K Number
K960522

Validate with FDA (Live)

Device Name
DIOPEXY PROBE
Manufacturer
Date Cleared
1996-07-19

(164 days)

Product Code
Regulation Number
886.4390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of this device is transscleral retinal photocoagulation (TSRPC) for retinopexy.

Device Description

The IRIS DioPexy Probe is a multiple-use fiber optic laser delivery device with integral optics, which terminates with a contact prism tip. It must be directly connected to a compatible IRIS OcuLight SL or SLx (810 nm-infrared) Photocoagulator. The prism tip is designed to deflect laser energy 90° from the axis of the shaft and be simultaneously act as a scleral depressor. The fiber optic cable is 3 meters (10 feet). The DioPexy Probe is shipped from the factory non-sterile in a hard plastic container. The device can be sterilized between each use in this container using Ethylene Oxide (EtO) gas in accordance with the user instructions.

AI/ML Overview

Here's an analysis of the provided text regarding the DioPexy Probe, focusing on acceptance criteria and supporting studies:

It is important to note that the provided 510(k) summary does not contain explicit performance acceptance criteria (e.g., target accuracy, sensitivity, specificity values) or detailed study results proving the device meets such criteria. The document focuses on demonstrating substantial equivalence to predicate devices based on intended use and technological characteristics, rather than reporting on specific quantitative performance metrics from a dedicated clinical study as would be expected for a diagnostic or AI-driven device.

Therefore, many of the requested sections below will indicate "Not Provided" or "Not Applicable" based on the information given.


Acceptance Criteria and Study Analysis for the DioPexy Probe

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (if stated explicitly)Reported Device Performance (if stated explicitly)
Functional Equivalence: Ability to produce chorioretinal adhesions for retinopexy.The device enables the ophthalmologist to produce chorioretinal adhesions during retinal detachment surgery using delivery of transscleral infrared laser for retinopexy. Physician observes visible pilot/aiming beam on the retina and monitors/titrates burn development. Chorioretinal adhesion also occurs without visible retinal uptake.
Safety: Device operates without causing undue harm.Implied by substantial equivalence to predicate devices (which are presumably safe). Ethylene Oxide (EtO) sterilization ensures sterility between uses.
Compatibility: Compatible with IRIS OcuLight SL or SLx photocoagulators.Stated as a requirement: "It must be directly connected to a compatible IRIS OcuLight SL or SLx (810 nm-infrared) Photocoagulator."
Sterilization: Can be sterilized between uses.Stated: "The device can be sterilized between each use in this container using Ethylene Oxide (EtO) gas in accordance with the user instructions."

No quantitative performance metrics (e.g., success rate of adhesion, complication rates) that would typically define "acceptance criteria" for a new medical device are explicitly provided in this 510(k) summary. The summary primarily focuses on establishing substantial equivalence based on the device's functional similarity and intended use to existing devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): Not provided. The 510(k) summary does not describe a specific clinical "test set" or study population beyond referencing general clinical practice for retinopexy.
  • Data Provenance: Not provided. The summary discusses the general principles of retinopexy and existing devices but does not mention specific clinical data collected for the DioPexy Probe itself, let alone its provenance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not provided.
  • Qualifications of Experts: Not provided.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable/Not provided. No specific test set or clinical study requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • MRMC Study: Not applicable. This document describes a laser delivery probe, not an AI-driven or diagnostic device. Therefore, no MRMC study or AI-assistance comparison was conducted or is relevant here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Standalone Performance: Not applicable. This device is a surgical instrument that relies entirely on human operation and interpretation (the physician monitoring the burn). It is not an algorithm or an autonomous device.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Not explicitly defined in terms of a study. For the predicate devices and the DioPexy Probe, the "ground truth" of successful retinopexy would typically be clinical assessment of chorioretinal adhesion and retinal reattachment, observed by an ophthalmologist, potentially confirmed by imaging over time, and ultimately by positive patient outcomes (prevention of severe visual loss). However, this is discussed at a conceptual level of medical practice rather than as the outcome of a specific clinical trial reported in this document.

8. The Sample Size for the Training Set

  • Sample Size (Training Set): Not applicable/Not provided. As this is not an AI/machine learning device, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable. As this is not an AI/machine learning device, there is no "training set" or ground truth establishment for it in that context.

{0}------------------------------------------------

K960522

510(k) Summary . III 19 1996 IRIS Medical Instruments, Inc. DioPexy Probe

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Theodore A. Boutacoff IRIS Medical Instruments, Inc. 340 Pioneer Way Mountain View. CA 94041 415-962-8100

Contact Person: (same as above)

Date Prepared: January 15,1996

Name of Device and Name/Address of Sponsor

DioPexy Probe

IRIS Medical Instruments, Inc. 340 Pioneer Way Mountain View, CA 94041

Classification Name

Ophthalmic Laser (Accessory) CFR Section: 886.4390 Product Code: 86 HOF

Predicate Devices

The IRIS DioPexy Probe for transscleral retinal photocoagulation (TSRPC) is substantially equivalent to four predicate devices currently marketed: (1) the MIRA MD-1000 Microdiathermy System (pre-amendment), (2) the CooperSurgical Frigitronics CE-2000 Ophthalmic Cryosurgical System (K872362, K891282), (3) the IRIS Medical Laser Indirect Ophthalmoscope delivery device (K903288), and (4) the IRIS Medical G-Probe (K915236).

{1}------------------------------------------------

Intended Use

The intended use of this device is transscleral retinal photocoagulation (TSRPC) for retinopexy.

Technological Characteristics and Substantial Equivalence

The DioPexy Probe enables the ophthalmologist to produce chorioretinal adhesions during retinal detachment surgery using delivery of transscleral infrared laser for retinopexy. Retinal detachments develop when a full thickness break in the retina allows vitreous fluid to dissect beneath the sensory retina causing it to separate from the retinal pigment epithelium (RPE)/choroid. To avoid progression to severe visual loss, a surgical procedure is necessary to reattach the retina. The principles of retinal detachment repair include retinopexy. Retinopexy is used to provide a seal around all retinal breaks to prevent further leakage of vitreous fluid into the subretinal space. Retinopexy creates an adhesion between the sensory retina and the RPE/choroid. This is stimulated by the post treatment inflammatory reaction at the level of the RPE/choroid and results in a chorioretinal scar. Current sources of energy used for retinopexy are heat (diathermy and laser) or freezing (cryotherapy).

The MIRA utilizes diathermy for retinopexy. Diathermy has been used for transscleral retinopexy for over 60 years. When used, a diathermy probe is placed on the scleral surface of the eye and heat generated at the tip of the probe is delivered transsclerally to the inside of the eye to the retina/choroid. Diathermy usually requires scleral dissection to minimize scleral damage caused by the heat conduction. The physician determines when sufficient energy is delivered by either observing the burn develop on the retina with a diagnostic indirect ophthalmoscope or by observing an external scleral reaction.

The Frigitronics utilizes cryotherapy for retinopexy. Cryotherapy is delivered transsclerally and was introduced in 1964. When used, a cryoprobe is placed on the surface of the sclera, the sclera is depressed, and cold generated at the tip of the probe creates an ice ball which conducts across the sclera. The physician observes the extent of ice ball progression inside the eye with a diagnostic indirect ophthalmoscope and when sufficient terminates energy delivery.

The IRIS Medical Laser Indirect Ophthalmoscope utilizes 810 nm diode laser photocoagulation for retinopexy. Diode laser

{2}------------------------------------------------

photocoagulation delivered transpupillary has been used since 1989 for retinopexy. When used, the LIO is placed on the physician's head and with the aid of a hand-held indirect lens is used to observe the treatment site. Only breaks which can be viewed directly, with or without the aid of scleral depression, can be treated. When the target is illuminated with the pilot/aiming beam, the physician activates the footswitch and delivers laser energy.

Transscleral application of diode laser has been shown to reduce intraocular pressure in the treatment of glaucoma using the IRIS Medical G-Probe (K915236) through transscleral cyclophotocoagulation (TSCPC). Transscleral photocoagulation of the sub-surface structures can be accomplished by placing a diode laser probe on the surface of the eye and allowing the infrared laser energy to be absorbed by the deep pigmented tissues. Typical laser parameters for TSCPC are 2000 milliseconds duration and 1750 milliwatts of power.

The DioPexy Probe is intended to be used to perform transscleral retinal photocoagulation (TSRPC) in retinopexy procedures to accomplish retinal detachment repair. When using the DioPexy Probe the retinal treatment site is always under direct observation by the treating physician using an indirect ophthalmoscope. The prism tip of the DioPexy Probe is placed on the outer surface of the sclera and the sclera is depressed. A low power visible pilot/aiming beam is projected transsclerally onto the retina. This visible aiming beam can be observed by the physician on the retina during treatment. Size and intensity of the burn can be monitored and titrated with the laser footswitch. The extent of the laser lesion depends upon the energy delivered to the treatment site. Typical laser parameters for TSRPC are 2000 milliseconds duration and 1100 milliwatts of power. When longer duration burns and lower powers are used, burn development is slower. Chorioretinal adhesion also occurs without visible retinal uptake, as with cryotherapy in bullous retinal detachments.

Device Description

The IRIS DioPexy Probe is a multiple-use fiber optic laser delivery device with integral optics, which terminates with a contact prism tip. It must be directly connected to a compatible IRIS OcuLight SL or SLx (810 nm-infrared) Photocoagulator. The prism tip is designed to deflect laser energy 90° from the axis of the shaft and be simultaneously act as a scleral depressor. The fiber optic cable is 3

{3}------------------------------------------------

meters (10 feet). The DioPexy Probe is shipped from the factory non-sterile in a hard plastic container. The device can be sterilized between each use in this container using Ethylene Oxide (EtO) gas in accordance with the user instructions.

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.