K Number
K955712

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-10-11

(301 days)

Product Code
Regulation Number
870.3680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The USCI® Balloon Bipolar Pacing Electrodes and the USCI® Special Care Electrodes are used to transmit an electrical stimulus from the pulse generator to the heart and/or to transmit an electrical signal from the heart to a recording device.

Device Description

The USCI® Balloon Bipolar Pacing Electrodes and USCI® Special Care Electrodes (temporary pacing leads) are flexible, insulated conductors, constructed of a woven filament and a polyurethane base with Benzalkonium Heparin coating and stainless steel electrodes.

AI/ML Overview

This submission is for a medical device, not an AI/ML algorithm. Therefore, the requested information (acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) is not applicable in the context of an AI/ML algorithm.

The document describes a 510(k) summary for a USCI® Bipolar Pacing Electrode with a benzalkonium heparin coating. The basis for substantial equivalence is shown through bench testing that compares the coated device to a predicate (uncoated) device.

Here's an interpretation of the "acceptance criteria" and "study" as they relate to this medical device submission:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Bench Test)Reported Device Performance (Coated Electrode)
Electrical Continuity: Ability to pace the heart.Demonstrated that the BH coated Bipolar Pacing Electrodes does not affect the ability of the catheter to pace.
Inflated Balloon Diameter: Within recommended diameter range.Confirmed that the BH coated Bipolar Pacing Electrode is within the recommended diameter range.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated in the provided text. It refers to "bench test data" and "electrical continuity tests" but does not quantify the number of devices tested.
  • Data Provenance: This is not applicable in the context of an AI algorithm. The "data" here refers to physical measurements and functional tests performed on the device itself as part of product development and regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable. Ground truth, in this context, would be the physical and electrical properties of the device, measured by standard engineering and medical device testing methods, not by expert interpretation of data.

4. Adjudication Method for the Test Set:

  • Not applicable. Bench testing results are typically objective measurements, not subject to adjudication by multiple human experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No. This is a medical device, not an AI algorithm, so an MRMC study is not relevant. The comparison is between a new device design and an existing predicate device using objective bench tests.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Not applicable. This is a physical medical device.

7. The Type of Ground Truth Used:

  • Bench Test Results/Performance Specifications: The "ground truth" is the established functional and dimensional requirements for such pacing electrodes, against which the performance of both the coated and uncoated devices is measured. These are objective engineering and performance specifications.

8. The Sample Size for the Training Set:

  • Not applicable. This is not an AI/ML submission that involves training data.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable.

{0}------------------------------------------------

OCT 11 1996

al and Regulatory Affairs Technology Park Dri Box 7025 rica, MA 01821 667-1300 (508) 670-4326

Image /page/0/Picture/2 description: The image shows the number K955712 at the top. Below the number is the USCI logo. The USCI logo is an oval with the letters USCI in the center.

510(k) SUMMARY FOR THE USCI® BIPOLAR PACING ELECTRODE 6.0

As required under Section 12, part (a)(i)(3A) of the Safe Medical Device Act of 1990, an adequate summary of any information respecting safety and effectiveness follows.

6.1 General Information

� Name and address of submitter:

Tel #: Fax #: Contact Person: Date of Preparation:

USCI, Div. of C.R. Bard,Inc. 1200 Technology Park Drive Billerica, MA 01821 (508) 667-1300 (508) 667-4301 Robert Miragliuolo December 11, 1995

◆ Device Names:

Trade Name(s):

Common Name:

Classification Name:

◆ Predicate Device Name: Trade Name:

USCI® Bipolar Balloon Pacing Electrode USCI® Special Care Electrode Temporary Transvenous Pacing Electrode Temporary Pacemaker Electrode

USCI® Bipolar Balloon Pacing Electrode

Image /page/0/Picture/16 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a three-dimensional appearance. The overall design is simple and eye-catching.

{1}------------------------------------------------

General Information con't 6.1

Device Description and Intended Use: �

The USCI® Balloon Bipolar Pacing Electrodes and USCI® Special Care Electrodes (temporary pacing leads) are flexible, insulated conductors, constructed of a woven filament and a polyurethane base with Benzalkonium Heparin coating and stainless steel electrodes.

The USCI® Balloon Bipolar Pacing Electrodes and the USCI® Special Care Electrodes are used to transmit an electrical stimulus from the pulse generator to the heart and/or to transmit an electrical signal from the heart to a recording device. These devices are sold sterile and are for one time use only. In addition to the pacing electrode with BH coating, each catheter is packaged with a needle cannula, an EKG clip, and a syringe to assist in the use of the pacing electrode.

{2}------------------------------------------------

Summary of Similarities and Differences 6.2

The coated USCI Bipolar Pacing Electrode is substantially equivalent to the uncoated USCI® Bipolar Balloon Pacing Electrode (covered under #K800298).

The similarities between the current devices, the USCI Bipolar Pacing Electrodes and the predicate device, the USCI® Bipolar Balloon Pacing Electrode, include identical materials: a latex balloon, a polyurethane shaft, and stainless steel electrodes.

The difference between the two devices is that the current USCT® Bipolar Pacing Electrode has been coated with benzalkonium heparin, which is identical to the coating used on the USCI® Safe-T-Cath Thermodilution catheter (#K791873), and the USCI® Temporary Atrio-Ventricular Electrode, (#K830094).

Results from bench test data demonstrate that the BH coated USCI Bipolar Pacing Electrodes perform similar to the predicate device, the uncoated USCI® Bipolar Balloon Pacing Electrode. Electrical continuity tests demonstrated that the BH coated Bipolar Pacing Electrodes does not affect the ability of the catheter to pace. The inflated balloon diameter test confirms that the BH coated Bipolar Pacing Electrode is within the recommended diameter range.

In conclusion, results from all the bench test data demonstrate that the BH coated Bipolar Pacing Electrodes perform similar to the predicate device, the uncoated Bipolar Balloon Pacing Electrode, therefore, a determination of substantial equivalence can be claimed.

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.