K Number
K954687

Validate with FDA (Live)

Date Cleared
1996-04-09

(181 days)

Product Code
Regulation Number
866.3510
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCESS® Rubella IgG assay aids in the diagnosis of Rubella infection and the determination of immunity.

Device Description

The ACCESS® Rubella IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to the Rubella virus in human serum, using the ACCESS® Immunoassay System.

AI/ML Overview

Here’s an analysis of the provided information, structured to address your request:

Acceptance Criteria and Device Performance for ACCESS® Rubella IgG Reagents

Acceptance CriteriaReported Device Performance
Relative Sensitivity (compared to HAI)98%
Relative Specificity (compared to HAI)99%
Concordance with predicate (Abbott IMx)90%
Within Site Precision (QC1)15%
Within Site Precision (QC2)6%
Within Site Precision (High Positive)9%
Total Precision (QC1)15%
Total Precision (QC2)6%
Total Precision (High Positive)11%

Study Details:

  1. Sample sizes used for the test set and the data provenance:

    • Comparison to HAI: 784 patient serum samples.
    • Comparison to Abbott IMx Rubella IgG: 670 patient serum samples.
    • Precision Studies: Not explicitly stated, but includes QC1, QC2, and a high positive sample (implying multiple measurements on these controls).
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The document refers to "clinical studies," which often implies prospective collection, but this is not confirmed.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not stated. The ground truth for the primary effectiveness claims (sensitivity and specificity) was established by HAI (Hemagglutination Inhibition Assay), which is described as a "standard laboratory reference method." This is a laboratory test, not an expert human consensus.
  3. Adjudication method for the test set:

    • Not applicable as the ground truth was established by a laboratory reference method (HAI), not human adjudication.
  4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    • No, an MRMC study was not done. This device is an automated immunoassay for antibody detection, not an imaging or diagnostic device requiring human interpretation alongside AI.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this study represents a standalone evaluation of the ACCESS® Rubella IgG assay. It is an automated immunoassay system, and its performance is assessed independently against reference methods and a predicate device. Human interpretation is not part of its operational mechanism.
  6. The type of ground truth used:

    • Expert Consensus: No.
    • Pathology: No.
    • Outcomes Data: No.
    • Reference Method/Laboratory Standard: The primary ground truth for sensitivity and specificity was HAI (Hemagglutination Inhibition Assay), described as a "standard laboratory reference method." For concordance, the ground truth was the performance of the Abbott IMx Rubella IgG, which is another commercially available, validated assay.
  7. The sample size for the training set:

    • Not applicable as this is an assay kit. Immunoassays typically do not have "training sets" in the same way machine learning models do. The assay is developed and optimized, but the concept of a training set for an algorithm is not relevant here.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no "training set" for this type of device. The assay development would involve extensive R&D, reagent optimization, and analytical validation using known positive and negative controls, but this is distinct from establishing ground truth for a machine learning training set.

{0}------------------------------------------------

SANOFI DIAGNOSTICS PASTEUR, INC. 1000 LAKE HAZELTINE DRIVE CHASKA, MINNESOTA 55318-1084 U.S.A. TEL: (612) 448-4848

K954687

APR - 9 1996

Image /page/0/Picture/3 description: The image shows a logo for Sanofi Pasteur Diagnostics. The logo is in a rectangular box with the word "Sanofi" in large, bold, white letters on a black background. Below that, the word "Pasteur" is in large, bold, white letters on a black background, with the word "Diagnostics" in smaller letters above it. The logo is simple and modern, and it is likely used to identify the company's products or services.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990.

1. General Information

Device Classification Name:Rubella Virus Serological Reagents
Device Trade Name:ACCESS® Rubella IgG Reagents
Applicant's Name and Address:Sanofi Diagnostics Pasteur, Inc.1000 Lake Hazelline DriveChaska, MN 55318
2. Predicate DeviceAbbott IMx Rubella IgGAbbott LaboratoriesDiagnostics DivisionAbbott Park, IL 60064

3. Device Description

The ACCESS® Rubella IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to the Rubella virus in human serum, using the ACCESS® Immunoassay System. The ACCESS® Rubella IgG assay aids in the diagnosis of Rubella infection and the determination of immunity.

4. Summary of Studies

In clinical studies, the ACCESS® Rubella IgG was compared to HAI on 784 patient serum samples. The relative sensitivity and specificity were 98% and 99%, respectively.

A comparison to the Abbott IMx Rubella IgG was done on 670 patient serum samples. The concordance was 90%.

Representative data for within site and total precision are 15%, 15% and 16% for the QC1 negative control, 6%, 6% and 6% for the QC2 low reactive control, and 9%, 11% and 12% for a high positive sample.

5. Conclusion

The Sanofi Diagnostics Pasteur ACCESS Rubella IgG is substantially equivalent to a standard laboratory reference method (HAI) and to another kit currently in commercial distribution for the determination of IgG antibodies to Rubella in human serum samples.

§ 866.3510 Rubella virus serological reagents.

(a)
Identification. Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).(b)
Classification. Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.