K Number
K961069

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-05-28

(71 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ArthroCare Urologic Multi-electrode Electrosurgery System is a high frequency electrosurgical device intended for use in general urological surgery to resect soft tissue and provide hemostasis.

Device Description

The ArthroCare Urologic Multi-electrode Electrosurgery System is a high frequency electrosurgical device. The System is comprised of three components: the UroWand, the UroWand Cable, and the Controller. The UroWand, UroWand Cable and the Controller are virtually identical to the ArthroCare Probe, Probe Cable and Controller determined substantially equivalent in K943450 for the ArthroCare Arthroscopic Electrosurgery System. The technology is also substantially equivalent to that of the ArthroCare Bipolar Loop Electrode System (K955531). The bipolar design of the ArthroCare UroWand incorporates the return electrode into the shaft of the device. The ArthroCare UroWand is designed to be operated in a physiological saline environment.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the ArthroCare Urologic Multi-electrode Electrosurgery System:

Summary of Acceptance Criteria and Supporting Study for ArthroCare Urologic Multi-electrode Electrosurgery System (K961069)

The provided document describes a premarket notification (510(k)) for the ArthroCare Urologic Multi-electrode Electrosurgery System. As such, the approach to "acceptance criteria" and "study" is focused on demonstrating substantial equivalence to predicate devices, rather than establishing de novo performance metrics with a dedicated clinical trial for a novel device.

The core acceptance criterion is substantial equivalence to existing legally marketed predicate devices in terms of safety and effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance

| Acceptance Criterion | Reported Device Performance (as demonstrated by comparison to predicates) |
| :------------------- | :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
| Safety: | |
| - Patient Ground Pad Requirement | Not Required: Bipolar design incorporates return electrode into the shaft, eliminating the need for a patient ground pad (unlike monopolar predicates). This is an improvement over predicates, implying enhanced safety by reducing risks associated with ground pads. |
| - Operating Environment | Physiological Saline Environment: Designed to operate in physiological saline, which is different from predicate products that require non-conductive environments (sterile water or glycine). This difference is presented as an advantage allowing equivalent efficacy with less power, implying no compromise in safety due to the new environment. |
| - Material Safety | Identical Materials: Materials used are identical to those of the ArthroCare Arthroscopic Electrosurgery System (K943450) and ArthroCare Bipolar Loop Electrode System (K955531). Predicate devices use stainless or titanium. This demonstrates established material safety. |
| - General Safety Standards | Meets Applicable Safety Requirements: The ArthroCare System meets applicable domestic and international safety requirements and standards. |
| Effectiveness: | |
| - Resection of Soft Tissue | Equivalent Performance (Bench & Animal Studies), Similar Depth of Alteration (Animal Studies): - Bench Testing: Demonstrated equivalent performance characteristics between the ArthroCare UroWand and predicate electrosurgical devices (ACMI Circon, Olympus America). This suggests equivalent ability to resect tissue. - In Vivo Arthroscopic Animal Studies: Demonstrated that the depth of soft tissue alteration with the ArthroCare Arthroscopy System (virtually identical to the urological system) is similar to that of a standard monopolar electrosurgery device. This indicates equivalent tissue manipulation capability. |
| - Hemostasis | Not explicitly detailed, but implied by 'equivalent performance characteristics' and 'substantial equivalence' to devices that provide hemostasis in urological surgery. Predicate devices are described as effective in "resecting soft tissue and provide hemostasis" The ArthroCare System is stated to be "intended for use in general urological surgery to resect soft tissue and provide hemostasis," assuming similar effectiveness to predicates. |

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated/quantified in terms of a specific number of cases or subjects.
    • Bench Testing: Implied to involve a comparison of device performance characteristics against predicate devices. The number of tests or specific samples is not provided.
    • In vivo Arthroscopic Animal Studies: Involved "animal studies," but the number of animals or specific experimental setup for the "test set" is not detailed.
  • Data Provenance:
    • Bench Testing: Presumably conducted by ArthroCare Corporation.
    • In vivo Arthroscopic Animal Studies: Conducted as "animal studies," likely a controlled laboratory setting.
    • Retrospective/Prospective: The document doesn't specify if the animal studies were prospective (designed specifically for this submission) or if it leverages retrospective data. Given the context of a 510(k) submission, new testing is often performed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not applicable in this context. The "ground truth" for this 510(k) submission is the established safety and effectiveness of the predicate devices. The studies (bench and animal) are designed to show that the new device performs similarly to these known, safe, and effective predicates, not to establish a new ground truth based on expert consensus for a novel condition. Independent experts are not mentioned in relation to establishing ground truth for the device's performance.

4. Adjudication Method for the Test Set

  • This is not applicable. There's no mention of human-reviewed test sets or adjudication processes in the way one would define it for diagnostic or screening device studies. The "adjudication" is essentially the comparison of performance against established predicate device characteristics.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. The document describes bench testing and animal studies comparing the device to existing technology, not human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as the device is an electrosurgical system, not an AI algorithm. Its performance is inherent to its physical and functional design, not an AI model.

7. The Type of Ground Truth Used

  • The "ground truth" implicitly used for comparison is the demonstrated safety and effectiveness of the legally marketed predicate devices (e.g., electrosurgery devices by Circon ACMI, Karl Storz, Olympus America).
    • For safety, this refers to their known history of safe use and compliance with standards.
    • For effectiveness, it refers to their known ability to "resect soft tissue and provide hemostasis" in urological procedures.
  • The studies (bench and animal) then serve to demonstrate the ArthroCare system performs equivalently against this established "ground truth" of the predicates.

8. The Sample Size for the Training Set

  • This information is not applicable. This is a hardware medical device submission; there is no "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set was Established

  • This information is not applicable for the same reasons as #8.

{0}------------------------------------------------

K961069

MAY 28 1996

ArthroCare Corporation Sunnyvale, California March 15, 1996

Summary of Safety and Effectiveness x

The ArthroCare Urologic Multi-electrode Electrosurgery System is a high frequency electrosurgical device intended for use in general urological surgery to resect soft tissue and provide hemostasis. The System is comprised of three components: the UroWand, the UroWand Cable, and the Controller. The UroWand, UroWand Cable and the Controller are virtually identical to the ArthroCare Probe, Probe Cable and Controller determined substantially equivalent in K943450 for the ArthroCare Arthroscopic Electrosurgery System. The technology is also substantially equivalent to that of the ArthroCare Bipolar Loop Electrode System (K955531).

Predicate electrode products and electrosurgical generators have been marketed since the 1950's by several different manufacturers such as Circon ACMI and Karl Storz. Other monopolar and bipolar electrosurgery devices have been cleared for market via 510/k)'s, such as the electrosurgery devices manufactured by Olympus America and Karl Storz-America, Inc. These products have been demonstrated to be safe and effective in resecting soft tissue during urological surgical procedures.

The ArthroCare Urologic Multi-electrode Electrosurgery System raises no new issues of safety or efficacy. The bipolar design of the ArthroCare UroWand incorporates the return electrode into the shaft of the device. This means that a patient ground pad is not required as with monopolar electrodes. The energy flows back to the Controller via the integral return electrode rather than through the patient's body to the grounding pad. In addition, the ArthroCare UroWand is designed to be operated in a physiological saline environment while the predicate products must be operated in a non-conductive environment, such as sterile water or glycine. This allows ArthroCare to achieve equivalent efficacy in tissue resection with less power requirements.

The materials used in the ArthroCare Urologic Multi-electrode Electrosurgery System are identical to those used in the ArthroCare Arthroscopic Electrosurgery System (K943450) and the ArthroCare Bipolar Loop Electrode System (K955531). Predicate devices use either stainless or titanium for the wire material of their electrosurgical electrodes. The materials used in the UroWand Cable are identical to those used in the Probe Cable (K943450 and K955531).

Bench testing has demonstrated equivalent performance characteristics between the ArthroCare UroWand and the ACMI Circon and Olympus America electrosurgical devices. In vivo arthroscopic animal studies have demonstrated that the depth of soft tissue alteration with the ArthroCare Arthroscopy System (which is

025

{1}------------------------------------------------

ArthroCare Corporation Sunnyvale, California Premarket Notification March 15, 1996

virtually identical to the ArthroCare Urological Multi-electrode Electrosurgery System), is similar to that of a standard monopolar electrosurgery device. In addition, the ArthroCare System meets applicable domestic and international safety requirements and standards.

By virtue of design, materials and function, the ArthroCare Urologic Multi-electrode Electrosurgery System is substantially equivalent to the predicate devices currently marketed in the United States.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).