K Number
K253444

Validate with FDA (Live)

Manufacturer
Date Cleared
2026-03-18

(167 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EUROPA™ Posterior Cervical Fusion Navigated Instruments are intended to be used in the preparation and placement of the EUROPA™ Posterior Cervical Fusion screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the stereotactic navigation system Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Description

The EUROPA™ Posterior Cervical Fusion Navigated Instruments are intended to be used with the EUROPA™ Posterior Cervical Fusion System screws. The EUROPA™ Posterior Cervical Fusion Navigated Instruments consist of non-sterile, re-usable instruments including probes, drivers, awls, drills, and taps that can be operated manually. These instruments are intended to be used with the Medtronic StealthStation® System and are manufactured from Stainless Steel per ASTM F899.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - EUROPA™ Posterior Cervical Fusion Navigated Instruments

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.03

March 18, 2026

MiRus, LLC
Anuradha Nagulapati
Senior Regulatory Affairs Engineer
1755 W. Oak Pkwy.
Suite 100
Marietta, Georgia 30062

Re: K253444
Trade/Device Name: EUROPA™ Posterior Cervical Fusion Navigated Instruments
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: February 16, 2026
Received: February 17, 2026

Dear Ms. Nagulapati:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K253444 - Anuradha Nagulapati
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

Page 3

K253444 - Anuradha Nagulapati
Page 3

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K253444

Device Name
EUROPA™ Posterior Cervical Fusion Navigated Instruments

Indications for Use (Describe)

The EUROPA™ Posterior Cervical Fusion Navigated Instruments are intended to be used in the preparation and placement of the EUROPA™ Posterior Cervical Fusion screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the stereotactic navigation system Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k)#: K253444

510(k) Summary

Prepared on: 2026-03-11

Contact Details

21 CFR 807.92(a)(1)

| Applicant Name | MiRus, LLC |
| Applicant Address | 1755 W. Oak Parkway Suite 100 Marietta GA 30062 United States |
| Applicant Contact Telephone | 470-428-8684 |
| Applicant Contact | Ms. Anuradha Nagulapati |
| Applicant Contact Email | anagulapati@mirusmed.com |

Device Name

21 CFR 807.92(a)(2)

| Device Trade Name | EUROPA™ Posterior Cervical Fusion Navigated Instruments |
| Common Name | Stereotaxic instrument |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Regulation Number | 882.4560 |
| Product Code(s) | OLO |

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K140454Navigated CD HORIZON SOLERA Screwdrivers and TapsOLO
K143628Navigated VERTEX SELECT InstrumentsOLO

Device Description Summary

21 CFR 807.92(a)(4)

The EUROPA™ Posterior Cervical Fusion Navigated Instruments are intended to be used with the EUROPA™ Posterior Cervical Fusion System screws. The EUROPA™ Posterior Cervical Fusion Navigated Instruments consist of non-sterile, re-usable instruments including probes, drivers, awls, drills, and taps that can be operated manually. These instruments are intended to be used with the Medtronic StealthStation® System and are manufactured from Stainless Steel per ASTM F899.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The EUROPA™ Posterior Cervical Fusion Navigated Instruments are intended to be used in the preparation and placement of the EUROPA™ Posterior Cervical Fusion screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the stereotactic navigation system Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Page 1 of 2

Page 6

Indications for Use Comparison

21 CFR 807.92(a)(5)

The Indications for Use of the subject device are substantially equivalent to those of the predicate devices.

Technological Comparison

21 CFR 807.92(a)(6)

The EUROPA™ Posterior Cervical Fusion Navigated Instruments have the same intended use, indications for use, and technological characteristics as the predicate systems, including the same design features, geometries, sizes, and materials. Performance data demonstrate that the EUROPA™ Posterior Cervical Fusion Navigated Instruments are substantially equivalent to the predicate systems.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

The EUROPA™ Posterior Cervical Fusion Navigated Instruments have been evaluated through comparative dimensional analyses showing equivalent critical geometry of the instruments and their array connection interfaces to the corresponding Medtronic StealthStation instrument counterparts.

The results of the engineering analyses show that the subject devices are substantially equivalent to the cleared predicates.

The EUROPA™ Posterior Cervical Fusion Navigated Instruments have the same intended use, indications for use, and technological characteristics as the predicate systems, including the same design features, geometries, sizes, and materials. Performance data demonstrate that the EUROPA™ Posterior Cervical Fusion Navigated Instruments are substantially equivalent to the predicate systems.

Page 2 of 2

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).