K Number
K952315

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-12-04

(567 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sathlaster is intended for use an adjunct to traditional Pulse Oximetry, in which the caregiver observes instantaneous Pulse Oximetry values as displayed on a Pulse Oximeter's face. SatMaster produces graphs and statistics describing a patient's Pulse Oximerry data trends over relatively long periods of time, including those times when the caregiver is not directly observing the Pulse Oximeter's display.

Device Description

SatMaster is the name of EMG Scientific's Computerized Pulse Oximetry Software product. This product consists of a Program Diskette, a Software License, a User Manual, and a Test Cable. Thus, SatMaster is a software-only "device," intended for use with an MS-DOS, PC-type computer provided by the end-user. Briefly, the SatMaster software provides a PC computer user the capability of transferring Pulse Oximetry data from a Pulse Oximeter, storing the data on the computer, displaying the data in various forms, annotating the data, calculating statistics, and printing the results.

AI/ML Overview

The provided document is a 510(k) Summary for a software device called "SatMaster," a Computerized Pulse Oximetry Software. This type of document is submitted to the FDA to demonstrate substantial equivalence to a legally marketed predicate device, rather than to establish new safety and effectiveness through a clinical study with acceptance criteria.

Therefore, the document does NOT contain the following information:

  • A table of acceptance criteria and reported device performance.
  • Sample sizes used for a test set or data provenance for such a test set.
  • Number of experts or their qualifications for establishing ground truth for a test set.
  • Adjudication method for a test set.
  • Results of a multi-reader multi-case (MRMC) comparative effectiveness study, or any effect size related to AI improvement.
  • Details of a standalone (algorithm only) performance study.
  • Type of ground truth used (e.g., pathology, outcomes data) for a test set.
  • Sample size for a training set.
  • How ground truth for a training set was established.

Analysis of the Provided Information:

The document focuses on demonstrating that SatMaster is substantially equivalent to a predicate device (Nellcor N-250 Pulse Oximeter, which is a combination of Nellcor N-200 Pulse Oximeter and Nellcor N-50 Powerbase).

Instead of clinical studies with acceptance criteria, the document provides:

  1. Device Description and Intended Use: SatMaster is a software-only device that transfers, stores, displays, annotates, calculates statistics, and prints pulse oximetry data from approved pulse oximeters. It is intended as an accessory to enhance a caregiver's use of Pulse Oximeters, providing trends over long periods.
  2. Comparison to Predicate Device:
    • Equivalence: SatMaster performs similar functions to the Nellcor N-50 Powerbase portion of the N-250, such as extracting data, displaying trend information, producing tabular lists, and generating histograms of SaO2 levels.
    • Differences: SatMaster is not a patient monitor, does not provide alarms or ECG/Pleth waveform output, does not power the oximeter, and provides additional reports and data-management functions not present in the predicate device.
  3. Safety and Effectiveness Argument:
    • The document argues that SatMaster's use should have no impact on the safety of the Pulse Oximetry monitoring process because the Pulse Oximeter itself (an approved device) performs the primary monitoring, and SatMaster only processes the data after collection.
    • It states that SatMaster software has been "designed and extensively tested to insure the correctness and accuracy of SatMaster's data retrieval, data management, data storage, statistical calculations, and reports." However, no details about these tests (e.g., methodology, sample sizes, specific metrics, or acceptance criteria) are provided.
    • It claims enhancements in safety and effectiveness compared to the predicate device, such as requiring patient ID entry (reducing mix-up risk) and generating more extensive reports.
    • It cites published articles describing research performed using SatMaster by various researchers, implying its clinical value. This is anecdotal evidence of device utility rather than a formal study proving specific performance metrics against defined acceptance criteria.

In summary, to directly answer the request, the provided 510(k) summary does not contain the information typically found in a study demonstrating a device meets acceptance criteria. It relies on a demonstration of substantial equivalence to an existing device and general statements about testing and clinical utility, rather than specific performance metrics against established thresholds.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a date and a series of numbers. The date reads "Dec. 4, 1796". Below the date, there is a sequence of numbers that reads "K952315".

Innovators of medical software solutions

Image /page/0/Picture/2 description: The image shows the word "SCIENTIFIC" in bold, black letters. The word is partially obscured by a black graphic to the left. The font is sans-serif and appears to be italicized.

1461 S. Livonia Avenue Los Angeles, California 90035-3317 FAX 310-551.1182 310-551.1181 shmuel@einstein.com Internet email:

510(k) Summary

Contact PersonShmuel Einstein
DateAugust 5, 1996
Device Name"SatMaster," Computerized Pulse Oximetry Software
510(k) File NumberK952315

The Device 1.

SatMaster is the name of EMG Scientific's Computerized Pulse Oximetry Software product. This product consists of a Program Diskette, a Software License, a User Manual, and a Test Cable. Thus, SatMaster is a software-only "device," intended for use with an MS-DOS, PC-type computer provided by the end-user.

Briefly, the SatMaster software provides a PC computer user the capability of transferring Pulse Oximetry data from a Pulse Oximeter, storing the data on the computer, displaying the data in various forms, annotating the data, calculating statistics, and printing the results.

Sathlaster has been in use outside the US for over 5 years. Prominent foreign users include the Queen Mary Hospital at the University of Hong Kong, Osaka Medical & Health Center for Mothers & Children, Adelaide Children's Hospital, Royal University Hospital, National University of Singapore, Helsingborgs Lasarett (Sweden), Royal Brisbanc Hospital (Australia), University of Malaysia -Department of Anaesthesiology, Universitair Zickenhuis Antwerpen, University of Calgary, Forschungsinstitut fur klinische Medizintchnik (Germany), Broadgreen Hospital NHS Trust (United Kingdom).

2. An Accessory

SatMaster is designed to be an accessory to enhance a caregiver's use of Pulse Oximeters. Pulse Oximeters, approved. Class II devices manufactured by several companies, are used to measure a patient's Oxygen Saturation (SaO2) and Pulse Rate. The current values of these clinical parameters are displayed on the Pulse Oximeter's front pancl. Many Pulse Oximeters contain a memory to store numerous hours of oximetry trend data.

328

{1}------------------------------------------------

August 5, 1996

Likewise, many Pulse Oximeters contain a Scrial Communications Port, so that the Pulse Oximeter may transfer data to a computer, a printer, or some other device. SatMaster would only be useful in conjunction with an oximeter having a Serial Communications Port, since SatMaster receives data from the Pulse Oximeter via a connection between the computer and the Pulse Oximeter's Serial Communications Port.

Some Pulse Oximeter's can be connected to various types of printing devices. These Pulse Oximeters, when connected to a printer, can produce graphs of SaO2 vs. time, histograms, and certain statistics.

Here is a list of the Pulse Oximeters that SatMaster has been designed to work with:

ManufacturerModelStock NumberClearance Date
NellcorN-200 Pulse OximeterK86378421 November 1986
NellcorSymphony N-3000 Pulse OximeterK94234712 May 1995
OhmedaBiox 3700 Pulse OximeterK85049428 March 1985
OhmedaBiox 3740 Pulse OximeterK8727727 October 1987

The Predicate Device 3.

The Predicate Device that SatMaster resembles is a Nelloor N-250 Pulse Oximeter. The Nellcor N-250 Pulse Oximeter is actually a composite device obtained by connecting a Nellcor N-200 Pulse Oximeter to the Nellcor N-50 Powerbase. The combination of these two devices is known as the N-250 Pulse Oximeter.

The FDA 510(k) clearances for these two Nelloor products are:

Nellcor N-200 Pulse OximeterK863784 (21 November 1986)
Nellcor N-50 PowerbaseK901399 (12 June 1990)

The features of these two devices that arc relevant to this discussion are as follows. The Nellcor N-200 Pulse Oximeter performs Pulse Oximetry monitoring on the patient, displays instantaneous values of the SaO2 and Pulse Rate, and stores the Pulse Oximetry data in its trend memory at 5 second intervals. The N-200 also produces audible alarms for conditions such as high and low SaO2, high and low Pulse Rate, etc.

{2}------------------------------------------------

Trend (see Figure 12), shows SaOt levels

The Nellcor N-50 Powerbase contains an LCD scroen, and a strip-chart printer. The unit can access the connected Nellcor N-200's trend memory and display a graph of SaO2 vs. time on the LCD screen. Using the N-50's built-in printer, the following graphs and statistics can be output:

  • Trend Graph of SaO2 and Pulse Rate vs. time 1.
  • A histogram showing percent of time at various SaO2 levels 2.
  • A Tabular list of SaO2 values at 30 second intervals 3.

Here is an example of the graphs and statistics produced by the Nellcor N-50:

Tabular printout (see Figure 11) provides a histogram for SaO2 and other tabular data for saturation and out

Image /page/2/Figure/8 description: The image shows a series of six different charts or diagrams, labeled 62 through 67. The first few charts, 62 through 66, appear to be some sort of data representation, possibly related to signal processing or data analysis. The last chart, 67, is a bar graph with values on the x-axis ranging from 70 to 100, and the y-axis labeled as 'STREK'.

Flours 11: N-90 Tabular Printout Ex

Image /page/2/Figure/10 description: The image shows a medical chart with the word "NELLCOR" vertically oriented on the left side. The chart includes numerical data, such as 68, 69, and 70, along with labels like "11/29", "100K", "50K", "RATE", and "00:59". The chart also contains a series of horizontal lines and vertical dotted lines, suggesting a graphical representation of patient data or monitoring information.

n (see Figure 13), shows a maximum of 12 hours of trend data

Image /page/2/Figure/13 description: The image shows two separate figures. The figure on the left has the word "NELLCOR" written vertically. The figure on the right is a bar graph that shows the percentage of time spent in a given range. The x-axis ranges from 70 to 100, and the y-axis ranges from 0 to 100k.

Image /page/2/Figure/14 description: The image contains text that appears to be from the Nello N-500 Operators Manual. The text indicates that the content is reproduced with permission from Nello. The text is in a dark font and is set against a light background, making it readable. The overall layout suggests it may be a header or source attribution for a document or excerpt.

Figure 13: N-50 Histogram Printout Ex

4. SatMaster's Equivalence to Predicate Device

SatMaster is substantially equivalent to the Predicate Device in the following areas:

  • The ability to extract Pulse Oximetry data from a Pulse Oximeter and to display Pulse Oximetry 1. trend information on the screen and in printed reports
  • The ability to produce a tabular list of Pulse Oximetry data values at various time intervals 2.
  • The ability to generate histograms showing percent of time at various SaO2 levels 3.

SatMaster is not intended to be a patient monitor. Therefore. SatMaster does not perform the functions of the Nellcor N-200 Pulse Oximeter. Rather, a computer running the SatMaster software performs some of the same functions a Nellcor N-50 Powerbase. Thus, a computer running the SatMaster software. when used with a Pulse Oximeter, provides some of the capabilities of the Nellcor N-250.

5. Differences between SatMaster and the Predicate Device

Although SatMaster is substantially equivalent to the Predicate Device, certain important differences should be noted:

    1. Whereas the N-250 combination serves as a patient oxygenation monitor, SatMaster performs no monitoring of the patient. Rather, SatMaster presents the user with reports based on the data
      3leo

{3}------------------------------------------------

that has been collected by the Pulse Oximeter, which is, of course, an approved monitoring device. Thus, by design, SatMaster does not perform the functions of the Nellcor N-200 Pulse Oximeter.

  • Since SatMaster is not intended to be a patient monitor, SatMaster has no provisions for 2. reporting of the Pulse Oximeter's alarms.
  • Although the Nellcor N-250 can be connected to an external ECG synchronization source, 3. SatMaster does not produce an ECG or Pleth waveform output, as the N-250 does.
  • The Nellcor N-50 Powerbase acts as a dock into which the Nellcor N-200 Pulse Oximeter is 4. inserted. The N-50 provides electrical power to the Nellcor N-200 Pulse oximeter. Likewise, several connectors are present in the N-50 to allow connection of the composite N-250 device to a variety of external devices.

Since SatMaster is a software system, SatMaster does not provide the Pulse Oximeter with electrical power, nor does it provide the Pulse Oximeter with any connections to external devices. Rather, the user's computer is connected to the Pulse Oximeter's Serial Data Port via a simple RS-232 cable.

  • SatMaster gencrates additional reports not provided by the N-250. Morcover, SatMaster-5. generated versions of the N-250's reports contain various enhancements.
  • SatMaster performs various data-management functions to aid in storing and organizing the 6. Pulse Oximetry data that has been obtained from the Pulse Oximeter. These-data management functions are designed to be easy for the user to use, and at the same time to increase patient safety.

6. Intended Use of SatMaster

Sathlaster is intended for use an adjunct to traditional Pulse Oximetry, in which the caregiver observes instantaneous Pulse Oximetry values as displayed on a Pulse Oximeter's face. SatMaster produces graphs and statistics describing a patient's Pulse Oximerry data trends over relatively long periods of time, including those times when the caregiver is not directly observing the Pulse Oximeter's display.

In a hospital setting. SalMaster would be used in association with ICU's. Operating and Recovery Rooms, Pulmonary Function Labs, Sleep Labs, and research.

In home-care settings. SatMaster would be used for sleep studies and home-oxygen evaluations.

It is expected that the Pulse Oximeter itself will be operated by competent personnel in accordance with the Pulse Oximcter manufacturer's FDA-approved instructions.

Once the patient monitoring process is complete, cither in the hospital or home setting, the Pulse Oximeter unit would normally be brought to the office where the computer resides. (Alternatively, a portable computer would be brought to the Pulse Oximeter.) After connecting the serial oximeter tocomputer data cable and invoking the SatMaster software, the accumulated Pulse Oximetry data would be transferred from the Pulse Oximeter to the computer. Once the data transfer is complete, it is no longer necessary for the computer to be connected to the Pulse Oximeter. The caregiver may now view the Pulse Oximetry data and statistics on the computer screen, annotate the data, and print various reports.

7. Safety and Effectiveness of SatMaster Compared to the Predicate Device

As noted above, SatMaster is not a patient monitor Rather, SatMaster is intended to be an Accessory to the use of Pulse Oximeters. It is expected that the Pulse Oximeter itself will be operated by competent personnel in accordance with the Pulse Oximeter manufacturer's FDA-approved instructions.

{4}------------------------------------------------

The safe and effective monitoring of a patient by a Pulse Oximeter while connected to an external device, such as a computer running SatMaster, is a documented feature of an approved Pulse Oximeter. Thus, the Pulse Oximeter-to-patient connection, and the Pulse Oximetry monitoring process, is unaffected by whether will be used to display or calculate statistics on the data. Therefore, using SatMaster should have no impact on the safety of the Pulse Oximetry monitoring process

The SatMaster software has been designed and extensively tested to insure the correctness and accuracy of SatMaster's data retrieval, data management, data storage, statistical calculations, and reports.

As mentioned above, certain features of SatMaster represent enhancements of safety and effectiveness when compared with the predicate device. For example, a strip chart generated by the predicate device does not bear the patient's name or ID number; the possibility exists that it could be mixed up with that of another patient. When using SatMaster, however, before any oximetry data is transferred to the computer, the operator must enter the patient's name and ID number into SatMaster's Patient Data Management facility. This identification information is then prominently shown on every subsequent screen and report.

Another example: the extensive reports and statistics generated by SatMaster can provide additional information to the caregiver over that produced by the predicate device.

SatMaster serves to cnhance the usage of Pulse Oximcers. Provided that a Pulse Oximeter is operated in accordance with the Pulse Oximeter's Manufacturer's instructions, SatMaster reports and statistics provide a convenient "bicture" of the data collected by Pulse Oximeters. In the opinion of a number of researchers, SatMaster's functionality may have clinical value in several areas, as can be seen from the following list of published articles describing research performed using SatMaster:

Durand M. McEvoy C. MacDonald K. Spontaneous desaturations in intubated very low birth weight infants (<1000 grams) with acute and chronic lung disease. Pediatric Pulmonology July 1992; 13: 136-142.

Durand M, McEvoy C, MacDonald K. Computerized pulse oximetry and pulmonary resistance in proterm ventilator dependent infants (<1000g) with chronic lung disease. Intensive Care Medicine June 1990: 16, Supplement 1: Abstract 265.

McEvoy C, Sardesai S, Macri C, Paul R, Durand M: Neonatal pulmonary mechanics and oxypenation following arnnioinfusion in labor: a randomized clinical trial. Pediatric Research April 1991; 29: 226A.

McEvoy C, Hewlett V, Sardesai S, Mendoza E. Durand M: Prone positioning decreases episodes of desaturation in infants with chronic lung disease. Journal of Pediatric Research April 1992: 31: 211A.

Durand M. McEvoy C, MacDonald K: Computerized pulse oximetry and pulmonary mechanics in intubated low birth weight infants with acute and chronic lung disease. Pediatric Research April 1990; 27: 300A.

Jones R, Lawson A, Gunawardene W, Roulson C, Brown A, Smith I: An evaluation of prolonged oximetric data acquisition. Anaesthesia and Intensive Care 1992; 20: 303-307.

MacDonald KD, Salaz E, Franceschini RE: Neonatal synchronous intermittent mandatory ventilation (SIMV) improves oxygenation. Respiratory Care 1991; 36; 811.

MacDonald KD, Salaz E, Wirtschafter DD: Effect of patient flow triggered ventilation (PFTV) on neonatal pulmonary mechanics. Respiratory Care 1991; 36 810

Thus SatMaster can be seen to be both safe and effective when used in accordance with its instructions.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).