K Number
K946127

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-10-08

(663 days)

Product Code
Regulation Number
868.5160
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ohmeda Excel Draw-Over Military Vaporizer System is specifically designed for use with the Ohmeda Single-Agent Draw-Over Vaporizer. It consists of an Excel 210 system with an attached draw-over vaporizer. The purpose is to provide a platform for the delivery of inhalation anesthetic agents using the draw-over vaporizer and supplemental gas supply.

Device Description

The Ohmeda Excel Draw-Over Military Vaporizer System is a modification of the Ohmeda Excel 210 that has been made to facilitate the Ohmeda Single-Agent Draw-Over Vaporizer and the militaries needs. The system consists of an Ohmeda Single Agent Draw Over Vaporizer and a kit that allows the vaporizer to be used with an Ohmeda Excel system. The kit consists of mounting attachments, O2 flush cover, monitoring ports, scavenging interconnection, and an extended length oxygen enrichment tube.

AI/ML Overview

This document is a 510(k) summary for the Ohmeda Excel Draw-Over Military Vaporizer System, seeking to establish substantial equivalence to previously cleared devices. It does not contain information about acceptance criteria or a study proving device performance against such criteria. The document focuses on regulatory classification, substantial equivalence arguments, and design modifications of the device.

Therefore, I cannot provide the requested information.

{0}------------------------------------------------

K946127

Image /page/0/Picture/2 description: The image shows the word "OHMEDA" in a bold, sans-serif font. The letters are large and black, and they appear to be slightly distressed or aged. The word is centered in the image, and the background is white.

THE BOC GROUP

Ohmeda Inc Ohmeda Drive PO Box 7550 Madison WI 53707-7550 608 221 1551

0

October 2, 1996

Subject:510(k) Summary of Safety and Effectiveness Information for the Ohmeda Excel Draw-Over
Military Vaporizer System
Proprietary:Ohmeda Excel Draw-Over Military Vaporizer System
Common:Anesthesia Machine
Classification:Primary:Gas Machine for Anesthesia Class II 21CFR868.5160 - 73BSZ
Secondary:Anesthetic Vaporizer (non heated) Class II 21CFR868.5880 - 73CAD
Breathing Gas Mixer Class II 21CFR868.5330 - 73BZR
Uncompensated Thorpe Tube FlowmeterClass II 21CFR868.2320 - 73CCN

OCT - 8 1996

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.

The Ohmeda Excel Draw-Over Military Vaporizer System is substantially equivalent to the following currently marketed Ohmeda devices:

  • Ohmeda "B' Series Gas Delivery System (currently known as the Ohmeda Excel Anesthesia System) 1. which was the subject of a cleared 510(k) premarket notification on 10/30/86 with FDA log number K863207.
  • Ohmeda PAC Portable Anesthesia Machine that was the subject of a cleared 510(k) premarket notification 2. on 7/1/85 with FDA log number K850648A.
  • Ohmeda Universal PAC Portable Anesthesia Machine that was the subject of a cleared 510(k) premarket లు notification on 9/7/90 with FDA log number K901117A.

The Ohmeda Excel Draw-Over Military Vaporizer System is specifically designed for use with the Ohmeda Single-Agent Draw-Over Vaporizer. It consists of an Excel 210 system with an attached draw-over vaporizer. The purpose is to provide a platform for the delivery of inhalation anesthetic agents using the draw-over vaporizer and supplemental gas supply.

The Ohmeda Excel Draw-Over Military Vaporizer System is a modification of the Ohmeda Excel 210 that has been made to facilitate the Ohmeda Single-Agent Draw-Over Vaporizer and the militaries needs. The system consists of an Ohmeda Single Agent Draw Over Vaporizer and a kit that allows the vaporizer to be used with an Ohmeda Excel system. The kit consists of mounting attachments, O2 flush cover, monitoring ports, scavenging interconnection, and an extended length oxygen enrichment tube.

The Ohmeda Excel Draw-Over Military Vaporizer System was designed to comply with the applicable portions of the following voluntary standards:

    1. ASTM F1161-1988 - "Standard Specification for Minimum Performance and Safety Requirements for Components and Systems of Continuous Flow Anesthesia Machines".
    1. MIL-STD-14728 - "Human Engineering Design Criteria for Military Systems, Equipment and Facilities".

Standards agencies, such as U.L. and E.T.L., do not provide specific certification services for this type of device.

The Ohmeda Excel Draw-Over Military Vaporizer System and the currently marketed devices are substantially equivalent in technological characteristics, theory of operation, and use of accessories.

§ 868.5160 Gas machine for anesthesia or analgesia.

(a)
Gas machine for anesthesia —(1)Identification. A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply.(2)
Classification. Class II (performance standards).(b)
Gas machine for analgesia —(1)Identification. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).(2)
Classification. Class II (performance standards).