. This is particularly noticeable for lesions in the posterior portions of the head which frequently require that the patient be treated in the prone position when using such floorstand-based systems. In addition, floorstand-based systems require that the gantry of the linear accelerator be specially rigged to protect against the possibility of accidental collision of the gantry with the floorstand during the execution of any treatment plan (see Lutz, et al, 1988). Couchmount systems, on the other hand, have the advantage of being able to treat a given patient with the potential of a full 360 degrees of gantry rotation and, thereby, allow the patient to be treated in a natural supine position, whilst also avoiding the need for gantry collision protection (see Colombo, et al, 1985, and Podgorsak, 1992).
The couchmount systems are desirable for the above-noted reasons, yet they have a disadvantage in that patient head fixation and stabilization within the coordinate system of the couch/gantry rotations of the Linac are generally not as accurate as floorstand-based systems. This is due to sagging and/or tilting of the couch which can occur in couchmount systems when weight, as weight of a patient's body, is applied to the couch tabletop after initial alignment. Such positional shifts are a source of error in couchmount systems. Attempts to solve this problem have been by cumbersome methods of bracing the couch tabletop (see Colombo, 1992), or the reliance on fixed, intersecting laser beams arranged as intersecting lines and emitted from laser alignment devices attached to the ceilings and walls of the room in which the linear accelerator is housed for the purpose of referencing the origin of the Linac couch/gantry axes of rotation. Such laser lights are commonly employed in the art of radiation therapy, and are also known in the art to frequently shift and require recalibration or alignment, and can introduce yet another source of error in target alignment within the couch/gantry coordinate system. Furthermore, such laser beams are usually 2-3 mm wide and may limit the accuracy of alignment because of the thickness of the alignment beams and parallax. This visual method relies on the use of positional laser beams for alignment and has the limitations noted herein. MIDCO has patented a technique which uses lasers for precise measurement of distances as a means for target volume localization within the couch/gantry coordinate system of a Linac-based radiosurgery/stereotactic radiotherapy system. This method is also a supplement to the aforementioned visual laser alignment method for couchmount systems.
Some manufacturers of radiosurgery equipment have also employed various techniques of target alignment verification (see Podgorsak, 1992, pp. 31-32). These alignment verification methods generally consist of the use of x-ray verification techniques. MIDCO utilizes a recognized in vivo portal x-ray imaging technique whereby images from the linear accelerator are taken with the patient on the treatment couch. We shall also describe this x-ray verification method as a part of this summary (see Serago, et al., 1991).
MIDCO, as other manufacturers, employs tertiary x-ray collimators in its radiosurgery system. These tertiary collimators consist of a set of 12, varying in bore diameters of 2.5 mm over a range of 12.5 to 40 mm. The collimator set consists of cerrobend filled inserts with a slightly tapered bore to correspond to beam divergence. The collimator inserts fit into the end of a tubular collimator housing that is attached to a collimator base plate which, upon being attached to the Linac gantry, can be fitted and variably adjusted to align the collimator tube to the central beam
{5}------------------------------------------------
of the Linac. This collimator set is substantially equivalent to other collimators on the market (510-K #K894263A). As designed, the collimator base permits the taking of portal x-rays with the collimators in place.
THE COUCHMOUNT SYSTEM
The Couchmount (patient support) System consists of:
- a) A couchmount adapter assembly
- b) A couchmount head holder and precision positioner
- c) A couchmount Precision Localization Box
- d) A couchmount laser measurement apparatus
- e) A couchmount test target assembly
These components are herein summarized and are designed for quick attachment and disconnection to the end of the couch table and support system. The couchmount head holder and precision positioner consist of a micrometer controlled assembly having 6 axes of adjustment (x, v. z. rotation, skew and tilt) which is adaptable to hold a variety of stereotactic frames. The Precision Localization Box fixes to the stereotactic frame and references the stereotactic frame's coordinate system. It can be used with room laser lights that are reflected off its sides (which have millimeter etchings) for visual alignment of stereotactic coordinates. It can also be used with anterior-posterior and lateral portal x-rays for in vivo stereotactic target confirmation (see Serago et. al. 1991; Siddon and Barth 1987). A unique feature of the Precision Localization Box is its use in conjunction with a patented laser measuring system (U.S. Patent number 5,553,112) that allows direct measurement of stereotactic targets. Lutz type phantom test targets are also possible with the system (Lutz, et al., 1988).
The laser measurement system is a patented improvement over other methods of stereotactic alignment, which rely primarily on the use of room laser lights or the use of phantom targets for referencing "hidden" stereotactic targets. This system has the additional advantage of permitting direct, and precise measurements of stereotactic targets by the use of a laser measuring device, having fractions of a millimeter of accuracy, which is attached to the Linac gantry and emits a laser beam which projects onto and reflects off the surfaces of a Precision Localization Box for the purpose of precisely measuring and centering isocenters within the couch/gantry coordinate system.
The cass Linac Radiosurgery Collimators and Couchmount Patient Support System are mechanically designed to allow quick and consistent assembly. In regular use, the system is assembled and tested with the Lutz-type phantom test target method to confirm the geometric accuracy of the system and target localization prior to any treatment.
SUMMARY OF TEST DATA
The overall accuracy of the mechanical movements and positioning of a given Linac radiosurgery system is of primary importance in the accurate delivery of a radiation dose to a target structure. For this reason, it was important to test the geometric accuracy of the Linac used to test the cass Collimators and Couchmount System. The initial accuracy of the test Linac for radiosurgery, a Varian 2100 C, was determined to be ≤1.0 mm for gantry rotational intersection with the rotational axis of the couch in all positions. The collimators and couchmount were
{6}------------------------------------------------
positioned according to guidelines in the cass Radiosurgery Hardware Setup Manual, and tests were done to determine the geometric accuracy of the complete radiosurgery system. The system was regularly assembled and disassembled over a one year period during which the system was tested.
The Lutz method (1988) of testing radiation beam accuracy was used to evaluate the collimators and couchmount as they would be integrated into similar Linacs for configuring a radiosurgery system. In these tests, random phantom targets throughout the extremes of the coordinate system of the Leksell frame and BRW frame were setup using the Lutz-type phantom pointer to simulate test targets. Radiographic images of the test targets were taken from various couch/gantry positions and examined for x, y, and z displacements (Dx, Dy, Dz). In summary, this data showed an error range of 0.48 mm ± 0.262 mm.
The Lutz phantom test is designed to test the mechanical accuracy of a given Linac radiosurgery system. These test results, using the cass collimators and couchmount, are an index of the accuracy that similarly configured systems would be capable of, given proper setup and configuration. These test results are considerably better than the established guidelines (≤1 mm) of many institutions using radiosurgery (see Podgorsak, et al. 1992) and substantially equivalent to predicate devices.
The overall accuracy of the system was tested according to guidelines recommended by Lutz, et al. (1988) and Serago, et al. (1991), in which a sheet of dosimetry film is placed in a humanoid phantom, and a target point is radiated and the isocenter of the multiple intersecting beams is compared to the target position of the phantom target point indicator. The target point is localized with both angiographic and CT imaging techniques. The accuracy of the Linac System used in these phantom tests was determined to be 0.5 mm ± 0.3 mm. This test is a composite test of the overall accuracy of the Linac radiosurgery system and incorporates errors due to CT image slice thickness and the accuracy of the angiographic localizer for determining hidden targets.
Port film tests were done according to the method recommended by Serago, et. al. (1991) (representing the Fischer System). The Laser Measurement System in conjunction with the Precision Localization Box was also used to check target alignment. Using this method, A-P and lateral port films were taken at various predetermined target sites set with the aid of the Laser Measurement System. The difference between the intended target and calculated target using this in vivo technique were determined. Total alignment error for in viva verification tests was 0.492 mm +/- 0.155 mm.
These test results are within the established published guidelines as noted by representatives of the Fischer System (Hartmann, et al., 1985, and Serago, et al., 1991), and the Radionics System (Winston and Lutz, 1988). Representatives of the Fischer System have reported localization accuracy of +/- 0.5 mm and representatives of the Radionics System have reported a localization accuracy of 0.5 mm +/- 0.2 mm. The cass Linac Couchmount System is substantially equivalent to predicate devices, and, with proper setup and quality control, is safe and effective. To date, a number of patients have been treated with this system, which has worked well in all instances.
The cass Couchmount System is designed to attach to the linear accelerator manufactured by Varian and Associates of 3045 Hanover Street, Mail Stop H-055, Palo Alto, California, 94304.
{7}------------------------------------------------
The system is adaptable to other linear accelerators by some slight modifications of the couchmount adaptor assembly. MIDCO submits a "Linac Site Information" questionnaire to each site for specifics on their linear accelerator and stereotactic hardware to assure proper configuration of the hardware for a specific site.
MIDCO provides each user institution with installation manuals for their cass Radiosurgery System and instructions on commissioning its system. A MIDCO installation team is present for the initial setup and alignments of the radiosurgery system, and specific training and instructions are given to the site's designated physicists and users. MIDCO personnel can be present during the first clinical case to insure proper use of the system. The user institution is encouraged to communicate to MIDCO any concerns it may have regarding the use of the System, and MIDCO provides the user with report for notifying MIDCO of any perceived problems or concerns.
{8}------------------------------------------------
Bibliography to the Summary of Safety and Effectiveness
- A) Articles related to Comparative Specifications of (Similar) Substantially Equivalent Devices
Colombo, F., et al., "External Stereotactic Irradiation by Linear Accelerator," Neurosurgery, 16: 154 (1985).- Relates to original radiosurgery system
Colombo, F., et al., "Linear Accelerator Radiosurgery of Cerebral Arteriovenous Malformations," Neurosurgery, 24: 833 (1989) .- Relates to original radiosurgery system
Friedman, W.A., & Bova, F.J., "The University of Florida Radiosurgery System," Surg Neurol, 32: 334 (1989) .- Relates to Phillips SRS 200, 510(k)#K891292
Hartmann, Gunther, et al., "Cerebral Radiation Surgery Using Moving Field Irradiation at a Linear Accelerator Facility," Int J Radiat Oncol Biol Phys, 11: 1185-1192 (1985) -- Relates to Fischer SRS system, 510(k)#K892425
Lutz, W., et al., "A System of Stereotactic Radiosurgery with a Linear Accelerator," Int J Radiat Oncol Biol Phys, 14: 373-381 (1988) -- Relates to Radionics X-Knife, 510(k)# K912630
Podgorsak, E., et al., "Radiosurgery with High Energy Photon Beams: A Comparison Among Techniques," Int J Radiat Oncol Biol Phys, 16: 857-868 (1989).
Rice, R.K .; Hansen, J.L .; Svensson, G.K .; "Measurements of Dose Distributions in Small Beams of 6 MV X-Rays," Phys Med Biol, 32: 1087 (1987).
Winston, K .; Lutz, W .; "Linear Accelerator as a Neurosurgical Tool for Stereotactic Radiosurgery," Neurosurgery (1988) -- Relates to Radionics X-Knife, 510(k)#K912630
- B) Articles related to Comparative in vivo Data (Hardware related studies)
Hartmann, Gunther, et al., "Precision and Accuracy of Stereotactic Convergent Beam Irradiations from a Linear Accelerator," Int J Radiat Oncol Biol Phys, 28: 481-492 (1993).
Pike, G., et al., "Dose Distributions in Radiosurgery," Medical Physics, 17: 296-304 (1990).
Serago, et al., "Stereotactic Target Point Verification of an X-Ray and CT Localizer," Int J Radiat Oncol Biol Phys, 20: 517-523 (1991).
Serago, et al., "Tissue Maximum Ratios (and Other Parameters) of Small Circular 4, 6, 10, 15 and 24 MV X-Ray Beams for Radiosurgery," Phys Med Biol, 37: 1-13 (1992).
Siddon, R.L .; Barth, L.N .; "Stereotactic Localization of Intracranial Targets," Int J Radiat Oncol Biol Phys, 13: 1241 (1987).
- C) Articles related to Clinical Studies
Barcia-Salorio, J., et al., "Radiosurgical Treatment of Carotid-Cavernous Fistula," Appl Neurophysiol, 45: 520 (1982).
Betti, O., et al., "Stereotactic Radiosurgery with the Linear Accelerator: Treatment of Arteriovenous Malformations," Neurosurgery, 24: 311 (1989).
{9}------------------------------------------------
Larson, D., et al., "Stereotactic Irradiation of Brain Tumors," Cancer, Feb. 1 Supplement, 792-799 (1990).
Larsson, B .; Leksell, L .; Rexed, B .; "The Use of High Energy Protons for Cerebral Surgery in Man," Acta Chir Scand, 125: 1 (1963).
Larsson, B .; Linden, K .; Sarby, B .; "Irradiation of Small Structures Through the Intact Skull," Acta Radio-Logica: Therapy, Physics and Biology, 13: 512 (1974).
Leavitt, D., et al., "Dynamic Field Shaping to Optimize Stereotactic Radiosurgery," Int J Radiat Oncol Biol Phys, 19, Supplement 1: 133 (1990).
Loeffler, J .; Alexander, E .; "The Role of Stereotactic Radiosurgery in the Management of Intracranial Tumors," Oncology, 4: 21-41 (1990).
Phillips, M.H., et al., "Comparison of Different Radiation Types and Irradiation Geometries in Stereotactic Radiosurgery," Int J Radiat Oncol Biol Phys, 118: 211-220 (1990).
Pozza, F., et al., "Low-Grade Astrocytomas: Treatment with Unconventional Fractionated External Beam Stereotactic Radiation Therapy," Radiology, 171: 565 (1989).
Saunders, W., et al., "Radiosurgery for Arteriovenous Malformations of the Brain Using a Standard Linear Accelerator: Rationale and Technique," Int J Radiat Oncol Biol Phys, 15: 441-447 (1988).
Sturm V., et al., "Stereotactic Percutaneous Single Dose Irradiation of Brain Metastases with a Linear Accelerator," Int J Radiat Oncol Biol Phys, 13: 279 (1987).
Note: There are now numerous articles in various radiotherapeutic and neurosurgical journals which relate to the methods and indications for radiosurgery, its principles and practice.
N/A