K Number
K954572

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-06-26

(268 days)

Product Code
Regulation Number
868.5905
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Maestro CRC is an optional accessory that provides remote control capabilities when used in coniunction with Respironics Aria CPAP and Virtuoso Smart CPAP Systems and is intended for use by a qualified clinician (any individual trained to perform sleep study diagnoses) in hospital/sleep laboratory settings where adult patients suffering from OSA are diagnosed and treated.

Device Description

The Maestro CRC (Figure 16-1) incorporates a control panel with a display screen and two user kevpads (for Function and Setting buttons, respectively) on top. Eight DC signal output ports (five of which are active: two of which will be usable with the Aria and Virtuoso Systems covered in this premarket notification), a power supply connector, and two communications connectors (one of which will be usable with the Aria and Virtuoso Systems covered in this premarket notification) are located on the back of the unit. A contrast adjustment control and a power on/off control accompanied by a Light Emitting Diode (LED) power on indicator are provided on its right side section. An annunciator and microprocessor are contained within the Maestro CRC enclosure. The enclosure measures 9 1/2" x 2" x 7 3/4" (W x H x D).

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the Maestro Clinical Remote Control, based on the provided text:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance
Electrical SafetyIEC 601-1 (General Requirements Safety of Medical Electrical Equipment)Found compliant and certified
Electromagnetic Compatibility (EMC)IEC 801-1 (EMC for Industrial Process and Measurement and Control Equipment. Part 1)Found compliant and certified
IEC 801-2 (Electrostatic Discharge Requirements)Found compliant and certified
IEC 801-3 (Radiated Electromagnetic Field Requirements)Found compliant and certified
IEC 801-4 (Electrical Fast Transients/Burst Requirements)Found compliant and certified
CISPR 11 (Limits and Methods of Measurement of Radio interference Characteristics of Industrial, Scientific, and Medical Equipment)Found compliant and certified
CISPR 16 (Specification for Radio Interference Measuring Apparatus and Measurement Methods)Found compliant and certified
Mechanical & Environmental EnduranceIEC 68-2-6 (Basic Environment Test Procedures. Part 2: Tests. Test Fc and Guidance: Vibration (Sinusoidal))Found compliant and certified
IEC 68-2-27 (Basic Environment Test Procedures. Part 2: Tests. Test Ea and Guidance: Shock)Found compliant and certified
IEC 68-2-37 (Basic Environment Test Procedures. Part 2: Tests. Test Fdc: Random Vibration Wide Band - Reproducibility Low)Found compliant and certified
MIL-STD-461D (Requirements for Control of Electromagnetic Interference Emissions and Susceptibility)Found compliant and certified
MIL-STD-462D (Measurements of Electromagnetic Interference Characteristics)Found compliant and certified
MIL-STD-810E (Environmental Test Methods)Found compliant and certified
Functional EquivalencePerformance in its intended environment is as safe and effective as predicate devices (Puritan-Bennett Companion 318 Nasal CPAP Diagnostics System (K910194) and Respironics REMstar Choice Clinical Remote Control (K920699))Verified through performance-related testing
Intended UseProvides remote control capabilities for Respironics Aria CPAP and Virtuoso Smart CPAP Systems for qualified clinicians in hospital/sleep laboratory settings for adult OSA patients. Not for life support.This is the stated intended use for which the device's performance was evaluated.
ClaimsProvides data monitoring, alphanumeric/graphical representation of patient data, interfaces with polysomnograph.Implied to be met by the device's design and features as described.

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify a sample size in terms of patient data or typical "test set" in the context of machine learning. The testing appears to be primarily focused on engineering and regulatory compliance of the device itself (electrical, EMC, mechanical, environmental) rather than a clinical performance study using patient data.
    • Data Provenance: Not applicable in the context of clinical data. The device itself was tested against international and military standards.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not applicable. The "ground truth" for this device's performance is defined by adherence to established engineering and safety standards (e.g., IEC, CISPR, MIL-STD). The compliance was likely assessed by qualified test engineers and regulatory personnel, but no specific "experts" or their qualifications for establishing ground truth in a clinical sense are mentioned.
  3. Adjudication Method for the Test Set:

    • Not applicable. The testing described is against objective, quantifiable technical standards, not involving subjective interpretation requiring adjudication among experts.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. The document describes testing for safety and effectiveness relative to predicate devices through engineering compliance, not a clinical trial with human readers assisted by AI. The device is a remote control, not an AI diagnostic tool.
  5. Standalone (Algorithm Only) Performance Study:

    • No, a standalone algorithm only performance study was not conducted in the context of AI. The Maestro Clinical Remote Control is a hardware and software system for remote operation and monitoring, not an AI algorithm for diagnosis. The "standalone" performance here refers to the device's ability to function independently according to its design specifications when connected to target CPAP systems.
  6. Type of Ground Truth Used:

    • Engineering and Regulatory Standards Compliance: The primary ground truth is established by the specified international and military technical standards (IEC, CISPR, MIL-STD). The device's ability to meet these objective criteria defines its "ground truth" for safety and effectiveness.
    • Functional Equivalence to Predicate Devices: The implicit ground truth for its functional performance is that it should operate in a manner that is "as safe and effective as" the predicate devices in its intended environment.
  7. Sample Size for the Training Set:

    • Not applicable. This device does not use machine learning, so there is no "training set."
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set.

{0}------------------------------------------------

Premarket Notification Section 16 - Summary of Safety and Effectiveness Maestro Clinical Remote Control

Image /page/0/Picture/2 description: The image contains a sequence of handwritten characters, specifically 'K954572'. The characters are written in a dark ink, and the handwriting style appears casual. The numbers are clearly distinguishable, and the letter 'K' is the first character in the sequence.

Image /page/0/Picture/3 description: The image shows a title card with the words "Section 16 Summary of Safety and Effectiveness" written on it. The text is written in a bold, sans-serif font and is centered on the card. The background of the card is a gray, textured pattern. The card is surrounded by a black border.

September 22, 1995

Image /page/0/Picture/5 description: The image shows the logo for RESPIRONICS INC. The logo consists of a triangle with a stick figure inside of it. The text "RESPIRONICS INC" is written in all caps below the triangle.

1001 Murry Ridge Drive, Murrysville, PA 15668

Official ContactRichard G. Confer
Manager, Regulatory Affairs
Respironics, Inc.
1001 Murry Ridge Drive
Murrysville, PA 15668
Classification NameNoncontinuous Ventilator
Common/Usual NameRemote Control
Proprietary NameMaestro Clinical Remote Control
Predicate DevicePuritan-Bennett Companion 318 Nasal CPAPDiagnostics System (K910194)
Respironics' REMstar Choice Clinical RemoteControl (K920699)

Predicate Device Information

The technological characteristics of the Maestro Clinical Remote Control (CRC) are substantially equivalent to the predicate devices listed above in terms of design, materials, energy source, and intended use. All three devices are intended for remote operation and monitoring of CPAP devices used in hospital/sleep laboratory settings to treat adult patients who suffer from Obstructive Sleep Apnea (OSA).

Section 16, Page 1

437

{1}------------------------------------------------

Substantial Equivalence

The Maestro Clinical Remote Control is substantially equivalent to remote and instrumentation portions of the Puritan-Bennett Companion® 318 Nasal CPAP Diagnostics System (K910194), and the Respironics REMstar Choice® Clinical Controller (K920699). These predicate devices are remote control and data monitoring devices intended for use with CPAP systems in the diagnosis and treatment of adult Obstructive Sleep Apnea. They are intended to be used by clinicians in hospital (sleep laboratory) environments and are not intended for life support or life-sustaining applications.

Testing was performed to demonstrate that the performance of the Maestro CRC in its intended environment is as safe and effective as that of the legally marketed predicate devices. The safety and effectiveness of the Maestro CRC was verified through performance related testing that consisted of Electrical Safety, Electromagnetic Compatibility, Mechanical and Environmental Testing. The Maestro CRC was found compliant and has been certified to the standards referenced in the "FDA Reviewer Guidance for Premarket Notifications."

Intended Use

The Maestro CRC is an optional accessory that provides remote control capabilities when used in coniunction with Respironics Aria CPAP and Virtuoso Smart CPAP Systems and is intended for use by a qualified clinician (any individual trained to perform sleep study diagnoses) in hospital/sleep laboratory settings where adult patients suffering from OSA are diagnosed and treated.

Claims

The Maestro CRC :

  • Provides data monitoring capabilities .
  • Presents alphanumeric and graphical representation of patient data .
  • . Interfaces with a polysomnograph data recording device

Benefit of using the Maestro CRC

The Maestro CRC enables the clinician to control and operate the Aria and Virtuoso Systems from a remote location, assuring that a patient receives uninterrupted sleep, necessary to diagnose and treat OSA. Using the Maestro CRC, a clinician can adjust CPAP device settings and monitor data from a control

{2}------------------------------------------------

room, rather than in the bedroom where these activities could potentially disturb the patient from sleep.

Device Description

The Maestro CRC (Figure 16-1) incorporates a control panel with a display screen and two user kevpads (for Function and Setting buttons, respectively) on top. Eight DC signal output ports (five of which are active: two of which will be usable with the Aria and Virtuoso Systems covered in this premarket notification), a power supply connector, and two communications connectors (one of which will be usable with the Aria and Virtuoso Systems covered in this premarket notification) are located on the back of the unit. A contrast adjustment control and a power on/off control accompanied by a Light Emitting Diode (LED) power on indicator are provided on its right side section. An annunciator and microprocessor are contained within the Maestro CRC enclosure. The enclosure measures 9 1/2" x 2" x 7 3/4" (W x H x D).

Note: The three auxiliary DC signal output ports and one auxiliary communications connector are not electrically connected and are, therefore, inactive. The auxiliary ports and connector are for future use and will not be covered in this premarket notification.

Image /page/2/Figure/6 description: The image shows a device with several labeled components. The labels include 'Display Screen', 'Function Buttons', and 'Setting Buttons' on the front. On the side, there are labels for 'Power On Light', 'On/Off Button', and 'Contrast Adjustment Control'. The top of the device is labeled with 'Auxiliary Communications Connector (Inactive)', 'Communications Connector', 'DC Signal Output Ports', 'Auxiliary DC Signal Output Ports (Inactive)', and 'Power Supply Connector'.

Maestro Clinical Remote Control Figure 16-1.

Image /page/2/Picture/9 description: The image shows the text "Section 16, Page 3". This text likely refers to a specific section and page number within a larger document. The text is written in a clear, bold font, making it easily readable. The background of the image is plain and does not contain any other elements.

{3}------------------------------------------------

DC Signal Outputs

The Maestro CRC may optionally be connected by cable to a polysomnograph machine. Data is transmitted as DC signals through the Maestro's DC signal output ports (Figure 16-2). The ports are labeled for their respective signals. The signals are Event Predictor, VT (Estimated Tidal Volume), Vest (Estimated Patient Flow), P (Pressure) and Vieak (Estimated Leak). Only the Event Predictor (for the Virtuoso) and Pressure (for the Aria and Virtuoso) signals are used when the Maestro CRC is connected to the Aria or Virtuoso. The remaining signal outputs are for the Duet and Quartet Systems, and will be covered in a future premarket notification applying to use of the Maestro CRC with these devices. Table 16-1 describes output signals available from the Maestro CRC when connected to the Aria and Virtuoso CPAP Systems.

Image /page/3/Figure/4 description: The image shows a diagram of a medical device with several ports labeled with their corresponding functions. The ports are labeled as Event Predictor, Estimated Tidal Volume (VT), Estimated Patient Flow (Vest), Pressure (P), and Estimated Leak (Vieak). A hand is pointing to a cable connected to the device, which is labeled as a cable to a polysomnograph.

Figure 16-2. DC Signal Outputs

{4}------------------------------------------------

EventPredictorVTV̅estPV̅Leak
Aria CPAP SystemN/AN/AN/AXN/A
Virtuoso SmartCPAP SystemXN/AN/AXN/A

Signal Outputs Table 16-1.

Note: There is no signal if a cable is connected to an output port that is not applicable to the Aria and Virtuoso devices. The availability of the DC signals from the Maestro CRC to the polysomnograph is a function of the specific CPAP device that is connected to the Maestro CRC. For example, connection to V est, VT, and V leak, when the Maestro CRC is connected to the Aria would result in no DC signal output to the polysomnograph.

Description of Signals

Since this premarket notification addresses use of the Maestro CRC with the Aria and Virtuoso Systems, only the signals applicable to the Aria and Virtuoso Systems (Event Predictor and Pressure) will be discussed.

  • Event Predictor This signal counts the number of recognizable energy . patterns created by an unstable airway within one minute. The Event Predictor output signal may provide an early warning that abnormal breathing events are about to occur.
  • Pressure (P) This signal permits continuous recording of the pressure . measured at the mask.

Gill

{5}------------------------------------------------

Principles of Operation

The Maestro CRC provides remote operation and monitoring of the Aria and Virtuoso Systems. Maestro also provides recording capabilities via outputs that connect to a polysomnograph for data recording. The data recording function is optional. The Maestro CRC can control the connected CPAP device whether or not it is connected to a polysomnograph.

When a patient is receiving therapy from the Aria or Virtuoso CPAP System, patient information/data is sent from the CPAP device to the Maestro CRC's memory. This information is updated at a frequency of 100 Hz. When the clinician uses the keypads to change settings or select functions, the microprocessor selects this information from the appropriate storage area for screen and/or polysomnograph display.

The Maestro CRC does not override or disable the pressure control at the CPAP device and cannot turn the CPAP device on or off. The following features of the Aria and Virtuoso Systems are disabled when connected to the Maestro:

  • Auto On Air flow will no longer automatically turn on when the patient . inhales through the connected patient circuit.
  • . Auto Off - Air flow will no longer automatically turn off when the mask is removed.
  • Audible Alerts (Mask Removal indicators) All alerts will now sound only . at the Maestro CRC so that the patient is not disturbed during the study.
  • Ramp/Reset Button Pressures can only be ramped and reset at the . Maestro CRC; however, in Virtuoso's Auto-CPAP mode, the Reset Button at the Virtuoso unit will reset the CPAP device during the leak test. Simply, a leak test is a 60 second test that allows the clinician to ensure that there are minimal or no leaks in the patient circuit.

{6}------------------------------------------------

Operations

When using the Maestro CRC, a clinician can perform a number of operations to control and retrieve information from the Aria or Virtuoso System. These operations include the adjustment of settings at the Maestro device while connected to either the Aria CPAP or Virtuoso Smart CPAP Systems. Since this premarket notification addresses only the Aria and Virtuoso Systems, only the operation of functions relating to these devices will be discussed.

CPAP system functions include modes of operation, parameter settings for each mode, and device parameters. Table 16-2 describes the modes available with the Aria and Virtuoso Systems, and a description of the modes follows.

CPAP ModeAuto-CPAP ModeSplit Nightwith Auto-CPAPMode
Aria CPAP SystemXN/AN/A
Virtuoso SmartCPAP SystemXXX
Table 16-2. Modes of the Aria and Virtuoso Systems
----------------------------------------------------------------

CPAP Mode

In the CPAP mode, the Aria and Virtuoso Systems produce fixed continuous positive airway pressure. The clinician can adjust settings to the following parameters in the CPAP mode:

  • Therapeutic Pressure
  • . Minimum Ramp Pressure
  • Ramp Time ●

Image /page/6/Picture/12 description: The image shows what appears to be a signature. The signature is composed of curved lines and strokes, forming a unique and stylized design. The lines are thick and bold, suggesting a deliberate and confident hand. The overall impression is that of a quick, yet distinctive autograph.

{7}------------------------------------------------

Auto-CPAP Mode

The Auto-CPAP mode has the ability to automatically determine upper airway instability and automatically generate a minimum pressure necessary to overcome airway collapse within a set pressure range defined by the Minimum Pressure and Maximum Pressure settings. The clinician can adjust settings to the following parameters in the Auto-CPAP mode:

  • Minimum Pressure .
  • Maximum Pressure .

Split Night with Auto-CPAP Mode

The term "Split Night" refers to a common sleep laboratory procedure in which a patient is diagnosed for half the night and titrated during the remaining half of the night. In this mode, the Virtuoso will deliver a minimum CPAP pressure for a preset time period referred to as the Delay Time. During Delay Time, the device will monitor and store data, but will not respond to airway instability when CPAP pressure is required. Once the Delay Time ends, the system will automatically respond to airway instability by increasing or decreasing the pressure to meet the patient's needs within designated pressure limits. The clinician can adjust settings to the following parameters in the Split Night with Auto-CPAP mode:

  • Delay Time .
  • Minimum Pressure .
  • . Maximum Pressure

The parameter settings for each mode are described below.

Therapeutic Pressure

The Therapeutic Pressure parameter applies only to the CPAP mode. Therapeutic Pressure is defined as the CPAP pressure level set by the clinician. The clinician can increase or decrease Therapeutic Pressure using the Setting Buttons.

{8}------------------------------------------------

Pressure adjustments can be made any time during therapy with a range of 3.0 to 20.0 cm H2O.

Ramp, Minimum Ramp Pressure, Ramp Time

The Ramp parameter applies only to the CPAP mode. · When activated, the Ramp feature initially reduces Therapeutic Pressure to the prescribed Minimum Ramp Pressure setting. Minimum Ramp Pressure is the pressure setting to which the device initially drops when Ramp is activated. Ramp incrementally increases to Therapeutic Pressure allowing the patient to fall asleep more comfortably. The time period required for the pressure to increase from Minimum Ramp Pressure to the Therapeutic Pressure is referred to as the Ramp Time.

The clinician can activate the Ramp and set both the Minimum Ramp Pressure and the Ramp Time by pressing the Ramp Function Button indicated by the main screen. Using the Setting Buttons, the Minimum Ramp Pressure can be set from 3.0 to 20.0 cm H2O and the Ramp Time can be set from 0 to 45 minutes.

Delay Time

The delay time parameter applies only to the Split Night with Auto-CPAP mode. In this mode, the Virtuoso will deliver a minimum CPAP pressure for a preset time period referred to as the Delay Time. Using the Setting Buttons, this parameter can be set from 30 minutes to 8 hours.

Minimum Pressure/Maximum Pressure

These Minimum Pressure/Maximum Pressure parameters apply only to the Auto-CPAP and Split Night with Auto-CPAP modes. Minimum and Maximum Pressure levels can each be set from 3 to 20 cm H2O. Minimum Pressure cannot be set higher than the Maximum Pressure setting, and Maximum Pressure cannot be set lower than the Minimum Pressure setting.

{9}------------------------------------------------

Reset

The Reset feature is only operable in the Auto-CPAP and Split Night with Auto-CPAP Modes on the Virtuoso System. When selected, Reset decreases the pressure to the Minimum Pressure setting and begins the self-titrating cycle again.

Maestro Device Parameters

Each Maestro function (Calibration, Mask Removal Indicator, View Graphs) is operable in all modes of the Aria and Virtuoso Systems.

Calibration

The clinician has the option to send calibration signals to the polysomnograph by pressing the "System Options/Cal" Function Button as indicated by the main screen. By pressing the "Calibrate Recorder" Button, the clinician can alternate between the baseline output signals and gain output signals allowing calibration of the polysomnograph.

Mask Removal Indicator

The Mask Removal Indicator uses an audible tone and a visual indicator to provide an alert if the mask, which delivers air pressure from the CPAP unit to the patient, is inadvertently removed or develops a significant leak during therapy. When using the Maestro CRC, all annunciators (visual and audible) are disabled at the CPAP device. The visual indicator is displayed on the Maestro's display screen and the audible tone is sounded through Maestro's annunciator. The audible tone can optionally be deactivated.

The clinician can access the audible settings of the Mask Removal Indicator by pressing the "System Options/Cal" Function Button as indicated by the main screen. Using the Setting Buttons, the clinician can activate or deactivate the audible tone and adjust its volume.

{10}------------------------------------------------

View Graphs

The clinician can view patient profile screens at any time during therapy by pressing the View Graphs Function Button indicated by the main screen. With this function, graphs and messages relating to patient profile data (Pressure Profile and Event Predictor Profile) are shown on the display screen. The graphs represent the pressure applied to the patient for the past hour.

User Operations

Set-up Procedures

The Maestro CRC communicates with the Aria or Virtuoso System by means of a Communications cable that connects the two devices via the RS-232 communication port. Optional connections from the Maestro CRC to a polysomnograph are made through the DC analog outputs that are applicable to the Aria and Virtuoso devices (i.e., Event Predictor and Pressure).

Start-up of the Maestro CRC

The clinician powers the Maestro CRC by turning the power on/off control switch to the on position. The LED adjacent to the power switch illuminates, indicating that the unit is powered, and an audible alert sounds. At start-up, the device performs a self-test. initializes the device software, and displays the start-up screen. An executable control function within the microprocessor manages normal and fault operational states, and identifies which Respironics CPAP System is connected. As the microprocessor completes the self-test and initialization, the Maestro CRC displays the start-up screen for approximately five seconds before the main screen appears.

Image /page/10/Picture/9 description: The image shows a signature in black ink on a white background. The signature appears to be stylized and illegible, with a series of connected strokes and loops. The strokes vary in thickness, suggesting the use of a pen or marker.

{11}------------------------------------------------

Using Display Menu Screens

The Maestro CRC Display Screen provides patient information as well as guidance in using the system. The screen also provides messages that keep the operator updated on the therapy being provided. Each menu screen will provide the following information:

  • . Screen title
  • Current mode .
  • Event Predictor* .
  • Current use of each Function and Setting Button .

*This parameter is only displayed in the Auto-CPAP mode when Maestro CRC is connected to the Virtuoso System.

In addition, a message area is provided in each menu screen for messages and directions that Maestro provides relating to selecting functions, setting parameters, and system status.

Selecting Functions

At the main screen, the clinician can use the Function Buttons located below the Display Screen to select the following Maestro CRC functions: System Options/Calibration, Ramp, View Graphs, and Change Mode. When the clinician selects a function, the appropriate menu appears on the Display Screen.

The Maestro CRC identifies the specific CPAP device model connected and the related functions available for control and display. The Maestro CRC will only provide menus for functions applicable to the CPAP System to which it is connected.

Adjusting Parameter Settings

The Setting Buttons, located at the right of the Display Screen, adjust the device parameter settings within the various function menus. Setting Buttons will relate to the parameters applicable to the current display screen. The menu provides onscreen display of the title for the current parameter of each pair of Setting Buttons.

Image /page/11/Picture/15 description: The image shows a signature in black ink on a white background. The signature is illegible, but it appears to be a cursive script with several loops and curves. The signature is compact and dense, with the individual letters not clearly distinguishable. The overall impression is that of a quick, stylized signature.

{12}------------------------------------------------

Accessories

There are no patient interface devices, such as masks, tubing, exhalation ports, or humidifiers, for use with the Maestro CRC. The Maestro CRC comes packaged with the following accessories:

  • Communications Cable (available in lengths of 25 ft. or 100 ft.) .
  • Power Supply (for connection to a 115 V or 230 V AC outlet) .
  • Power Cord (115 V or 230 V, depending on country of use) .
  • . Operating Instructions
  • Stand Assembly for positioning the Maestro CRC unit in a convenient . position for viewing and operation

Image /page/12/Figure/9 description: This image shows a collection of items, including a Maestro Clinical Remote Control, Operating Instructions, a Power Cord (North American cord shown), a Power Supply, a Communications Cable (available in lengths of 25 ft. (7.6 m), or 100 ft. (30.5 m), and a Stand Assembly. The remote control is a handheld device with a screen and buttons. The operating instructions are a booklet. The power cord and supply are used to power the remote control. The communications cable is used to connect the remote control to another device. The stand assembly is used to hold the remote control.

{13}------------------------------------------------

Safety and Effectiveness

Testing was performed to demonstrate that the performance of the Maestro CRC in its intended environment is as safe and effective as that of the legally marketed predicate devices.

The safety and effectiveness of the Maestro Clinical Remote Control was verified through performance-related testing that consisted of Electrical Safety, Electromagnetic Compatibility, and Mechanical and Environmental testing. The Maestro CRC was found compliant and has been certified to the following standards referenced in the "FDA Reviewer Guidance for Premarket Notifications."

  • IEC 601-1: General Requirements Safety of Medical Electrical Equipment; .
  • IEC 801-1: Electromagnetic Compatibility for Industrial Process and Measurement and . Control Equipment. Part 1;
  • IEC 801-2: Electrostatic Discharge Requirements; .
  • IEC 801-3: Radiated Electromagnetic Field Requirements; .
  • IEC 801-4: Electrical Fast Transients/Burst Requirements; .
  • CIPSR 11: Limits and Methods of Measurement of Radio interference Characteristics of . Industrial, Scientific, and Medical Equipment;
  • CISPR Specification for Radio Interference Measuring Apparatus and . CISPR 16: Measurement Methods;
  • IEC 68-2-6: Basic Environment Test Procedures. Part 2: Tests. Test Fc and Guidance: . Vibration (Sinusoidal);
  • IEC 68-2-27: Basic Environment Test Procedures. Part 2: Tests. Test Ea and Guidance: . Shock;
  • IEC 68-2-37: Basic Environment Test Procedures. Part 2: Tests. Test Fdc: Random . Vibration Wide Band - Reproducibility Low;
  • MIL-STD-461D: Requirements for Control of Electromagnetic Interference Emissions and . Susceptibility;
  • . MIL-STD-462D: Measurements of Electromagnetic Interference Characteristics; and
  • MIL-STD-810E: Environmental Test Methods .

The Maestro Clinical Remote Control is intended specifically for use with the Respironics Aria CPAP and Virtuoso Smart Systems that provide Continuous Airway Pressure (CPAP) for the treatment of adult OSA. CPAP devices, with and without remote control, have been on the market since 1984 and have been shown to be effective in treating this disease.

BiPAP® and REMstar Choice® are registered trademarks of Respironics, Inc. Companion® is a registered trademark of Puritan-Bennett. Inc.

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).