HEMAGEN CMV IGG KIT
K944438 · Hemagen Diagnostics, Inc. · LFZ · Feb 20, 1997 · Microbiology
Device Facts
| Record ID | K944438 |
| Device Name | HEMAGEN CMV IGG KIT |
| Applicant | Hemagen Diagnostics, Inc. |
| Product Code | LFZ · Microbiology |
| Decision Date | Feb 20, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.3175 |
| Device Class | Class 2 |
Intended Use
This enzyme-linked immunosorbent assay (ELISA) is intended to determine an individual’s serologic status with respect to IgG antibodies to CMV. When the assay is used in the qualitative mode, a reactive result may indicate current or past infection with CMV. When used in the semi-quantitative mode, this test can detect significant antibody rises associated with seroconversion, reinfection, or reactivation of latent disease. This product is not FDA-cleared for use in screening blood or plasma donors. The performance of this assay has not been established for neonates and pregnant women. Results from Immunocompromised patients should be interpreted with caution.
Device Story
Hemagen CMV IgG Kit is an ELISA for qualitative or semi-quantitative detection of IgG antibodies to CMV in human serum. Purified CMV antigens are immobilized on microwell plates. Patient serum is incubated; specific antibodies bind to antigens. Horseradish peroxidase-conjugated secondary antibody binds to patient IgG; TMB substrate added; color change measured via EIA plate reader. Optical density (OD) is used to determine antibody presence. Standard curve generated using low, medium, and high calibrators (AU/mL). Cutoff value determined via ROC analysis. Used in clinical laboratories to assess serologic status, seroconversion, reinfection, or reactivation. Results assist clinicians in diagnosing CMV infection status.
Clinical Evidence
Bench testing only. Comparison study (N=252) vs. predicate showed 100% relative sensitivity (98.0-100% CI) and 98.4% relative specificity (91.7-99.7% CI). Alternate site evaluation (9 samples, 3 sites) showed 100% agreement. CDC CMV/HSV panel (N=100) showed 100% concordance with expected results. Interference testing (lipemic/hemolytic samples) showed no significant effect. Prozone/Hook effect testing confirmed high positive results for high-titer sera. Cross-reactivity testing with rubella, varicella-zoster, and herpes simplex showed no impact on specificity.
Technological Characteristics
ELISA-based immunoassay. Purified CMV antigens immobilized on microwell plates. Horseradish peroxidase-conjugated secondary antibody; TMB substrate. Optical density measurement via EIA plate reader. Standardized to characterized material (AU/mL). Cutoff determined by ROC analysis. Equivocal zone +/- 10%.
Indications for Use
Indicated for individuals requiring serologic status determination for IgG antibodies to cytomegalovirus (CMV). Not for blood/plasma donor screening. Performance not established for neonates or pregnant women; interpret with caution for immunocompromised patients.
Regulatory Classification
Identification
Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.
Predicate Devices
- Gull Laboratories, Inc. CMV IgG ELISA Test (Product No. CME 100)
Related Devices
- K955360 — INCSTAR CYTOMEGALOVIRUS IGG FAST ELISA ASSAY · Incstar Corp. · Jul 22, 1996
- K955361 — INCSTAR CYTOMEGALOVIRUS IGG ELISA ASSAY · Incstar Corp. · Jul 22, 1996
- K981163 — DIAMEDIX IS-CMV IGG TEST SYSTEM · Diamedix Corp. · Sep 21, 1998
- K971335 — QUANTA LITE CMV IGG ELISA · Inova Diagnostics, Inc. · Dec 1, 1997
- K131605 — ELECSYS PRECICONTROL CMV IGG, ELECSYS CMV IGG IMMUNOASSAY · Roche Diagnostics · Feb 28, 2014
Submission Summary (Full Text)
{0}
FEB-13-97 13:33 FROM:HEMAGEN DIAGNOSTICS
ID:6178903748
PAGE 2/9
K944438
510(k) Summary
FEB 20 1997
1 Submitter's Name/Contact Person
Joseph M. Califano, Regulatory Affairs Manager
Address
Hemagen Diagnostics, Inc.
34-40 Bear Hill Road
Waltham, MA, 02154
Phone: (617) 890-3766
Fax: (617) 890-3748
e-mail: jcalifano@hemagen.com
Date Prepared
9 September 1994
Date Amended
13 Feb 1997
2 Name of Device
Anti-cytomegalovirus test kit
Proprietary name:
Hemagen CMV IgG Kit
3 Comparative Device
Gull Laboratories, Inc. CMV IgG ELISA Test
Product No. CME 100 (96 determinations)
Page 1
{1}
FEB-13-97 13:34 FROM:HEMAGEN DIAGNOSTICS
ID:6178903748
PAGE 3/9
## 4 Description of Device
An enzyme-linked immunosorbent assay (ELISA) designed for the qualitative or semi-quantitative detection of circulating IgG antibodies to cytomegalovirus in human serum.
The ELISA methodology is commonly used for serum antibody evaluations. Purified antigens from Cytomegalovirus have been attached to the inner surfaces of the microwell plate. During the initial incubation step, antibodies in patient serum bind specifically to the immobilized antigen and remain in place after a wash step.
A second antibody, which is conjugated to the enzyme horseradish peroxidase, is used to recognize the gamma-chain region of the bound anti-cytomegalovirus antibodies. In the wells where the second antibody remains bound, the enzyme catalyzes a color change in the substrate, tetramethylbenzidine (TMB). After the reaction is stopped, the color is read in an EIA plate reader.
## 5 Intended Use of Device
This enzyme-linked immunosorbent assay (ELISA) is intended to determine an individual’s serologic status with respect to IgG antibodies to CMV. When the assay is used in the qualitative mode, a reactive result may indicate current or past infection with CMV. When used in the semi-quantitative mode, this test can detect significant antibody rises associated with seroconversion, reinfection, or reactivation of latent disease. This product is not FDA-cleared for use in screening blood or plasma donors. The performance of this assay has not been established for neonates and pregnant women. Results from Immunocompromised patients should be interpreted with caution.
Page 2
{2}
FEB-13-97 13:34 FROM:HEMAGEN DIAGNOSTICS
PAGE 4/9
## 6.(A) Technological Characteristics
The Hemagen CMV IgG Kit and the Gull Laboratories CMV IgG ELISA Test are both based on similar technologies: they are both enzyme-linked immunosorbent assays. The proposed device and the predicate device utilize optical density as a measure of antibody presence, with an established cutoff point between a positive and a negative reaction.
The cutoff for the proposed device is based upon the comparative performance with the predicate device. The optimal cutoff value was selected utilizing receiver operating characteristic (ROC) methods. The cutoff is subject to an equivocal zone of +/- 10 %.
The low calibrator supplied with the proposed device is set at the cutoff activity level. The activity of patient specimens is determined by using the standard curve to find the corresponding activity level for the optical density measured.
The proposed device is supplied with three calibrators (low, medium, and high) that have been standardized to a characterized material. A standard curve is generated by plotting the optical densities obtained for each of the calibrators as a function of the calibrator activity level, AU/mL. The activity of patient specimens is determined directly from this standard curve.
Page 3
{3}
FEB-13-97 13:34 FROM:HEMAGEN DIAGNOSTICS
ID:6179993745
PAGE 5/9
## 6.(B) Performance Data
### I. Comparison Testing
The Hemagen kit and the comparative device were used to test serum specimens from normal blood donors. A total of 252 samples were evaluated. The comparison data for these specimens are summarized in Table 1:
Table 1: Summary of all specimens, N=252
| PROPOSED | COMPARATIVE DEVICE | | | |
| --- | --- | --- | --- | --- |
| | POS | NEG | IND | TOTAL |
| POS | 188 | 1 | 0 | 189 |
| NEG | 0 | 63 | 0 | 63 |
| IND | 0 | 0 | 0 | 0 |
| Total | 188 | 64 | 0 | 252 |
The relative sensitivity was found to be 100 % (189/188). 0.86 Confidence interval of 98.0 to 100 %
The relative specificity was found to be 98.4 % (63/64). 0.86 Confidence interval of 91.7 to 99.7 %
### II. Alternate Site Evaluations
A panel consisting of nine "blind" serum samples was evaluated by three independent laboratories following a formal protocol:
i. Four CMV IgG negatives
ii. Five CMV IgG positives
The samples were evaluated in parallel with both the predicate and proposed devices in triplicate on three different days at each laboratory.
#### Agreement between sites
All 3 sites reported 100 % agreement for all 9 of the samples with both the proposed and predicate devices.
### III. Interfering Substances
Lipemic and hemolytic samples were evaluated with the Hemagen CMV IgG Kit following NCCLS Proposed Guideline, "Interference Testing in Clinical Chemistry" Document EP7-P ISSN 0273-3099. Samples with hemoglobin concentrations of ≤ 500 mg/dL and lipid concentrations of ≤ 3,000 mg/dL did not have any significant effect on the assay results.
Page 4
{4}
FEB-13-97 13:35 FROM:HEMAGEN DIAGNOSTICS
ID:6178903748
PAGE 6/9
## IV. Prozone and "Hook Effect"
The Hemagen CMV IgG Kit was used to assay a high titered serum sample to determine if the kit would return unexpectedly low values. The results of this evaluation indicate that the kit gives appropriately high positive results with high titered sera.
## V. Cross reactivity
Five samples which were positive for cytomegalovirus antibodies and five samples which were negative for cytomegalovirus antibodies with another commercially available assay were evaluated for the presence of other viral antibodies. IgG antibodies to rubella virus, varicella-zoster virus, and herpes simplex virus did not affect the specificity of the Hemagen CMV IgG Kit.
## VI. Evaluation of the CDC CMV/HSV serum panel
The CMV/HSV Evaluation Panel consists of 100 (50 pairs) blind coded serum samples from clinically evaluated patients. All of the 100 samples were evaluated and characterized with the Hemagen CMV IgG Kit in accordance with the draft package insert. The results were submitted to CDC for scoring. The CDC reported that the Hemagen CMV IgG Kit had correctly characterized 66 of 66 samples as "CMV positives," and 34 of 34 samples as negatives.
Page 5
{5}
FEB-13-97 13:35 FROM:HEMAGEN DIAGNOSTICS
ID:6178903748
PAGE 7/9
## VII. Semi-Quantitation Evaluations
### A. Paired sera precision studies
Four fold dilutions of CMV reactive serum samples were utilized to evaluate within-run and between-run precision of the Hemagen CMV IgG Kit.
#### Intra-assay precision
Four positive samples were selected. A 1:4 dilution of each sample was prepared. Each neat and 1:4 Dilution was assayed 10 consecutive times for each of the five samples.
| | Dilution | Mean AU/mL | Std. Dev | % CV¹ | AU Ratio |
| --- | --- | --- | --- | --- | --- |
| Sample 1 | Neat | 175 | 12.0 | 6.9 | |
| Sample 1 | 1:4 | 45 | 4.6 | 10.4 | 3.90 |
| Sample 2 | Neat | 147 | 5.6 | 3.8 | |
| Sample 2 | 1:4 | 42 | 1.9 | 4.6 | 3.50 |
| Sample 3 | Neat | 129 | 6.8 | 5.3 | |
| Sample 3 | 1:4 | 30 | 0.9 | 2.9 | 4.30 |
| Sample 4 | Neat | 118 | 5.2 | 4.4 | |
| Sample 4 | 1:4 | 30 | 2.1 | 7.0 | 3.93 |
#### Inter-assay precision
Five positive samples were selected. A 1:4 dilution of each sample was prepared. Each neat and 1:4 dilution was assayed twice a day for 5 different days.
| | Dilution | Mean AU/mL | Std. Dev | % CV¹ | AU Ratio |
| --- | --- | --- | --- | --- | --- |
| Sample 1 | Neat | 156 | 10.8 | 6.9 | |
| Sample 1 | 1:4 | 39 | 4.1 | 10.6 | 4.00 |
| Sample 2 | Neat | 133 | 15.4 | 11.6 | |
| Sample 2 | 1:4 | 37 | 3.9 | 10.5 | 3.59 |
| Sample 3 | Neat | 144 | 15.1 | 10.5 | |
| Sample 3 | 1:4 | 38 | 5.0 | 13.0 | 3.79 |
| Sample 4 | Neat | 122 | 6.7 | 5.5 | |
| Sample 4 | 1:4 | 29 | 3.7 | 12.8 | 4.21 |
| Sample 5 | Neat | 133 | 13.9 | 10.5 | |
| Sample 5 | 1:4 | 33 | 4.3 | 13.1 | 4.03 |
1. These values include the well-to-well variation inherent in the plastic strips, which can range up to 5 %, according to the plastic strip supplier.
Page 6
{6}
FEB-13-97 13:35 FROM:HEMAGEN DIAGNOSTICS
ID:6178903745
PAGE 8/9
## B. AU Ratio Establishment
A total of 43 high titered serum samples were selected. Multiple 2 and 4 fold dilutions were performed. Based on the results, it was determined that AU ratios of > 3.2 are indicative of a significant rise in antibody titer.
## 7 Conclusions
The results of the comparative studies and the fact that both the Hemagen kit and the predicate kit utilize an enzyme-linked Immunosorbent assay method support the claim that the Hemagen kit is a safe and effective *in vitro* diagnostic test and is substantially equivalent to the comparative device.
Page 7