K Number
K960964

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-09-27

(200 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ohmeda 7900 Anesthesia Ventilator provides mechanical ventilation for patients during surgery, as well as adiustment and monitoring various patient parameters-

Device Description

The 7900 Anesthesia Ventilator is a microprocessor based, electronically driven ventilator with a built in monitoring system for inspired oxygen, ainway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for compensation of compression losses, fresh gas contribution, and small leakage in the breathing absorber, bellows and system. User settings and microprocessor calculations catterns. User interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices.

AI/ML Overview

The provided document is a 510(k) summary for the Ohmeda 7900 Anesthesia Ventilator, dated February 29, 1996. It primarily focuses on demonstrating substantial equivalence to a predicate device (Ohmeda 7800 Anesthesia Ventilator) and adherence to various medical device standards.

This document does not contain information responsive to your request about acceptance criteria and a study proving a device meets those criteria, particularly within the context of AI/ML device performance or clinical studies involving human readers and AI assistance.

Specifically, the document lacks the following information that you requested:

  1. A table of acceptance criteria and the reported device performance: While it lists various standards the device complies with, it doesn't provide specific quantitative acceptance criteria for clinical performance (e.g., sensitivity, specificity, accuracy) or device performance parameters from a study.
  2. Sample size used for the test set and the data provenance: There is no mention of a test set, study participants, or data origin.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: As there's no clinical performance study described, there's no mention of experts or ground truth establishment.
  4. Adjudication method for the test set: Not applicable as no specific test set or study is described.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance: This document predates widespread AI in medical devices and does not describe any MRMC studies or AI assistance.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as it's not an AI/ML device.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable as it's not an AI/ML device.
  9. How the ground truth for the training set was established: Not applicable.

The document mainly focuses on:

  • Substantial equivalence to a predicate device.
  • Description of the device's function and monitored parameters.
  • Compliance with various medical device standards (e.g., ASTM, ISO, IEC, UL) which typically refer to electrical safety, electromagnetic compatibility, and basic functional requirements, not clinical performance metrics in a study.
  • Mention of a "rigorous software development process" and software validation, but this is a general statement about software quality assurance, not a specific clinical performance study.

In summary, the provided text does not contain the type of acceptance criteria and study details relevant to AI/ML device performance that your request is looking for.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "OHMEDA" in bold, black letters. A symbol resembling a snake is placed below the letters "OHM". The letters are closely spaced together, and the overall impression is one of a bold, vintage-style logo or heading.

K960964

THE BOC GROUP

February 29, 1996

Ohmeda Inc Ohmeda Drive PO Box 7550 Madison WI 53707-7550 608 221 1551

.

606 22
Subject:510(k) Summary of Safety and Effectiveness Informationfor the Ohmeda 7900 Anesthesia Ventilator.
Proprietary:Ohmeda 7900 Anesthesia Ventilator
Common:Continuous Ventilator
Classification:Ventilator, Continuous Class II - 21CFR868.5895 - 73CBK

The 510(k) summary of safety and effectiveness information is accordance with the requirements of SMDA 1990 and 1992.

The Ohmeda 7900 Anesthesia Ventilator is substantially equivalent to the currently marketed Ohmeda 7800 Anesthesia Ventilator.

The Ohmeda 7900 Anesthesia Ventilator provides mechanical ventilation for patients during surgery, as well as adiustment and monitoring various patient parameters-

Patient breathing waveform during volume mode or pressure mode VE Exhaled minute volume VTE Exhaled tidal volume O2 Percent of oxygen inspired fimin Units of breath rate per minute Pmax Maximum airway pressure during a patient breath Pmean Mean airway pressure, calculated every patient breath or 10 seconds whichever occurs first Pplateau Airway pressure measured at the end of inspiratory pause time.

The 7900 Anesthesia Ventilator is a microprocessor based, electronically driven ventilator with a built in monitoring system for inspired oxygen, ainway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for compensation of compression losses, fresh gas contribution, and small leakage in the breathing absorber, bellows and system. User settings and microprocessor calculations catterns. User interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices.

The software for the Ohmeda 7900 Anesthesia Ventilator has been developed following a rigorous software development process and has been fully specified and validated by Ohmeda.

The Ohmeda 7900 Anesthesia Ventilator complies with the following standards:

  • ASTM F1101 Sections 5.2, 5.3.2, 5.4.3, 5.5, 5.7.1, 6.2.2, 6.2.3, 6.6 and 6.8.1 1.
    1. ISO 10651 - Sections 50.3.1, 51.10 and 51.7
    1. IEC 601-1 - 1988, EN 60601-1
  • IEC 601-1-2, EN 60601-1-2 4.
    • CISPR 11 Group 1 (EN55011) ●
    • . EN 55011: "Limits and Methods of Measurement of Radio Interference Characteristics of Industrial, Scientific and Medical (ISM) Equipment"
    • IEC 801-2, 8kV air, 3kV contact
    • IEC 801-3, 3V/m
    • IEC 801-4, 2kV power line
    • IEC 801-5, 2kV line to earth, 1 kV line to line
    1. EN 60601-1-4
  • ર. DIN 13252/06 - 1984
    1. E-DIN 132252/01 - 1991
    1. prEN 740/12 - 1994
  • UL 2601 1994 9.
    1. IEC 68-2-3 Test Ca, IEC 68-2-1 Test Ab-Cold, IEC 68-2-2 Test Bb-Hot, IEC 68-2-34, IEC 68-2-34, IEC 68-2-37

The Ohmeda 7800 Anesthesia Ventilator and the Ohmeda 7900 Anesthesia Ventilator are substantially equivalent in technological characteristics, theory of operation and measurement methods. The Ohmeda 7900 Anesthesia Ventilator is the next generation of ventilator, utilizing the technological advances of today to bring Ohmeda ventilators to the next century.

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).