The Braun Hemostasis Valve Introducer Kits are designed to facilitate percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.
Device Story
Braun Hemostasis Introducer Kits consist of hemostasis valve introducer sheath, dilator, and optional guidewire/obturator. Used by physicians in clinical settings to facilitate percutaneous catheter insertion into vasculature. Sheath design prevents blood backflow during procedure. Available in straight and contralateral curve configurations, 5-9 French sizes, 10-60 cm lengths. Device provides access path for diagnostic or interventional catheters; enables clinician to maintain vascular access while minimizing blood loss.
Clinical Evidence
Bench testing only. Performance metrics include: hub-to-body tensile strength (14.5-18.4 lbs), shape recovery (85-88%), kink resistance (no kinks at 170° for 5F), and leak testing (no leaks at 100 psi for sheath, 6 psi for gasket).
Technological Characteristics
Materials: PEBA sheath. Construction: Braided body. Sizes: 5-9 French. Lengths: 10-60 cm. Configurations: Straight and contralateral curve. Features: Hemostasis valve, dilator, optional guidewire/obturator. Energy source: None (mechanical).
Indications for Use
Indicated for patients requiring percutaneous catheterization of the vasculature, facilitating device introduction while maintaining hemostasis.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
K222184 — Introducer Sheath Kits · Beijing Demax Medical Technology Co.,Ltd · Feb 21, 2023
K203670 — Extra Large Check-Flo Introducer · Cook Incorporated · Feb 16, 2021
K242248 — Introducer Sheath Kits · Beijing Demax Medical Technology Co.,Ltd · Aug 22, 2024
K172331 — Type I, Type II, Type III, Type IV · Lepu Medical Technology (Beijing) Co., Ltd. · Mar 2, 2018
K961745 — BRAUN PERCUTANEOUS INTRODUCER SET · B.Braun Medical, Inc. · Sep 17, 1996
Submission Summary (Full Text)
{0}
K960406
# II. SUMMARY OF SAFETY AND EFFECTIVENESS
Applicant Name & Address: B. Braun Medical Inc., Cardiovascular Division, 13000 Hwy 55 Minneapolis, MN 55441-3813 (612) 559-3370
Contact: Barbara Atzenhoefer, Regulatory Affairs Manager
Date Prepared: January 22, 1996
Common or Usual Name: Braun Hemostasis Introducer Kits
Device Classification Name: Catheter Introducer
## Substantial Equivalence:
K895044* Balkin Up & Over Cook
K945396 Braun Fluoro-Tip™ Hemostasis Introducer Kits B. Braun Cardiovascular
Device Description: Braun Hemostasis Introducer Kits consist of a hemostasis valve introducer sheath packaged with a corresponding dilator. One or more of the following accessories may also be packaged with the sheath/dilator; guidewire, and/or obturator. The introducer sheaths are available in both straight and contralateral curve versions. They will be available in 5 - 9 French sizes, in lengths from 30 - 60 cm in the curved version and 10 - 60 cm in the straight version. Each will be packaged with an appropriately sized dilator. Optional items to be kitted with the sheath and dilator include a guidewire, and/or obturator. The components, their materials and accessories, are similar to those previously cleared under Braun 510(k)s; K894446, K930057, K931743, K945396, and K950882.
Statement of Intended Use: The Braun Hemostasis Valve Introducer Kits are designed to facilitate percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.
Summary/Comparison of Technological Characteristics:
| Characteristic | Braun Hemostasis Introducer Kits | Cook Up & Over Balkin Contralateral- K895044 | Braun Fluoro-Tip K945396 |
| --- | --- | --- | --- |
| Available French Sizes | 5 - 9 French | Curved versions 5.5, 6 & 7 fr Straight versions of the Check-Flo introducer sheaths are available in 4-9 french | 7 - 10 fr |
| Available Lengths | curved - 30 - 60 cm
straight - 10 - 60 cm | curved - 40 cm
straight - 7 - 13 cm | 10 - 32 cm |
| Body Stock Design | Braided | non-braided | non-braided, double layer sheath |
| Sheath Material | peba | Teflon® | FEP (Teflon®) |
| Hub to Body Tensile | 5 fr - 14.5 lbs avg
9 fr - 18.4 lbs avg | 7.2 lbs avg | 7 fr - 14 lbs avg |
| Shape Recovery | 5 fr 85% recovery @ 5 sec
86% recovery @ 60 sec
9fr 85% recovery @ 5 sec
88% recovery @ 60 sec | 72% recovery @ 5 sec
77% recovery @ 60 sec | Not applicable |
| Sheath Leak Test | No leaks at 100 psi | No leaks at 100 psi | Not performed |
| Gasket leak Test | No leaks at 6 psi | No leaks at 6 psi | No leaks at 10 psi |
| Kink Resistance | 5 fr - No kinks at 170°
9 fr - Kinked at avg of 107° | Kinked at an avg of 76° | Not performed |
| Radiopacity | More visible than Cook | Less visible than Braun | Not compared against Cook |
3
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.