K Number
K253318

Validate with FDA (Live)

Date Cleared
2026-01-30

(122 days)

Product Code
Regulation Number
866.3480
Age Range
0 - 97
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Clungene RSV Antigen Rapid Test is a lateral flow immunoassay intended for the qualitative detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. The test is intended for in vitro diagnostic use as an aid in the diagnosis of RSV infections.

Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or patient management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an FDA-cleared molecular assay.

Device Description

The Clungene RSV Antigen Rapid Test is a lateral flow immunoassay based on the principle of the double-antibody sandwich technique. The nasopharyngeal swab from individuals is processed by the extraction buffer in buffer tube. After the extracted sample is added into the specimen well, the respiratory syncytial virus (RSV) nucleoprotein antigen in the sample will react with anti-RSV antigen antibody conjugated with color micro particles to form an antigen-antibody labeled complex. This complex migrates on the membrane via capillary action until the test line, where it will be captured by the pre-coated anti-RSV antigen antibody. If the specimen contains RSV nucleoprotein antigen, a colored line will appear in the test line (T) region indicating a positive result. If the specimen does not contain RSV nucleoprotein antigen, a colored line will not appear in this region indicating a negative result. To serve as a procedural control, a colored line will always appear at the control line (C) region indicating that proper volume of specimen has been added and membrane wicking has occurred.

The test result should be read at 15-20 minutes after adding the extracted sample is added into the specimen well. One line in the control line region (C), and another line in the test line region (T) indicates a positive result, regardless of color intensity. When only One line is present in the control region(C) and no line is present in the test line region (T) results are negative result. If the control line(C) fails to appear, the result is invalid, and the operator should review the procedure and repeat the test with a new test cassette.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Clungene RSV Antigen Rapid Test

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.08.02
Silver Spring, MD 20993
www.fda.gov

January 30, 2026

Hangzhou Clongene Biotech Co., Ltd.
℅ Jenny Xia
LSI International, Inc.
504 East Diamond Ave., Suite H
Gaithersburg, MD 20877

Re: K253318
Trade/Device Name: Clungene RSV Antigen Rapid Test
Regulation Number: 21 CFR 866.3480
Regulation Name: Respiratory syncytial virus serological reagents
Regulatory Class: Class I
Product Code: GQG
Dated: September 29, 2025
Received: September 30, 2025

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K253318 - Jenny Xia Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K253318 - Jenny Xia Page 3

Sincerely,

JOSEPH BRIGGS -S

Joseph Briggs, Ph.D.
Deputy Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K253318

Device Name: Clungene RSV Antigen Rapid Test

Indications for Use (Describe)

The Clungene RSV Antigen Rapid Test is a lateral flow immunoassay intended for the qualitative detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. The test is intended for in vitro diagnostic use as an aid in the diagnosis of RSV infections.

Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or patient management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an FDA-cleared molecular assay.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) SUMMARY

K253318

  1. Date: January 29, 2026

  2. Submitter: Hangzhou Clongene Biotech Co., Ltd.
    No.1 Yichuang Road, Yuhang District,
    Hangzhou 311121 Zhejiang, China

  3. Contact person: Jenny Xia
    LSI International Inc.
    504 East Diamond Ave., Suite H
    Gaithersburg, MD 20877
    Telephone: 301-525-6856
    Fax: 301-916-6213
    Email: jxia@lsi-consulting.org

  4. Device Name: Clungene RSV Antigen Rapid Test

  5. Classification: Class II

Product CodeCFR #Panel
GQG21 CFR 866.3480Respiratory Syncytial Virus Serological Reagents Microbiology
  1. Predicate Devices:
    K240280, Nano-Check RSV Test

  2. Intended Use

The Clungene RSV Antigen Rapid Test is a lateral flow immunoassay intended for the qualitative detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. The test is intended for in vitro diagnostic use as an aid in the diagnosis of RSV infections.

Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or patient management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an FDA-cleared molecular assay.

  1. Device Description

The Clungene RSV Antigen Rapid Test is a lateral flow immunoassay based on the principle of the double-antibody sandwich technique. The nasopharyngeal swab from individuals is processed by the extraction buffer in buffer tube. After the extracted sample is added into the specimen well, the respiratory syncytial virus (RSV) nucleoprotein antigen in the sample will react with anti-RSV antigen antibody conjugated with color micro particles to form an antigen-antibody labeled complex. This complex migrates on the membrane via capillary action until the test line, where

Page 6

it will be captured by the pre-coated anti-RSV antigen antibody. If the specimen contains RSV nucleoprotein antigen, a colored line will appear in the test line (T) region indicating a positive result. If the specimen does not contain RSV nucleoprotein antigen, a colored line will not appear in this region indicating a negative result. To serve as a procedural control, a colored line will always appear at the control line (C) region indicating that proper volume of specimen has been added and membrane wicking has occurred.

The test result should be read at 15-20 minutes after adding the extracted sample is added into the specimen well. One line in the control line region (C), and another line in the test line region (T) indicates a positive result, regardless of color intensity. When only One line is present in the control region(C) and no line is present in the test line region (T) results are negative result. If the control line(C) fails to appear, the result is invalid, and the operator should review the procedure and repeat the test with a new test cassette.

  1. Substantial Equivalence Information

A summary comparison of features of the Clungene RSV Antigen Rapid Test and the predicate device is provided in the following table.

ContentsAmountDescription
Test cassettes25Each sealed foil pouch containing one test device and one desiccant
Buffer tubes25Tube sealed with foil film containing 0.3 mL of extraction buffer
Sterile Swabs25Single use nasopharyngeal swabs for specimen collection
Positive Control Swab1Non-infectious recombinant RSV nucleocapsid protein is dried on the swab tip
Negative Control Swab1Non-infectious recombinant Group C Streptococci is dried on the swab tip
Work Station1/
Instructions for Use1/
Quick Reference Instructions1/
Device & Predicate DevicePredicate device (K240280)Candidate device
Nano-Check RSV TestClungene RSV Antigen Rapid Test
Regulation Number21 CFR 866.3480Same
Regulatory ClassClass IISame
Product CodeGQGSame
Intended UseThe Nano-Check RSV Test is a rapid immunochromatographicThe Clungene RSV Antigen Rapid Test is a lateral flow immunoassay intended for

Page 7

assay for the qualitative detection of respiratory syncytial virus (RSV) nucleoprotein antigen in anterior nasal swab specimens from patients with signs and symptoms of respiratory infections. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients aged 6 months to 6 years old, and adults over 60 years of age. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as an FDA-cleared molecular assay.the qualitative detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. The test is intended for in vitro diagnostic use as an aid in the diagnosis of RSV infections. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or patient management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an FDA-cleared molecular assay.
Indication TypePrescription Use OnlySame
TechnologyImmunochromatographic assaySame
Test Result TypeQualitativeSame
Specimen TypeAnterior nasal swabNasopharyngeal swab
Test TargetNucleoprotein of RSVSame
InstrumentationNoneSame
Device FormatCassetteSame
Result InterpretationVisual determination of the presence or absence of colored line at the Test line and a colored line at the Control line indicate the presence or absence of RSVSame

Page 8

Time to Result15 minutes15-20 minutes
  1. Performance Summary

Analytical Performance

a. Limit of Detection (LoD)

To determine the Limit of Detection (LoD) of the Clungene RSV Antigen Rapid Test, four respiratory syncytial virus (RSV) strains were tested. Contrived samples were prepared by spiking each RSV strains (two RSV A and two RSV B strains) into pooled negative nasal fluid matrix, which had been confirmed as RSV negative by RT-PCR prior to use.

The study was conducted in two phases. In the initial screening phase, serial 10-fold dilutions of each RSV strain were prepared in the negative matrix and tested in triplicate using three lots of the test kit. The lowest concentration demonstrating 100% positivity (3 out of 3 replicates) for each strain was identified and was selected for further refinement.

In the LoD determination phase, 2-fold dilutions were prepared around the preliminary LoD concentration identified in the screening phase. Operators tested 20 replicates of each dilution using three lots of the test kit. The final LoD for each RSV strain was established as the lowest concentration achieving at least 95% positivity.

The results are presented in the table below.

RSV SubtypeStrainLOD (TCID50/mL)Positive Rate
ARSV-A (2006 Isolate)1.5×10²60/60 (100%)
RSV-A (ATCC-2012-10)6.4×10³60/60 (100%)
BRSV-B (ATCC-2012-11)1.9×10²60/60 (100%)
RSV-B (B WV/14617/85)2.2×10³60/60 (100%)

b. Inclusivity

An inclusivity study was performed to demonstrate that the Clungene RSV Antigen Rapid Test can detect a broad range of RSV strains. In addition to the strains evaluated in the LoD study, five strains of RSV A and four strains of RSV B were tested in the study. Serial 10-fold dilutions of these RSV strains were prepared with the negative matrix and tested in triplicate using three lots of the test kit. The lowest concentration at which the device demonstrated 100% positivity (3 out of 3 replicates) was chosen for further study. Further characterization was performed by evaluating 2-fold dilutions of the RSV strains around the preliminary LoD. Operators tested 3 replicates using three lots of the test kit. The lowest concentration of each strain that resulted in 100% detection (3/3) is presented in the table below.

RSV SubtypeStrainLowest Concentration with 100% Detection
ARSV-A (A2)1.6×10³ TCID50/mL
RSV-A(Long)8.0×10² PFU/mL
RSV-A (1/2015 Isolate #1)6.3×10¹ TCID50/mL
RSV-A (2/2015 Isolate 2)1.2×10¹ TCID50/mL

Page 9

BRSV-A (A2023/06-12NSMM)7.0×10² TCID50/mL
RSV-B(CH93(18)-18)3.2×10² TCID50/mL
RSV-B (B1)7.0×10² TCID50/mL
RSV-B (18537)1.2×10³ TCID50/mL
RSV-B (9320)1.4×10³ TCID50/mL

c. High-dose Hook Effect

A high-dose hook effect study was conducted to determine if a Hook Effect would be observed at high analyte concentration (i.e., a false negative at high concentrations of RSV). Four different RSV strains were diluted to four or five different concentrations in pooled negative nasal fluid matrix and tested in three replicates using three lots of the Clungene RSV Antigen Rapid Test. Concentrations ranged from 10,000x LoD (the maximum virus concentration possible, undiluted from the viral stock) to 1x LoD diluted in negative clinical matrix. All tested samples demonstrated 100% positivity, as expected. The Clungene RSV Antigen Rapid Test did not display a Hook Effect for the tested RSV concentrations.

RSV StrainConcentration (TCID50/mL)Positive Agreement (RSV positive/Total tested)
RSV-A (2006 Isolate)1.5×10⁶(10000×LoD)100% (9/9)
1.5×10⁵ (1000×LoD)100% (9/9)
1.5×10⁴ (100×LoD)100% (9/9)
1.5×10³ (10×LoD)100% (9/9)
1.5×10² (1×LoD)100% (9/9)
RSV-A (ATCC-2012-10)6.4×10⁶ (1000×LoD)100% (9/9)
6.4×10⁵ (100×LoD)100% (9/9)
6.4×10⁴ (10×LoD)100% (9/9)
6.4×10³ (1×LoD)100% (9/9)
RSV-B (ATCC-2012-11)3.8×10⁴ (200×LoD)100% (9/9)
1.9×10⁴ (100×LoD)100% (9/9)
1.9×10³ (10×LoD)100% (9/9)
1.9×10² (1×LoD)100% (9/9)
RSV-B (B WV/14617/85)8.9×10⁵ (400×LoD)100% (9/9)
2.2×10⁵ (100×LoD)100% (9/9)
2.2×10⁴ (10×LoD)100% (9/9)
2.2×10³ (1×LoD)100% (9/9)

d. Specificity / Cross Reactivity and Microbial Interference

The potential cross-reactivity and microbial interference of the Clungene RSV Antigen Rapid Test were evaluated using a panel of common organisms. Cross reactivity was assessed by testing each organism with RSV negative samples, and

Page 10

microbial interference was assessed by testing each organism with samples containing RSV at concentrations 2x LoD. A total of 56 organisms were tested across three device lots. All RSV negative samples gave negative results, and all RSV positive samples gave positive results at the tested concentrations, showing no cross-reactivity or microbial interference with the Clungene RSV Antigen Rapid Test.

MicroorganismConcentration TestedCross Reactivity (negative / total test)Microbial Interference (positive / total test)
Adenovirus Type 1,Adenoid 711.6×10⁶ TCID50/mL9/99/9
Adenovirus Type 2,Adenoid 6 (NIAID 202-001-014)1.1×10⁷ TCID50/mL9/99/9
Adenovirus Type 3,G.B.8.9×10⁸ TCID50/mL9/99/9
Adenovirus Type B78.9×10⁵ TCID50/mL9/99/9
SARS-CoV-2, USAWA1/20203.8×10⁵ TCID50/mL9/99/9
Human coronavirus OC431.6×10⁵ TCID50/mL9/99/9
Human coronavirus 229E2.8×10⁵ TCID50/mL9/99/9
Human coronavirus NL631.8×10⁵ TCID50/mL9/99/9
Human herpesvirus 5 (Human Cytomegalovirus), Merlin, ATCC-2011-31.2×10⁵ TCID50/mL9/99/9
Human Parainfluenza virus type 1, C358.9×10⁵ TCID50/mL9/99/9
Human Parainfluenza virus type 2, Greer8.9×10⁵ TCID50/mL9/99/9
Human Parainfluenza virus type 38.5×10⁶ TCID50/mL9/99/9
Human Parainfluenza virus type 4A, M-251.6×10⁵ TCID50/mL9/99/9
Measles virus, Edmonston2.1×10⁵ TCID50/mL9/99/9
Mumps virus, 11.2×10⁶ TCID50/mL9/99/9
Rhinovirus 1A, 20602.2×10⁶ PFU/mL9/99/9
Herpes simplex virus 1, Maclntyre1.4×10⁶ TCID50/mL9/99/9
Varicellovirus Human herpesvirus 3,Ellen2.9×10⁶ TCID50/mL9/99/9
Epstein-Barr Virus, P-33.2×10⁷ Copies/mL9/99/9
Rotavirus A, WA (TCadapted)5.0×10⁶ TCID50/mL9/99/9
Enterovirus 68, 2007 Isolate1.5×10⁵ TCID50/mL9/99/9
Enterovirus 71, H8.9×10⁶ TCID50/mL9/99/9
MERS-CoV, Florida/USA2_Saudi Arabia_20148.0×10⁵ TCID50/mL9/99/9

Page 11

MicroorganismConcentration TestedCross Reactivity (negative / total test)Microbial Interference (positive / total test)
Influenza A (H1N1),Brisbane/59/073.6×10⁵ TCID50/mL9/99/9
Influenza A (H1N1),A/California/7/20092.3×10⁷ TCID50/mL9/99/9
Influenza A (H3N2),Brisbane/10/075.3×10⁵ TCID50/mL9/99/9
Influenza A (H3N2),Victoria/361/113.8×10⁵ TCID50/mL9/99/9
Influenza A (H3N2),Hong Kong/8/681.3×10⁵ TCID50/mL9/99/9
Influenza A (H3N2),Wisconsin/67/051.1×10⁵ TCID50/mL9/99/9
Influenza B(Yamagata),Yamagata/16/881.3×10⁵ TCID50/mL9/99/9
Influenza B (Yamagata),Victoria/504/006.3×10⁵ TCID50/mL9/99/9
Influenza B(Victoria),Malaysia/2506/041.9×10⁶ TCID50/mL9/99/9
Influenza B (Victoria),Victoria/2/877.6×10⁵ TCID50/mL9/99/9
Influenza B (Victoria) B/Florida/78/20152.8×10⁶ TCID50/mL9/99/9
Human Metapneumovirus 16 Type A1 , IA10-20036.3×10⁵ TCID50/mL9/99/9
Coxsackievirus, B4, J.V.B. (Benschoten)1.6×10⁶ TCID50/mL9/99/9
Bordetella pertussis8.7×10⁷ CFU/mL9/99/9
Chlamydophila pneumonia, TW-1839.1×10⁶ IFU/mL9/99/9
Mycoplasma pneumoniae M129-B74.2×10⁵ CFU/mL9/99/9
Pseudomonas aeruginosa2.8×10⁸ CFU/mL9/99/9
Staphylococcus aureus9.5×10⁸ CFU/mL9/99/9
Haemophilus influenzae8.0×10⁷ CFU/mL9/99/9
Streptococcus pneumoniae1.0×10⁷ CFU/mL9/99/9
Streptococcus pyogenes1.3×10⁸ CFU/mL9/99/9
Streptococcus salivarius1.3×10⁷ CFU/mL9/99/9
Streptococcus mutans1.2×10⁶ CFU/mL9/99/9
Legionella pneumophila1.6×10⁶ CFU/mL9/99/9
Klebsiella pneumoniae5.8×10⁷ CFU/mL9/99/9
Staphylococcus epidermidis1.2×10⁹ CFU/mL9/99/9

Page 12

MicroorganismConcentration TestedCross Reactivity (negative / total test)Microbial Interference (positive / total test)
Mycobacterium tuberculosis1.2×10⁷ CFU/mL9/99/9
Candida albicans3.6×10⁶ CFU/mL9/99/9
Corynebacterium diphtheriae1.4×10⁶ CFU/mL9/99/9
Moraxella catarrhalis2.2×10⁸ CFU/mL9/99/9
Lactobacillus sp.5.0×10⁷ CFU/mL9/99/9
Neisseria meningitidis8.8×10⁷ CFU/mL9/99/9
Escherichia coli EAEC2.6×10⁷ CFU/mL9/99/9

e. Endogenous and Exogenous Interfering Substances

Potentially interfering substances that may be present in respiratory specimens or that may be artificially introduced into the specimen were evaluated with the Clungene RSV Antigen Rapid Test. The interferants were prepared in pooled negative nasal fluid (PNF) at their recommended concentration. Negative samples were prepared using virus-negative pooled nasal fluid (PNF) specimens containing each interfering substance, and positive samples were prepared by combining RSV-A at 2x LoD in PNF with each interfering substance individually. All samples were evaluated in triplicate per lot using three different lots of devices to confirm that the interfering substance did not produce false result in either presence or absence of RSV. None of the tested potential interfering substances showed false positive or false negative test results using the Clungene RSV Antigen Rapid Test with RSV positive and negative samples at the concentrations specified below.

SubstanceConcentration TestedDiluted with positive sample (positive/ total test)Diluted with negative sample (negative / total test)
Purified mucin protein2.5 mg/mL9/99/9
Whole Blood4%9/99/9
Phenylephrine HCl15% v/v9/99/9
Oxymetazoline HCl15% v/v9/99/9
Sodium chloride with preservatives15% v/v9/99/9
Beclomethasone0.5 mg/mL9/99/9
Budesonide Nasal Spray15% v/v9/99/9
Mometasone furoate0.5 mg/mL9/99/9
Dexamethason5 mg/mL9/99/9
Flunisolide5 mg/mL9/99/9
Triamcinolone acetonide5 mg/mL9/99/9
Fluticasone propionate5 mg/mL9/99/9
Histamine dihydrochloride10 mg/mL9/99/9
Benzocaine3 mg/mL9/99/9

Page 13

SubstanceConcentration TestedDiluted with positive sample (positive/ total test)Diluted with negative sample (negative / total test)
Menthol3 mg/mL9/99/9
Zanamivir5 mg/mL9/99/9
Ribavirin5 mg/mL9/99/9
Oseltamivir Phosphate5 mg/mL9/99/9
Peramivir100 μg/mL9/99/9
Mupirocin10 mg/mL9/99/9
Tobramycin5 μg/mL9/99/9
D-Biotin3500 ng/mL9/99/9
Remdesivir0.2 mg/mL9/99/9
Molnupiravir2.5 mg/mL9/99/9
Paracetamol50 μg/mL9/99/9
Euphorbium Nasal Spray15% v/v9/99/9
Sinna Nasal Spray15% v/v9/99/9
Quantum HEALTH TheraZinc15% v/v9/99/9
Leukocytes5×10⁶ cells/mL9/99/9

f. Precision/Reproducibility Study

A reproducibility study was performed to evaluate the reproducibility of the Clungene RSV Antigen Rapid Test. The study was performed at three external testing sites consisting of four samples at various virus concentrations including near the respective LoD (i.e., true negative, high negative, low positive, and moderate positive). These samples were tested by six untrained operators at 5 replicates per day for each of three lots over 5 days for a total of 450 replicates per concentration and a total of 1,800 data points collected. The results were 100% agreement between expected and read results within replicates, by lot, by operator, by day, between sites, and overall.

SampleExpected Result/Total TestedTotal % Agreement with Expected Results (95% Confidence Intervals)
Site ASite BSite C
Negative swab sample150/150 (100%)150/150 (100%)150/150 (100%)450/450 (100%) (99.2%-100%)
RSV high negative swab sample (0.1×LOD)150/150 (100%)150/150 (100%)150/150 (100%)450/450 (100%) (99.2%-100%)
RSV low positive swab sample (1×LOD)150/150 (100%)150/150 (100%)150/150 (100%)450/450 (100%) (99.2%-100%)
RSV moderate positive swab sample (3×LOD)150/150 (100%)150/150 (100%)150/150 (100%)450/450 (100%) (99.2%-100%)

Page 14

A lot-to-lot precision study was conducted at one external site. Test samples were contrived in negative nasal clinical matrix (NCM) and spiked with RSV-A strain to target four concentrations: true negative (NCM, unspiked), high negative (0.1x LoD), low positive (1x LoD), and moderate positive (3x LoD). The test samples were prepared by coating the viral solutions onto the swabs. The samples were randomized and blind-coded. The study was performed over twelve non-consecutive days by six untrained operators using three lots. Each operator performed two replicates at each concentration per day using one lot of the device. Two separate operators tested each lot every other day (i.e. operator 1 tested lot 1 on days 1, 3, 5, 7, 9, and 11, while operator 4 tested lot 1 on days 2, 4, 6, 8, 10, and 12). Each sample concentration was tested a total of 72 times (2 replicates x 12 days x 3 lots). Results are summarized below.

SampleExpected Result/Total TestedTotal % Agreement with Expected Results (95% Confidence Intervals)
Lot 1Lot 2Lot 3
Negative swab sample24/24 (100%)24/24 (100%)24/24 (100%)72/72 (100%) (94.9%-100%)
RSV high negative swab sample (0.1×LOD)24/24 (100%)24/24 (100%)24/24 (100%)72/72 (100%) (94.9%-100%)
RSV low positive swab sample (1×LOD)24/24 (100%)24/24 (100%)24/24 (100%)72/72 (100%) (94.9%-100%)
RSV moderate positive swab sample (3×LOD)24/24 (100%)24/24 (100%)24/24 (100%)72/72 (100%) (94.9%-100%)

g. Internal Controls

The Clungene RSV Antigen Rapid Test has a built-in internal procedural control which monitors that sufficient volume of the sample is added and the reagent flow across the membrane occurred. Formation of a pinkish-red line in the control region on the strip confirms proper sample application and that the reagents are functioning appropriately. If no visible signal appears on the control line, the test result is invalid, and the operator is instructed in the labelling to test the sample again with another test device.

h. External Controls

The Clungene RSV Antigen Rapid Test contains one positive external control swab and one negative external control swab that allows for monitoring of the performance of the assay. The positive control swab contains noninfectious recombinant RSV nucleocapsid protein, and the negative control swab contains noninfectious recombinant Group C Streptococci.

Lot-to-lot precision of the external positive and negative control swabs was evaluated by testing ten replicates each of three lots of the external controls. For each external

Page 15

control lot, all positive and negative controls produced 100% agreement with the expected results, as summarized below.

External ControlControl Lot No.Test Result/Total Tests% Agreement to Expected Result
# Negative# Positive
External Positive Control10/1010/10100%
20/1010/10100%
30/1010/10100%
External Negative Control110/100/10100%
210/100/10100%
310/100/10100%

i. Specimen Stability

Specimen stability was assessed using a negative sample (consisting of pooled negative human nasal fluid) and a contrived low positive sample (containing RSV-A strain at 2x LoD in the pooled negative human nasal fluid). Contrived samples on the swabs were stored at 2-8°C, room temperature, 35°C, and -20°C and tested at the following timepoints: 1, 2, 3, and 4 hours for room temperature and 35°C; 6, 12, 24, and 48 hours for 2-8°C; and 1, 3, and 5 days for -20°C. Three replicates of each concentration were tested in accordance with the package insert at each timepoint for each condition using three device lots by three operators.

Swabs were also stored at -20°C for 24 hours, then transferred to room temperature for thawing, constituting one freeze-thaw cycle. The above was repeated to make two freeze-thaw cycles. After each freeze-thaw cycle, tests were performed in triplicate using three device lots by three operators.

The results obtained in these studies are summarized below.

Storage TemperatureTesting time post-storageTest Results of Positive Swab SpecimensTest Results of Negative Swab Specimens
Lot 1Lot 2Lot 3Lot 1Lot 2Lot 3
/0h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
2-8℃6h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
12h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
24h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
48h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
Room temperature1h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
2h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
3h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
4h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
35℃1h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
2h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
3h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
4h+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)

Page 16

Storage TemperatureTesting time post-storageTest Results of Positive Swab SpecimensTest Results of Negative Swab Specimens
Lot 1Lot 2Lot 3Lot 1Lot 2Lot 3
-20℃Day 1+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
Day 3+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
Day 5+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
Freeze-thaw cyclesTest Results of Positive Swab SpecimensTest Results of Negative Swab Specimens
Lot 1Lot 2Lot 3Lot 1Lot 2Lot 3
0 cycles+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
1 cycle+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)
2 cycles+ (3/3)+ (3/3)+ (3/3)- (3/3)- (3/3)- (3/3)

j. Prospective Clinical Study

The performance of the Clungene RSV Antigen Rapid Test in detecting RSV nucleoprotein antigen was evaluated in a multi-center, prospective study across five testing sites using nasopharyngeal (NP) swab samples collected from patients with signs and symptoms consistent with RSV. The study was conducted between November 2024 to April 2025. Informed consent was obtained for all patients prior to testing. Two NP swabs were collected from each subject. One swab (for comparator testing) was collected by the study operator and placed into transport media for comparator testing and the other swab was collected by the study operator and tested immediately with the Clungene RSV Antigen Rapid Test on site. The performance of the Clungene RSV Antigen Rapid Test with NP swabs was estimated based on the comparison with results obtained with an FDA cleared molecular test for RSV (positive percent agreement (PPA) and negative percent agreement (NPA)).

A total of 722 subjects with RSV symptoms were enrolled in the study and a total of 722 nasopharyngeal swab specimens were considered evaluable. Specimens were collected from patients between 0 to 97 years old. Of these, 20.2% (146/722) were from subjects less than 6 years old, 32.1% (232/722) were from patients 6 to 21 years old, and 47.6% (344/722) were from patients 22 years or older. There were 55.5% (401/722) females and 44.5% (321/722) males.

The clinical performance of the Clungene RSV Antigen Rapid Test, expressed as PPA and NPA, compared to the comparator method is presented below.

Clungene RSV Antigen Rapid TestComparator RT-PCR
PositiveNegative
Positive985
Negative13606
Total111611

Page 17

Positive Percent Agreement (PPA)88.3% (95% CI: 81.0%-93.0%)
Negative Percent Agreement (NPA)99.2% (95% CI: 98.1%-99.6%)

Expected Values/Reference Range

A patient sample is expected to be negative for RSV.

  1. Conclusion

Based on the data submitted in this traditional 510(k) submission, the Clungene RSV Antigen Rapid Test has been shown to be substantially equivalent in terms of intended use, technological characteristics, and assay performance to the predicate device.

§ 866.3480 Respiratory syncytial virus serological reagents.

(a)
Identification. Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.