K Number
K961238

Validate with FDA (Live)

Date Cleared
1996-06-28

(88 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Anatomical Region: Head, Body, Extremity, Spine, Neck, TMJ, and Heart
Nuclei excited: Hydrogen
Diagnostic Use: 2D/3D Imaging, MR Angiography

Device Description

The Software Version 12.0/12.1 Upgrade for the MRT-50A/MRT-150A adds new sequences and software enhancements to the existing MRT-50A/150A software version. Version 12.1 also adds a new hardware to increase the number of slices available for FastSE sequences.

AI/ML Overview

The provided 510(k) summary describes a software upgrade for an MRI device and focuses on establishing substantial equivalence to a previously cleared version. It does not contain a typical "study" in the sense of a clinical trial with acceptance criteria for diagnostic performance metrics (e.g., sensitivity, specificity, AUC). Instead, the performance is demonstrated by showing that safety and imaging parameters are either unchanged, improved, or remain within acceptable limits.

Here's an analysis of the provided information based on your requested categories:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" for diagnostic performance in the way newer medical device submissions might. Instead, it demonstrates that the device's technical characteristics and imaging performance either meet or exceed those of the predicate device (V11.9) or established NEMA guidelines.

CategoryAcceptance Criteria (Implied)Reported Device Performance
Safety Parameters
Maximum static field strengthB0 unchanged from predicate device.MRT-50A: 0.5T (No change) MRT-150A: 1.5T (No change)
Rate of change of magnetic fielddB/dt unchanged from predicate device.7.5T/sec, τ = 1000ms (No change for both models)
Radio frequency power depositionSAR remains below cleared levels of predicate device.0.3W/kg max SAR (MRT-50A) 0.8W/kg max SAR (MRT-150A) SAR remains below the cleared levels.
Acoustic Noise levelsMeasured in accordance with NEMA guidelines.MRT-50A: 80 dB MRT-150A: 105 dB worst case, 85 - 100 dB typical
Imaging Performance Parameters
Specification volumeComparable to predicate device.MRT-50A: Head: 14cmX14cmX14cm, Body: 125mm diameter cross section X 250mm longMRT-150A: 10cm dsv, 20cm dsv (No change implicitly, standard specs)
Signal to noise ratioPhantom images and clinical images presented for new sequences must show acceptable SNR.Sample phantom images and clinical images were presented for all new sequences. Signal-to-noise ratios for each phantom image were given. Clinical studies were presented for all new sequences. (Implies acceptable SNR was demonstrated though specific values are not provided in the summary)
Uniformity in three orthogonal planesUnchanged from predicate device.Unchanged
Slice Profile in three orthogonal planesUnchanged from predicate device.Unchanged
Geometric Distortion in three orthogonal planesUnchanged from predicate device (due to no change in gradient system).Unchanged (V12.0 and V12.1 do not change the gradient system, affecting linearity and thus geometric distortion).
Slice ThicknessNot applicable / Unchanged if applicable.Not Applicable (It implies no direct change to slice thickness capabilities from the V11.9, only new sequences utilizing previous thickness capabilities).
Inter-slice Cross-talkNot applicable / Unchanged if applicable.Not Applicable (presumably similar reasoning as slice thickness).
Overall EquivalenceSubstantially equivalent to predicate device."All safety and effectiveness attributes remain essentially unchanged." and "Version 12.0/12.1 Upgrade for the MRT-50A/150A is substantially equivalent to the V11.9 Software release for the MRT-50A/150A."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "test set sample size" in the context of a clinical performance study. The performance is evaluated based on:

  • Phantom images: These are standardized tests using physical phantoms, not patient data.
  • Clinical images: The document states "Clinical studies were presented for all new sequences," but does not provide details on the number of cases, patient demographics, data provenance (country of origin), or whether they were retrospective or prospective. Given the era (1996), these would likely have been internal validation cases rather than large, multi-center trials with detailed reporting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the summary. The "clinical images" presented were likely evaluated by internal radiologists or medical experts to confirm image quality and diagnostic utility, but no details on their number or qualifications are given.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided as there is no formal clinical trial or reader study described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A MRMC comparative effectiveness study was not conducted. This submission is for MRI software and hardware upgrades, not an AI or CAD (Computer-Aided Detection) system. Therefore, there is no AI assistance component or associated effect size reported.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is an MRI system, which inherently requires human operation and interpretation. It is not an algorithm designed for standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the safety and performance parameters listed:

  • Safety parameters (static field, dB/dt, SAR): Ground truth is based on physical measurements and engineering specifications, often against established standards or internal benchmarks.
  • Acoustic Noise: Measured in accordance with NEMA guidelines.
  • Imaging Performance Parameters (SNR, Uniformity, etc.): Ground truth is established through physical phantom measurements and subjective assessment of clinical images by imaging experts (implied). There is no mention of pathology or outcomes data for performance evaluation in this summary.

8. The sample size for the training set

This information is not applicable/not provided. The software upgrade does not describe a machine learning or AI component that would require a "training set" in the modern sense. The software development and optimization would have involved iterative testing and refinement, but not a distinct "training set" as understood for AI models today.

9. How the ground truth for the training set was established

This information is not applicable/not provided for the reasons stated in point 8.

{0}------------------------------------------------

TOSHIBA

الموالي الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع المو

TOSHIBA AMERICA MRI, INC.

K961238

280 UTAH AVENUE. SOUTH SAN FRANCISCO. CA 94080 PHONE: (415) 872-2722

IN 2 8 1006

510(k) Summary

SUMMARY OF SAFETY AND EFFECTIVENESS MRT-50A/150A SOFTWARE UPGRADE 12,0 and MRT-150A SOFTWARE UPGRADE 12,1

    1. Model Name: MRT-50A/MRT-150A Magnetic Resonance Device Accessory Device Name: MRT-50A/150A Upgrade 12.0, MRT-150A Upgrade 12.1 Trade/Proprietary Name:
  • #2936923 2. Establishment Registration:
  • TOSHIBA AMERICA MRI, INC. 3. Name and Address: 280 Utah Avenue South San Francisco, CA 94080
    • Contact Person: Steven M. Kay (714) 730-5000
  • DATE OF SUBMISSION: 4. March 27, 1996

DEVICE DESCRIPTION 5.

The Software Version 12.0/12.1 Upgrade for the MRT-50A/MRT-150A adds new sequences and software enhancements to the existing MRT-50A/150A software version. Version 12.1 also adds a new hardware to increase the number of slices available for FastSE sequences.

DEVICE CHARACTERISTICS: 6.

The V12.0/12.1 product is an upgrade to the MRT-50A/150A software release V11.9. V12.0/12.1 includes hardware upgrades, software changes and new sequences. The hardware upgrade in V12.1 increases the number of slices available for FastSE sequences. Version 12.0 software changes provide improved FastSE sequences and Multislab MRA. Version 12.1 software changes provide improved Fat Suppression sequences and STC Imaging.

SAFETY PARAMETERS 7.

MRT-50AMRT-150A
Maximum static field strength:0.5T1.5T
There are no changes to B0
Rate of change of magnetic field:7.5T/sec, $\tau$ = 1000ms7.5T/sec, $\tau$ = 1000ms
There are no changes to dB/dt
Radio frequency power deposition:0.3W/kg max SAR0.8W/kg max SAR
S.A.R. remains below the cleared levels.
Acoustic Noise levels:80 dB105 dB worst case85 - 100 dB typical

Acoustic noise data was measured in accordance with NEMA guidelines.

{1}------------------------------------------------

510(k) Summary (cont'd)

SUMMARY OF SAFETY AND EFFECTIVENESS MRT-50A/160A SOFTWARE UPGRADE 12,0 and MRT-150A SOFTWARE UPGRADE 12,1

8. IMAGING PERFORMANCE PARAMETERS

MRT-50AMRT-150A
Specification volume:Head: 14cmX14cmX14cmBody: 125mm diameter cross section X 250mm long10cm dsv20cm dsv
Signal to noise ratio:Sample phantom images and clinical images were presented for all new sequences. Signal-to-noise ratios for each phantom image were given. Clinical studies were presented for all new sequences.
Uniformity in three orthogonal planes:Unchanged
Slice Profile in three orthogonal planes:Unchanged
Geometric Distortion in three orthogonal planes:UnchangedV12.0 and V12.1 do not change the gradient system, especially with regard to linearity.It is the gradient linearity that affects geometric distortion.
Slice Thickness:Not Applicable
Interlice Snaring.Not Annlinghle

9. INTENDED USE

: 1

Anatomical Region:Head, Body, Extremity, Spine, Neck, TMJ, and Heart
Nuclei excited:Hydrogen
Diagnostic Use:2D/3D Imaging, MR Angiography

10. EQUIVALENCY INFORMATION

Version 12.0/12.1 Upgrade for the MRT-50A/150A is substantially equivalent to the V11.9 Software release for the MRT-50A/150A. The 510(k) submission for V11.9 received FDA clearance on January 26, 1995 (K941864). Some of the scanning techniques require that the Super Performance Package be installed on the MRI system. Version 11.1 with the Super Performance Package was cleared for the MRT-50A in K924808. Version 11.1 was cleared for the MRT-150A in K914012. The MRT-150A with the Super Performance Package received market clearance in K925286.

Version 12.0/12.1 provides an improved selection of FastSE sequences, adds the FastIR technique and improves the performance of the Vascular Imaging Package. All safety and effectiveness attributes remain essentially unchanged.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.