K Number
K960551

Validate with FDA (Live)

Date Cleared
1996-03-06

(27 days)

Product Code
Regulation Number
862.1370
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DSL Active hGH ELISA Kit was developed for the quantitative measurement of human growth hormone in human serum. The measurement of growth hormone in serum is used as a diagnostic aid in the evaluation of hGH deficiency disorders or acromegaly.

Device Description

Mouse monoclonal antibody "capture antibody" is immobilized to the inner surface of microtiter plate wells. Growth hormone in the standards or serum samples is "sandwiched" between the capture antibody and the anti-hGH monoclonal antibody conjugated to horseradish peroxidase enzyme.

AI/ML Overview

This document is a "SUMMARY OF SAFETY AND EFFECTIVENESS" for a diagnostic device, specifically the "DSL 10-1900 hGH ELISA Kit". It describes the device, its intended use, and a study comparing it to a predicate device to demonstrate substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in a quantitative format for the new device's performance. Instead, it compares the new device to a predicate device (Nichols Institute hGH IRMA) to show "substantial equivalence." The reported performance relates to this comparison.

Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (Comparison to Nichols Institute hGH IRMA)
Strong linear correlation with predicate deviceRegression analysis: Y = 1.06X + 0.25
High correlation coefficient with predicate deviceCorrelation coefficient: 0.99
Similar intended use as predicate device"These kits have the same intended use."

2. Sample size used for the test set and the data provenance

  • Sample Size: 41 patient samples
  • Data Provenance: The document does not specify the country of origin. It states "patient samples were collected," implying they are retrospective or a convenience sample collected specifically for the study. It doesn't indicate if they were collected prospectively.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. For an ELISA kit, "ground truth" would typically be established by laboratory reference methods or expert clinical interpretation of patient conditions, but the text focuses on comparing the new ELISA to another established ELISA.

4. Adjudication method for the test set

This information is not provided in the document, as the study is a direct comparison of two assays rather than a human-reader-based assessment requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This document describes an in-vitro diagnostic (IVD) device (ELISA kit) for measuring a biomarker, not an AI or imaging device that would typically involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The entire study described is a standalone performance of the DSL hGH ELISA kit compared to the Nichols Institute hGH IRMA kit. Both are automated or semi-automated laboratory assays, not requiring human "in-the-loop" interpretation in the same way an AI imaging algorithm might.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document implies that the Nichols Institute hGH IRMA served as the de-facto "ground truth" or reference method against which the new DSL hGH ELISA Kit was compared to establish substantial equivalence. It's a comparison against an existing, presumably validated, diagnostic assay.

8. The sample size for the training set

This information is not provided in the document. For an ELISA kit, training would typically involve assay development and optimization (e.g., reagent concentrations, incubation times), not a "training set" in the machine learning sense. The 41 samples discussed are for the equivalence comparison, not for training.

9. How the ground truth for the training set was established

This information is not applicable as there is no mention of a "training set" or "ground truth" for it in the context of this ELISA kit's development and validation in this document. The "ground truth" concept applies more directly to the comparison study described in point 7.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo that appears to be an abstract design. The logo is composed of thick, black lines forming shapes that resemble stylized letters. The overall design is compact and has a modern, minimalist aesthetic.

K960551/

Diagnostic Systems Laborotones 145 Medical Center Boulevil 1/etister lexas 77508 1217 116 Tel 713.332.967 Fox 713.554.429

Customer Assistonce Cer Tel 800.231.79 Fax 713.338.189

SUMMARY OF SAFETY AND EFFECTIVENESS

DSL 10-1900 hGH ELISA Kit. Name of Device: Enzyme-Linked Immunosorbent Assay, hGH. Classification Name: Analyte Name: Human Growth Hormone. Regulatory Class: l

Submitter: John Willis Diagnostic Systems Laboratories, Inc. 445 Medical Center Blvd. Webster. Texas 77598 Phone: (713) 332-9678

Date: February 7, 1996

The DSL Active hGH ELISA Kit was developed for the quantitative measurement of human growth hormone in human serum. Mouse monoclonal antibody "capture antibody" is immobilized to the inner surface of microtiter plate wells. Growth hormone in the standards or serum samples is "sandwiched" between the capture antibody and the anti-hGH monoclonal antibody conjugated to horseradish peroxidase enzyme.

The measurement of growth hormone in serum is used as a diagnostic aid in the evaluation of hGH deficiency disorders or acromegaly.

The DSL hGH ELISA is substantially equivalent to the Nichols Institute hGH IRMA. These kits have the same intended use.

To demonstrate substantial equivalence between the two assays, 41 patient samples were collected and assayed simultaneously by both methods. Samples were chosen with low medium and high levels of hGH. Linear regression analysis of the results obtained for the comparison yield the following results: V= 1.06x + 0.25 with a correlation coefficient of 0.99.

§ 862.1370 Human growth hormone test system.

(a)
Identification. A human growth hormone test system is a device intended to measure the levels of human growth hormone in plasma. Human growth hormone measurements are used in the diagnosis and treatment of disorders involving the anterior lobe of the pituitary gland.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.