K Number
K954124

Validate with FDA (Live)

Date Cleared
1996-03-19

(200 days)

Product Code
Regulation Number
886.5916
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RGP Spherical contact lenses made from O->PERM 30A (sterafocon A) which contain a visibility tint are indicated for daily wear use. The lenses are indicated for visual acuity in patients with with non-diseased eyes. The spherical lenses are being applied for from +12.00 to -20.00 Diopters up to 4.00 Diopters of astigmatism where it does not interfere with visual acuity.

RGP Toric contact lenses made from 0->PERM 30A (sterafocon A) which contain a visibility tint are also indicated for daily wear. The lenses are indicated for visual acuity in patients with with non-diseased eyes. The toric lenses are being applied for from +12.00 to -20.00 Diopters up to 6.00 Diopters of astigmatism where it does not interfere with visual acuity.

Device Description

O->PERM 30A (sterafocon A) is a RGP silicone-acrylate polymer which can be lathed into spherical or toric contact lenses. These hemispherical shells when placed on the eye act as a refracting medium to focus light rays on the retina.

AI/ML Overview

Here's an analysis of the provided text regarding the O->PERM 30A (sterafocon A) contact lens, structured to address your specific questions.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Metric (Criteria)Reported Device Performance
SafetyAdverse Reactions (none reported)"There were no lens related Adverse Reactions reported in the study."
Slit Lamp Findings (typical, no treatment required)"All Slit Lamp Findings were typical... No slit lamp findings required treatment."
Bioburden Level (<100 CFU/lens for alternate site manufacturers)Not directly assessed in the clinical study, but a quality control measure for manufacturing.
EffectivenessVisual Acuity (20/30 or better for majority)Trial Lens: 94.4% (68 out of 72 eyes) achieved 20/30 or better. Control Lens: 88.2% (30 out of 34 eyes) achieved 20/30 or better.
Wearing Time (typical for daily wear: 10-14 hours/day)"The average wearing time for all patients participating in the Study was typical of daily wear patients generally: 10 to 14 hours per day."
Discontinuations (normal for contact lenses)Trial Cell: 16.7% (6 out of 36) Control Cell: 17.6% (3 out of 17)
Symptoms/Complaints (no unanticipated, no treatment required)"There were no unanticipated observations... and none... required treatment..."
Lens Replacements (mostly for fitting/loss, normal reasons)"Lens replacements during the study were mostly for fitting related reasons, or other normal contact lens wearing related reasons (such as loss)."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set):
    • The study aimed for approximately 15 patients per investigator.
    • The study had a "random distribution of 2:1 in the trial lens : control lens."
    • Based on reported visual acuity data:
      • Trial lens: 72 eyes (likely 36 patients, assuming 2 eyes per patient)
      • Control lens: 34 eyes (likely 17 patients, assuming 2 eyes per patient)
    • Total participants in effectiveness assessment: Approximately 53 patients (106 eyes).
    • In the discontinuation section, it states: "6 out of 36 (16.7%) in the trial cell, and 3 out of 17 (17.6%) in the control cell," reinforcing the approximate patient count.
  • Data Provenance: The data is from a prospective clinical study conducted in the USA (specifically mentioning "Salvatori Ophthalmics Manufacturing Inc. of Sarasota, Florida" as a collaborator and "The South West Independent TRB" for IRB approval).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number or specific qualifications of experts used to establish ground truth for the clinical study. It refers to "Investigators" who conducted the patient examinations and followed the protocol established by Optical Polymer Research, Inc. (OPR). Given the context of a contact lens study, these investigators would typically be ophthalmologists or optometrists experienced in fitting and evaluating contact lens wear.

4. Adjudication Method for the Test Set

The document does not mention a specific adjudication method (e.g., 2+1, 3+1). The study design appears to be focused on individual investigator assessments against predefined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is for a medical device (contact lens) and not an AI-powered diagnostic or assistive tool. Therefore, the concept of human readers improving with AI assistance is not applicable here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, this is a study for a physical medical device (contact lens), not an algorithm or AI. Standalone algorithm performance is not applicable.

7. The Type of Ground Truth Used

The ground truth used in this study is based on clinical observations and measurements by qualified healthcare professionals (Investigators), combined with patient-reported outcomes. This includes:

  • Visual acuity measurements.
  • Slit lamp examination findings.
  • Patient wearing time.
  • Reports of adverse reactions, symptoms, and complaints.
  • Reasons for lens replacements or discontinuations.

8. The Sample Size for the Training Set

This document describes a clinical study for a physical medical device. There is no mention of a training set in the context of machine learning or AI. The term "training set" is not relevant to this type of device submission.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for an AI or algorithm in this submission, this question is not applicable.

{0}------------------------------------------------

510(k) Notification for O->PERM 30A (sterafocon A) Feb. 15, 1996

MAR 1 9 Janes SECTION 11 SUMMARY of SAFETY AND EFFECTIVENESS

  • I . General Information:
    :

  • A. Generic (USAN) Name: sterafocon A

  • B. Device Trade Name: O->PERM 30A

  • C. Sponsor's Name and Address : Dr. Paul Schuman Optical Polymer Research, Inc. 5921 N.E. 38th Street Gainesville, FL 32609 (904) 378-1027

  • D. 510(k) Number: K954124

  • E. Date Notice of approval sent to Sponsor:

  • F. First OPR Approved Alternate Site Manufacturer and Distributor of Lenses made from the 0->PERM 30 A (sterafocon A) optical blanks (Pending FDA approval):

Salvatori Ophthalmics Mfg. Inc. 6416 Parkland Drive Sarasota, FL 34243 (813) 751-0383

II. Indications

RGP Spherical contact lenses made from O->PERM 30A (sterafocon A) which contain a visibility tint are indicated for daily wear use. The lenses are indicated for visual acuity in patients with with non-diseased eyes. The spherical lenses are being applied for from +12.00 to -20.00 Diopters up to 4.00 Diopters of astigmatism where it does not interfere with visual acuity.

RGP Toric contact lenses made from 0->PERM 30A (sterafocon A) which contain a visibility tint are also indicated for daily wear. The lenses are indicated for visual acuity in patients with with non-diseased eyes. The toric lenses are being applied for from +12.00 to -20.00 Diopters up to 6.00 Diopters of astigmatism where it does not interfere with visual acuity.

IJI. Device Description

O->PERM 30A (sterafocon A) is a RGP silicone-acrylate polymer which can be lathed into spherical or toric contact lenses. These hemispherical shells when placed on the eye

{1}------------------------------------------------

act as a refracting medium to focus light rays on the retina.

  • IV. Alternative Practices or Procedures
    Alternative practices or procedures are the use of other available previously approved RGP silicone-acrylate contact lenses or soft (hydrophilic) contact lenses or prescription spectacles as deemed best by the patient's eye care practitioner.

  • V. Technical Summaries

    • A. Nonclinical
        1. Toxicology

Optical Polymer Research, Inc. had all the pre-clinical testing studies necessary for IRB review for Class II daily wear RGP contact lenses. The South West Independent TRB granted their approval to OPR's protocol to start a clinical study on O->PERM 30 A (steratocon A) for daily wear on June 6, 1994.

Conclusion: Approved by IRB

    1. Microbiology
      To assure that a low bioburden level is maintained by alternate site lens manufacturers, a yearly bioburden test will be done using a validated bioburden test method. A <100 CFU/lens will be the specification for bioburden. Implementation of written quality assurance measures will be assured by all alternate sites.

Conclusion: Approved by IRB

    1. Tiens compatibility studies: Used FDA approved RGP contact lenses solutions. Some preliminary bench work to assure compatibility.
      Conclusion: Approved by IRB
    1. Lens Stability: not required with the RGP material.
    1. Preservative Uptake Tests: Are not required with the RGP material.
  • B. Clinical

This study was run with the help of Salvatori Ophthalmics Manufacturing Inc. of Sarasota, Florida. The clinical study started after obtaining IRB approval.

Optical Polymer Research, Inc. (OPR) will own the material O->PERM 30A (sterafocon A) and has submitted a protocol for the acceptance of alternate site finishing labs. All labs will be subject to testing

{2}------------------------------------------------

and, if they pass, acceptance.

The equipment manufacturers of lathes and polishing equipment may be different between the various contact lens labs but the finished product will be equivalent in all instances. The recommended manufacturing materials, manufacturing methods, and quality control methods employed by the companies are specified by OPR.

Optical Polymer Research, Inc. was last FDA inspected on August 15/17, 1995

  • C. Patient Selection:
    Fach patient was be selected according to the following criteria:

    1. They must fill in and sign an informed consent form (as approved by the IRB) before any trial lens is There are to be no exceptions. dispensed.
    1. Have no excessive corneal astigmatism that would interfere with the correction of visual acuity.
    1. The patients eyes should be correctable to at least 20/30 for distance as a minimum standard.
    1. They must have non-diseased eyes which are normal or which have a preexisting ocular condition which would not be expected to interfere with the patient's ability to wear contact lenses successfully. A normal eye is defined as one having all the following characteristics:
    • a. No evidence of lid infection.
    • b. No structural lid abnormality
    • c. No conjunctival abnormality or infection
    • d. A cornea which is clear with no edema, staining, vascularization or abnormal opacities, all as observed by slit lamp examination.
    • e. No iritis.
    • f. No other active disease that would contraindicate the wearing of contact lenses or has caused the patient to terminate contact lenses at some previous time.
    1. Use no ocular medication.
    1. Have no known sensitivity to solutions currently used by the investigator for lens care.
    1. Have binocular correction with contact lenses.

The Food and Drug Administration is aware that certain patient characteristics such as mild edema or staining or other minor physical conditions may not necessarily preclude the patient from participating in the study: however, these conditions should be well documented in the preliminary or initial visit and followed during the study.

{3}------------------------------------------------

D. Study Period

The study period was defined as three (3) months continuous wear. The patients were seen by the Investigators at 7 days, 14 days, 1 month, 2 months and 3 months following the initial dispense date.

  • F. Study Population
    The study population was to consist of randomly selected patients, approximately fifteen (15) per investigator with a random distribution of 2:1 in the trial lens : control lens .

    1. Safety appears to be well within Normal Limits
      The safety of the O->PERM 30 A polymer (sterafocon A) is demonstrated by the study to be well within normal limits for a contact lens of this class.
  • a. Adverse Reactions
    There were no lens related Adverse Reactions reported in the study.

  • b. Slit Lamp Findings
    All Slit Lamp Findings were typical of that which is generally expected from patients wearing No slit lamp contact lenses in this class. findings required treatment.

    1. Effectiveness
      The effectiveness of the O->PERM 30 A (sterafocon A) is demonstrated by the study to be well within Normal Limits for a contact lens in this class.
  • a. Visual Acuity
    The majority of the patients completing the study period exhibited a final VA of 20/30 or better: 68 eyes out of 72 (94.4%) for the trial lens and 30 eyes out of 34 (88.2%) for the control lens.

  • b. Wearing Time
    The average wearing time for all patients participating in the Study was typical of daily wear patients generally: 10 to 14 hours per day.

  • c. Discontinued Patients
    Patient discontinuations were normal for contact lens patients in general with 6 out of 36 (16.78)

{4}------------------------------------------------

in the trial cell, and 3 out of 17 (17.6%) in the control cell.

  • d. Symptoms Problems and Complaints
    There were no unanticipated observations in this category and none of the observations noted required treatment or gave any cause for a determination of a problem directly attributable to the wearing of the trial lenses.

  • e. Replacements
    Lens replacements during the study were mostly for fitting related reasons, or other normal contact lens wearing related reasons (such as loss).

  • I . Potential Adverse Effects of the Device on Health
    None known

  • II. Conclusion drawn from Study
    It can be concluded from the clinical study results that the O->PERM 30 A (sterafocon A) polymer is Safe and Effective for daily wear use in patients with non diseased eyes.

  • III. FDA Recommendation
    Tt is recommended that the FDA approve the O->PERM 30 A (sterafocon A) for Daily wear use.

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.