Intended Use

The Distal Radius Plate System is designed for internal fixation of the distal radius. The distal radius plates are available in left and right volar configurations and in two sizes, small and large.

The Humerus Plate System is designed for internal fixation of the proximal humerus. The humerus plates are available in right and left configurations and in three hole and five hole length.

Cortical and cancellous bone screws are required to secure the plates to the bone. A variety of screw are available for this application in order to accommodate different fracture types and locations, bone density and surgeon preference.

Device Description

The ACE Humerus Plates and Radius Plates are anatomically contoured bone plates which are attached to the bone with cortical or cancellous bone screws. The humerus plates will offered in two configurations for both the left side and the right side. The plates are offered in three hole and five hole configurations.

The ACE Radius plates are also offered in a right and left configuration both for volar fixation and dorsal fixation. The plates are offered in a large and a small size.

AI/ML Overview

Here's an analysis of the provided information, focusing on acceptance criteria, study details, and ground truth establishment:

Disclaimer: Based on the provided text, the device is a medical implant (bone plates) and not an AI/Software as a Medical Device (SaMD). Therefore, many of the requested categories related to AI performance, ground truth establishment for AI, and MRMC studies are not applicable. I will provide information relevant to the provided text for a physical medical device.

Acceptance Criteria and Device Performance (Based on Physical Device Testing)

Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance (ACE Upper Extremity Plates)
Mechanical StrengthComparable static bending strength to predicate devices."The ACE Plates were comparable in bending strength to the predicate device"
BiocompatibilityManufactured from a biocompatible material commonly used in orthopedics."manufactured of a biocompatable titanium alloy, commonly used in the orthopedic industry."
Intended UseSuitable for internal fixation of the distal radius and proximal humerus."designed for internal fixation of the distal radius" and "designed for internal fixation of the proximal humerus."

Study Details (Applicable to Physical Device Testing)

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify the exact number of plates tested. It mentions "Three types of plates (proximal humerus, volar distal radius and dorsal distal radius) were tested," implying at least one sample of each type. It's likely more than one per type were tested for statistical validity, but this is not explicitly stated.
    • Data Provenance: Not explicitly stated. This was a laboratory study conducted by Ace Medical Company or a contracted lab. There is no mention of "clinical data" or "patient data," indicating it was a benchtop mechanistic test.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For a physical device's mechanical testing, "ground truth" is typically established by engineering standards and measurement accuracy, not expert consensus.

  3. Adjudication method for the test set: Not applicable. This refers to consensus among experts, which is not relevant for benchtop mechanical testing. Results are typically determined by measurement and adherence to predefined engineering specifications or comparison thresholds.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This study refers to an AI/Software as a Medical Device (SaMD), not a physical orthopedic implant.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This refers to an AI/Software as a Medical Device (SaMD).

  6. The type of ground truth used: For mechanical testing, the "ground truth" is defined by established engineering principles and the performance of predicate devices. The study compared the static bending strength of ACE plates to predicate devices (Synthes and DePuy plates). Therefore, the "ground truth" for performance was the established mechanical properties of these legally marketed, equivalent devices.

  7. The sample size for the training set: Not applicable. This refers to an AI/Software as a Medical Device (SaMD). The design and development of physical implants do not typically involve "training sets" in the AI sense.

  8. How the ground truth for the training set was established: Not applicable. This refers to an AI/Software as a Medical Device (SaMD).


Summary of Non-Applicable Sections for a Physical Device:

The provided text describes a submission for physical bone plates, not an AI/Software as a Medical Device (SaMD). Therefore, many of the requested items related to AI device performance evaluation, such as:

  • Sample size for the training set
  • How ground truth for the training set was established
  • Number of experts for ground truth
  • Adjudication method for experts
  • MRMC comparative effectiveness study
  • Standalone algorithm performance

...are not applicable to this 510(k) summary. The study described focuses on benchtop mechanical testing to demonstrate substantial equivalence to predicate devices, a standard practice for physical orthopedic implants.

{0}------------------------------------------------

K955472

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ACE UPPER EXTREMITY PLATE SYSTEM

Ace Medical Company
2260 East El Segundo Blvd.
El Segundo, CA 90245
Phone (310) 615-0066

APR

Contact Person:Nancy Giezen
Senior Regulatory Affairs Specialist
Paul Doner
Director, Regulatory/Clinical Affairs
Proprietary Name:ACE Humerus Plates
ACE Radius Plates
Common Name:Bone Plate
Classification Name:Single / Multiple Component Metallic Bone Fixation Appliance and Accessories, 21 CFR 888.3030
Proposed Regulatory Class:Class II
Classification Panel:Orthopaedic Device Panel
Device Product Code:87HRS Plate, Fixation, Bone

Device Description:

The ACE Humerus Plates and Radius Plates are anatomically contoured bone plates which are attached to the bone with cortical or cancellous bone screws. The humerus plates will offered in two configurations for both the left side and the right side. The plates are offered in three hole and five hole configurations.

The ACE Radius plates are also offered in a right and left configuration both for volar fixation and dorsal fixation. The plates are offered in a large and a small size.

{1}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ACE UPPER EXTREMITY PLATE SYSTEM

Intended Use:

The Distal Radius Plate System is designed for internal fixation of the distal radius. The distal radius plates are available in left and right volar configurations and in two sizes, small and large.

The Humerus Plate System is designed for internal fixation of the proximal humerus. The humerus plates are available in right and left configurations and in three hole and five hole length.

Cortical and cancellous bone screws are required to secure the plates to the bone. A variety of screw are available for this application in order to accommodate different fracture types and locations, bone density and surgeon preference.

Substantial Equivalence:

The ACE Humerus Plate and Radius Plate systems are considered substantially equivalent to the DePuy ORIF Small Fragment Plates and the Synthes Small Plates and Small Titanium Plates.

Performance Data:

The ACE Humerus and Radius plates were tested in static bending and compared to the Synthes (USA) and the DePuy plates. Three types of plates (proximal humerus, volar distal radius and dorsal distal radius) were tested by applying static cantilever bending forces on the shaft/buttress interface of the plate. The ACE Plates were comparable in bending strength to the predicate device

Safety and Effectiveness:

The upper extremity plates are manufactured of a biocompatable titanium alloy, commonly used in the orthopedic industry. The plates were shown to exhibit strength characteristics comparable to the predicate device. Thirdly, no new indications have been introduced with these devices. Therefore, the ACE upper extremity plates are considered to be a safe and effective means of fracture fixation.

N/A