K Number
K954497

Validate with FDA (Live)

Date Cleared
1996-03-06

(161 days)

Product Code
Regulation Number
862.3870
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use in the qualitative analysis of cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

Device Description

FRONTLINE® CANNABIS is a homogeneous immunochromatographic assay for use in the qualitative analysis of cannabinoids in human urine. The assay is based on Gold Labeled Optical-read Rapid Immuno Assay (GLORIA) technology. Each test strip contains monoclonal antibodies reactive to a 11-nor-Δ9-THC-carboxylic acid derivative labeled with colloidal gold (conjugate) and 11-nor-Δ9-THC-carboxylic acid polyhapten, bulking agents, stabilizers and preservatives (solid phase). When immersed in urine, the test strip absorbs the volume of fluid necessary for the chromatographic reaction to occur. By capillary action, the urine passes through a compartment containing soluble conjugate which specifically binds to the cannabinoid analyte. Excess conjugate is retained by the solid phase in a separate compartment. Only that conjugate with bound cannabinoid analyte passes to the detection pad where a red-colored conjugate-analyte complex is viewed. The color developed on the detection pad is compared visually with a scale provided on the test strip vial. The intensity of the color developed correlates with the concentration of the analyte in the sample and the user may classify the analyte sample as: negative or positive (≥50 ng/mL).

AI/ML Overview

Here's an analysis of the provided text regarding the FRONTLINE® CANNABIS device, structured according to your requested information categories:

Acceptance Criteria and Device Performance Study

The acceptance criteria for the FRONTLINE® CANNABIS device are not explicitly stated as numerical thresholds (e.g., "sensitivity must be >X%"). Instead, the conclusion of the performance study ("FRONTLINE® CANNABIS is substantially equivalent to other commercially available cannabinoid test systems") implies that the device's performance, as measured against predicate devices and confirmation methods, met an implicit standard of equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

Given the lack of explicit numerical acceptance criteria, the table below reflects the reported performance without specific "acceptance criteria" columns. The "Performance Metric" and "Reported Device Performance" are derived from the comparison studies.

Performance Metric (Implicit Acceptance Criteria: Substantial Equivalence)Reported Device Performance (vs. EIA & Confirmation)Reported Device Performance (vs. FPIA & Confirmation)
Agreement with Predicate Device (Positive Samples)243 samples FRONTLINE® positive153 samples FRONTLINE® positive
Agreement with Predicate Device (Negative Samples)500 samples FRONTLINE® negative240 samples FRONTLINE® negative
False Positives14 samples3 samples
False NegativesNot explicitly stated (implied 0 for this comparison if 243 were true positives)2 samples
Unconfirmed Samples4 samples1 sample

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set):
    • Comparison to EIA: 761 metabolized urine samples
    • Comparison to FPIA: 399 metabolized urine samples
  • Data Provenance: The document does not specify the country of origin for the samples. It states that the samples were "metabolized urine samples evaluated at three sites" (for EIA comparison) and "evaluated at two sites" (for FPIA comparison). The type of data is retrospective, as it refers to "metabolized urine samples" that were collected and then evaluated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document refers to "confirmation analysis" and compares the FRONTLINE® device to "EIA and the confirmation method" and "FPIA and the confirmation method." This implies that a more definitive, likely laboratory-based, method was used as the ground truth. However, the document does not specify the number of experts involved in establishing this ground truth, nor their qualifications. It mentions only "the confirmation method."

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method for the test set. The term "confirmation analysis" suggests that the results from the predicate device and the FRONTLINE® device were compared against a higher-standard confirmation method, which implicitly serves as the adjudicator of true positive/negative status. However, no details about how discrepancies were resolved or if human adjudicators were involved are provided.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The study focuses on the standalone performance of the device compared to predicate devices and a confirmation method, not on human readers' improvement with or without AI assistance. The FRONTLINE® CANNABIS device itself is a "non-instrumented assay" requiring visual comparison, but the study design is not an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance study was done. The FRONTLINE® CANNABIS device is described as a "non-instrumented assay" where "The color developed on the detection pad is compared visually with a scale provided on the test strip vial." The performance data presented are for the device itself determining positivity or negativity, with this visual interpretation being part of the device's operational design. The comparison is between the device's output and established methods (EIA, FPIA, and confirmation), making it a standalone evaluation of the device.

7. The Type of Ground Truth Used

The ground truth used was a "confirmation method" alongside predicate devices (EIA and FPIA). While the exact nature of this "confirmation method" is not detailed, in urine drug testing, confirmation typically refers to more definitive analytical techniques like Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS), which are considered gold standards for identifying and quantifying substances.

8. The Sample Size for the Training Set

The document does not mention a training set or any machine learning/AI model training. This device appears to be an immunochromatographic assay, not an AI-driven one.

9. How the Ground Truth for the Training Set Was Established

As there is no training set mentioned or implied for this type of device, this question is not applicable.

{0}------------------------------------------------

K954497

510(k) SUMMARY - FRONTLINE® CANNABIS

Submitter Name:Boehringer Mannheim GmbH
Submitter Address:Sandhofer Str. 116D-68298 Mannheim, Germany
Contact Person:Richard Naples, DirectorRegulatory and Government Affairs
Phone Number:(317) 576-3858
Fax Number:(317) 845-2324
Date Prepared:February 29, 1996
Device Trade Name:FRONTLINE® CANNABIS
Device Common Name:Immunochromatographic test for cannabinoids
Classification Name:Cannabinoids test system(per 21 CFR §862.3870)
Predicate Devices:Emit® d.a.u. Cannabinoid 100 ng Assay;TDx® Cannabinoids
Device Description:FRONTLINE® CANNABIS is a homogeneousimmunochromatographic assay for use in the qualitativeanalysis of cannabinoids in human urine. The assay is basedon Gold Labeled Optical-read Rapid Immuno Assay(GLORIA) technology. Each test strip contains monoclonalantibodies reactive to a 11-nor-Δ9-THC-carboxylic acidderivative labeled with colloidal gold (conjugate) and 11-nor-Δ9-THC-carboxylic acid polyhapten, bulking agents,stabilizers and preservatives (solid phase). When immersedin urine, the test strip absorbs the volume of fluid necessaryfor the chromatographic reaction to occur. By capillaryaction, the urine passes through a compartment containingsoluble conjugate which specifically binds to the cannabinoidanalyte. Excess conjugate is retained by the solid phase in aseparate compartment. Only that conjugate with boundcannabinoid analyte passes to the detection pad where a red-colored conjugate-analyte complex is viewed. The colordeveloped on the detection pad is compared visually with ascale provided on the test strip vial. The intensity of thecolor developed correlates with the concentration of theanalyte in the sample and the user may classify the analytesample as: negative or positive (≥50 ng/mL).
Intended Use:For use in the qualitative analysis of cannabinoids in human urineat a cutoff concentration of 50 ng/mL.

l

ub

:

{1}------------------------------------------------

Device Technological Characteristics and Comparison to Predicate Devices:

FRONTLINE® is based on Gold Labeled Optical-read Rapid Immuno Assay (GLORIA) technology. Each test strip conrains monoclonal antibodies reactive to a 1 1-nor-A9-THCcarboxylic acid derivative labeled with colloidal gold (conjugate) and 11-nor-A9-THC-carboxylic acid polyhapten, bulking agents, stabilizers and preservatives (solid phase). When immersed in urine, the test strip absorbs the volume of fluid necessary for the chromatographic reaction to occur. By capillary action, the urine passes through a compartment containing soluble conjugate which specifically binds to the cannabinoid analyte. Excess conjugate is retained by the solid phase in a separate compartment. Only that conjugate with bound cannabinoid analyte passes to the detection pad where a red-colored conjugate-analyse complex is viewed. The color developed on the detection pad is compared visually with a scale provided on the test strip vial. FRONTLINE@ CANNABIS is a non-instrumented assay and requires no user calibration.

Emit® d.a.u. Cannabinoid 100 ng Assay is an enzyme immunoassay (EIA) based on the competition between analyte in in the sample and analyse labeled with an enzyme for antibody binding sites. Results are measured with a suitable instrument based on absorbance changes compared to those generated by defined calibrators.

TDx@ Cannabinoids is a fluorescence polarization immunoassay (FPIA) in which the change in fluorescence polarization due to antibody binding correlates with the concentration of analyte in the sample. Results are measured with a suitable instrument and compared to those generated by defined calibrators.

Performance Data:

Conclusion:

Comnarison to EIA: A sotal of 761 metabolized urine samples were evaluated at three sites. 243 samples were FRONTLINE® positive and 500 were FRONTLINE® negative when compared to EIA and the confirmation method. Fourteen samples were classified as false positive and four samples were unavailable for confirmation analysis.

Comparison to PPIA: A total of 399 metabolized urine samples were evaluated at two sites. 153 samples were FRONTLINE® positive and 240 were FRONTLINES negative when compared to FPIA and the confirmation method. Three samples were classified as false positive, two samples were classified as false negative, and one sample was unavailable for confirmation analysis.

FRONTLINE® CANNABIS is substantially equivalent to other commercially available cannabinoid test systems.

ટ વે

§ 862.3870 Cannabinoid test system.

(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).