(59 days)
AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning of endovascular interventions.
AV Vascular is a post-processing software application intended for visualization, assessment, and quantification of vessels in computed tomography angiography (CTA) and magnetic resonance angiography (MRA) data with a unified workflow for both modalities.
AV Vascular includes the following functions:
-
Advanced visualization: the application provides all relevant views and interactions for CTA and MRA image review: 2D slides, MIP, MPR, curved MPR (cMPR), stretched MPR (sMPR), path-aligned views (cross-sectional and longitudinal MPRs), 3D volume rendering (VR).
-
Vessel segmentation: automatic bone removal and vessel segmentation for head/neck and body CTA data, automatic vessel centerline, lumen and outer wall extraction and labeling for the main branches of the vascular anatomy in head/neck and body CTA data, semi-automatic and manual creation of vessel centerline and lumen for CTA and MRA data, interactive two-point vessel centerline extraction and single-point centerline extension.
-
Vessel inspection: enable inspection of an entire vessel using the cMPR or sMPR views as well as inspection of a vessel locally using vessel-aligned views (cross-sectional and longitudinal MPRs) by selecting a position along a vessel of interest.
-
Measurements: ability to create and save measurements of vessel and lumen inner and outer diameters and area, as well as vessel length and angle measurements.
-
Measurements and tools that specifically support pre-procedural planning: manual and automatic ring marker placement for specific anatomical locations, length measurements of the longest and shortest curve along the aortic lumen contour, angle measurements of aortic branches in clock position style, saving viewing angles in C-arm notation, and configurable templated
-
Saving and export: saving and export of batch series and customizable reports.
This summarization is based on the provided 510(k) clearance letter for Philips Medical Systems' AV Vascular device.
Acceptance Criteria and Device Performance for Aorto-iliac Outer Wall Segmentation
| Metrics | Acceptance Criteria | Reported Device Performance (Mean with 98.75% confidence intervals) |
|---|---|---|
| 3D Dice Similarity Coefficient (DSC) | > 0.9 | 0.96 (0.96, 0.97) |
| 2D Dice Similarity Coefficient (DSC) | > 0.9 | 0.96 (0.95, 0.96) |
| Mean Surface Distance (MSD) | < 1.0 mm | 0.57 mm (0.485, 0.68) |
| Hausdorff Distance (HD) | < 3.0 mm | 1.68 mm (1.23, 2.08) |
| ∆Dmin (difference in minimum diameter) | > 95% |∆Dmin| < 5 mm | 98.8% (98.3-99.2%) |
| ∆Dmax (difference in maximum diameter) | > 95% |∆Dmax| < 5 mm | 98.5% (97.9-98.9%) |
The reported device performance for all primary and secondary metrics meets the predefined acceptance criteria.
Study Details for Aorto-iliac Outer Wall Segmentation Validation
-
Sample Size used for the Test Set and Data Provenance:
- Sample Size: 80 patients
- Data Provenance: Retrospectively collected from 7 clinical sites in the US, 3 European hospitals, and one hospital in Asia.
- Independence from Training Data: All performance testing datasets were acquired from clinical sites distinct from those which provided the algorithm training data. The algorithm developers had no access to the testing data, ensuring complete independence.
- Patient Characteristics: At least 80% of patients had thoracic and/or abdominal aortic diseases and/or iliac artery diseases (e.g., thoracic/abdominal aortic aneurysm, ectasia, dissection, and stenosis). At least 20% had been treated with stents.
- Demographics:
- Geographics: North America: 58 (72.5%), Europe: 3 (3.75%), Asia: 19 (23.75%)
- Sex: Male: 59 (73.75%), Female: 21 (26.25%)
- Age (years): 21-50: 2 (2.50%), 51-70: 31 (38.75%), >71: 45 (56.25%), Not available: 2 (2.5%)
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Number of Experts: Three
- Qualifications: US-board certified radiologists.
-
Adjudication Method for the Test Set:
- The three US-board certified radiologists independently performed manual contouring of the outer wall along the aorta and iliac arteries on cross-sectional planes for each CT angiographic image.
- After quality control, these three aortic and iliac arterial outer wall contours were averaged to serve as the reference standard contour. This can be considered a form of consensus/averaging after independent readings.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- The provided document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to measure human reader improvement with AI assistance. The study focused on the standalone performance of the AI algorithm compared to an expert-derived ground truth.
-
Standalone (Algorithm Only Without Human-in-the-Loop Performance):
- Yes, the performance data provided specifically describes the standalone performance of the AI-based algorithm for aorto-iliac outer wall segmentation. The algorithm's output was compared directly against the reference standard without human intervention in the segmentation process.
-
Type of Ground Truth Used:
- Expert Consensus/Averaging: The ground truth was established by averaging the independent manual contouring performed by three US-board certified radiologists.
-
Sample Size for the Training Set:
- The document states that the testing data were independent of the training data and that developers had no access to the testing data. However, the exact sample size for the training set is not specified in the provided text.
-
How the Ground Truth for the Training Set Was Established:
- The document implies that training data were used, but it does not describe how the ground truth for the training set was established. It only ensures that the testing data did not come from the same clinical sites as the training data and that algorithm developers had no access to the testing data.
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
Philips Medical Systems Nederland B.V.
Arbel Shezaf
Regulatory Affairs Manager
Veenpluis 6
Best, 5684 PC
Netherlands
Re: K253735
Trade/Device Name: AV Vascular
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: November 24, 2025
Received: November 24, 2025
Dear Arbel Shezaf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
January 22, 2026
Page 2
Philips Medical Systems Nederland B.V.
Arbel Shezaf
Regulatory Affairs Manager
Veenpluis 6
Best, 5684 PC
Netherlands
January 22, 2026
Re: K253735
Trade/Device Name: AV Vascular
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: November 24, 2025
Received: November 24, 2025
Dear Arbel Shezaf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Imaging Software Team
Assistant Director
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K253735
Please provide the device trade name(s).
AV Vascular
Please provide your Indications for Use below.
AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning of endovascular interventions.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Page 5
510(k) SUMMARY
Philips Medical Systems' AV Vascular
| Company's Name and Address | Philips Medical Systems Nederland B.V.Veenpluis 65684 PC BestThe Netherlands |
|---|---|
| Contact Person | Arbel ShezafRegulatory Affairs ManagerMATAM building 34, Haifa, 3100202, Israel+972544487597arbel.shezaf@philips.com |
| Date | January 22, 2026 |
| Device Trade name | AV Vascular |
| Classification Name | Automated radiological image processing software |
| Primary Product Code | QIH |
| Secondary Product Code | LLZ |
| Classification | 21 CFR 892.2050 |
| Primary Predicate Device | Pie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) |
| Secondary Predicate Device | Philips Medical Systems' The Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) |
| Reference Device | Philips Medical Systems' Spectral CT Applications (K150665) |
Device Description
AV Vascular is a post-processing software application intended for visualization, assessment, and quantification of vessels in computed tomography angiography (CTA) and magnetic resonance angiography (MRA) data with a unified workflow for both modalities.
AV Vascular includes the following functions:
-
Advanced visualization: the application provides all relevant views and interactions for CTA and MRA image review: 2D slides, MIP, MPR, curved MPR (cMPR), stretched MPR (sMPR), path-aligned views (cross-sectional and longitudinal MPRs), 3D volume rendering (VR).
-
Vessel segmentation: automatic bone removal and vessel segmentation for head/neck and body CTA data, automatic vessel centerline, lumen and outer wall extraction and labeling for the main branches of the vascular anatomy in head/neck and body CTA data, semi-automatic and manual creation of vessel centerline and lumen for CTA and MRA data, interactive two-point vessel centerline extraction and single-point centerline extension.
-
Vessel inspection: enable inspection of an entire vessel using the cMPR or sMPR views as well as inspection of a vessel locally using vessel-aligned views (cross-sectional and longitudinal MPRs) by selecting a position along a vessel of interest.
-
Measurements: ability to create and save measurements of vessel and lumen inner and outer diameters and area, as well as vessel length and angle measurements.
-
Measurements and tools that specifically support pre-procedural planning: manual and automatic ring marker placement for specific anatomical locations, length measurements of the longest and shortest curve along the aortic lumen contour, angle measurements of aortic branches in clock position style, saving viewing angles in C-arm notation, and configurable templated
Page 6
measurements.
- Saving and export: saving and export of batch series and customizable reports.
Indications for Use
AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning of endovascular interventions.
Verification and Validation
Software verification and validation activities were performed to verify that the software meets the product requirements.
Verification and Validation
Verification was performed according to the verification plan. Product requirement specifications were tested and found to meet the requirements.
Validation
Validation was performed according to the validation plan. User requirement specifications were tested and found to meet the requirements. The validation results provide evidence that the product meets its intended use and user requirements.
Performance Data
Performance testing was conducted for subclavian artery extraction on body cases (model-based algorithm), aorto-iliac wall segmentation (AI-based algorithm) and ring marker placement (model-based algorithm).
The datasets represent the targeted US patient population in terms of key demographics, clinical and technical characteristics.
Subclavian Artery Centerline Extraction
CT angiography scans of the body were collected for testing. The automatically extracted subclavian artery centerlines were compared with a reference standard.
Ring Marker Placement
CT angiography scans of the body were collected for testing, with the majority coming from patients with aortic or iliac aneurysms. Automatically placed ring markers at the ostia of the renal and superior mesenteric arteries were evaluated by expert vascular interventionalists for their clinical acceptability without any adjustment.
Aorto-iliac Outer Wall Segmentation
To validate the performance of the aorto-iliac wall segmentation, body CT angiography scans were collected from 80 patients.
The aorto-iliac outer wall contour automatically generated by the AI-based algorithm in the subject device was compared with a reference standard in terms of Dice similarity coefficient (DSC), mean surface distance (MSD), and Hausdorff distance (HD). The measurement accuracy (minimum and maximum diameter, Dmin and Dmax) was also evaluated.
- Testing Data, Demographics, & Independence from Training Data
The demographic information of the 80 patients included for performance testing is:
| Demographics | Number of patients (percentage) |
|---|---|
| Geographics | North America: 58 (72.5%)Europe: 3 (3.75%)Asia: 19 (23.75%)None of these testing data were from the same clinical sites which provided the training data. |
| Sex | Male: 59 (73.75%)Female: 21 (26.25%) |
| Age (years) | 21-50: 2 (2.50%)51-70: 31 (38.75%) |
Page 7
| >71: 45 (56.25%)Not available: 2 (2.5%) |
|---|
The performance testing data were collected retrospectively from 7 clinical sites in the US, 3 European hospitals, and one hospital in Asia. All performance testing datasets were acquired from clinical sites distinct from those which provided the algorithm training data, and the algorithm developers had no access to the testing data, ensuring complete independence between training and testing data. At least 80% of the included patients had thoracic and/or abdominal aortic diseases and/or iliac artery diseases (e.g., thoracic/abdominal aortic aneurysm, ectasia, dissection, and stenosis). At least 20% of patients had been treated with stents. The composed testing data were considered representative of the intended patient population in the US.
- Reference Standard (i.e., Truthing Process)
For each of the CT angiographic images, three US-board certified radiologists (i.e., truthers) independently performed manual contouring of the outer wall along the aorta and iliac arteries on cross-sectional planes. The order of the cases was randomized for each truther to reduce bias. After quality control, these three aortic and iliac arterial outer wall contours were averaged as the reference standard contour.
- Performance & Acceptance Criteria
For primary performance metrics (2D and 3D DSC, MSD, and HD) the performance statistics and acceptance criteria are:
| Metrics | Acceptance criteria | Mean (98.75% confidence intervals) |
|---|---|---|
| DSC | 3D DSC > 0.9 | 0.96 (0.96, 0.97) |
| 2D DSC > 0.9 | 0.96 (0.95, 0.96) | |
| MSD [mm] | MSD < 1.0 mm | 0.57 (0.485, 0.68) |
| HD [mm] | HD < 3.0 mm | 1.68 (1.23, 2.08) |
For secondary performance metrics (Dmin and Dmax), the performance statistics and acceptance criteria are:
| Metrics | Acceptance criteria | Proportion (95% confidence intervals) |
|---|---|---|
| ∆Dmin | >95% |∆Dmin|<5 mm | 98.8% (98.3-99.2%) |
| ∆Dmax | >95% |∆Dmax|<5 mm | 98.5% (97.9-98.9%) |
- Clinical Subgroups & Confounders
The performance of automatic aorto-iliac outer wall segmentation was evaluated internally using independent testing CT angiographic data acquired with multi-vendor scanners (GE Healthcare, Siemens, Philips, Toshiba/Canon) from patients of which at least 80% had aortic and/or iliac artery aneurysm, dissection, stents, calcification, stenosis, and/or plaque. Primary performance metrics (DSC, MSD, and HD) were analyzed across subgroups (defined by sex, age group, data source, vessel, pathology, reconstruction method, slice thickness, and manufacturer). The algorithm's performance was consistent across multiple subgroups and comparable to the inter-truther variability. The outer wall segmentation accuracy, however, might be reduced in cases of giant aneurysms (diameter > 10 cm), which is rare in the intended patient population.
Page 8
Therefore, the performance of all three algorithms is state of the art and on par with the expectation.
Substantial Equivalence
AV Vascular is as safe and effective as its predicate device. Both devices have similar intended use and indications for use, technological characteristics, and principles of operation. The minor technological differences between AV Vascular and its predicate device raise no new issues of safety or effectiveness.
Performance data demonstrate that Philips Medical Systems' AV Vascular is safe and effective. Thus, AV Vascular is substantially equivalent to Pie Medical Imaging's 3mensio. Table 1 below summarizes the substantive feature/technological similarities and differences between the subject and predicate devices.
Table 1. Substantial Equivalence
| Comparison Feature | Subject DevicePhilips Medical Systems' AV Vascular | Predicate DevicePie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) | Secondary Predicate DevicePhilips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Reference DevicePhilips Medical Systems' Spectral CT Applications (K150665) | Comparison between the subject and predicate devices (identical/different) |
|---|---|---|---|---|---|
| Device Class | Class II | Class II | Class II | Class II | Identical |
| Classification Panel | Radiology | Radiology | Radiology | Radiology | Identical |
| Product Code | QIH, LLZ | LLZ | JAK | JAK, LLZ | Similar to primary predicate device |
| Regulation Description | Medical image management and processing system | Medical image management and processing system | Computed tomography x-ray system | Computed tomography x-ray systemMedical image management and processing system | Identical to primary predicate device |
| Regulation Number | 892.2050 | 892.2050 | 892.1750 | 892.1750892.2050 | Identical to primary predicate device |
| Indications for Use | AV Vascular is indicated to assist users in the visualization, assessment and quantification of vascular anatomy | 3mensio Workstation is a standalone software for medical image analysis intended for advanced | The Multimodality Advanced Vessel Analysis (MM AVA) application is intended for visualization, assessment and | The Philips Spectral CT Applications support viewing and analysis of images at energies selected | Different |
Page 9
| Comparison Feature | Subject DevicePhilips Medical Systems' AV Vascular | Predicate DevicePie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) | Secondary Predicate DevicePhilips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Reference DevicePhilips Medical Systems' Spectral CT Applications (K150665) | Comparison between the subject and predicate devices (identical/different) |
|---|---|---|---|---|---|
| on CTA and/or MRA datasets, in order to assess patients with suspected or diagnosed vascular pathology and to assist with pre-procedural planning for endovascular interventions. | visualization and quantitative analysis for diagnostic and/or for assistance during treatment in the field of cardiology or radiology by means of enabling visualization and measurement of structures of the heart and vessels for:• Pre-operational planning and sizing for cardiovascular interventions and surgery• Postoperative evaluation• Support of clinical diagnosis by quantifying dimensions in coronary arteries• Support of clinical diagnosis by quantifying calcifications (calcium scoring) in the coronary arteriesTo facilitate the above, the 3mensio Workstation provides general functionality such as: | quantification of vascular datasets. | from the available spectrum in order to provide information about the chemical composition of the body materials and/or contrast agents. The Spectral CT Applications provide for the quantification and graphical display of attenuation, material density, and effective atomic number. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.The Spectral enhanced Advanced Vessel Analysis (sAVA) application is intended to assist clinicians in viewing and evaluating CT images, for the inspection of contrast-enhanced vessels. |
Page 10
| Comparison Feature | Subject DevicePhilips Medical Systems' AV Vascular | Predicate DevicePie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) | Secondary Predicate DevicePhilips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Reference DevicePhilips Medical Systems' Spectral CT Applications (K150665) | Comparison between the subject and predicate devices (identical/different) |
|---|---|---|---|---|---|
| • Segmentation of cardiovascular structures• Automatic and manual centerline detection• Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched CMRP, Slabbing, MIP, AIP, MinIP• Measurement and annotation toolsReporting tools | The Spectral enhanced Comprehensive Cardiac Analysis (sCCA) application is intended to assist clinicians in viewing and evaluating cardiovascular CT images.The Spectral enhanced Tumor Tracking (sTT) application is intended to assist clinicians in viewing and evaluating CT images, for the inspection of tumors. |
Clinical Characteristics
| Intended body part | Head and neck, body, peripherals | Heart and vessels | Head and neck, body, peripherals | Head and neck, body, peripherals | Identical to secondary predicate device |
| Type of scans | CTA and MRA | CT Angiography | CTA and MRA | CT Angiography | Identical to secondary predicate device |
Technological features
| Spectral capabilities | Yes | No | No | Yes | Identical to reference device |
| Subclavian Artery | Automatic vessel centerline extraction of the head, neck and | Automatic vessel centerline extraction | Automatic vessel centerline | Automatic vessel centerline | Different |
Page 11
| Comparison Feature | Subject DevicePhilips Medical Systems' AV Vascular | Predicate DevicePie Medical Imaging's 3mensio Structural Heart / 3mensio Vascular (K153736) | Secondary Predicate DevicePhilips Medical Systems' Multimodality Advanced Vessel Analysis (MM AVA) application (K203216) | Reference DevicePhilips Medical Systems' Spectral CT Applications (K150665) | Comparison between the subject and predicate devices (identical/different) |
|---|---|---|---|---|---|
| Centerline Extraction | body including shoulder region | extraction of the head and neck | extraction of the head and neck | ||
| Aorto-iliac Outer Wall Segmentation | Automatic non-AI based vessel contouring of major vessels.Automatic AI-based vessel contouring specifically for the aorta and common and external iliac arteries. | No | Automatic non-AI based vessel contouring of major vessels, including the aorta and common and external iliac arteries. | Automatic non-AI based vessel contouring of major vessels, including the aorta and common and external iliac arteries. | Different |
| Review Marker tool | Review Marker tool including manual ring markers, and non-AI based automatic ring markers for the ostia of the SMA, LRA and RRA | Review Marker tool including manual ring markers | Review Marker tool | Review Marker tool | Different |
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).