(167 days)
The Stealth AXiS™ Surgical System, with the Stealth AXiS™ ENT clinical application, is intended for precise positioning of surgical instruments and as an aid for locating anatomical structures in open, minimally invasive, and percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
This can include, but is not limited to, the following procedures:
- Functional endoscopic sinus surgery (FESS)
- Endoscopic skull base procedures
- Lateral skull base procedures
The Stealth AXiS ENT Clinical Application works in conjunction with the Stealth AXiS Surgical System, which consists of clinical software, surgical instruments, a referencing system, and platform/computer hardware. The Stealth AXiS™ ENT Clinical Application helps guide surgeons during ENT procedures such as functional endoscopic sinus surgery (FESS), endoscopic skull base procedures, and lateral skull base procedures. The system tracks the position of instruments in relation to the surgical anatomy, known as localization, and then identifies this position on preoperative or intraoperative images of a patient.
N/A
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.04
Silver Spring, MD 20993
www.fda.gov
March 16, 2026
Medtronic Navigation, Inc.
Maulik Shah
Senior Regulatory Affairs Specialist
200 Medtronic Dr.
Lafayette, Colorado 80026
Re: K253395
Trade/Device Name: Stealth AXiS™ ENT clinical application
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: PGW
Dated: November 18, 2025
Received: November 18, 2025
Dear Maulik Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K253395 - Maulik Shah Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K253395 - Maulik Shah Page 3
Sincerely,
SHUCHEN PENG -S
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K253395
Please provide the device trade name(s). Stealth AXiS™ ENT Clinical Application (9736551)
Please provide your Indications for Use below.
The Stealth AXiS™ Surgical System, with the Stealth AXiS™ ENT clinical application, is intended for precise positioning of surgical instruments and as an aid for locating anatomical structures in open, minimally invasive, and percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
This can include, but is not limited to, the following procedures:
- Functional endoscopic sinus surgery (FESS)
- Endoscopic skull base procedures
- Lateral skull base procedures
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Page 5
Stealth AXiS™ ENT Clinical Application 510k Summary
510(k) Summary
November 18, 2025
I. Company:
Medtronic Navigation, Inc.
200 Medtronic Drive
Lafayette, CO 80026
Telephone Number: (720) 890-3160
Contact: Maulik Shah
Senior Regulatory Affairs Specialist
Telephone Number: (929) 350-7138
Email: maulik.shah2@medtronic.com
Carey Brenner (Alternate)
Principal Regulatory Affairs Specialist
Telephone Number: (720) 890-2185
Email: carey.j.brenner@medtronic.com
II. Proprietary Trade Name:
Stealth AXiS™ ENT Clinical Application
III. Common Name:
Ear, Nose, and Throat Stereotaxic Instrument
IV. Classification Name:
Stereotaxic Instrument (21 CFR 882.4560)
V. Classification:
Class II
VI. Product Code:
PGW (Ear, Nose, and Throat Stereotaxic Instrument)
VII. Predicate Devices:
The legally marketed predicate devices are identified below:
| Subject Device | Primary Predicate Device |
|---|---|
| Stealth AXiS ENT Clinical Application | StealthStation S8 ENT Software v1.3 K200723 S.E June 26, 2020 |
VIII. Device Description:
The Stealth AXiS ENT Clinical Application works in conjunction with the Stealth AXiS Surgical System, which consists of clinical software, surgical instruments, a referencing system, and platform/computer hardware. The Stealth AXiS™ ENT Clinical Application helps guide surgeons during ENT procedures such as functional endoscopic sinus surgery (FESS), endoscopic skull base procedures, and lateral skull base procedures. The system tracks the
Page 6
position of instruments in relation to the surgical anatomy, known as localization, and then identifies this position on preoperative or intraoperative images of a patient.
IX. Indications for Use:
The Stealth AXiS™ Surgical System, with the Stealth AXiS™ ENT clinical application, is intended for precise positioning of surgical instruments and as an aid for locating anatomical structures in open, minimally invasive, and percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
This can include, but is not limited to, the following procedures:
- Functional endoscopic sinus surgery (FESS)
- Endoscopic skull base procedures
- Lateral skull base procedures
X. Comparison of Technological Characteristics:
A comparison of the technological characteristics of the subject and the predicate device is provided in the table below.
Table 1. Technological Comparison of Stealth AXiS ENT Clinical Application v1.0 (Subject) and StealthStation S8 ENT Application v1.3 (Predicate)
| Technological Characteristic | Stealth AXiS ENT Application v1.0 Subject Device | StealthStation S8 ENT Application v1.3 (K200723) Predicate Device | Discussion |
|---|---|---|---|
| Product Code | PGW | PGW | Same |
| Regulation Number | 882.4560 | 882.4560 | Same |
| Classification | Class II | Class II | Same |
| Intended Use / Indications for Use | The Stealth AXiS™ Surgical System, with the Stealth AXiS™ ENT clinical application, is intended for precise positioning of surgical instruments and as an aid for locating anatomical structures in open, minimally invasive, and percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy. This can include, but is not limited to, the following procedures: | The StealthStation S8 System, with the StealthStation ENT software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy. This can include, but is not limited to, the following procedures: Functional endoscopic sinus surgery (FESS) | Equivalent The intended use/ indications for use of the predicate and subject device is the same. The predicate and subject devices share the same fundamental purpose – to assist clinicians through surgical navigation by aiding in precisely locating anatomical structures relative to the surgical instrument location. The subject device adds minimally invasive procedures. The inclusion of minimally invasive procedures does not represent a change in clinical application; it is added to reflect |
Page 7
| Functional endoscopic sinus surgery (FESS) Endoscopic skull base procedures Lateral skull base procedures | Endoscopic skull base procedures Lateral skull base procedures | the ENT procedures that are performed through less invasive approaches for procedures. The subject device operates under the same principles as the predicate device. Therefore, the subject and predicate have the same intended use | |
|---|---|---|---|
| System Accuracy Requirement | Under representative worst-case configuration, the Stealth AXiS Surgical System with Stealth AXiS ENT application, has demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in trajectory angle accuracy with a mean error ≤ 2.0 degrees | Under representative worst-case configuration, the StealthStation S8 System with StealthStation ENT Software v1.3 has demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in trajectory angle accuracy with a mean error ≤ 2.0 degrees | Identical |
| Imaging Modalities | X-Ray based MR based Nuclear Medicine based | X-Ray based MR based Nuclear Medicine based | Identical |
| View (Display) Features | 3D, 2D Anatomic Orthogonal (coronal, sagittal, axial) Trajectory 1 and 2 Target Guidance Trajectory Guidance Probes Eye Look Ahead Video Input | 3D, 2D Anatomic Orthogonal (coronal, sagittal, axial) Virtual Endoscopic Trajectory 1 and 2 Target Guidance Trajectory Guidance Probes Eye Look Ahead Video Input | Identical |
| Exam-to-Exam Registration Features | Pre-Merge (previously Identity Merge) Registration Manual Merge Registration Automatic Merge Registration | Identity Merge Registration Manual Merge Registration Automatic Merge Registration | Identical |
| Patient Registration Features | Trace registration Touch registration PointMerge® registration (referred to as Landmark registrations) | Trace registration Touch registration | Equivalent The Patient Registration feature, PointMerge, available in the predicate StealthStation S8 ENT Clinical Application v1.3, has been removed from the subject device. Removing a Patient Registration feature does not impact on the intended use or technological characteristics of the device. |
| Planning Features | Plan Entry and Target Selection Model Building Advanced Visualization | Plan Entry and Target Selection Model Building Advanced Visualization | Equivalent Note: The ENT clinical application is capable of 2D and 3D models, same as the predicate, so the feature has been renamed "Model Building" to capture all capabilities. |
Page 8
| Compatible Medtronic EM Instrumentation | Medtronic instruments tracked via Electromagnetic localization technology located within the instrument and patient trackers | Medtronic instruments tracked via Electromagnetic localization technology located within the instrument and patient trackers | Identical |
|---|---|---|---|
| Scanner Interface Technology (to imaging devices) | Network Connectivity CD, DVD, USB DICOM Import DICOM Export | Network Connectivity CD, DVD, USB DICOM Import DICOM Export | Identical |
XI. Discussion of the Non-Clinical Testing:
Testing demonstrates that the software will perform as intended according to the outlined design requirements. The following testing was conducted on the Stealth AXiS™ ENT Clinical Application to verify that the device will perform as intended meeting all the design inputs:
-
Under representative worst-case configuration, the Stealth AXiS Surgical System with Stealth AXiS ENT application, has demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in trajectory angle accuracy with a mean error ≤ 2.0 degrees.
-
Software Verification and Validation testing using a release candidate of the Stealth AXiS ENT Clinical application verifying the software requirements are met and software performs as intended.
-
Software Verification and Validation testing using a release candidate of the Stealth AXiS ENT Clinical application verifying that the software performs as intended when running on the Stealth AXiS Surgical System.
-
Usability Testing was conducted in accordance with FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices" demonstrating that the usability and human factors requirements were adequately met.
XII. Discussion of the Clinical Testing
No clinical testing was performed.
XIII. Conclusion:
The indications for use have been slightly reworded to explicitly acknowledge minimally invasive procedures. The inclusion of minimally invasive procedures in the intended use does not represent a change in the clinical application; it is added to reflect the neurosurgical procedures that are performed through less invasive approaches. All the core navigation, imaging, planning, and display features remain identical. The Stealth AXiS ENT clinical application has been shown through comparison, testing, and clinical literature to be substantially equivalent to the identified predicate device.
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).