K Number
K961404

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-06-27

(77 days)

Product Code
Regulation Number
870.2850
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MX960 reusable pressure transducer is used to convert hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals which can then be displayed using separate monitoring equipment.

Device Description

The MX960 reusable pressure transducer is a non-fluid pathway device. A pressure waveform is conveyed through a disposable dome which is coupled to the reusable pressure transducer during pressure monitoring. Both the transducer and dome have mutually exclusive diaphragms. Accordingly, fluid contact is limited to the dome portion of the system. The materials which comprise the MX960 have been aggressively tested per the ANSVAAMI/ISO 10993 "Biological Evaluation of Medical Devices" and the "Tripartite Biocompatibility Guidance for Medical Devices". All materials have successfully met these standards.

AI/ML Overview

The Medex, Inc. MX960 Reusable Pressure Transducer underwent performance testing against the ANSI/AAMI BP22-1994 standard for blood pressure transducers.

Here's a breakdown of the acceptance criteria and the study details:

1. Acceptance Criteria and Reported Device Performance

PARAMETERAAMI SPECIFICATION (Acceptance Criteria)MX960 Performance (Reported Device Performance)
Sensitivity - µV/V/mmHgDisclosure5.0 Nominal
Unbalance - mmHg0 ± 750 ± 50
Zero Drift - mmHgDisclosure≤ 2 in 4 hrs after 5 min
Input Impedance - OHMS< 3000630 Nominal
Output Impedance - OHMS> 200300 ± 10%
Calibration Requirement - mmHgNot Specified100 ± 3
Accuracy - mmHg< ± 1% or < ±3%meets or exceeds AAMI
Compensated Temperature Range - °C15 to 4015 to 40
Storage Temperature Range - °C-25 to 70-25 to 70
Temperature Coefficient of Sensitivity - %/°CDisclosure0 ± 0.1
Temperature Coefficient of Offset - mmHg/°CDisclosure0 ± 0.3
Operating Pressure - mmHg-30 to 30030 to 300
Overpressure - mmHg-400 to 4000-400 to 4000
Volumetric Displacement - µL/100mmHgNot SpecifiedNot Specified
Symmetry -1 ± 5%1 ± 5%
Defibrillator Withstand - per AAMI5 in 5 min at 360J5 in 5 min at 360J
Leakage Current of 115 VAC at 60 Hz - µA< 10< 5
Light Sensitivity - per AAMIDisclosure< 1mmHg shift
Mechanical Shock Withstand - per AAMIDisclosure3 falls from 1m
Phase Shift - per AAMI< 5°< 5°

2. Sample size used for the test set and the data provenance:

  • The document does not explicitly state the sample size used for the test set in terms of the number of individual devices or the duration/number of tests performed for each parameter. It indicates that the MX960 was "aggressively tested."
  • The data provenance is not specified (e.g., country of origin, retrospective or prospective). It is an in-house performance test against a recognized standard.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This device is a physical transducer, not an AI or diagnostic imaging device that requires expert interpretation for ground truth. The "ground truth" for its performance is established by standardized measurements against the ANSI/AAMI BP22-1994 specifications using calibrated testing equipment. Therefore, no human experts are directly involved in establishing "ground truth" in the way they would be for image analysis or disease diagnosis.

4. Adjudication method for the test set:

  • Not applicable, as this is a performance test against objective engineering specifications, not a diagnostic task requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is a performance test for a reusable pressure transducer, not a device involving AI or human interpretation in a diagnostic context.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This is a standalone performance test of a physical medical device. The "algorithm" here is the transducer's internal mechanism, and its performance is measured directly.

7. The type of ground truth used:

  • The ground truth is based on the ANSI/AAMI BP22-1994 standard specifications for blood pressure transducers. This standard defines the acceptable ranges and limits for various operational and performance parameters.

8. The sample size for the training set:

  • Not applicable. This is a physical device performance test, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable for the reasons mentioned above.

{0}------------------------------------------------

K961404

510(k) Summary

Image /page/0/Picture/3 description: The image shows the logo for "medexinc.". The logo is black and white. The logo has a symbol on the left side that looks like a stylized flower or star. The text "medexinc." is in a bold, sans-serif font.

This summary regarding 510(k) safety and effectiveness and being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

§ 807.92 (a)(1) Submitter's (and Contact) Names, Address, Telephone No., Summary Date

  • · John Toomey Senior Project Engineer Medex, Inc. 6250 Shier-Rings Road Dublin, Ohio 43017 (614) 791 5415
  • · 04.09.96

& 807.92 (a)(2) Device Name (Including Trade Name), Common Name, Classification Name

  • · MX960
  • · Reusable Pressure Transducer
  • · Transducer, Blood Pressure, Extravascular (CV, 74 DRS)

$ 807.92 (a)(3) Legally Marketed Predicate Device to Which Equivalence is Claimed

  • · Medex, Inc.'s MX860 NOVATRANS IT reusable pressure transducer ( K95 1 129).

$ 807.92 (a)(4) Description of the Premarket Notification Device

  • · The MX960 reusable pressure transducer is a non-fluid pathway device. A pressure waveform is conveyed through a disposable dome which is coupled to the reusable pressure transducer during pressure monitoring. Both the transducer and dome have mutually exclusive diaphragms. Accordingly, fluid contact is limited to the dome portion of the system.
  • · The materials which comprise the MX960 have been aggressively tested per the ANSVAAMI/ISO 10993 "Biological Evaluation of Medical Devices" and the "Tripartite Biocompatibility Guidance for Medical Devices". All materials have successfully met these standards.

§ 807.92 (a)(5) Intended Use

  • · The MX960 reusable pressure transducer is used to convert hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals which can then be displayed using separate monitoring equipment.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Medex Inc. The logo consists of a symbol on the left, followed by the text "medexinc." The symbol on the left appears to be four stylized figures arranged in a diamond shape. The text "medexinc" is in a bold, sans-serif font, and there is a registered trademark symbol after the "c".

S 807.92 (a)(6) Technical Characteristics Summary

PropertyPredicate MX860(MX860)
Intended UseThe MX860 reusable pressure transducer's intended use is to convert hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals which can then be displayed using separate monitoring equipment.The MX860 reusable pressure transducer's intended use is to convert hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals which can then be displayed using separate monitoring equipment.
Pressure Sensor ComponentPiezoresistive Wheatstone bridge networkPiezoresistive Wheatstone bridge network
Isolated Fluid PathYesYes
Performance Specific AAMIMeets or exceeds ANSI/AAMI BP22-1994Meets or exceeds ANSI/AAMI BP22-1994
Transducer HousingPolycarbonatePolycarbonate
Housing Diamter0.657"0.657"

§ 807.92 (b)(1), (b)(3) Performance Testing Assessment

  • · The MX960 was aggressively tested against the American National Standard for Blood Pressure Transducers, ANSI/AAMI BP22-1994. In all cases, the MX960 meets or exceeds AAMI specifications. The following table list the AAMI requirements, and the respective performance of the MX860 and the MX960.
PARAMETERPREDICATE MX860MX960AAMI SPECIFICATION
Sensitivity - µV/V/mmHg5.0 Nominal5.0 NominalDisclosure
Unbalance - mmHg0 ± 500 ± 500 ± 75
Zero Drift - mmHg≤ 2 in 4 hrs after 5 min≤ 2 in 4 hrs after 5 minDisclosure
Input Impedance - OHMS630 Nominal630 Nominal< 3000
Output Impedance - OHMS300 ± 10%300 ± 10%> 200
Calibration Requirement - mmHg100 ± 3100 ± 3Not Specified
Accuracy - mmHgmeets or exceeds AAMImeets or exceeds AAMI< ± 1% or < ±3%
Compensated Temperature Range - °C10 to 4515 to 4015 to 40
Storage Temperature Range - °C25 to 70-25 to 70-25 to 70
Temperature Coefficient of Sensitivity - %/°C0 ± 0.10 ± 0.1Disclosure
Temperature Coefficient of Offset - mmHg/°C0 ± 0.30 ± 0.3Disclosure
Operating Pressure - mmHg30 to 30030 to 300-30 to 300
Overpressure - mmHg-700 to 4000-400 to 4000-400 to 4000
Volumetric Displacement - µL/100mmHgNot SpecifiedNot SpecifiedNot Specified
Symmetry -1 ± 5%1 ± 5%1 ± 5%
Defibrillator Withstand - per AAMI5 in 5 min at 360J5 in 5 min at 360J5 in 5 min at 360J
Leakage Current of 115 VAC at 60 Hz - µA< 3< 5<10
Light Sensitivity - per AAMI< 1mmHg shift< 1mmHg shiftDisclosure
Mechanical Shock Withstand - per AAMI3 falls from 1m3 falls from 1mDisclosure
Phase Shift - per AAMI< 5°< 5°< 5°

§ 870.2850 Extravascular blood pressure transducer.

(a)
Identification. An extravascular blood pressure transducer is a device used to measure blood pressure by changes in the mechanical or electrical properties of the device. The proximal end of the transducer is connected to a pressure monitor that produces an analog or digital electrical signal related to the electrical or mechanical changes produced in the transducer.(b)
Classification. Class II (performance standards).