K Number
K960548

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-04-26

(78 days)

Product Code
Regulation Number
872.3820
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AH® PLUS™ Root Canal Sealer is used for permanent sealing of root canals following established endodontic procedures. The predicate device, AH26® Root Canal Sealer has the same intended use.

Device Description

AH® PLUS™ Root Canal Sealer is a two-component paste: paste root canal sealer based on epoxy-amine resin chemistry. This easy-to-mix sealer adapts closely to the walls of the prepared root canal and provides outstanding long-term dimensional stability with minimal shrinkage upon setting.

The final product consists of two components, the epoxy resin paste (Paste A) and the amine-containing paste (Paste B) portions which are mixed prior to insertion into the root canal.

AI/ML Overview

The provided text describes a 510(k) summary for a medical device called AH® PLUS™ Root Canal Sealer. It focuses on demonstrating equivalence to a predicate device (AH26® Root Canal Sealer) based on material composition, physical properties, and biological testing.

However, the document does not contain the kind of detailed information requested in the prompt regarding acceptance criteria, study methodologies for AI/machine learning device performance, or human reader studies.

Specifically, the prompt asks for:

  1. A table of acceptance criteria and reported device performance: The document states that the physical properties "meet or exceed those standards found in ISO Specification 6876 and ADA Specification 57" and that it's comparable to the predicate in "setting time, flow, shrinkage, and microleakage physical property values." However, it does not provide the specific numerical acceptance criteria or the reported device performance values in a table.
  2. Sample size for test set and data provenance: Not mentioned.
  3. Number of experts and qualifications for ground truth: Not mentioned, as this is not a study assessing expert performance or an AI’s ability to interpret data.
  4. Adjudication method: Not applicable.
  5. MRMC comparative effectiveness study, effect size: Not applicable, as this is not an AI-assisted diagnostic device.
  6. Standalone performance study: Not applicable in the context of an AI device. The document does describe pre-clinical and biological testing of the sealer's components and mixed product.
  7. Type of ground truth: Not applicable in the requested context. The "ground truth" here is the adherence to ISO/ADA standards and biological safety assessments.
  8. Sample size for training set: Not applicable (not an AI device).
  9. How ground truth for training set was established: Not applicable (not an AI device).

Based on the provided text, the device is a root canal sealer, not an AI or machine learning-based device that would typically have the acceptance criteria and study designs described in the prompt. The "testing" mentioned refers to chemical and biological safety assessments and physical property comparisons to established standards and a predicate device.

To directly answer your request based only on the provided text, here's what can be extracted, acknowledging the limitations:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategoryStandard/Predicate ComparisonReported Device Performance
Physical PropertiesISO Specification 6876 and ADA Specification 57"Meet or exceed" these standards
Comparison to AH26® Root Canal Sealer (predicate device)Comparable in setting time, flow, shrinkage, microleakage
MutagenicityAmes Mutagenicity Test (Non-Mutagenic)Paste A, amines in Paste B, and mixed cured sealer found to be non-mutagenic
Acute Oral ToxicityID50 >5000 mg/kg (for resins in Paste A) / Non-toxic (for Paste B)Resins in Paste A: ID50 >5000 mg/kg (in rats) Paste B: Non-toxic (in rats)
CytotoxicityGrowth Inhibition Test (Non-cytotoxic)Mixed cured sealer found to be non-cytotoxic
SensitizationISO 10993-10 (Weak sensitizer) / OECD guidelines (Non-sensitizer)Mixed cured sealer: Weak sensitizer (ISO), Non-sensitizer (OECD)
ImplantationNo macroscopic visible response / Moderate microscopic response / No irritant effectNo macroscopic visible response Moderate microscopic response No other evidence of an irritant effect (in rabbits)

Missing Information (as per prompt's specific requirements, due to the nature of the device and document):

  • Sample size used for the test set and the data provenance: Specific sample sizes for biological and physical tests are not provided in this summary. Data provenance (country, retrospective/prospective) is not mentioned.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is not an AI/diagnostic imaging study requiring expert adjudication of images/data.
  • Adjudication method for the test set: Not applicable.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable (not an AI algorithm).
  • The type of ground truth used: The "ground truth" in this context is adherence to established international standards (ISO, ADA), and results from standardized biological safety tests (Ames, Growth Inhibition, Sensitization, Implantation, Toxicity studies).
  • The sample size for the training set: Not applicable (not an AI device).
  • How the ground truth for the training set was established: Not applicable (not an AI device).

{0}------------------------------------------------

K960548

DENTSPLY

TSPLY Internation

Wast College Avenue Box 870

510(k) SUMMARY

Page 1 of 2

NAME & ADDRESS: DENTSPLY International, 570 West College Avenue, York, PA 17404 P. Jeffery Lehn CONTACT: (717) 849-4343 (717) 845-7511 FAX: PHONE: DATE PREPARED: April 18, 1996

TRADE OR PROPRIETARY NAME: AH® PLUS™ ROOT CANAL SEALER

Root Canal Sealer COMMON OR USUAL NAME:

21 CFR, 872.3820 CLASSIFICATION NAME: Root Canal Filling Resin

AH26® Root Canal Sealer Pre-1976 Device LEGALLY MARKETED DEVICE:

DEVICE DESCRIPTION: AH® PLUS™ Root Canal Sealer is a two-component paste: paste root canal sealer based on epoxy-amine resin chemistry. This easy-to-mix sealer adapts closely to the walls of the prepared root canal and provides outstanding long-term dimensional stability with minimal shrinkage upon setting.

The final product consists of two components, the epoxy resin paste (Paste A) and the amine-containing paste (Paste B) portions which are mixed prior to insertion into the root canal.

AN® PLUS™ Root Canal Sealer and AH26® Root Canal Sealer are both two-component systems that react via an epoxide/amine chemical reaction to cause setting. They are both used in conjunction with the same auxiliary materials in the root canal (e.g., gutta percha points).

The physical properties of AH® PLUS® Root Ganal Sealer meet or exceed those standards found in ISO Specification 6876 and ADA Specification 57. AH® PIJS® Root Canal Sealer and AH26® Root Canal Sealer are comparable in setting time, flow, shrinkage, and microleakage physical property values.

INTENDED USE: AH® PLUS™ Root Canal Sealer is used for permanent sealing of root canals following established endodontic procedures. The predicate device, AH26® Root Canal Sealer has the same intended use.

{1}------------------------------------------------

510(k) SUMMARY (cont'd.)

Page 2 of 2

TECHNOLOGICAL CHARACTERISTICS: The components of AH® PLUS™ Root Canal Sealer are either found in other DENTSPLY legally marketed predicate devices or their safety is documented in Material Safety Data Sheets.

AH® PLUS™ Root Canal Sealer has undergone extensive pre-clinical testing.

Paste A, the amines contained in Paste B, and the mixed cured sealer were evaluated using the Ames Mutagenicity Test and found to be non-mutagenic.

The resins contained in Paste A were found to have an IDso in rats of >5000 mg/kg in studies reported in MSDSs. Paste B was evaluated for Acute Oral Toxicity in rats and found to be non-toxic. The mixed cured sealer was evaluated using the Growth Inhibition Test and found to be non-cytotoxic

The mixed, cured sealer was evaluated in guinea pigs for its sensitization potential and found to be a weak sensitizer according to ISO 10993-10 and a non-sensitizer according to OECD guidelines.

Implantation tests in rabbits on the mixed sealer resulted in no macroscopic visible response and a moderate degree of microscopic response. There was no other evidence of an irritant effect of the sealer.

CONCLUSION

The results of our testing show that, the fact that AH PLUS™ Root Canal Sealer has different technological characteristics than the predicate device, does not raise different questions of safety and efficacy.

We believe this is supported by the ingredient toxicity testing, the performance data, and the final product testing of AH PLUS™ Root Canal Sealer.

§ 872.3820 Root canal filling resin.

(a)
Identification. A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.(b)
Classification. (1) Class II if chloroform is not used as an ingredient in the device.(2) Class III if chloroform is used as an ingredient in the device.
(c)
Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a root canal filling resin described in paragraph (b)(2) of this section that was in commercial distribution before May 28, 1976. Any other root canal filling resin shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.