Intended Use

MENTOR® Advent™ A/B System is a dual function ultrasonic ophthalmic imaging system. The A-scan is used to measure the axial length of the eye. The system includes various programs for calculating the optical power of the IOL to be implanted during cataract surgery. Anterior chamber depth and lens thickness of the eye are also measured and displayed. The B-scan is used to image the internal structure of the eye including the opaque media and posterior pathology for the purpose of diagnosing pathological or traumatic conditions in the eve.

Device Description

MENTOR® Advent™ A/B System is an easy to use ophthalmic ultrasonic system, offering capabilities and versatility for both A-scan and B-scan operations.

The A-Scan biometric unit of the Advent System has the following features: Operational modes - Automatic and Manual Probe alignment - LED fixation light and audio feedback Multiple eye categories - Normal, Aphakic, Cataractous, Dense Cataractous and Pseudophakic IOL calculations - programs include the latest in SRK, Holladay and Hoffer Q Patient storage - up to 16 patients, including right and left eye User interface - standard computer keyboard Display - 9" high resolution white phosphor CRT Accessories - printer, video output

The B-Scan imaging unit of the Advent System has the following features: Scanning - 50 degrees sector scan probe, allowing eye view from wide variety of angles

Normal mode - requires selection of parameters Windowed mode - allows user selectable scanning angle Controls - depth, gain, contrast and brightness Display - 9" high resolution white phosphor CRT User interface - standard computer keyboard

Accessories - camera, printer. VCR

MENTOR Advent™ System uses the following transducer operations and characteristics:

A-Scan: There is a single ultrasound mode where the transducer is pulsed at a repetition frequency of approximately 5 Hz. The transducer is a non-scanning crystal with a frequency of approximately 8 Mhz. This system uses the same ultrasound transducer as in the predicate device.

B-Scan: There are two ultrasound modes, Normal and Windowed. The transducer is a single crystal mechanical sector probe with a center frequency of 11 MHz. It is housed in an oil-filled sealed chamber. The transducer sweeps 8 frames/sec. Sweep angle is 50°, pulse repetition frequency is 5.5 KHz and there are 275 transmitted lines per scan. In the Windowed mode, the size and position of the user selected window affects the pulse rate and spacing of the pulse, and thus the radiated field. Ultrasonic power measurements reported later have been made in the worst case condition,

AI/ML Overview

This 510(k) summary explicitly states that the determination of substantial equivalence is not based on an assessment of any performance data, clinical or non-clinical. Therefore, there is no study described in the provided text that proves the device meets any acceptance criteria.

Response based on provided text:

  1. A table of acceptance criteria and the reported device performance: Not provided. The submission states, "The determination of substantial equivalence is not based on an assessment of any performance data, clinical or non-clinical."

  2. Sample sized used for the test set and the data provenance: Not applicable as no performance study was conducted.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no performance study was conducted.

  4. Adjudication method for the test set: Not applicable as no performance study was conducted.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device described is an ophthalmic ultrasonic system, not an AI-assisted device. Additionally, no performance studies were conducted.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as no performance study was conducted.

  7. The type of ground truth used: Not applicable as no performance study was conducted.

  8. The sample size for the training set: Not applicable. This is a medical device, not an AI/ML algorithm requiring a training set in the conventional sense. No performance studies were conducted.

  9. How the ground truth for the training set was established: Not applicable. This is a medical device, not an AI/ML algorithm. No performance studies were conducted.

{0}------------------------------------------------

510(k) SUMMARY

K960765

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

K960765 The assigned 510(k) number is:

AUG - 8 1996

Submitted By:

David Downey Director, Quality Assurance & Regulatory Affairs Mentor O&O. Inc. 3000 Longwater Drive Norwell, MA 02061 Tel: (617) 871-8184 Fax: (617) 871-7785

Date Prepared: January 27, 1996

Device Name

Proprietary Name: MENTOR Advent™ A/B System

Common Name: A-B Scanner

Classification Name: Ultrasonic System, Imaging, Ophthalmic (86 HPR)

Substantial Equivalence Claim

The substantial equivalence device selected for comparison to this device is Teknar A/B III. Its 510(k) control number is K871536.

Device Description

MENTOR® Advent™ A/B System is an easy to use ophthalmic ultrasonic system, offering capabilities and versatility for both A-scan and B-scan operations.

The A-Scan biometric unit of the Advent System has the following features: Operational modes - Automatic and Manual Probe alignment - LED fixation light and audio feedback Multiple eye categories - Normal, Aphakic, Cataractous, Dense Cataractous and Pseudophakic IOL calculations - programs include the latest in SRK, Holladay and Hoffer Q Patient storage - up to 16 patients, including right and left eye User interface - standard computer keyboard Display - 9" high resolution white phosphor CRT Accessories - printer, video output

510(k) Summary: Mentor O&O Advent A/B System

{1}------------------------------------------------

<360165 p. 2 or 4

The B-Scan imaging unit of the Advent System has the following features: Scanning - 50 degrees sector scan probe, allowing eye view from wide variety of angles

Normal mode - requires selection of parameters Windowed mode - allows user selectable scanning angle Controls - depth, gain, contrast and brightness Display - 9" high resolution white phosphor CRT User interface - standard computer keyboard

Accessories - camera, printer. VCR

MENTOR Advent™ System uses the following transducer operations and characteristics:

A-Scan: There is a single ultrasound mode where the transducer is pulsed at a repetition frequency of approximately 5 Hz. The transducer is a non-scanning crystal with a frequency of approximately 8 Mhz. This system uses the same ultrasound transducer as in the predicate device.

B-Scan: There are two ultrasound modes, Normal and Windowed. The transducer is a single crystal mechanical sector probe with a center frequency of 11 MHz. It is housed in an oil-filled sealed chamber. The transducer sweeps 8 frames/sec. Sweep angle is 50°, pulse repetition frequency is 5.5 KHz and there are 275 transmitted lines per scan. In the Windowed mode, the size and position of the user selected window affects the pulse rate and spacing of the pulse, and thus the radiated field. Ultrasonic power measurements reported later have been made in the worst case condition,

Intended Use

MENTOR® Advent™ A/B System is a dual function ultrasonic ophthalmic imaging system. The A-scan is used to measure the axial length of the eye. The system includes various programs for calculating the optical power of the IOL to be implanted during cataract surgery. Anterior chamber depth and lens thickness of the eye are also measured and displayed. The B-scan is used to image the internal structure of the eye including the opaque media and posterior pathology for the purpose of diagnosing pathological or traumatic conditions in the eve.

{2}------------------------------------------------

K 960 165 P. 306

A

Substantial Equivalence Comparison

ﻪ ﻟـ ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ

MENTOR Advent™A/B SystemTeknar A/B III
IndicationThe A-scan is used to measure theaxial length of the eye. The systemincludes various programs forcalculating the optical power of theIOL to be implanted duringcataract surgery. Anterior chamberdepth and lens thickness of the eyeare also measured and displayed.The B-scan is used to image theinternal structure of the eyeincluding opaque media andposterior pathology for thepurpose of diagnosing pathologicalor traumatic conditions in the eye.Same
Transducer/Probe DesignA-Scan:Non-scanning crystal, 8 MHzfrequency, transducer in directcontact with eye surface.Same
B-Scan:Single crystal mechanical sectorprobe with 11 MHz centerfrequency focused at 25 mm.Housed in an oil-filled sealedchamber. 8 frames per sec. sweep,50°sector scan.Single crystal mechanical sectorprobe with 10 MHz centerfrequency focused at 18 mm.Housed in an oil-filled sealedchamber. 15 frames per sec. sweep50°sector scan.
Probe PhysicalsA-Scan:Weight - 5 oz., Length - 3.875",Diameter - 0.40".Same
B-Scan:Weight - 1.7 oz., Length - 4.75",Diameter - 0.625".Weight - 2.6 oz., Length - 6.00",Diameter - 0.625".
Scanning ModeA-Scan:Automatic & ManualAutomatic, Manual & Diagnostic
B-Scan:Normal and WindowedNormal
A-Scan Eye CategoriesNormal, Aphakic, Cataractous,Dense Cataractous andPseudophakic.Same
A-Scan IOL CalculationsSoftware programs for SRK,Holladay and Hoffer Qcalculations.Same
A-Scan InitializedVelocity1532 m/sec - vitreous and aqueous1642 m/sec - normal lens1629 m/sec - cataractous lensSame
A-Scan Probe AlignmentLED fixation light and audiofeedbackSame
MENTOR® Advent™ A/B SystemTeknar A/B III
B-Scan ControlsDepth, gain, contrast and brightness.Same
B-Scan Depth20/50/70 mm50/70 mm
B-Scan ScaleCalibrated to 1550 meters/sec.Same

510(k) Summary: Mentor O&O Advent A/B System

{3}------------------------------------------------

K960765 P. 4 of 4

The determination of substantial equivalence is not based on an assessment of any performance
data, clinical or non-clinical.

D~n D
Submitter's Signature

2-15-96
Date

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.