K Number
K961375

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-06-18

(70 days)

Product Code
Regulation Number
862.1340
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DIA SCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip for semiquantitative urinalysis. DIA SCREEN® Reagent Strips have been developed to allow health care professionals to test semiguantitatively for patient urine levels of specific gravity. ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, billirubin, and urobilinogen by visual comparison with a color chart.

Device Description

DIA SCREEN® Reagent Strips are plastic strips to which are affixed reagent areas. Several product configurations are included with this submission, but all configurations are limited to one, some or all of the following tests: specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen,

The reagent test areas on the DIA SCREEN® Reagent Strips are ready to use upon removal from the bottle and the entire reagent strip is disposable after use. The strips are to be read visually, requiring no additional laboratory equipment for testing.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The device is considered substantially equivalent to existing marketed products, with the primary acceptance criterion being a high degree of correlation with these predicate devices and confirmatory tests. The general acceptance criterion for this device, which is a visual reagent test strip, is established through comparison to predicate devices, where "90% of all observations are within one color block."

Analyte (Test)Acceptance Criteria (Agreement within 1 Color Block with Predicate Devices or Confirmatory Test)DIA SCREEN Lot 00226A PerformanceDIA SCREEN Lot 00226B Performance
Specific Gravity≥ 90%96.5%96.5%
Ketone≥ 90%100%100%
Glucose≥ 90%100%100%
Protein≥ 90%100%100%
Blood≥ 90%98.3%98.3%
Nitrite≥ 90%100%100%
pH≥ 90%100%100%
Ascorbic Acid≥ 90%98.2%100%
Bilirubin≥ 90%100%100%
Urobilinogen≥ 90%100%100%

All analytes meet or exceed the "90% of all observations are within one color block" criterion when adjusted for confirmatory test results.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The number of urine samples varied slightly by analyte, but generally the study used:
    • 58 samples for Specific Gravity, Ketone, Glucose, Blood, and pH.
    • 57 samples for Protein, Nitrite, and Ascorbic Acid.
    • 48 samples for Bilirubin.
    • 54 samples for Urobilinogen.
  • Data Provenance: The study used fresh urine samples obtained at the medical facility (University of Minnesota Department of Medical Technology where the evaluation was conducted). This indicates prospective data collection from a clinical setting within the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number of individuals who performed the visual readings or those who conducted the confirmatory tests as "experts" in the sense of specialized clinical diagnosticians establishing ground truth.

However, the study was conducted "under the direction of Nancy Brunzel, C.L.S., M.T., at the University of Minnesota Department of Medical Technology."

  • Qualifications of Nancy Brunzel: C.L.S., M.T. (Certified Laboratory Scientist, Medical Technologist). These qualifications indicate expertise in clinical laboratory procedures and analysis.
  • Implication for "Experts": While not explicitly stated how many readers were involved, it's reasonable to infer that the visual readings were performed by competent laboratory personnel, likely medical technologists or similar professionals, under Ms. Brunzel's supervision. The "ground truth" was established not solely by visual reading but significantly by confirmatory reference methods when discrepancies arose (see point 7).

4. Adjudication Method for the Test Set

The primary "adjudication" or reconciliation method was a confirmatory testing strategy.

  • The study compared two lots of DIA SCREEN® strips against Ames Multistix 10 SG and Behring Rapignost strips.
  • "When data between the two products did not agree within one color block, an alternate procedure (confirmation test)" was used.
  • This implies a 2-way comparison (DIA SCREEN vs. Predicate) with a third, independent, and typically more accurate confirmatory method acting as the adjudicator for discrepancies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
  • This device is a visual dip-and-read non-AI based test strip. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here. The study focuses on the performance of the DIA SCREEN® strips themselves against predicate devices and confirmatory tests, as read apparently by trained laboratory personnel.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

  • No, a standalone (algorithm only) performance study was not done.
  • This device relies on visual comparison with a color chart by a human reader. It is not an automated or AI-driven diagnostic system; it explicitly states "by visual comparison with a color chart" and "requiring no additional laboratory equipment for testing."

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the test set was established using a combination of:

  • Predicate device results: Initial comparison was made against Ames Multistix 10 SG and Behring Rapignost.
  • Confirmatory (Reference) laboratory methods: For any discrepancies (results not within one color block) between the DIA SCREEN® strips and the predicate devices, a specific, established confirmatory test was employed. Examples include:
    • Specific Gravity: Refractometer
    • Ketone: Acetest
    • Glucose: Chemstrip Reagent Strip (from Boehringer Mannheim)
    • Protein: SSA (Sulfosalicylic Acid test)
    • Blood: Microscopic Evaluation
    • Nitrite: Microscopic Evaluation
    • pH: pH Meter
    • Ascorbic Acid: Urispec Reagent Strip (from Henry Schein)
    • Bilirubin: Ictotest
    • Urobilinogen: Chemstrip Reagent Strip (from Boehringer Mannheim)

This multi-faceted approach, heavily relying on established reference methods for discrepant results, serves as a strong form of ground truth.

8. The Sample Size for the Training Set

The document describes a clinical evaluation study (which serves as the "test set" in this context) but does not explicitly mention a separate "training set" for the DIA SCREEN® Reagent Strips. This is common for predicate-based 510(k) submissions, especially for visual assays. The implicit "training" for such devices often comes from the established clinical utility of the predicate devices and the internal development and quality control processes of the manufacturer to ensure the strips produce expected reactions.

9. How the Ground Truth for the Training Set Was Established

As no specific training set is mentioned in the provided text, the method for establishing its ground truth is also not described. The focus of this 510(k) submission is on demonstrating substantial equivalence through a direct comparison study (test set) against already established and approved predicate devices.

{0}------------------------------------------------

K961375

Summary of Safety and Effectiveness

April 4, 1996

The following information is provided as a summary of safety and effectiveness information for the DIA SCREEN® Reagent Strip System.

  • Common Name: Visual Reagent Test Strip for Urinalysis (a)
    Trade/Proprietary Name: DIA SCREEN® Reagent Strips, in many different configurations with as many different product codes. DIA SCREEN® is the registered trademark of Dia-Screen Incorporated, a wholly owned subsidiary of Genesis Labs, Inc.

  • Establishment Registration Number 2183670 [b]
    Genesis Labs, Inc. 5182 West 76th Street Minneapolis, MN 55439-2904

  • Contact Person: Bert Walter (c)

  • Intended Use [d]

The DIA SCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip for semiquantitative urinalysis. DIA SCREEN® Reagent Strips have been developed to allow health care professionals to test semiguantitatively for patient urine levels of specific gravity. ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, billirubin, and urobilinogen by visual comparison with a color chart.

  • Product Description [e]
    DIA SCREEN® Reagent Strips are plastic strips to which are affixed reagent areas. Several product configurations are included with this submission, but all configurations are limited to one, some or all of the following tests: specific gravity, ketone, glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen,

The reagent test areas on the DIA SCREEN® Reagent Strips are ready to use upon removal from the bottle and the entire reagent strip is disposable after use. The strips are to be read visually, requiring no additional laboratory equipment for testing.

{1}------------------------------------------------

The directions must be followed exactly. Accurate timing is essential to provide optimal results. The reagent strips must be kept in the original bottle containing the desiccant with the cap tightly closed to maintain reagent activity. To obtain optimal results, it is necessary to use fresh, well-mixed, uncentrifuged urine.

Dia-Screen Corporation, a subsidiary of Genesis Labs, Inc., previously received determinations of substantial has equivalence for complete lines of urine reagent strips which have been marketed under the Bioscan (K940043) and DIA SCREEN (K952971) label.

The DIA SCREEN® Reagent Strips will be manufactured by Genesis Labs for distribution by Dia-Screen. The intended use of the strips which are the subject of this new premarket notification is exactly the same as the strips which were reviewed previously (K940043 and K952971).

  • Substantial Equivalence ്രി
    The intended use of the strips which are the subject of this new premarket notification is exactly the same as the strips which were reviewed previously (K940043 and K952971).

The design and function of the new DIA SCREEN® Reagent Strips are the same as the design and function of both the Bioscan and DIA SCREEN Reagent Strips which have been previously reviewed under premarket notifications K940043 and K952971.

The DIA SCREEN® Reagent Strips for Urinalysis are substantially equivalent to Ames Multistix 10 SG, Behring Rapignost. Bioscan and Boehringer Mannheim Chemstrip urine test strip systems (90% of all observations are within one color block).

Characteristics of the Ames Multistix 10 SG and Behring Rapignost systems are compared with the DIA SCREEN® system in the following table:

{2}------------------------------------------------

Strip NameDistributorReagents for:This 510 (k)DIA SCREEN®DIA SCREEN®Previous 510 (k)DIA SCREEN®DIA SCREEN®Ames Multistix 10 SGAmesRapignostBehring
SpecificGravityBromthymol blueDetergentNABromthymol bluePoly (methyl vinyl-ethermaleic anhydrideNA
KetoneSodium nitroprussideSodium nitroprussideSodium nitroprussideBufferSodium nitroprussideGlycineBuffer
GlucoseGlucose oxidasePeroxidasePotassium iodideGlucose oxidasePeroxidasePotassium iodideGlucose oxidasePeroxidasePotassium iodideBufferGlucose oxidasePeroxidaseTolidine hydrochlorideBuffer
ProteinTetrabromphenol blueCitric acidTetrabromphenol blueCitric acidTetrabromphenol blueBufferTetrabromphenol blueBuffer
Blood2,5 Dimethylhexane-2,5-dihydroperoxide3,3',5,5'-TetramethylbenzidineNADiisopropylbenzene-dihydroperoxide3,3',5,5'-TetramethylbenzidineCumene hydroperoxideTetramethylbenzidinedihydrochlorideBuffer
NitriteHydroxy(3)-1,2,3-tetrahydrobenzo-(h)quinolinep-Arsanilic acidNAp-Arsanilic acid1,2,3,4-Tetrahydro-benzo(h)quinoline-3-olBuffer4-Arsanilic acidN-(Naphthyl)-ethylenediammoniumDihydrochlorideBuffer
pHBromthymol blueMethyl redNABromthymol blueMethyl redBromthymol blueCresol redMethyl red
Ascorbic AcidPhosphomolybdic acidNANA2,6-Dichloro-phenol-indophenolBuffer salts
This 510 (k)Previous 510 (k)
Strip NameDIA SCREEN®DIA SCREEN®Ames Multistix 10 SGRapignost
DistributorDIA SCREEN®DIA SCREEN®AmesBehring
Reagents for:
Bilirubin2,4 Dichlorobenzene- diazonium saltNA2,4-Dichloroaniline- diazonium saltBuffer2,4-Dichlorobenzene- diazonium saltBuffer
UrobilinogenDimethylaminobenzaldehydeNADiethylaminobenzaldehydeFluorodiazoniumtetra- fluoroborateBuffer
This 510 (k)Previous 510 (k)N - MultistixRapignost
Strip NameDIA SCREEN®DIA SCREENAmesBehring
DistributorDIA SCREEN®DIA SCREENAmesBehring
Packaged withDesiccantYesYesYesYes
Control AvailableYesYesYesYes
Time required toread strips30 to 60 seconds60 seconds30 to 60 seconds60 seconds
StorageBetween 15 - 30 °C (59°-86° F). Do Not Store inrefrigerator or freezer. Donot expose to moisture,heat or light.Between 15 - 30 °C(59°-86° F). Do NotStore in refrigerator orfreezer. Do not exposeto moisture, heat orlightBetween 15°- 30° C (59°-86° F) Do Not Store in DirectSunlightBetween 15 - 30° C (59°- 86°F). Do Not Store inrefrigerator or freezer. Donot expose to moisture, heator light.
510 (k)This submissionK952971
Expected Values:
Specific GravityRandom urines vary from1.001-1.035. Twenty-fourhour urines from normaladults with normal dietsand fluid intake will havea specific gravity of1.016-1.022NARandom urines vary from1.001-1.035. Twenty-fourhour urines from normaladults with normal diets andfluid intake will have aspecific gravity of 1.016-1.022NA
KetoneShould not be detected innormal urine.Should not be detectedin normal urine.Normal specimens yieldnegative results.Should not be detected innormal urine
GlucoseNone found in normalurine. Concentrations of100 mg/dL may beconsidered abnormal.None found in normalurine. Concentrationsof 100 mg/dL may beconsidered abnormal.None found in normal urine.Concentration of 100 mg/dLmay be consideredabnormal.None found in normal urine.Concentrations of 100 mg/dLmay be consideredabnormal.
Strip NameDistributorExpected values:This 510 (k)DIA SCREEN®DIA SCREEN®Previous 510 (k)DIA SCREEN®DIA SCREEN®Ames Multistix 10 SGAmesRapignostBehring
ProteinNormal specimensordinarily contain someprotein (0-4 mg/dL).Persistent results of traceor higher indicatesignificant proteinuria.Normal specimensordinarily containsome protein (0-4mg/dL). Persistentresults of trace orhigher indicatesignificant proteinuria.Normal specimens ordinarilycontain some protein (0-4mg/dL). A color matchingany "+" block indicatessignificant proteinuriaNormal specimens ordinarilycontain some protein (0-4mg/dL). Persistent results of30 mg/dL or higher indicatesignificant proteinuria
BloodThe practical detectionlimit of this test is 5 to 10erythrocytes per microliterof urineNAThe significance of the tracereaction may vary amongpatients. Clinical judgmentis requiredThe practical detection limitof this test is 5 to 10erythrocytes per microliter ofurine
NitriteAny degree of pink colorafter 30 secondsindicates clinicallysignificant bacteriaNANormally, no nitrite isdetectable in urineConcentrations between .05and 0.1 mg/dL produce apink color, indicatingsignificant bacteria
pHNormal urine has a pH of6 and urine pH valuesgenerally range from 5 to8NABoth the normal andabnormal pH range is from 5to 9 pH unitsNormal urine has a pH of 6and urine pH valuesgenerally range from 5 to 9
Ascorbic AcidUrinary excretion ofascorbic acid can resultfrom the intake of vitaminCNANAAny color change after 60seconds indicates a positiveresult
BilirubinNo bilirubin is detectablein urine of healthypersonsNANormally, no bilirubin isdetectable in healthypersonsNo bilirubin is detectable inurine of healthy persons
UrobilinogenNormal range is 0.1 to 1.0mg/dLNANormal range is 0.1 to 1.0Ehrlich units per 100 mLNormal range is 0.5 to 1.0mg/dL

ﻤﺠﻌ

{3}------------------------------------------------

ﺳﺴ

{4}------------------------------------------------


{5}------------------------------------------------

ﺳﻌﻪ

{6}------------------------------------------------

Clinical Laboratory Evaluation of DIA SCREEN® Strips

The DIA SCREEN® Reagent Strip for Urinalysis is a dip-and-read test strip intended for use as an in-vitro diagnostic aid using urine specimens. DIA SCREEN® Reagent Strips have been developed to allow health care professionals to semiquantitatively determine levels of specific gravity, ketone. glucose, protein, blood, nitrite, pH, ascorbic acid, bilirubin, and urobilinogen in patient urine by visual comparison with a color chart of each concentration range. No additional reagents or laboratory equipment are required. These reagent strips are packaged in a plastic vial containing a desiccant. The test strips must be maintained tightly capped in the plastic vial to assure reagent activity. The directions-for-use must be followed exactly.

An independent laboratory evaluation of the new DIA SCREEN® Reagent Strips for Urinalysis was conducted under the direction of Nancy Brunzel, C.L.S., M.T., at the University of Minnesota Department of Medical Technology. The purpose of the laboratory evaluation was to establish the performance of the DIA SCREEN® Reagent Strip for Urinalysis when compared to "510 (k) approved" marketed urinalysis strip systems.

The study consisted of comparing the performance of two (2) manufactured lots of DIA SCREEN® Reagent Strips with one (1) lot of a marketed product (Ames Multistix 10 SG and Behring Rapignost).

Fresh urine samples were obtained at the medical facility. The comparison strips, Ames Multistix 10 SG and Behring Rapignost strips, and abnormal urine controls were furnished by Dia-Screen. When data between the two products did not agree within one color block, an alternate procedure (see table below) was used to test the sample.

AnalyteConfirmation TestManufacturer
Specific GravityRefractometerNA
KetoneAcetestAmes
GlucoseChemstrip Reagent StripBoehringer Mannheim
ProteinSSANA
BloodMicroscopic EvaluationNA
NitriteMicroscopic EvaluationNA
pHpH MeterNA
Ascorbic AcidUrispec Reagent StripHenry Schein
BilirubinIctotestAmes
UrobilinogenChemstrip Reagent StripBoehringer Mannheim

{7}------------------------------------------------

Fresh, well-mixed, and uncentrifuged urine from a mixed patient population was reacted with the DIA SCREEN® Reagent Strips and the results were read by visual comparison with the DIA SCREEN® color chart. Results of these readings were compared to results measured by Multistix 10 SG and Rapignost Strips.

A summary of the data obtained from the clinical evaluation described above is provided in the following tables (pages 9 and 10):

{8}------------------------------------------------

DIA SCREEN versus Ames and Rapignost Clinical Correlation

Tables I and II present the percentage of responses for the indicated lots of the DIA SCREEN and Ames and Rapignost Reagent Strips that are within the same color block and that are within plus or minus one color block. Table III presents a similar analysis for the two lots of the DIA SCREEN Reagent Strips.

DIA SCREEN Lot 00226A versus Ames Lot D397095 and Table I. Rapignost Lot 633962A

Same Color Block
Specific Gravity24/5841.4%
Ketone52/5889.6%
Glucose46/5879.3%
Protein41/5771.9%
Blood45/5877.6%
Nitrite56/5798.2%
pH23/5839.6%
Ascorbic Acid46/5780.7%
Bilirubin (lot 0306A)44/4891.7%
Urobilinogen49/5490.7%
Within 1 Color Block
53/5891.4%
58/58100%
58/58100%
57/57100%
53/5891.4%
57/57100%
58/58100%
56/5798.2%
46/4895.8%
50/5492.6%

Table II. DIA SCREEN Lot 00226B versus Ames Lot D397095 and Rapignost Lot 633962A

Same Color BlockWithin 1 Color Block
Specific Gravity25/5843.1%52/5889.6%
Ketone51/5887.9%58/58100%
Glucose49/5884.5%58/58100%
Protein43/5775.4%57/57100%
Blood46/5879.3%53/5891.4%
Nitrite56/5798.2%57/57100%
pH27/5846.5%58/58100%
Ascorbic Acid49/5786.0%57/57100%
Bilirubin (lot 0306B)44/4891.7%46/4895.8%
Urobilinogen49/5490.7%50/5492.6%

{9}------------------------------------------------

Table III. DIA SCREEN Lot 00226A versus DIA SCREEN Lot 00226B

Same Color Block
Specific Gravity54/5893.1%Within 1 Color Block
Ketone57/5898.3%୧୫।୧୫100%
Glucose55/5894.8%58/58100%
Protein55/5796.5%100%
Blood57/5898.3%57157100%
Nitrite57/57100%ടുക്കു100%
pH52/5889.6%57157100%
Ascorbic Acid54/5794.7%58158100%
Bilirubin (lots 0306A/B)48/48100%56/5798.2%
Urobilinogen54/54100%48/48100%
54/54100%

{10}------------------------------------------------

DIA SCREEN versus Ames and Rapignost Clinical Correlation

Adjusted to use confirmatory test results for discrepant results.

Tables IV and V present the percentage of responses of the DIA SCREEN strip within one color block of the Ames and Rapignost Test Strips or the confirmatory test method when discrepancies of greater than one color block arose. 96.5% agreement within one color block was obtained for all analytes.

DIA SCREEN Lot 00226A versus Ames Lot D397095 and Table IV. Rapignost Lot 633962A

Within 1 Color Block
Specific Gravity55/5896.5%
Ketone58/58100%
Glucose58/58100%
Protein57/57100%
Blood57/5898.3%
Nitrite57/57100%
pH58/58100%
Ascorbic Acid56/5798.2%
Bilirubin (lot 0306A)48/48100%
Urobilinogen54/54100%

DIA SCREEN Lot 00226B versus Ames Lot D397095 and Table V. Rapignost Lot 633962A

Within 1 Color Block
Specific Gravity55/58 96.5%
Ketone58/58 100%
Glucose58/58 100%
Protein57/57 100%
Blood57/58 98.3%
Nitrite57/57 100%
pH58/58 100%
Ascorbic Acid57/57 100%
Bilirubin (lot 0306B)48/48 100%
Urobilinogen54/54 100%

§ 862.1340 Urinary glucose (nonquantitative) test system.

(a)
Identification. A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.(b)
Classification. Class II.